Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    ATTUNE Total Knee System, ATTUNE Cementless Knee System, LPS Limb Preservation System, Sigma High Performance
    (HP) Partial Knee System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ATTUNE™ Total Knee System Indications for Use
    Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.

    ATTUNE™ Cementless Knee System Indications for Use
    The ATTUNE™ Cementless CR and PS Femoral Components are intended for cementless use within the ATTUNE™ Total Knee Replacement System.
    Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present).

    LPS TM Limb Preservation System Indications for Use
    The DePuy LPS System is intended for use in replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include:

    • · malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) requiring extensive resection and replacement;
    • · patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis, requiring extensive resection and replacement;
    • · revision cases for a failed previous prosthesis requiring extensive resection and replacement;
    • · severe trauma requiring extensive resection and replacement.
      The LPS System is also intended for use in bone loss post-infection, where the surgeon has elected to excise the bone and replacement is required.
      The S-ROM tibial tray and the non-porous coated straight and bowed stems are intended for cemented use only.
      The porous-coated metaphyseal sleeves are intended for either cemented or cementless applications.

    Sigma™ High Performance (HP) Partial Knee System Indications for Use
    The SIGMA™ High Performance Partial Knee System is indicated for single compartmental knee replacement in skeletally mature individuals with osteoarthritis, post-traumatic arthritis of the tibiofemoral surfaces or a history of gout or pseudogout. All components are intended for CEMENTED USE ONLY.

    Device Description

    ATTUNE™ Total Knee System: A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.

    ATTUNE™ Cementless Knee System: A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant with or without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.

    LPS - Limb Preservation System: The DePuy LPSTM Limb Preservation System is designed for the replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia. The DePuy LPS system offers a variety of component options (including, but not limited to, proximal femoral bodies, segmental components, distal femoral components, femoral stems, tibial stems, proximal tibial components, hinged tibial insert bearings, metaphyseal sleeves, and adapters). The components, which can be used in conjunction with certain components from other systems, are for treatment of patients presenting bone loss and deformity associated with bone tumors resection, trauma, infection, and difficult revision arthroplasty. A total femoral replacement is possible in those cases where no part of the femur can be salvaged.

    Sigma High Performance (HP) Partial Knee System: The DePuy SIGMATM High Performance Partial Knee System is a single compartmental knee prosthesis, composed of individually packaged femoral, and tibial components designed to be used in various combinations to replace the natural articular surfaces of the knee joint. The unicompartmental femoral components are Co-Cr-Mo metal implants. The metal backed tibial components are Co-Cr-Mo with polyethylene inserts. The all-polyethylene unicompartmental tibial component are manufactured from polyethylene. The unicompartmental femoral components are designed for individuals who require a higher than normal degree of flexion (up to 155°).

    AI/ML Overview

    This document (K233980) is a 510(k) premarket notification for several DePuy knee systems. It primarily focuses on adding updated MRI compatibility information and standardizing language in the Instructions for Use (IFU) and labels. The core claim for substantial equivalence relies on non-clinical performance data related to MRI safety and bacterial endotoxin testing, rather than a clinical study of device performance in a human-in-the-loop or standalone AI context.

    Therefore, the requested information regarding acceptance criteria, study design for device performance, sample sizes for test and training sets, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment cannot be fully provided from this document as it pertains to a different type of device (knee implants) and regulatory submission (510(k) for labeling changes related to MRI safety, not an AI/software as a medical device performance study).

    However, I can extract the acceptance criteria for the non-clinical performance tests that were conducted, and the reported performance as implied by the conclusion of substantial equivalence.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    For MRI safety, the acceptance criteria are implicit in meeting the standards listed. The reported "performance" is that the devices meet these standards, thus proving MRI safety.

    Acceptance Criteria (Implicit)Reported Device Performance (Implied)
    Conformance to ASTM F2503-23 (Standard practice for marking medical devices for safety in MR environment)Met
    Conformance to ASTM F2182-19E2 (Measurement of Radio Frequency Induced Heating On or Near Passive Implants during MR)Met
    Conformance to ASTM F2052-21 (Measurement of Magnetically Induced Displacement Force on Medical Devices in the MR Environment)Met
    Conformance to ASTM F2213-17 (Measurement of Magnetically Induced Torque on Medical Devices in the MR Environment)Met
    Conformance to ASTM F2119-07 (Evaluation of MR Image Artifacts from Passive Implants)Met
    Conformance to ANSI/AAMI ST 72:2019 (Bacterial Endotoxin Testing)Met

    2. Sample sizes used for the test set and the data provenance

    • Sample Size: Not applicable in the context of clinical or AI performance. The "test set" here would refer to the physical devices tested for MRI compatibility and bacterial endotoxin. The document does not specify the number of individual devices subjected to these non-clinical tests.
    • Data Provenance: Not applicable for a typical AI/software study. The "data" here are the results from physical and chemical testing of the devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This is not applicable as there is no "ground truth" established by experts in the context of an AI or diagnostic device study. The ground truth for these tests is defined by the physical properties measured according to established ASTM and ANSI standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This is not applicable as this is not an image-based or diagnostic AI study requiring human adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical tests were conducted to demonstrate substantial equivalence." (pages 6, 9, 12, 15)

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable as the device is a knee implant, not an algorithm or software. The "performance" assessment is of the physical and material properties of the implant itself in relation to MRI safety.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the non-clinical tests is established by the physical and chemical properties measured according to the specified ASTM and ANSI international standards for MRI compatibility and bacterial endotoxin levels.

    8. The sample size for the training set

    • Not applicable. There is no AI model or training set involved in this regulatory submission.

    9. How the ground truth for the training set was established

    • Not applicable. There is no AI model or training set involved in this regulatory submission.

    In summary: This FDA 510(k) submission for DePuy knee systems is related to changes in labeling for MRI compatibility, not an AI/software/diagnostic device. The "study" described focuses on non-clinical bench testing to demonstrate MRI safety and bacterial endotoxin compliance, adhering to recognized industry standards, rather than clinical performance or AI algorithm validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193549
    Manufacturer
    Date Cleared
    2020-04-16

    (118 days)

    Product Code
    Regulation Number
    888.3530
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    SIGMA High Performance (HP) Partial Knee System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INDICATIONS
    The SIGMA High Performance Partial Knee System is indicated for single compartmental knee replacement in skeletally mature individuals with osteoarthritis, post-traumatic arthritis of the tibiofemoral surfaces or a history of gout or pseudogout. All components are intended for CEMENTED USE ONLY.

    Device Description

    The DePuy SIGMA® High Performance Partial Knee System is a single compartmental knee prosthesis, composed of individually packaged femoral and tibial components designed to be used in various combinations to replace the natural articular surfaces of the knee joint.

    The unicompartmental femoral components are Co-Cr-Mo metal implants, available with or without a porous coating. The metal backed tibial components are Co-Cr-Mo and polyethylene and are available without a porous coating. The all-polyethylene unicompartmental tibial component manufactured from polyethylene.

    The unicompartmental femoral components are designed for individuals who require a higher than normal degree of flexion (up to 155°).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the SIGMA High Performance (HP) Partial Knee System. While the document outlines information about the device itself and its equivalence to predicate devices, it does not contain the acceptance criteria or a study that specifically proves the device meets those criteria in the context of an AI/ML-driven device's performance.

    The document states:

    • "SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE": This lists a number of engineering and material tests (Biocompatibility, UHMWPE Material Property Characterization, Range of Motion/Constraint, Contact Area/Stress, Wear Testing, Pull-Off, Bacterial Endotoxin testing).
    • "SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION - Clinical testing was not necessary to demonstrate substantial equivalence."

    This indicates that the submission is focused on demonstrating substantial equivalence based on material properties, design, and manufacturing processes, rather than the performance of an AI/ML algorithm. Therefore, the requested information regarding acceptance criteria, device performance, sample sizes for test/training sets, ground truth establishment, expert adjudication, or MRMC studies is not present in the provided text, as these are typically associated with the evaluation of AI/ML software performance.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1