Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133114
    Manufacturer
    Date Cleared
    2014-07-03

    (276 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Micropuncture Introducer Set is intended for the placement of wire guides up to .038 inch diameter into the peripheral vascular system when a small 21 gage needle stick is desired.

    Device Description

    The Micropuncture Introducer Sets contains a coaxial catheter, a wire guide, and a needle. The device is utilized to gain access to the vasculature using the Seldinger technique.

    AI/ML Overview

    Here's an analysis of the provided document regarding the acceptance criteria and supporting studies for the Micropuncture® Introducer Set:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document provides a general statement that "The predetermined acceptance criteria were met" for each test. Specific numerical acceptance criteria were not detailed in this summary.

    Test ConductedAcceptance CriteriaReported Device Performance
    Insertion force testingPredetermined acceptance criteria (not specified numerically)Met
    Coaxial catheter tensile testingHub would not be dislodged during insertion and withdrawal (not specified numerically)Met
    Wire guide resistance to damage testingNo damage or defects after repeated flexing (not specified numerically)Met
    Wire guide resistance to fracture testingNo signs of fracture (not specified numerically)Met
    Wire guide tensile strength testingDistal tip withstands 4.0N before failureMet
    Wire guide torque response and torque strengthOutput rotation within 90 degrees of input rotation; withstands 10 rotations without failureMet
    Biocompatibility testing (various assays)In conformance with ISO 10993-1:2009 (not specified numerically)Met

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not provide information on the sample sizes used for any of the tests. The data provenance (e.g., country of origin, retrospective/prospective) is also not mentioned.

    3. Number of Experts Used to Establish Ground Truth and Their Qualifications

    This information is not applicable as the document describes performance testing for a medical device (introducer set), not a diagnostic algorithm or image analysis tool requiring expert ground truth for interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reason as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The document describes laboratory-based performance testing, not studies involving human readers or AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The device described is a physical medical instrument, not a software algorithm.

    7. The Type of Ground Truth Used

    The concept of "ground truth" as typically applied to diagnostic algorithms is not directly applicable here. The tests performed are laboratory-based engineering and material science tests. The "ground truth" would be the direct measurement of physical properties (e.g., force, rotation, tensile strength) or biological responses (e.g., cytotoxicity).

    8. The Sample Size for the Training Set

    Not applicable. The device is a physical product, not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1