(47 days)
This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
The VASSALLO® GT 018 G12 and G30 Peripheral Guide Wires in this submission are a steerable guide wire with a maximum diameter of 0.018 inches (0.45mm) in available lengths of 190cm and 300cm. The devices have a solid core with a hydrophilic coated coil-type distal end. The entire coil is radiopaque to facilitate selection of the blood vessel and confirmation of the position of the guide wire's distal end by fluoroscopy. The core shaft surface is coated with Polytetrafluoroethylene (PTFE). About 2 cm of the distal end can be shaped. A detachable extension wire (hereafter "extension wire") is available to connect with the proximal end of the guide wire with a length of less than 300 cm. A Torque device is included in the same package.
The provided document is a 510(k) summary for the VASSALLO® GT 018 G12 and G30 Peripheral Guide Wires. This type of submission is for medical devices that claim substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive clinical trials. Therefore, the information typically requested regarding AI/human reader studies, ground truth establishment for training, and detailed performance metrics as one might find for an AI-powered diagnostic device, is largely not applicable here.
The document describes non-clinical testing performed to demonstrate that the device meets acceptance criteria and performs similarly to its predicate device.
Here's the information extracted from the document:
1. A table of acceptance criteria and the reported device performance
The document states that the VASSALLO GT 018 G12 and G30 met all acceptance criteria for the non-clinical tests performed. Specific acceptance values are not provided in this summary, but the types of tests and the conclusion that the device met the criteria are listed.
| Test Performed | Acceptance Criteria (Not explicitly stated in values in the document, but overall met) | Reported Device Performance |
|---|---|---|
| Dimensional Verification | (Implicitly, within specified manufacturing tolerances) | Met all acceptance criteria |
| Visual Inspection | (Implicitly, free from defects) | Met all acceptance criteria |
| Tensile Strength / Tip Pull | (Implicitly, adequate strength) | Met all acceptance criteria |
| Torque Strength | (Implicitly, adequate strength to transmit torque) | Met all acceptance criteria |
| Torqueability | (Implicitly, ability to be torqued and steered) | Met all acceptance criteria |
| Coating Adhesion/Integrity | (Implicitly, sufficient adhesion and integrity) | Met all acceptance criteria |
| Catheter Compatibility / Lubricity | (Implicitly, compatible with catheters and adequate lubricity) | Met all acceptance criteria |
| Corrosion Resistance | (Implicitly, resistant to corrosion) | Met all acceptance criteria |
| Kink Resistance | (Implicitly, resistant to kinking) | Met all acceptance criteria |
| Tip Flexibility | (Implicitly, appropriate flexibility) | Met all acceptance criteria |
| Radiopacity | (Implicitly, visible under fluoroscopy) | Met all acceptance criteria |
2. Sample size used for the test set and the data provenance
The document does not specify sample sizes for each non-clinical test. This is typically found in the detailed test reports, not the summary. The tests are non-clinical laboratory tests, meaning they were performed in a controlled lab environment and are not based on human or animal data provenance (e.g., country of origin, retrospective/prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable as the tests are non-clinical, mechanical, and material property evaluations. There is no "ground truth" to be established by human experts in the context of diagnostic interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable as there is no diagnostic image or data interpretation by experts that would require an adjudication method. The tests are objective measurements of physical properties.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. This device is a physical medical device (peripheral guide wire), not an AI diagnostic or assistance tool. Therefore, MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable. This device is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
This information is not applicable for the non-clinical tests described. The "ground truth" for these tests would be the established engineering specifications, material properties, and performance benchmarks that the device is designed to meet.
8. The sample size for the training set
This information is not applicable. There is no "training set" in the context of this physical medical device.
9. How the ground truth for the training set was established
This information is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2023
Filmec Co. Ltd. % Meagan Fagan Principal Consultant CardioMed Device Conultants, LLC 1783 Forest Drive Suite 254 Annapolis. Maryland 21401
Re: K232578
Trade/Device Name: Vassallo GT 018 G12; Vassallo GT 018 G30 Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: August 24, 2023 Received: August 25, 2023
Dear Meagan Fagan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely, Lydia S. Glaw - J Digitally signed by Lydia S.
Glaw -S Glaw -S Glaw -S Date: 2023.10.11 15:05:50 S -04'00 Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K232578
Device Name
VASSALLO GT 018 G12 and VASSALLO GT 018 G30
Indications for Use (Describe)
This product is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, fo:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510(k) Summary
(as required by 21 CFR 807.92)
Image /page/4/Picture/3 description: The image shows the logo for Filmec. The logo consists of a red abstract shape on the left, resembling a stylized letter 'A' or an upward-pointing arrow. To the right of the shape is the word "Filmec" in black, with a red square above the 'i'. The logo is simple and modern, with a clear contrast between the red and black elements.
PERIPHERAL GUIDE WI VASSALLO® GT 018 G12 and G30
510(k) K232578
| Date Prepared: | August 24, 2023 |
|---|---|
| Applicant: | FILMECC CO., LTD.1703 Wakita-cho, Moriyama-kuNagoya-shi, Aichi 463-0024JapanTEL: +81-52-768-1212, FAX: +81-52-768-1222 |
| Contact: | Noriaki TsuzukiRegulatory Affairs / Vigilance SectionFILMECC CO., LTD.1703 Wakita-cho, Moriyama-kuNagoya-shi, Aichi 463-0024JapanTEL: +81-52-768-1212, FAX: +81-52-768-1222e-mail: noriaki.tsuzuki@filmecc.com |
| Trade Name: | VASSALLO® GT |
| Device Classification: | Class II per 21 CFR §870.1330 |
| Classification Name: | Catheter, Guide, Wire |
| Product Code: | DQX - Catheter Guide Wire |
| Predicate Devices: | Filmecc VASSALLO GT (K203533) |
| Reference Devices | Filmecc VASSALLO GT 018 Floppy (K213949) |
INTENDED USE/INDICATIONS FOR USE:
VASSALLO® GT
These products are intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.
DEVICE DESCRIPTION:
The VASSALLO® GT 018 G12 and G30 Peripheral Guide Wires in this submission are a steerable guide wire with a maximum diameter of 0.018 inches (0.45mm) in available lengths of 190cm and 300cm.
FILMECC CO., LTD.
{5}------------------------------------------------
The devices have a solid core with a hvdrophilic coated coil-type distal end. The entire coil is radiopaque to facilitate selection of the blood vessel and confirmation of the position of the guide wire's distal end by fluoroscopy.
The core shaft surface is coated with Polytetrafluoroethylene (PTFE). About 2 cm of the distal end can be shaped. A detachable extension wire (hereafter "extension wire") is available to connect with the proximal end of the quide wire with a length of less than 300 cm. A Torque device is included in the same package.
COMPARISON WITH PREDICATE DEVICE:
Comparisons of the VASSALLO® GT 018 G12 and G30 and predicate device show that the technological characteristics of the subject device such as the components, design, materials, sterilization method, shelf life and operating principle are similar, if not identical, to the currently marketed predicate device, The intended use of the subject devices and its predicate are the same.
| Name ofDevices | VASSALLO® GT 018G12 and G30 | VASSALLO GTFloppy and G40 | VASSALLO GT 018Floppy |
|---|---|---|---|
| Subject | Predicate | Reference | |
| 510(k) | TBD | K203533 | K213949 |
| Intended Useand Indications | This product is intended to facilitate the placement and exchange of diagnostic andtherapeutic devices during intravascular procedures. This device is intended for peripheralvascular use only. | ||
| Nominal OD | 0.45mm(0.018in) | 0.36mm(0.014in) | 0.45mm(0.018in) |
| Overall Length | 190, 300cm | ||
| Outer Coil | Platinum-Nickel | Platinum-Nickel or Platinum-Nickel/Stainless Steel | Platinum-Nickel/StainlessSteel |
| Tapered CoreWire | Stainless Steel | ||
| Tip Shape | Straight (shapeable) | ||
| Coating | Hydrophilic, Hydrophobic | ||
| Sterilization | Ethylene Oxide, SAL 10-6 |
NON-CLINICAL TESTING/PERFORMANCE DATA:
Non-clinical laboratory testing was performed on the VASSALLO® GT 018 G12 and G30 to determine substantial equivalence to the predicate device. The following testing/assessments were performed:
- Dimensional Verification ●
- Visual Inspection ●
- Tensile Strength / Tip Pull ●
- Torque Strength
- Torqueability .
- Coating Adhesion/Integrity ●
- Catheter Compatibility / ● Lubricity
- Corrosion Resistance ●
- Kink Resistance
- Tip Flexibility ●
- Radiopacity ●
FILMECC CO., LTD.
{6}------------------------------------------------
The in vitro bench tests demonstrated that the VASSALLO GT 018 G12 and G30 met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate devices.
BIOCOMPATIBILITY:
Testing was performed to assess biocompatibility of the predicate VASSALLO GT. The following tests were performed:
- Cytotoxicity ●
- Sensitization ●
- Intracutaneous Irritation ●
- Acute Systemic Toxicity
- Material Mediated . Pyrogenicity
- Hemolvsis
- Partial Thromboplastin Time
- In Vivo Thromboresistance ●
- Complement Activation (SC5b-9)
The results from this testing are being leveraged for the subject device and support the biocompatibility of the VASSALLO GT 018 G12 and G30.
STERILIZATION:
The minor dimensional differences of the subject device do not represent a greater sterilization challenge than the predicate VASSALLO GT quide wires. As such, no additional sterilization was required, and the subject device was adopted into the product family for which a sterilization cycle was validated in accordance with AAMI TIR28.
CONCLUSION:
The VASSALLO GT 018 G12 and G30 have the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate device. Performance data demonstrates that the device functions as intended. The conclusions drawn from the nonclinical tests demonstrate that the subject VASSALLO GT 018 is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Therefore, the VASSALLO GT 018 G12 and VASSALLO GT 018 G30 are substantially equivalent to the predicate devices.
§ 870.1330 Catheter guide wire.
(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.