Search Results
Found 1 results
510(k) Data Aggregation
(203 days)
Wallaby 017 Micro Catheter
The Wallaby 017 Micro Catheter is intended to access peripheral and neuro vasculature for the controlled selective infusion of physician-specified therapeutic devices such as embolization materials such as contrast media as well as delivery of embolic coils.
The Wallaby 017 Micro Catheter is a single-use, vascular catheter consisting of a single lumen, variable stiffness, composite catheter. The device is a microcatheter with an inner diameter (ID) of 0.017", designed with a working length of 150 cm. The device has three different tip configurations: straight (0°), 45°, 90°, and is steam shapeable by the user. The device is supplied as a kit with an introducer sheath, shaping mandrel, and mandrel card provided with a single catheter. The distal tip of the Wallaby 017 Micro Catheter is visible under fluoroscopy and the distal shaft of each catheter is designed with an external hydrophilic coating to reduce friction during use. The proximal end of each microcatheter incorporates a strain relief and a standard luer adapter to facilitate the attachment of accessories. Each catheter has a semi-rigid proximal shaft which transitions into a flexible distal shaft to facilitate the advancement of the catheter in tortuous vasculature.
The Wallaby 017 Micro Catheter is a non-active, surgically invasive device intended for limited duration use within the vasculature.
Here's a summary of the acceptance criteria and the study details for the Wallaby 017 Micro Catheter, based on the provided document.
It's important to note that the document describes a medical device rather than an AI/ML powered device. Therefore, many of the typical questions for AI studies (like sample size for test/training sets, ground truth methodology with experts, adjudication, or MRMC studies) are not applicable here. The "performance" being evaluated is the physical and functional performance of the catheter itself, not the performance of an algorithm.
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this medical device are based on passing various bench tests and biocompatibility evaluations. The full "acceptance criteria" are implied by "met all pre-defined acceptance criteria" for each test. Specific numerical thresholds for each criterion are not provided in this summary but would have been defined in detailed test protocols.
Table of Acceptance Criteria and Reported Device Performance:
Test Category | Specific Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|---|
Design Verification | Visual Inspection | Met pre-defined acceptance criteria | Met all pre-defined acceptance criteria. |
Dimensional Inspection | Met pre-defined acceptance criteria | Met all pre-defined acceptance criteria. | |
Simulated Use | Performs as intended, met all criteria | Performs as intended and met all pre-defined acceptance criteria under simulated use conditions. | |
Physician Validation (Usability) | Performs as intended, equivalent to comparator | Performs as intended and demonstrates equivalency to the comparator device under simulated use conditions. | |
Delivery and Retrieval | Met pre-defined acceptance criteria | Met all pre-defined acceptance criteria. | |
Tip Stiffness | Met acceptance criteria | Met the acceptance criteria. | |
Tip Shaping | Met pre-defined acceptance criteria | Met the pre-defined acceptance criteria. | |
System Tensile (hub, shaft, tip) | Met minimum tensile strength requirement | Met the predefined acceptance criteria. | |
Elongation to Failure | Met all pre-defined acceptance criteria | Met all pre-defined acceptance criteria. | |
Torque To Failure | Met pre-defined acceptance criteria | Met all pre-defined acceptance criteria. | |
Coating Integrity | Met pre-defined acceptance criteria | Met all pre-defined acceptance criteria. | |
Coating Lubricity | Met all pre-defined acceptance criteria | Met all pre-defined acceptance criteria. | |
Catheter Dynamic and Static Burst/Leak | Does not leak/burst, compatible with accessories | Evaluated to verify the device does not leak, burst, and is compatible with accessories per ISO 10555-1 and ISO 594-1. | |
Kink Resistance | Met acceptance criteria | Met acceptance criteria. | |
Particulate | Met pre-defined acceptance criteria, comparable to reference device | Met all pre-defined acceptance criteria and was comparable to the reference device. | |
Corrosion Resistance | Corrosion resistant per ISO 10555-1 | Corrosion resistant per ISO 10555-1. | |
Radiopacity | Marker band visibility under fluoroscopy, met criteria | Met the pre-defined acceptance criteria. | |
DMSO and Liquid Embolic Compatibility | Met pre-defined acceptance criteria | Met the pre-defined acceptance criteria. | |
Biocompatibility | Cytotoxicity (MTT - L-929) | Non-cytotoxic | Non-cytotoxic. |
Intracutaneous Irritation | Non-irritant | Non-irritant. | |
Sensitization (Guinea Pig Maximization) | Non-sensitizing | Non-sensitizing. | |
Acute Systemic Toxicity | Non-toxic (no abnormal clinical signs) | Non-toxic. | |
Rabbit Pyrogen | Non-pyrogenic | Non-pyrogenic. | |
Complement Activation - SC5b-9 Assay | Not a potential activator of complement system | Not a potential activator of complement system. | |
Hemolysis — Direct Contact and Extract Method | Non-hemolytic | Non-hemolytic. | |
Thromboresistance Evaluation | No adverse effects, thrombus score ≤ 3 | Thromboresistance of test device similar to control device. | |
Chemical Characterization (Physiochemical) | Extractables/leachables similar to reference, pass | Pass. | |
Sterilization/Shelf Life | Sterility Assurance Level | SAL of 10^-6^ (ISO 11135:2014) | Verified to ensure a sterility assurance level (SAL) of 10^-6^. |
Shelf Life (12 months) | Device and packaging remain functional | Established that the device and packaging remain functional for the 12-month shelf-life. |
Study Details (Applicable for a medical device cleared via 510(k))
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- The document does not specify exact sample sizes (N-values) for each bench test. For example, "The device was evaluated..." doesn't indicate if this was one unit or multiple units.
- The data provenance is from non-clinical bench testing performed by Wallaby Medical for regulatory submission in the United States (as indicated by the FDA 510(k) process). It is prospective in the sense that the testing was conducted specifically to support this regulatory submission.
- No human-collected test sets (like medical images) are involved, so country of origin of data in that sense is not applicable.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable as this is a physical medical device, not an AI/ML algorithm. "Ground truth" for physical device performance is established through standardized engineering and biological tests against predefined specifications and industry standards (e.g., ISO standards).
- For the "Physician Validation (Usability)" test, it states "The device was evaluated in a simulated anatomy model by physicians." The number and qualifications of these physicians are not detailed in this summary.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This concept pertains to expert review for AI/ML algorithm performance. For physical device testing, results are typically objective measurements or observations against predefined pass/fail criteria.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a non-AI medical device without human "readers" in the context of an AI study.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. The device itself is standalone in the sense that it is a physical product, but the concept of "standalone algorithm performance" does not apply.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- "Ground truth" for this device's performance is established by engineering specifications, industry standards (e.g., ISO 10555-1, ISO 594-1, ISO 11135:2014, ISO 10993 series, USP ), and comparison to legally marketed predicate devices. The results of the bench tests and biocompatibility evaluations, when conforming to these standards and specifications, are the "ground truth" for proving the device's safety and effectiveness.
-
The sample size for the training set:
- Not applicable. There is no AI/ML model for which to define a training set.
-
How the ground truth for the training set was established:
- Not applicable. There is no AI/ML model or training set.
Ask a specific question about this device
Page 1 of 1