(244 days)
The Digital Thermometer is intended to measure oral temperature of human body. It can be used in home environment or at a doctor's clinic. It is indicated for use by people of all ages.
The Digital Thermometer consists two parts. The thermometer probe part is a compact, small, light-weight, and battery-powered unit with thermometer sensor for oral temperature measurements. The thermometer probe must be connected to a compatible mobile phone device through its headphone jack for displaying measured temperatures.
The mobile application could be conducted for operation and display temperature measurement when the Digital Thermometer connects with mobile device. The mobile application can be conducted on iOS operating system and Android operating system of mobile device.
The mobile application could perform the functionality on mobile device screen including measurement activity, historical data activity, trend activity, basic setting and initial setting activity.
The provided text describes a digital thermometer, Model 2180, and its substantial equivalence to a predicate device, the Basal Digital Thermometer (K141505). The information primarily focuses on the device's characteristics, intended use, and compliance with various standards.
Here's an analysis of the acceptance criteria and the study information as requested, based only on the provided text:
1. A table of acceptance criteria and the reported device performance
The document lists several standards and requirements that the Digital Thermometer complies with, implying these act as acceptance criteria. The reported performance is that the device "complies" or "met the endpoints."
| Acceptance Criteria / Standard | Reported Device Performance |
|---|---|
| Measurement Accuracy | |
| ±0.1°C (35.5°C to 42.0°C) | Digital Thermometer: ±0.1°C (35.5°C to 42.0°C) |
| ±0.2°C (other range) | Digital Thermometer: ±0.2°C (other range) |
| Display Resolution | |
| 0.1°C (°F) | Digital Thermometer: 0.1°C (°F) |
| Displayed Temperature Range | |
| (Implicit from ASTM E1112-00) | Complies with ASTM E1112-00 displayed temperature range requirements |
| Environmental (Operating) | |
| 10°C~40°C (50 ~ 104°F) | Digital Thermometer: 10°C~40°C (50 ~ 104°F) |
| 15%-95% RH | Digital Thermometer: 15%-95% RH |
| Environmental (Storage) | |
| -20°C ~50°C (-4.0 ~ 122°F) | Digital Thermometer: -20°C ~50°C (-4.0 ~ 122°F) |
| 15%-95% RH | Digital Thermometer: 15%-95% RH |
| Biocompatibility (Cytotoxicity) | |
| No Cytotoxicity (ISO 10993-5) | Met endpoints for cytotoxicity |
| Biocompatibility (Sensitization & Irritation) | |
| No Sensitization & Irritation-negligible (ISO 10993-10) | Met endpoints for sensitization & irritation-negligible |
| Electrical Safety (AAMI / ANSI ES60601-1) | Complies with general requirements, protection against electrical/mechanical hazards, excessive temperatures, etc. |
| Electromagnetic Compatibility (AAMI / ANSI/ IEC 60601-1-2) | Complies with radiated emission, electrostatic discharge immunity, RF electromagnetic field immunity, etc. |
| Human Interface and Mechanical Test (ASTM E1112-00) | Complies with ASTM E1112-00 requirements |
| Software Guidance (Premarket Submissions for Software) | Complies with guidance requirements including level of concern, software description, V&V documentation. |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the sample size used for any specific test set, nor does it mention data provenance (e.g., country of origin, retrospective/prospective). It generally states that the device was "verified and validated" or "complies" with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The document refers to compliance with established standards (e.g., ASTM E1112-00 for accuracy), which inherently have their own defined test methodologies, but details about specific expert involvement in this device's testing ground truth establishment are absent.
4. Adjudication method for the test set
The document does not describe any adjudication method (e.g., 2+1, 3+1) for establishing ground truth for a test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable as the device is a digital thermometer, not an AI-assisted diagnostic tool involving human readers. No MRMC study was mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device is a standalone digital thermometer that provides a temperature reading. Its performance criteria (e.g., accuracy) were assessed in this context. The "software validation" mentioned indicates evaluation of the algorithm within the device and mobile application, which can be seen as a standalone performance assessment for its accuracy and functionality.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the performance of the thermometer, the "ground truth" would be established by reference standards or calibrated equipment used in accordance with the specified testing methodologies of standards like ASTM E1112-00. The text does not detail the specifics of such "ground truth" generation for its internal testing beyond stating compliance with the standard.
8. The sample size for the training set
This information is not applicable. The device is a digital thermometer, not a machine learning model that requires a training set in the conventional sense.
9. How the ground truth for the training set was established
This information is not applicable, as there is no "training set" for this type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized, sans-serif font, and the color is blue. The "K" and "jump" are connected, and there is a registered trademark symbol to the right of the word "jump".
510(k) Summary
1. Submitter
K-jump Health Co., Ltd.
No. 56, Wu Kung 5th Rd. New Taipei Industrial Park New Taipei City 24890, Taiwan
Contact person: JM Lin, Regulatory Affairs Representative +886-2-2299-1378 ext. 237 Phone: Fax: +886-2-2299-1385 Email: jm@kjump.com.tw
Date Prepared: January 25, 2017
2. Name of Device
Trade Name: Digital Thermometer, Model 2180 Common/Usual Name: Digital Thermometer Regulation Number: 21 CFR 880.2910 Classification Name: Clinical electronic thermometer Regulatory Class: II Product Code: FLL
3. Predicate Device
| Device Name | 510(k) Number | Decision Date |
|---|---|---|
| K-jump Health Co., Ltd.Basal Digital Thermometer | K141505 | 8/26/2014 |
This predicate has not been subject to a design-related recall.
{1}------------------------------------------------
4. Device Description
The Digital Thermometer consists two parts. The thermometer probe part is a compact, small, light-weight, and battery-powered unit with thermometer sensor for oral temperature measurements. The thermometer probe must be connected to a compatible mobile phone device through its headphone jack for displaying measured temperatures.
The mobile application could be conducted for operation and display temperature measurement when the Digital Thermometer connects with mobile device. The mobile application can be conducted on iOS operating system and Android operating system of mobile device.
The mobile application could perform the functionality on mobile device screen including measurement activity, historical data activity, trend activity, basic setting and initial setting activity.
5. Indications for Use
The Digital Thermometer is intended to measure oral temperature of human body. It can be used in home environment or at a doctor's clinic. It is indicated for use by people of all ages.
6. Technological Characteristics
The Digital Thermometer uses a thermistor sensor to detect the thermal energy from human body. The sensor in the thermometer detects temperature change and transfers to the proper electronic signal. The electronic signal could be received and converted to the temperature reading. The measured data will be displayed on the screen of a mobile device. The technological characteristics of Digital Thermometer have no significant difference with predicate device.
7. Comparison of technological characteristics with Predicate Device
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized blue font. There is a registered trademark symbol to the right of the word jump.
The Digital Thermometer is substantially equivalent to the predicate devices, K141505, the Basal Digital Thermometer. The comparison of their technological characteristics is summarized in the table below.
| Characteristics | Digital Thermometer | Basal Digital Thermometer |
|---|---|---|
| 510(k) Number | K170262 | K141505 |
| Intended Use | Similar | Similar |
| Technology | ||
| Measurement method | Thermistor detection | Thermistor detection |
| Feature | Compatible with iOS / AndroidMobile Device via headphone jackconnectionMeasured data display on thescreen of mobile device | Compatible with iOS/AndroidMobile Device viaheadphone/NFC connectionMeasured data transfer tomobile device |
| Function | ||
| Measuring Range | 32.0°C~43.9°C | 32.00°C~43.90°C |
| C/F Switchable | Yes | Yes |
| Performance | ||
| Display | 3 1/2 digitson the screen of a mobile device | 4 1/2 digitson the LCD of medical device |
| Display Resolution | 0.1°C (°F) | 0.01°C (°F) |
| Measuring accuracy | ±0.1°C (35.5°C to 42.0°C);±0.2°C (other range) | ±0.1°C (34.00°C to 42.00°C);±0.2°C (other range) |
| Low battery detection | Yes | Yes |
| Memory | 1 set memory | 40 set memory |
| Environmental | ||
| Operating condition | 10°C~40°C (50 ~ 104°F)15%-95% RH | 5°C~40°C (41.0 ~ 104°F)15%-95% RH |
| Storage condition | -20°C ~50°C (-4.0 ~ 122°F)15%-95% RH | -25°C ~55°C (-13.0 ~ 131°F)15%-95% RH |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for K-Jump. The logo is blue and features a stylized "K" with a dot above it, resembling a person jumping. The word "Jump" is written in a cursive font next to the "K", and there is a registered trademark symbol next to the word "Jump".
| Characteristics | Digital Thermometer | Basal Digital Thermometer | |
|---|---|---|---|
| Material | Component materials | Stainless Steel,Acrylonitrile butadiene styrene | Stainless Steel,Acrylonitrile butadiene styrene |
| Biocompatibility | No CytotoxicityNo Sensitization &Irritation-negligible | No CytotoxicityNo Sensitization &Irritation-negligible | |
| Power | Internal power(one SR41 battery) | Internal power(one CR2032 battery) | |
| Physical | Dimensions | 87.5 x 14.2 x 8.7 mm | 125.5 x 34 x 9.6 mm |
| Weight | 6g (with battery) | 24g (with battery) |
> Discussion of Main Differences:
The Digital Thermometer connects to the mobile device through headphone jack. The measured data will be displayed on the screen of a mobile device. The predicate device could display measured data on the device and transfer measured data to a mobile device.
While connected with mobile device, the usage of Digital Thermometer is identical to the predicate device. Risk analysis, software validation and associated verification have been performed and demonstrated that the display feature of the subject device is acceptable;
8. Performance Summary
The performance of the Digital Thermometer is verified and validated to comply with following recognized standards and does not raise any safety and effectiveness issues.
-
- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,, C1:2009/(R)2012 and A2:2010/(R)2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized blue font. There is a registered trademark symbol to the right of the word jump.
The Digital Thermometer complies with to applicable ANSI/AAMI ES60601-1 requirements including general requirements, protection against electrical hazards, protection against mechanical hazards, protection against excessive temperatures, hazardous situations and fault conditions, and constructions.
-
- AAMI / ANSI/ IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
The Digital Thermometer complies with applicable ANSI/AAMI/IEC 60601-1-2 requirements including radiated emission test, electrostatic discharge immunity test, radiated RF electromagnetic field immunity test, and power frequency magnetic field immunity test.
- AAMI / ANSI/ IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
-
- ASTM E1112-00 (Reapproved 2011), Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
The Digital Thermometer complies applicable ASTM E1112 requirements including displayed temperature range, maximum permissible laboratory error, ambient conditions for operation and storage, human interface and mechanical test.
- ASTM E1112-00 (Reapproved 2011), Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
-
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
The Digital Thermometer complicable the guidance requirements including level of concern, software description, software requirements specification, architecture design chart, software design specification, verification and validation documentation.
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
-
- Biocompatibility:
AAMI / ANSI / ISO 10993-1:2009(R)2013 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process;
- Biocompatibility:
AAMI / ANSI / ISO 10993-5:2009(R)2014 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity and
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized blue font. To the right of the word "K-Jump" is the registered trademark symbol.
AAMI / ANSI / ISO 10993-10:2010(R)2014 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
The Digital Thermometer complies applicable biocompatibility requirements and the device met the endpoints for cytotoxicity, sensitization & irritation-negligible
The Digital Thermometer is verified and validated by evaluating safety, EMC, device-specific standard ASTM E1112-00, software validation and biocompatibility testing without raising any safety and effectiveness issues.
9. Conclusions
The Digital Thermometer has the similar intended use, materials, technological characteristics and the same fundamental scientific technology with the predicate device. Moreover, both devices comply with same performance standards. All information described above can demonstrate the subject device is substantial equivalent to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, abstract design connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 28, 2017
K-jump Health Co., Ltd. JM Lin Regulatory Affairs Representative No. 56. Wu Kung 5th Rd., New Taipei Industrial Park New Taipei City, 24890 TAIWAN
Re: K170262
Trade/Device Name: Digital Thermometer, Model 2180 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 30, 2017 Received: August 31, 2017
Dear JM Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
§ 880.2910 Clinical electronic thermometer.
(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.