Search Results
Found 2 results
510(k) Data Aggregation
(8 days)
The Total Bilirubin assay is used for the quantitation of total bilirubin in human serum or plasma. Measurement of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Total Bilirubin is an in vitro diagnostic assay for the quantitative determination of total bilirubin in human serum or plasma. The Total Bilirubin assay is a clinical chemistry assay in which the conjugated bilirubin is oxidized to biliverdin. The resulting decrease in absorbance at 444 nm is directly proportional to the concentration of total bilirubin.
The provided 510(k) summary describes the Abbott Laboratories TBil (Total Bilirubin) assay, an in vitro diagnostic device for the quantitative determination of total bilirubin in human serum or plasma.
This submission is a traditional 510(k) for a new device, not an AI/ML powered device. The document explicitly states that the device is an in vitro diagnostic assay and describes its chemical principle (oxidation of conjugated bilirubin to biliverdin). Therefore, many of the questions regarding AI/ML device validation (e.g., sample sizes for test/training sets, number of experts, adjudication methods) are not applicable.
Here's an analysis of the provided information, focusing on the relevant acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria/Metric | Reported Performance of TBil Assay | Predicate Device (Johnson & Johnson Vitros™ Total Bilirubin) Performance |
---|---|---|---|
Method Comparison | Correlation Coefficient | 0.994 | N/A (implied benchmark for correlation) |
Slope | 1.055 | N/A (implied benchmark for slope = 1) | |
Y-intercept | 0.009 mg/dL | N/A (implied benchmark for Y-intercept = 0) | |
Precision | Total %CV (Level 1/Panel 101) | 5.6% | N/A (stated as "similar Performance Characteristics" for both) |
Total %CV (Level 2/Panel 102) | 4.3% | N/A | |
Linearity | Upper Limit | Up to 23.38 mg/dL | N/A (difference between assay ranges noted, but specific predicate range not given) |
Sensitivity | Limit of Quantitation | 0.13 mg/dL | N/A |
Study Proving Device Meets Acceptance Criteria:
A "comparative performance study" was conducted using the Abbott AEROSET™ System. This study compared the performance of the Abbott TBil assay with the predicate device, the Total Bilirubin on the Johnson & Johnson Vitros™ System. The results, as presented in the table above, demonstrated "acceptable correlation" and other performance characteristics, leading to the conclusion of "substantial equivalence."
2. Sample Size for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not explicitly state the sample size used for the comparative performance studies (method comparison or precision studies).
- Data Provenance: The document does not specify the country of origin of the data. It implies the studies were conducted by Abbott Laboratories. The studies appear to be prospective in nature, as they involve actively "conducting" comparative performance studies and precision studies for the new device.
3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts
This is an in vitro diagnostic assay
based on a chemical reaction, not an AI/ML device relying on expert image or data interpretation. Therefore, the concept of "experts establishing ground truth" in the context of clinical interpretation is not applicable. The "ground truth" for method comparison would be the results obtained from the predicate device (Johnson & Johnson Vitros™ System), which is also a quantitative measurement system.
4. Adjudication Method for the Test Set
As this is a quantitative chemical assay and not an AI/ML device with subjective interpretations, adjudication methods are not applicable. The comparison is based on numerical agreement between the new device and the predicate device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices that assist human readers in interpreting complex data (e.g., medical images). The Abbott TBil assay is an automated in vitro diagnostic assay that provides a quantitative measurement.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
The device is an automated in vitro diagnostic assay. Its performance, as described by linearity, sensitivity, and precision, represents its standalone performance without human interpretation in the loop beyond operating the instrument and collecting the sample. The "method comparison" also evaluates its standalone performance against another standalone device.
7. The Type of Ground Truth Used
The "ground truth" for the method comparison was the quantitative measurements obtained from the legally marketed predicate device (Johnson & Johnson Vitros™ System for Total Bilirubin). For precision, the ground truth refers to the intrinsic variability of the assay itself, demonstrating reproducibility around a given (control) concentration.
8. The Sample Size for the Training Set
This is an in vitro diagnostic assay that performs a chemical reaction, not an AI/ML device that requires a training set. Therefore, the concept of a "training set" in the context of machine learning is not applicable.
9. How the Ground Truth for the Training Set was Established
As there is no training set for an AI/ML algorithm, this question is not applicable.
Ask a specific question about this device
(18 days)
The Total Bilirubin assay is used for the quantitation of total bilirubin in human serum or plasma. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.
Total Bilirubin is an in vitro diagnostic assay for the quantitative determination of Total Bilirubin in human serum or plasma. The Total Bilirubin assay is a clinical chemistry assay in which the analyte in the sample reacts with diazotised sulphanilic acid to produce an acid azobilirubin, the absorbance of which is proportional to the concentration of bilirubin in the sample.
The information provided describes the performance of the Abbott Laboratories TBil (Total Bilirubin) assay. Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state "acceptance criteria" with predefined thresholds for each performance characteristic. Instead, it demonstrates substantial equivalence to a predicate device (Roche® Cobas Mira® Plus Automated Chemistry System Total Bilirubin assay) by showing comparable performance. The reported performance metrics are presented below:
Performance Characteristic | Reported Device Performance (TBil Assay) | Equivalence Justification/Predicate Performance |
---|---|---|
Correlation with Predicate | Correlation coefficient = 0.9925 | Substantially equivalent to Roche Cobas Mira Plus Automated Chemistry System Total Bilirubin assay, indicating similar clinical results. |
Method Comparison (Slope) | Slope = 0.873 | Part of the method comparison with the predicate device, contributing to the demonstration of substantial equivalence. |
Method Comparison (Y-intercept) | Y-intercept = - 0.072 mg/dL | Part of the method comparison with the predicate device, contributing to the demonstration of substantial equivalence. |
Precision (Within-run, %CV) | Not explicitly stated separately, but included in total %CV. | Studies were conducted using two levels of control material. |
Precision (Between-run, %CV) | Not explicitly stated separately, but included in total %CV. | Studies were conducted using two levels of control material. |
Precision (Between-day, %CV) | Not explicitly stated separately, but included in total %CV. | Studies were conducted using two levels of control material. |
Precision (Total %CV, Level 1/Panel 111) | 4.8% | Considered acceptable. |
Precision (Total %CV, Level 2/Panel 112) | 5.1% | Considered acceptable. |
Linearity (Upper Limit) | 20.0 mg/dL | Considered acceptable for the intended use. |
Limit of Quantitation (Sensitivity) | 0.264 mg/dL | Considered acceptable for the intended use. |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample size used for the comparative performance studies (method comparison and precision studies). It mentions that "comparative performance studies were conducted using the ALCYON™ Analyzer" and "precision studies were conducted using the Total Bilirubin assay... using two levels of control material."
- Sample Size: Not explicitly stated for specific studies.
- Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). However, the study focuses on the analytical performance of the device in a laboratory setting.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This type of information is not relevant or provided for this device. The device is an in vitro diagnostic assay for quantitative determination of a chemical analyte (Total Bilirubin). The ground truth for such devices is established by comparing the device's measurements to a reference method or validated predicate device, not by expert interpretation of images or clinical cases.
4. Adjudication Method for the Test Set
Not applicable. As mentioned above, this is an in vitro diagnostic assay, not a subjective diagnostic interpretation. The comparison is objective, based on measured values and statistical correlation.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This device is an automated in vitro diagnostic assay for chemical analysis, not an AI-assisted diagnostic tool that would involve human readers interpreting cases.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, in essence. The studies described are of the device's standalone analytical performance. The "comparative performance studies" and "precision studies" evaluate the assay itself, operating as an algorithm within the ALCYON™ Analyzer, comparing its output to a predicate device and assessing its inherent variability and limits. There is no human-in-the-loop component in the direct measurement and quantification of bilirubin by the device.
7. The Type of Ground Truth Used
The "ground truth" for this in vitro diagnostic assay is implicitly the measurements obtained from the Roche Cobas Mira Plus Automated Chemistry System Total Bilirubin assay (K910591), which served as the predicate device. The performance of the new device (TBil) was compared against this established method.
8. The Sample Size for the Training Set
Not applicable/provided. This device is a traditional chemical assay, not a machine learning or AI algorithm that requires a "training set" in the computational learning sense. The assay's parameters would have been developed and optimized based on chemical principles and validation studies, not through a data-driven training process as typically understood for AI.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As explained in point 8, there is no "training set" in the context of this traditional in vitro diagnostic assay.
Ask a specific question about this device
Page 1 of 1