Search Results
Found 1 results
510(k) Data Aggregation
(47 days)
VASSALLO GT 018 G12; VASSALLO GT 018 G30
This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
The VASSALLO® GT 018 G12 and G30 Peripheral Guide Wires in this submission are a steerable guide wire with a maximum diameter of 0.018 inches (0.45mm) in available lengths of 190cm and 300cm. The devices have a solid core with a hydrophilic coated coil-type distal end. The entire coil is radiopaque to facilitate selection of the blood vessel and confirmation of the position of the guide wire's distal end by fluoroscopy. The core shaft surface is coated with Polytetrafluoroethylene (PTFE). About 2 cm of the distal end can be shaped. A detachable extension wire (hereafter "extension wire") is available to connect with the proximal end of the guide wire with a length of less than 300 cm. A Torque device is included in the same package.
The provided document is a 510(k) summary for the VASSALLO® GT 018 G12 and G30 Peripheral Guide Wires. This type of submission is for medical devices that claim substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials. Therefore, the information typically requested regarding AI/human reader studies, ground truth establishment for training, and detailed performance metrics as one might find for an AI-powered diagnostic device, is largely not applicable here.
The document describes non-clinical testing performed to demonstrate that the device meets acceptance criteria and performs similarly to its predicate device.
Here's the information extracted from the document:
1. A table of acceptance criteria and the reported device performance
The document states that the VASSALLO GT 018 G12 and G30 met all acceptance criteria for the non-clinical tests performed. Specific acceptance values are not provided in this summary, but the types of tests and the conclusion that the device met the criteria are listed.
Test Performed | Acceptance Criteria (Not explicitly stated in values in the document, but overall met) | Reported Device Performance |
---|---|---|
Dimensional Verification | (Implicitly, within specified manufacturing tolerances) | Met all acceptance criteria |
Visual Inspection | (Implicitly, free from defects) | Met all acceptance criteria |
Tensile Strength / Tip Pull | (Implicitly, adequate strength) | Met all acceptance criteria |
Torque Strength | (Implicitly, adequate strength to transmit torque) | Met all acceptance criteria |
Torqueability | (Implicitly, ability to be torqued and steered) | Met all acceptance criteria |
Coating Adhesion/Integrity | (Implicitly, sufficient adhesion and integrity) | Met all acceptance criteria |
Catheter Compatibility / Lubricity | (Implicitly, compatible with catheters and adequate lubricity) | Met all acceptance criteria |
Corrosion Resistance | (Implicitly, resistant to corrosion) | Met all acceptance criteria |
Kink Resistance | (Implicitly, resistant to kinking) | Met all acceptance criteria |
Tip Flexibility | (Implicitly, appropriate flexibility) | Met all acceptance criteria |
Radiopacity | (Implicitly, visible under fluoroscopy) | Met all acceptance criteria |
2. Sample size used for the test set and the data provenance
The document does not specify sample sizes for each non-clinical test. This is typically found in the detailed test reports, not the summary. The tests are non-clinical laboratory tests, meaning they were performed in a controlled lab environment and are not based on human or animal data provenance (e.g., country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the tests are non-clinical, mechanical, and material property evaluations. There is no "ground truth" to be established by human experts in the context of diagnostic interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as there is no diagnostic image or data interpretation by experts that would require an adjudication method. The tests are objective measurements of physical properties.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. This device is a physical medical device (peripheral guide wire), not an AI diagnostic or assistance tool. Therefore, MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable. This device is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
This information is not applicable for the non-clinical tests described. The "ground truth" for these tests would be the established engineering specifications, material properties, and performance benchmarks that the device is designed to meet.
8. The sample size for the training set
This information is not applicable. There is no "training set" in the context of this physical medical device.
9. How the ground truth for the training set was established
This information is not applicable.
Ask a specific question about this device
Page 1 of 1