K Number
K133114
Manufacturer
Date Cleared
2014-07-03

(276 days)

Product Code
Regulation Number
870.1340
Panel
CV
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Micropuncture Introducer Set is intended for the placement of wire guides up to .038 inch diameter into the peripheral vascular system when a small 21 gage needle stick is desired.

Device Description

The Micropuncture Introducer Sets contains a coaxial catheter, a wire guide, and a needle. The device is utilized to gain access to the vasculature using the Seldinger technique.

AI/ML Overview

Here's an analysis of the provided document regarding the acceptance criteria and supporting studies for the Micropuncture® Introducer Set:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a general statement that "The predetermined acceptance criteria were met" for each test. Specific numerical acceptance criteria were not detailed in this summary.

Test ConductedAcceptance CriteriaReported Device Performance
Insertion force testingPredetermined acceptance criteria (not specified numerically)Met
Coaxial catheter tensile testingHub would not be dislodged during insertion and withdrawal (not specified numerically)Met
Wire guide resistance to damage testingNo damage or defects after repeated flexing (not specified numerically)Met
Wire guide resistance to fracture testingNo signs of fracture (not specified numerically)Met
Wire guide tensile strength testingDistal tip withstands 4.0N before failureMet
Wire guide torque response and torque strengthOutput rotation within 90 degrees of input rotation; withstands 10 rotations without failureMet
Biocompatibility testing (various assays)In conformance with ISO 10993-1:2009 (not specified numerically)Met

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not provide information on the sample sizes used for any of the tests. The data provenance (e.g., country of origin, retrospective/prospective) is also not mentioned.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications

This information is not applicable as the document describes performance testing for a medical device (introducer set), not a diagnostic algorithm or image analysis tool requiring expert ground truth for interpretation.

4. Adjudication Method for the Test Set

This information is not applicable for the same reason as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The document describes laboratory-based performance testing, not studies involving human readers or AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. The device described is a physical medical instrument, not a software algorithm.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically applied to diagnostic algorithms is not directly applicable here. The tests performed are laboratory-based engineering and material science tests. The "ground truth" would be the direct measurement of physical properties (e.g., force, rotation, tensile strength) or biological responses (e.g., cytotoxicity).

8. The Sample Size for the Training Set

Not applicable. The device is a physical product, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reason as point 8.

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).