Search Results
Found 86 results
510(k) Data Aggregation
(60 days)
SIGMA Spectrum Infusion Pump with Master Drug Library
The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, or epidural. The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.
The SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to, hospitals and outpatient care areas.
The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
SIGMA Spectrum is a large volume pump within the SIGMA Spectrum infusion system used by clinicians at the patient bedside to control the delivery of medications from a bag. The pump moves fluid from the bag to the patient via specified administration sets using a peristaltic pumping action. The pump offers various programmable delivery modes to address specific patient care needs. The delivery modes available to support the patient are determined by how the pump and its associated drug library are configured.
The pump provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725. The system includes a software controlled, electromechanical pump used for the controlled administration of fluids including pharmaceutical drugs, blood, blood products, and mixtures of required patient therapy through administration sets at clinician's selectable rates and volumes.
The pump is intended for the controlled administration of fluids through the following clinically accepted routes of administration: intravenous, arterial, subcutaneous, and epidural. The pump is intended to be used in conjunction with legally marketed and compatible administration sets, as indicated in the device labeling, and medications provided by the user. The subject device is suitable for patient care in hospitals and outpatient health care facilities.
The Master Drug Library (MDL) is a stand-alone (not embedded in the pump) software application installed on a hospital-provided computing platform and used to create a drug library file. MDL facilitates the generation, configuration, and management of a facility-specific drug library file for dedicated infusion pumps. The drug library file is intended to be distributed to all compatible infusion pumps in the hospital.
This submission includes software design and labeling changes to address the issues leading to recalls Z-0530-2022 and Z-2103-2023.
This FDA 510(k) clearance letter pertains to an infusion pump, not an AI/ML powered medical device. Therefore, many of the requested categories in your prompt (such as "Number of experts used to establish the ground truth," "Adjudication method," "MRMC study," "Standalone performance," "Type of ground truth," and "Training set sample size/ground truth establishment") are not applicable to this type of medical device submission.
The document primarily focuses on demonstrating substantial equivalence to a predicate device through a comparison of technical characteristics and verification of performance against established requirements.
Here's an analysis based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" for the overall device in a quantifiable manner that would typically be found in an AI/ML context (e.g., specific sensitivity, specificity, or AUC targets). Instead, it demonstrates compliance with a range of technical specifications, which inherently act as acceptance criteria for the design and performance. The "Reported Device Performance" is implied by the statement that "Non-clinical testing met all acceptance criteria."
Below is a table summarizing key technical characteristics that function as performance criteria for the infusion pump. Since the subject device is deemed "substantially equivalent" to the predicate, and no new performance claims are made that deviate from the predicate, their performance characteristics are identical as presented.
Characteristic | Acceptance Criteria (Subject Device & Predicate K230022) | Reported Device Performance (Subject Device) |
---|---|---|
Pumping Mechanism | Linear peristaltic design | Linear peristaltic design |
Modes of Delivery | Continuous, Intermittent | Continuous, Intermittent |
Routes of Admin. | Intravenous, Arterial, Subcutaneous, Epidural | Intravenous, Arterial, Subcutaneous, Epidural |
User Interface Display | Color LCD | Color LCD |
AC Power Input | 115 VAC ±15%, 50 - 60 Hz / 300 mA Max | 115 VAC ±15%, 50 - 60 Hz / 300 mA Max |
AC Power Output | 9 VDC/1200 mA, short circuit protected | 9 VDC/1200 mA, short circuit protected |
Operating Temp (Std/WBM) | 15.6 to 32.2°C (60 to 90°F), 20-90% RH non-condensing | 15.6 to 32.2°C (60 to 90°F), 20-90% RH non-condensing |
Operating Temp (802.11b/g) | 15.6 to 26.7°C (60 to 80°F), 20-90% RH non-condensing | 15.6 to 26.7°C (60 to 80°F), 20-90% RH non-condensing |
Storage Temp. | -10 to +49°C (14 to 120°F), 10-90% RH non-condensing | -10 to +49°C (14 to 120°F), 10-90% RH non-condensing |
Single Fault Bolus | Max 0.56 mL | Max 0.56 mL |
Anti-Free Flow System | Set-based, utilizing IV set slide clamp | Set-based, utilizing IV set slide clamp |
Low Battery Alarm | ≤15 minutes of battery power remaining | ≤15 minutes of battery power remaining |
Air-In-Line Detection | >2.5 cm air bubbles (140 μL in Baxter sets); >1 mL accumulated air over 15 min (room temp); >1.5 mL accumulated air over 15 min (15.5°C) | Meets criteria |
Downstream Occlusion Alarms | User adjustable Low (41 kPa ±27 kPa), Medium (89 kPa ±41 kPa), High (131 kPa ±62 kPa) | User adjustable, meets specified values |
Max Downstream Occlusion Press. | 207 kPa (30 psi) | 207 kPa (30 psi) |
Flow Rate Range | 0.5 to 999 mL/hr | 0.5 to 999 mL/hr |
Low-Flow Continuity | Max period of no-flow is 90 seconds at 0.5 mL/hr | Max period of no-flow is 90 seconds at 0.5 mL/hr |
Volumetric Accuracy (DEHP sets) - 0.5-1.9 mL/hr | ±0.1 mL/hr (over 1 hr, up to 96 hrs) | ±0.1 mL/hr (over 1 hr, up to 96 hrs) |
Volumetric Accuracy (DEHP sets) - 2.0-999 mL/hr | ±5% (over 1 hr, up to 96 hrs) | ±5% (over 1 hr, up to 96 hrs) |
Volumetric Accuracy (Non-DEHP sets) - 10-125 mL/hr | ±10% ( |
Ask a specific question about this device
(90 days)
DePuy Knee Prosthesis System Universal Stem Extensions and Universal Femoral Metaphyseal Sleeves; DePuy Sigma
PS Femoral Components; DePuy Sigma Cruciate Retaining (C/R) Porocoat Femoral Components; S-ROM NOILES
Rotating Hinge Knee System; DePuy P.F.C. SIGMA Total Knee System; DePuy SIGMA Total Knee System
ATTUNE™ Revision Knee System: Candidates for total knee replacement include patients with A severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or rheumatod arthritis Moderate valgus, varus, or flexion deformities Avascular necrosis of the femoral condyle A previous unsuccessful knee replacement, osteotomy, or other knee procedure ATTUNE Revision Knee System implants are designed for use in total knee arthroplasty for patients with: Absence or loss of both cruciate ligaments Moderate varus-valgus or flexion instability that requires a bearing surface with increased constraint in the clinical judgment of the surgeon Bone loss that requires supplemental fixation in the clinical judgment of the surgeon The porous-coated metaphyseal sleeves are intended for either cementless applications. ANY NON POROUS-COATED COMPONENTS ARE INTENDED FOR CEMENTED USE ONLY.
DePuy Knee Prosthesis System Universal Stem Extensions and Universal Femoral Metaphyseal Sleeves: The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are intended for use with the PFC, PFC Sigma, Sigma TC3 Revision Knee, or S-ROM knee prosthesis in total knee replacement surgery for patients suffering from severe pain and disability due to permanent structural damage resulting from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, pseudogout, trauma or failed prior surgical intervention. These devices are for cemented use only. The DePuy Universal Femoral Metaphyseal Sleeve and Universal Stem components are also intended for use with the DePuy LPS prosthesis for replacement of the mid-shaft portion of the femur, proximal, distal femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include: Malignant tumors (e.g., osteosarcomas, gian cell tumors, bone tumors) requiring extensive resection and replacement; patient conditions of noninflammatory degenerative join disease (NIDD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g. rheumatoid arthritis, requiring extensive resection and replacement; revision for failed previous prosthesis cases requiring extensive resection and replacement. The LPS prosthesis is also intended for use in bone loss post-infection, where the surgeon has elected to excise the bone and replacement is required. The Universal Stem and the Universal Metaphyseal Sleeve components are intended for cemented use only.
DePuy Sigma PS Femoral Components and DePuy Sigma Cruciate Retaining (C/R) Porocoat Femoral Components: Candidates for total knee replacement include patients with a severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant. Total knee replacement may be considered for younger patients if, in the surgeon, an unequivocal indication for total knee replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and knee joint loading can be assured. This includes severely crippled patients with multiple joint involvement for whom a gain in knee mobility may lead to significant improvement in the quality of their lives. THE SIGMA C/R POROCOAT® FEMORAL COMPONENTS ARE INTENDED FOR CEMENTED OR CEMENTLESS USE AS THE FEMORAL COMPONENT OF A TOTAL KNEE REPLACEMENT SYSTEM. THE SIGMA PS FEMORAL COMPONENTS ARE INTENDED FOR CEMENTED USE AS THE FEMORAL COMPONENTS OF A TOTAL KNEE REPLACEMENT SYSTEM.
S-ROM™ NOILESTM Rotating Hinge Knee: The S-ROM NOILES Rotating Hinge Knee is indicated in cases for cement use in patients who have reached skeletal maturity and for whom the surgeon has decided to resect both cruciate ligaments due to the following conditions: 1. Severe instability, gross deformity and/or bone loss. 2. Failure of a previous knee reconstruction procedure. 3. Trauma or tumor resection. 4. Absent or markedly insufficient collateral ligaments.
DePuy P.F.C. TM SIGMA™ Total Knee Prosthesis and DePuy SIGMA™ Total Knee Prosthesis: The DePuy SIGMA™ and P.F.C. ™SIGMA™ Total Knee Prosthesis are intended for use in total knee replacement surgery for patients suffering from severe pain and disability due to permanent structural damage resulting from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, pseudogout, trauma or failed prior surgical intervention. The DePuy SIGMA™ and P.F.C. ™SIGMA™ Total Knee Prosthesis are intended for cement use only.
ATTUNE™ Revision Knee System: A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component consists of a metal tibial base without porous coating, and a locking polyethylene insert. Some metal components have modular stems, porous and non porous-coated sleeves and/or modular augments. The patella component is an all polyethylene design Total knee arthroplasty may include supplemental fixation through stems, sleeves, and/or modular augments where bone loss requires said fixation in the opinion of the surgeon. Total knee arthroplasty may also include more constrained bearing surfaces where necessary to provide stability where musculoligamentous supporting structures are insufficient.
DePuy Knee Prosthesis System Universal Stem Extensions and Universal Femoral Metaphyseal Sleeves: A Total Knee Prosthesis System is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant, with or without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial tray with or without porous coating, and a polyethylene insert and locking components. Some metal components have modular stems, sleeves and/or modular wedges. The patella component may be of an all polyethylene design or may be a metal backed polyethylene component. The wobble bit is an instrument used to aid implant assembly.
DePuy Sigma PS Femoral Components and DePuy Sigma Cruciate Retaining (C/R) Porocoat Femoral Components: A SIGMA™ Total Knee System is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant, with or without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial tray with or without porous coating, and a polyethylene insert and locking components. Some metal components have modular stems, sleeves and/or modular wedges or augments. The patella component is an all polyethylene design.
S-ROM™ NOILESTM Rotating Hinge Knee System: The S-ROM NOILES Rotating Hinge Knee System is a tricompartmental total knee replacement for both primary and revision cases. The S-ROM NOILES Rotating Hinge Knee System includes the femoral component with hinge pin, the tibial plateau assembly, and the distal femoral augmentation blocks. Replacement bumpers for the femoral component assembly and replacement hinge bearings for the tibial plateau assembly are also available.
DePuy P.F.C. TM SIGMA™ Total Knee Prosthesis and DePuy SIGMA™ Total Knee Prosthesis: The DePuy SIGMA™ and P.F.C.™ SIGMA® Total Knee Prosthesis is a total knee prosthesis composed of individually packaged femoral, tibial base, tibial insert and patellar components designed to replace the natural articular surface of the knee joint or after a failed previous implant. Some femoral and tibial components can be used with modular stems, porous and non-porous coated sleeves and/or modular augments when supplemental fixation is required in the judgement of the surgeon. The natural patella may or may not be resurfaced.
The provided text describes several knee prosthesis systems from DePuy, specifically outlining their intended use, indications for use, and a summary of non-clinical tests conducted to demonstrate substantial equivalence to previously marketed predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance:
The devices in this application (ATTUNE™ Revision Knee System, DePuy Knee Prosthesis System Universal Stem Extensions and Universal Femoral Metaphyseal Sleeves, DePuy Sigma PS Femoral Components, DePuy Sigma Cruciate Retaining (C/R) Porocoat Femoral Components, S-ROM™ NOILES™ Rotating Hinge Knee System, DePuy P.F.C.™ SIGMA™ Total Knee Prosthesis, DePuy SIGMA™ Total Knee Prosthesis) are being submitted with a modification to labeling to include updated MRI compatibility information and modernized/standardized language in the Instructions for Use (IFU) and labels. There are no changes in design, manufacturing, principle of operation, indication, or intended use of the devices themselves. Therefore, the "acceptance criteria" and "reported device performance" in this context relate to Magnetic Resonance (MR) Safety standards.
Acceptance Criteria (Standard) | Reported Device Performance (Summary from Non-Clinical Testing) |
---|---|
ASTM F2503-23: Standard practice for marking medical devices and other items for safety in the magnetic resonance environment | Testing performed to determine MR Safety, including marking devices for the MR environment. |
ASTM F2182-19E2: Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants during Magnetic Resonance | Testing performed to determine MR Safety, specifically measuring radio frequency induced heating. |
ASTM F2052-21: Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment | Testing performed to determine MR Safety, specifically measuring magnetically induced displacement force. |
ASTM F2213-17: Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment | Testing performed to determine MR Safety, specifically measuring magnetically induced torque. |
ASTM F2119-07: Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants | Testing performed to determine MR Safety, specifically evaluating MR image artifacts. |
ANSI/AAMI ST 72:2019: Bacterial endotoxin testing | The proposed devices also meet the requirement of bacterial endotoxin testing as specified in ANSI/AAMI ST 72:2019. |
2. Sample size used for the test set and the data provenance:
The document explicitly states: "No clinical tests were conducted to demonstrate substantial equivalence." The "tests" mentioned are non-clinical (bench testing) and refer to compliance with ASTM standards for MR safety and ANSI/AAMI for bacterial endotoxin. The document does not provide specific sample sizes for these non-clinical tests or their data provenance (e.g., country of origin, retrospective/prospective). It merely states that the tests were performed according to the specified standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as "No clinical tests were conducted." The ground truth for this submission is established through compliance with recognized industry standards for MR safety and bacterial endotoxin testing, not through expert-reviewed clinical data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This is not applicable as "No clinical tests were conducted." The assessment is based on the results of non-clinical, standard laboratory tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The submission is for knee joint prostheses and concerns MR safety labeling, not an AI-assisted diagnostic or treatment system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable. The submission does not involve an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The "ground truth" for this submission is the compliance with established international and national standards for medical device safety, specifically:
- ASTM standards for Magnetic Resonance (MR) Environment safety (F2503-23, F2182-19E2, F2052-21, F2213-17, F2119-07).
- ANSI/AAMI ST 72:2019 for bacterial endotoxin testing.
8. The sample size for the training set:
This is not applicable as "No clinical tests were conducted." There is no "training set" in the context of this submission, which relies on non-clinical engineering and biological safety testing against established standards.
9. How the ground truth for the training set was established:
This is not applicable, as explained in point 8.
Ask a specific question about this device
(94 days)
ATTUNE Total Knee System, ATTUNE Cementless Knee System, LPS Limb Preservation System, Sigma High Performance
ATTUNE™ Total Knee System Indications for Use
Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.
ATTUNE™ Cementless Knee System Indications for Use
The ATTUNE™ Cementless CR and PS Femoral Components are intended for cementless use within the ATTUNE™ Total Knee Replacement System.
Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant (provided that adequate bone is present).
LPS TM Limb Preservation System Indications for Use
The DePuy LPS System is intended for use in replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include:
- · malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) requiring extensive resection and replacement;
- · patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis, requiring extensive resection and replacement;
- · revision cases for a failed previous prosthesis requiring extensive resection and replacement;
- · severe trauma requiring extensive resection and replacement.
The LPS System is also intended for use in bone loss post-infection, where the surgeon has elected to excise the bone and replacement is required.
The S-ROM tibial tray and the non-porous coated straight and bowed stems are intended for cemented use only.
The porous-coated metaphyseal sleeves are intended for either cemented or cementless applications.
Sigma™ High Performance (HP) Partial Knee System Indications for Use
The SIGMA™ High Performance Partial Knee System is indicated for single compartmental knee replacement in skeletally mature individuals with osteoarthritis, post-traumatic arthritis of the tibiofemoral surfaces or a history of gout or pseudogout. All components are intended for CEMENTED USE ONLY.
ATTUNE™ Total Knee System: A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.
ATTUNE™ Cementless Knee System: A Total Knee Prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant with or without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial base without porous coating, and a polyethylene insert and locking components. The patella component may be of an all polyethylene design.
LPS - Limb Preservation System: The DePuy LPSTM Limb Preservation System is designed for the replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia. The DePuy LPS system offers a variety of component options (including, but not limited to, proximal femoral bodies, segmental components, distal femoral components, femoral stems, tibial stems, proximal tibial components, hinged tibial insert bearings, metaphyseal sleeves, and adapters). The components, which can be used in conjunction with certain components from other systems, are for treatment of patients presenting bone loss and deformity associated with bone tumors resection, trauma, infection, and difficult revision arthroplasty. A total femoral replacement is possible in those cases where no part of the femur can be salvaged.
Sigma High Performance (HP) Partial Knee System: The DePuy SIGMATM High Performance Partial Knee System is a single compartmental knee prosthesis, composed of individually packaged femoral, and tibial components designed to be used in various combinations to replace the natural articular surfaces of the knee joint. The unicompartmental femoral components are Co-Cr-Mo metal implants. The metal backed tibial components are Co-Cr-Mo with polyethylene inserts. The all-polyethylene unicompartmental tibial component are manufactured from polyethylene. The unicompartmental femoral components are designed for individuals who require a higher than normal degree of flexion (up to 155°).
This document (K233980) is a 510(k) premarket notification for several DePuy knee systems. It primarily focuses on adding updated MRI compatibility information and standardizing language in the Instructions for Use (IFU) and labels. The core claim for substantial equivalence relies on non-clinical performance data related to MRI safety and bacterial endotoxin testing, rather than a clinical study of device performance in a human-in-the-loop or standalone AI context.
Therefore, the requested information regarding acceptance criteria, study design for device performance, sample sizes for test and training sets, number and qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth establishment cannot be fully provided from this document as it pertains to a different type of device (knee implants) and regulatory submission (510(k) for labeling changes related to MRI safety, not an AI/software as a medical device performance study).
However, I can extract the acceptance criteria for the non-clinical performance tests that were conducted, and the reported performance as implied by the conclusion of substantial equivalence.
Here's what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance
For MRI safety, the acceptance criteria are implicit in meeting the standards listed. The reported "performance" is that the devices meet these standards, thus proving MRI safety.
Acceptance Criteria (Implicit) | Reported Device Performance (Implied) |
---|---|
Conformance to ASTM F2503-23 (Standard practice for marking medical devices for safety in MR environment) | Met |
Conformance to ASTM F2182-19E2 (Measurement of Radio Frequency Induced Heating On or Near Passive Implants during MR) | Met |
Conformance to ASTM F2052-21 (Measurement of Magnetically Induced Displacement Force on Medical Devices in the MR Environment) | Met |
Conformance to ASTM F2213-17 (Measurement of Magnetically Induced Torque on Medical Devices in the MR Environment) | Met |
Conformance to ASTM F2119-07 (Evaluation of MR Image Artifacts from Passive Implants) | Met |
Conformance to ANSI/AAMI ST 72:2019 (Bacterial Endotoxin Testing) | Met |
2. Sample sizes used for the test set and the data provenance
- Sample Size: Not applicable in the context of clinical or AI performance. The "test set" here would refer to the physical devices tested for MRI compatibility and bacterial endotoxin. The document does not specify the number of individual devices subjected to these non-clinical tests.
- Data Provenance: Not applicable for a typical AI/software study. The "data" here are the results from physical and chemical testing of the devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This is not applicable as there is no "ground truth" established by experts in the context of an AI or diagnostic device study. The ground truth for these tests is defined by the physical properties measured according to established ASTM and ANSI standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This is not applicable as this is not an image-based or diagnostic AI study requiring human adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. The document explicitly states: "No clinical tests were conducted to demonstrate substantial equivalence." (pages 6, 9, 12, 15)
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This is not applicable as the device is a knee implant, not an algorithm or software. The "performance" assessment is of the physical and material properties of the implant itself in relation to MRI safety.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the non-clinical tests is established by the physical and chemical properties measured according to the specified ASTM and ANSI international standards for MRI compatibility and bacterial endotoxin levels.
8. The sample size for the training set
- Not applicable. There is no AI model or training set involved in this regulatory submission.
9. How the ground truth for the training set was established
- Not applicable. There is no AI model or training set involved in this regulatory submission.
In summary: This FDA 510(k) submission for DePuy knee systems is related to changes in labeling for MRI compatibility, not an AI/software/diagnostic device. The "study" described focuses on non-clinical bench testing to demonstrate MRI safety and bacterial endotoxin compliance, adhering to recognized industry standards, rather than clinical performance or AI algorithm validation.
Ask a specific question about this device
(86 days)
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. The intended routes of administration consist of the following clinically accepted routes; intravenous, arterial, subcutaneous, epidural or irrigation of fluid space. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used in conjunction with legally marketed and compatible intravenous administration sets and medications provided by the user.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is suitable for a variety of patient care environments such as, but not limited to hospitals and outpatient care areas.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to reduce operator interaction through guided programming, thereby helping to reduce errors. The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used by trained healthcare professionals.
The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is a large volume infusion pump system that provides for safe and effective delivery of fluids into a patient in a controlled manner, as identified in 21 CFR, 880.5725. The pump is a software controlled, electromechanical device used for the infusion of pharmaceutical drugs, blood, blood products and mixtures of required patient therapy through administration sets at user selectable rates and volumes. The feedback-controlled, motorized pumping mechanism is of linear peristaltic design and uses inlet and outlet valves for flow control. The pump utilizes a primary and secondary processor to assure safe operation while providing infusion pump capabilities for a wide range of applications.
The pump is specifically manufactured and calibrated for the application of a manufacturer's brand of standard gravity administration sets, as indicated in the device labeling. For use, the administration set is loaded into the infusion pump. After acceptance of program parameters, the pump is started and fluid is propelled by the peristaltic action of the pumping mechanism against the outside surface of the administration set tubing. The pump is controlled to create smooth fluid dynamics, precision volumetric accuracy and uniformity of flow rate. None of the pump materials contact the administration set's fluid path.
The infusion pump is small in comparison to the traditional Large Volume Parenteral (LVP) infusion pumps currently on the market. It is designed to be used in a variety of patient care environments such as, but not limited to hospitals and outpatient care areas using an IV pole mounted configuration.
The Master Drug Library (MDL) Editor is a software application that allows the generation, configuration and management of a downloadable drug library into a SIGMA Spectrum infusion pump.
The drug library can be loaded directly into the SIGMA Spectrum infusion pump through a wireless network host or through an Infrared Data Association (IrDA) device. The MDL Editor software operates on a Microsoft Windows® platform.
Using the MDL Editor software application, a facility can provide preprogrammed delivery profiles, advisories and limits for a corresponding drug that is intended for a specific use classification or clinical care area, thus reducing the risk of medication errors.
The MDL Editor software application allows the ability to generate both standard or customized drug and fluid reports by clinical care area. The MDL Editor software application also provides a feature to restrict/limit the access of data to only appropriate personnel, providing additional security and rights to specific users.
The provided text is a 510(k) summary for the Baxter SIGMA Spectrum Infusion Pump with Master Drug Library. Its purpose is to demonstrate substantial equivalence to a predicate device, not to detail the full acceptance criteria and study proving its performance.
The document does not contain any information about acceptance criteria or a study that proves the device meets specific performance criteria related to AI or a clinical effectiveness study.
Instead, the document focuses on:
- Identifying the device and its predicate.
- Describing the device's function as an infusion pump.
- Stating its indications for use.
- Explaining that the specific 510(k) notification is to "update the algorithm and labeling related to the upstream occlusion alarm of the pump" and that "Performance testing for the software was completed."
Therefore, I cannot provide the requested information based on the given text. The prompt asks for details that would typically be found in a detailed validation report or clinical study summary, which is not present in this 510(k) summary.
Ask a specific question about this device
(767 days)
ROTEM sigma Thromboelastometry System
The ROTEM sigma thromboelastometry system is a fully integrated and automated in vitro diagnostic system designed to monitor and analyze a patient's coagulation status by measuring the viscoelastic properties of a 3.2% citrated venous or arterial whole blood sample. The ROTEM sigma system is indicated for use with adult patients 21 years and older where a semi-quantitative evaluation of their blood coagulation properties is desired, in the point of care and laboratory settings. Coagulation evaluations on the ROTEM sigma instrument, together with the ROTEM sigma complete + hep cartridge, are used to assess peri-operative hemorrhage and/or thrombosis in cardiovascular surgery and liver transplantation. The single use, multichannel cartridge ROTEM sigma complete + hep contains the following assays:
INTEM C is a semi-quantitative assay used to monitor coagulation via the intrinsic pathway in citrated whole blood samples.
EXTEM C is a semi-quantitative assay used to monitor coagulation via the extrinsic pathway in citrated whole blood samples.
FIBTEM C is a semi-quantitative assay used to monitor coagulation via the extrinsic pathway in citrated whole blood samples, after blocking platelet contribution to clot firmness.
HEPTEM C is a semi-quantitative assay used to monitor coagulation via the intrinsic pathway in citrated whole blood samples, after inactivating heparin.
Results from the ROTEM sigma should not be the sole basis for a patient diagnosis; ROTEM sigma results should be considered along with a clinical assessment of the patient's condition and other laboratory tests.
For in vitro Diagnostic Use.
For professional use only.
The ROTEM sigma is an in vitro diagnostic (IVD) whole blood hemostasis system intended for use in the evaluation of coagulopathies in Point of Care (POC) or laboratory settings. It uses rotational thromboelastometry to provide semiquantitative information about the coagulation state of a blood sample. The ROTEM sigma system records the kinetic changes in a sample of 3.2% citrated whole blood during clot formation, as well as when the sample clot retracts and/or lyses.
Several parameters are measured and reported for this purpose. The graphical presentation reflects the various physiological results, which describe the interaction between coagulation factors and inhibitors, fibrinogen, platelets, and the fibrinolysis system. Additionally, the effect of certain drugs influencing hemostasis, in particular some anticoagulants (e.g. heparin), can be detected.
The ROTEM sigma technology uses rotational thromboelastometry that is based on a fixed cylindrical cup and an oscillating vertical axis. The axis is supported by a high precision ball bearing and oscillates through an angle of 4.75°. The oscillation of the axis is driven by a motor that is connected to the axis via a spring. For the measurement, the channel's measurement axis engages the plastic pin in the cup of the disposable heated cartridge holding the blood sample. The oscillation is detected optically via a mirror plate at the upper end of the axis, which reflects the light from a diode light source onto a light sensitive sensor. If no clotting takes place, the pin movement is not restricted. As a clot forms and attaches itself between the pin and cup surfaces, the pin movement becomes increasingly restricted. The result is a balance between the spring tension and the tension of the clot. As the clot becomes firmer, the oscillation amplitude of the axis is reduced.
The ROTEM sigma assays are based on the principle of either
- intrinsic coagulation activation with or without the presence of heparin, or
- extrinsic coagulation activation with or without the presence of platelet inhibitors.
Here's an analysis of the provided text to extract information about the acceptance criteria and the study proving the device meets them.
It's important to note that this document is a 510(k) summary for an In Vitro Diagnostic (IVD) device, specifically a Thromboelastometry System (ROTEM sigma). The "acceptance criteria" here refer to performance specifications (e.g., precision, reproducibility, method comparison) for the device's measurements, not typical "acceptance criteria" for an AI/ML model's diagnostic accuracy (like sensitivity/specificity against a ground truth). Similarly, "expert consensus" or "adjudication" in the context of IVDs refers to establishing the accepted values of controls or comparing to a reference method, not to human image interpretation.
Acceptance Criteria and Device Performance for ROTEM sigma Thromboelastometry System
The ROTEM sigma Thromboelastometry System is an in vitro diagnostic device for assessing blood coagulation. Its performance is demonstrated through various analytical studies. The acceptance criteria are implicit in the presentation of the study results, where the variability and agreement with reference methods are shown to be within acceptable ranges for an IVD.
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is an IVD device, the "acceptance criteria" are typically defined by ranges of acceptable precision (CV%), reproducibility (SD/CV%), and correlation with a predicate device (slope, intercept, R-value). The document presents the results of these studies, and the implicit acceptance criteria are met if these results fall within expected performance ranges for such devices.
Summary of Key Performance Parameters (Implicit Acceptance Criteria and Reported Performance)
Performance Study | Parameter | Acceptance Criteria (Implicit) | Reported Device Performance (Worst Case/Overall) |
---|---|---|---|
Precision | %CV (Within-Laboratory) | Should be low, indicating consistent results for identical samples. | Highest %CV observed: |
(Whole Blood) | INTEM C: CT, A5, A10, A20, MCF | Generally 240 replicates per control type). |
* **Whole Blood (Normal, Contrived Hypocoagulable, Contrived Hypercoagulable):** Run in triplicate, in one (1) day, on five (5) instruments, resulting in 15 replicates per sample type per lot (3 lots => 45 replicates per sample type).
* **Lysis Precision (Normal Whole Blood & Hyperfibrinolysis Blood):** Fifteen (15) replicates per sample type per cartridge lot (3 lots => 45 reps per type).
* **Lot-to-Lot Variability (Normal Donor Whole Blood):** Thirty (30) replicates per cartridge lot (3 lots => 90 replicates).
- Reproducibility (External Clinical Sites):
- Three (3) external clinical sites.
- Four (4) ROTEM sigma instruments per site.
- Three (3) lots of controls (ROTEM sigma ROTROL N and P).
- Run in triplicate, twice a day for five (5) days, resulting in thirty (30) replicates per control per site (90 replicates per control across all sites).
- Interference:
- Various Interferents (Heparins, Acids, Ticagrelor): Eight (8) replicates at three (3) interferent levels for normal and hypocoagulable whole blood samples.
- Lupus Anticoagulant: Eleven (11) donors, each run on three (3) instruments with one (1) replicate per instrument.
- Reference Intervals: One hundred twenty (120) whole blood samples from healthy donors.
- Method Comparison:
- Number of samples for each assay parameter: INTEM C (144-148), EXTEM C (183-187), FIBTEM C (183), HEPTEM C (182).
- Data provenance: Patient samples from the "intended use populations and contrived samples." Conducted at four (4) clinical sites. No specific country of origin is mentioned beyond "internal precision study" and "external clinical sites" (implicitly within the regulatory region/country this submission applies to). The studies are prospective in the sense that they involve planned laboratory testing and data collection.
- Arterial vs. Venous Study: Seventy-four (74) matched venous and arterial citrated whole blood samples. Performed at two (2) external clinical sites.
Data Provenance (Summary):
The studies include data from internal testing and multiple external clinical sites. The samples include healthy donors, patient samples (intended use populations), and contrived samples (e.g., hypocoagulable, hypercoagulable). The studies are analytical performance studies, not clinical outcome studies. These are prospective laboratory studies designed to evaluate technical performance. No specific countries are listed, but the context of an FDA 510(k) submission suggests either U.S. or international data accepted by the FDA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
For an IVD device like the ROTEM sigma, "ground truth" is established by controlled laboratory methods and comparison to existing, cleared predicate devices, rather than human interpretation or expert consensus in the typical sense for image-based AI.
- Precision and Reproducibility: Ground truth is inherent in the preparation of controlled samples (normal, contrived hypocoagulable, contrived hypercoagulable, control materials) with known characteristics. The goal is to demonstrate the device's ability to consistently measure these known or expected characteristics. No external human experts are involved in creating this "ground truth" beyond standard laboratory practices and quality control.
- Method Comparison: The "ground truth" or reference standard for comparison is the predicate device, the ROTEM delta (K083842, K101533). The performance is assessed by correlating the ROTEM sigma's measurements with those of the legally marketed predicate device. This is a comparative study, not one establishing an absolute "ground truth" with human experts.
There is no mention of experts establishing ground truth in the way one would for an AI in medical imaging (e.g., radiologists annotating images).
4. Adjudication Method for the Test Set
For IVD analytical performance studies, traditional "adjudication" (e.g., 2+1, 3+1 for discordant reads) is not applicable. The measurements are quantitative. For method comparison, if there were significant discrepancies between the new device and the predicate, further investigation would occur to understand the cause, but it's not an adjudication process of human interpretation. The consistency across multiple instruments and sites is shown in the reproducibility studies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. An MRMC study is typically for evaluating the diagnostic performance of software, often AI, sometimes with and without human assistance (e.g., radiologists reading images). This document describes the analytical performance of an in vitro diagnostic instrument that measures blood coagulation properties. It does not involve human readers interpreting "cases" in a diagnostic context. Therefore, there's no "effect size of how much human readers improve with AI vs without AI assistance" to report here, as AI is not assisting human interpretation in this context; it is the measurement device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
The ROTEM sigma is inherently a "standalone" device in its primary function. It's an automated system that measures coagulation parameters. The performance studies (precision, reproducibility, method comparison) are all "algorithm only" in the sense that they evaluate the device's ability to produce consistent and accurate measurements on its own. Human intervention during operation involves sample loading and system maintenance, but the measurement process itself is automated ("fully integrated and automated").
7. The Type of Ground Truth Used
The "ground truth" in this context refers to established values or reference methods:
- Established Reference Materials/Controls: For precision, reproducibility, and lot-to-lot variability, the "ground truth" is based on the known or expected values of the control materials and the stability of the whole blood samples used.
- Predicate Device (ROTEM delta): For method comparison, the ROTEM delta serves as the reference standard or "ground truth" for evaluating the clinical equivalence of the ROTEM sigma's measurements.
- Healthy Donor Samples: For reference interval establishment, samples from healthy individuals are used to define the typical range of values, which then serve as a clinical "ground truth" for normalcy.
- Contrived Samples: For some studies (precision, method comparison), samples are artificially manipulated (e.g., hypocoagulable, hypercoagulable) to represent different physiological states. The "ground truth" for these is the known manipulation.
There is no "pathology" or "outcomes data" ground truth in the direct sense of a diagnostic accuracy study. The device provides semi-quantitative measurements that aid in diagnosis, but it does not provide a definitive diagnosis itself.
8. The Sample Size for the Training Set
This document describes the validation of a medical device, not the training of an AI/ML model. Therefore, there is no "training set" in the context of machine learning. The device determines coagulation parameters based on a physical principle (rotational thromboelastometry) and built-in algorithms, not from a data-driven training process in the AI sense.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" for an AI/ML model, this question is not applicable.
Ask a specific question about this device
(239 days)
SIGMA Sterilization Pouch and Roll
The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam Sterilizer and via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:
- · Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 30 minutes.
- · Gravity steam at 132°C (270°F) for 15 minutes; Drying time of 30 minutes.
- · Gravity steam at 135°C (275°F) for 10 minutes; Drying time of 30 minutes.
- · Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
- · Pre-vacuum steam at 135°C (275°F) for 3 minutes; Drying time of 16 minutes.
· EO sterilization cycle is 4 hours at 55°C (131°F) with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L.
Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the pouch has undergone either a steam or EO sterilization process.
The SIGMA sterilization pouches and rolls are inserted into the Pouch/Roll, sealed, and then sterilized in the Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization.
The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the steam/or EO Sterilization cycles.
The chemical indicators ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas. The color of Chemical Indicator changes from Blue to Greenish Black, when exposed to Steam. And the color changes from red to yellow, when exposed to EO gas.
The Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.
The provided document is a 510(k) premarket notification for the SIGMA Sterilization Pouch and Roll. It does not describe an AI/ML medical device, but rather a sterilization packaging product. Therefore, the questions related to AI/ML device performance metrics such as sample size for test set, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are not applicable.
The document focuses on demonstrating substantial equivalence to a predicate device (K180661) through non-clinical performance testing.
Here's an analysis of the acceptance criteria and study as presented for this non-AI/ML medical device:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria and reported device performance are extensively detailed across multiple tables, primarily Table 2 ("Summary of the Proposed and Predicate Devices Technological Characteristics") and Table 3 ("Summary of Non-Clinical Testing"). I will extract the key information from these tables.
Feature | Test/Standard | Acceptance Criteria | Reported Device Performance (Proposed Device) | Result |
---|---|---|---|---|
Sterilant Penetration / Drying Time / Aeration | ||||
Steam Sterilization - SAL | ANSI/AAMI/ISO 17665-1:2006/(R)2013, ANSI/AAMI/ISO TIR 17665-2:2009 (R2016) | Meet the requirement of SAL 10-6, test BI (Steam processed): No bacterial growth | Using half-cycle and full-cycle analysis. Test BI: No bacterial growth | Pass |
Steam Sterilization - Drying Time | ANSI/AAMI/ISO 17665-1:2006/(R)2013, ANSI/AAMI/ISO TIR 17665-2:2009 (R2016) | Weight difference before sterilization and after drying shall not exceed 0% | Weight difference = 0% | Pass |
Steam Sterilization - Visual Drying | (Standard not explicitly listed for this specific bullet, but falls under the general standards above) | (Implied visual dryness) | Visual are drying. | Pass |
EO Gas Sterilization - SAL | AAMI / ANSI / ISO 11135:2014 | Meet the requirement of SAL 10-6, test BI (EO processed): No bacterial growth | Using half-cycle analysis. Test BI: No bacterial growth | Pass |
EO Gas Sterilization - Residuals | ISO 10993-7:2008 (R) 2012 | EO 400 (kgf/mm2), MD > 450 (kgf/mm2) | CD = 458, MD = 462 (min. value after Steam Sterilization) | Pass |
Tear Resistance Test (Plastic Film) | ASTM D 1004 -13 | CD > 300 (kgf/mm2), MD > 350 (kgf/mm2) | CD = 443, MD = 453 (min. value after Steam Sterilization) | Pass |
Thickness Variation | ASTM F 2251-13 | ±0.02 | ±0.003 | Pass |
Internal Pressurization Failure Resistance | ASTM F1140/F1140M-13 | Pressurize to 80% of burst value and hold for 30s. Burst value > 5.0 (kPa) | Average Burst pressure = 5.8 (kPa) (after Accelerated Aging) | Pass |
Dye Penetration Test (Seal Leaks) | ASTM F1929-15 | No leakage across the seal width of sterile barrier system | No Channels identified on package | Pass |
Seal Strength | ASTM F88/F88M-15 | Strength > 356 (gf /25.4mm) | Site 1= 588, Site 2= 598, Site 3= 909, Site 4= 1016 (min. value of test results) | Pass |
Accelerated Aging (Steam) | ASTM F 1980-07 | Incubated for 22.5 days under controlled conditions (Temp: 55°C / RH: 50%) simulating 6 months post-steam sterilization storage. | a. Test articles (no product included and steam sterilized 135°C/15min) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 6 months. b. Test articles (product included and steam sterilized 135°C/15min) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 6 months. (Implied that package integrity was maintained after aging, as it "Pass"ed) | Pass |
Accelerated Aging (EO Gas, Sterility) | ASTM F 1980-07 | Incubated for 137 days under controlled conditions (Temp: 55°C / RH: 50%) simulating 3 years post-EO gas sterilization storage. | a. Test articles (no product included) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 3 years. b. Test articles (product included and EO sterilized) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 3 years. (Implied that package integrity was maintained after aging, as it "Pass"ed) | Pass |
Accelerated Aging (General Shelf Life) | ASTM F 1980-07 | Incubated for 137 days under controlled conditions (Temp: 55°C / RH: 50%) simulating 3 years storage. | a. Test articles (no product included) were accelerated aged at 55°C, and 75% RH to simulate the real time aging of 3 years. b. Test articles (product included) were accelerated aged at 55°C, and 75% RH to simulate the real time aging of 3 years. (Implied that package integrity was maintained after aging, as it "Pass"ed) | Pass |
Bubble Emission (Gross Leaks) | ASTM D3078-02(R2013) | No Leakage | No Leakage | Pass |
Internal Pressurization (Gross Leaks) | ASTM F2096-11 | No Leakage | No Leakage | Pass |
Microbial Barrier Test | DIN 58953-6 | CFU 1.7 | Average LRV = 3.31 | Pass |
Chemical Indicator Efficacy Testing (Type 1 Process Indicators) | ||||
Steam CI Color Change (Low Exposure) | AAMI/ANSI/ISO 11140-1:2014 | 121°C / 3.0 min & 134°C / 0.5 min: Unacceptable result (Color should not change significantly) | 121°C / 3.0 min & 134°C / 0.5 min: the result of color is Blue (No change) | Pass |
Steam CI Color Change (High Exposure) | AAMI/ANSI/ISO 11140-1:2014 | 121°C / 10.0 min & 134°C / 2.0 min: Acceptable result (Color should change to endpoint color) | 121°C / 10.0 min & 134°C / 2.0 min: the result of color is from Blue to Greenish Black (Endpoint color) | Pass |
Steam CI Color Change (Dry Heat Control) | AAMI/ANSI/ISO 11140-1:2014 | Dry heat 140°C / 30 min: Unacceptable result (Color should not change significantly) | Dry heat 140°C / 30 min: the result of color is Blue (No change) | Pass |
EO Gas CI Color Change (Absence of EO) | AAMI/ANSI/ISO 11140-1:2014 | Absence of EO gas / 90 min: Unacceptable result (Color should not change significantly) | Absence of EO gas / 90 min: the result of color is Red (No change) | Pass |
EO Gas CI Color Change (Low Exposure) | AAMI/ANSI/ISO 11140-1:2014 | EO gas Teat / 2 min: Unacceptable result (Color should not change significantly) | EO gas Teat / 2 min: the result of color is Red (No change) | Pass |
EO Gas CI Color Change (High Exposure) | AAMI/ANSI/ISO 11140-1:2014 | EO gas Teat / 20 min: Acceptable result (Color should change to endpoint color) | EO gas Teat / 20 min: the result of color is Yellow (Endpoint color) | Pass |
Chemical Indicator Shelf Life | AAMI/ANSI/ISO 11140-1:2014 | All performance attributes should maintain the original color: 3 years shelf life | The real-time test was carried out from October 15, 2007 to December 15, 2010. Test group exposed to Steam maintained Dark Green; test group exposed to EO maintained Yellow; Control group maintained original color. | Pass |
Regarding the sections that are not applicable to non-AI/ML devices (points 2-9):
-
Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable, as this is a physical medical device. The testing involves laboratory evaluations of material properties and sterilization efficacy, not interpretation of data. The document does not specify general "sample sizes" in terms of number of patients/cases, but rather the number of units tested for specific physical and chemical properties. The provenance is likely from the manufacturer's testing facilities in Taiwan.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" in the context of AI/ML refers to expert-labeled data. For this device, "ground truth" is established by adherence to recognized international standards (e.g., ASTM, ISO, AAMI) and their specified pass/fail criteria for physical and chemical properties.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are typically for resolving discrepancies in expert labeling of AI/ML data. The "acceptance criteria" here are objective measurements against defined standards.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a standalone physical product, not an AI-assisted diagnostic or therapeutic device.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no AI algorithm. The device's performance is inherently standalone in its function as a sterilization pouch.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to an AI device. For this physical device, the "ground truth" for its performance is determined by adherence to established international and national standards for sterilization packaging, material science, and chemical indicators (e.g., microbial barrier, tensile strength, peel strength, chemical indicator color change).
-
The sample size for the training set: Not applicable. This is a physical medical device, not an AI/ML model for which a training set would be required.
-
How the ground truth for the training set was established: Not applicable. Same reason as point 8.
Ask a specific question about this device
(207 days)
MECTA Sigma
The MECTA Sigma ECT device is for use in treating catatonia or severe major depressive episode (MDE) associated with major depressive disorder (MDD) or bipolar disorder (BPD) in patients age 13 years or older who are treatment-resistant or who require a rapid response due to the severity of their psychiatric or medical condition.
Not Found
The provided document is an FDA 510(k) clearance letter for the MECTA Sigma Electroconvulsive therapy (ECT) device. It does not contain information about acceptance criteria, device performance, or studies that prove the device meets specific criteria (such as a clinical trial report or a comprehensive performance study).
The document states that the device is substantially equivalent to legally marketed predicate devices. This equivalence determination by the FDA typically means that the new device has the same intended use and the same technological characteristics as the predicate, or, if it has different technological characteristics, that the new device does not raise different questions of safety and effectiveness.
Therefore, I cannot provide the requested information based on the given text. A 510(k) clearance letter is a regulatory document, not a scientific study report detailed enough to answer the specific questions about acceptance criteria, study design, and performance metrics.
Ask a specific question about this device
(118 days)
SIGMA High Performance (HP) Partial Knee System
INDICATIONS
The SIGMA High Performance Partial Knee System is indicated for single compartmental knee replacement in skeletally mature individuals with osteoarthritis, post-traumatic arthritis of the tibiofemoral surfaces or a history of gout or pseudogout. All components are intended for CEMENTED USE ONLY.
The DePuy SIGMA® High Performance Partial Knee System is a single compartmental knee prosthesis, composed of individually packaged femoral and tibial components designed to be used in various combinations to replace the natural articular surfaces of the knee joint.
The unicompartmental femoral components are Co-Cr-Mo metal implants, available with or without a porous coating. The metal backed tibial components are Co-Cr-Mo and polyethylene and are available without a porous coating. The all-polyethylene unicompartmental tibial component manufactured from polyethylene.
The unicompartmental femoral components are designed for individuals who require a higher than normal degree of flexion (up to 155°).
The provided text describes a 510(k) premarket notification for a medical device, the SIGMA High Performance (HP) Partial Knee System. While the document outlines information about the device itself and its equivalence to predicate devices, it does not contain the acceptance criteria or a study that specifically proves the device meets those criteria in the context of an AI/ML-driven device's performance.
The document states:
- "SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE": This lists a number of engineering and material tests (Biocompatibility, UHMWPE Material Property Characterization, Range of Motion/Constraint, Contact Area/Stress, Wear Testing, Pull-Off, Bacterial Endotoxin testing).
- "SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION - Clinical testing was not necessary to demonstrate substantial equivalence."
This indicates that the submission is focused on demonstrating substantial equivalence based on material properties, design, and manufacturing processes, rather than the performance of an AI/ML algorithm. Therefore, the requested information regarding acceptance criteria, device performance, sample sizes for test/training sets, ground truth establishment, expert adjudication, or MRMC studies is not present in the provided text, as these are typically associated with the evaluation of AI/ML software performance.
Ask a specific question about this device
(441 days)
PFC SIGMA Knee System
Candidates for total or unicompartmental knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, or a failed previous implant. In candidates for unicompartmental knee arthroplasty, only one side of the medial or lateral compartment) is affected.
The Sigma® C/R Porocoat® Femoral Components are intended for cementless use as the femoral component of a Total Knee Replacement system.
In the US, all other porous coated components have been cleared for CEMENTED USE ONL Y. Any Non-Porous coated components are intended for CEMENTED USE ONLY.
A total knee prosthesis is composed of individually packaged femoral, tibial and patellar components designed to replace the natural articular surface of the knee joint. The femoral component is a metal implant, with or without a porous coating. The tibial component may be an all polyethylene component or comprised of a metal tibial tray with or without porous coating, and a polyethylene insert and locking components. Some metal components have modular stems, sleeves and/or modular wedges. The patella component may be of an all polyethylene design or may be a metal backed polyethylene component.
This 510(k) submission is to provide MRI compatibility labeling for the PFC SIGMA Knee System. No clinical studies were conducted as the submission is based on non-clinical testing for MRI compatibility.
Here's a breakdown based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test Type | Acceptance Criteria | Reported Device Performance |
---|---|---|
Magnetically Induced Displacement Force (ASTM F2052-14) | No excessive movement or dislodgement of the device in the MR environment. (Implied by safety concerns addressed in the FDA guidance and ASTM standards) | Testing concluded no safety issues related to magnetic field interactions under specific conditions. |
Torque (ASTM F2213-06) | No excessive rotation or twisting of the device in the MR environment. (Implied by safety concerns addressed in the FDA guidance and ASTM standards) | Testing concluded no safety issues related to magnetic field interactions under specific conditions. |
Radiofrequency (RF) Heating (ASTM F2182-11a) | No unsafe temperature increase in the device due to RF energy from the MR scanner that could impact patient safety. (Implied by safety concerns addressed in the FDA guidance and ASTM standards) | Testing concluded no safety issues related to magnetic field interactions under specific conditions. |
Image Artifacts (ASTM F2219-07) | Image quality should not be unacceptably degraded by device-induced artifacts, or the effects assessed and characterized. (Implied by concerns about imaging) | Effects of the implants on image quality were evaluated. |
MR Conditional Labeling (ASTM F2503-13) | The device should meet criteria to be labeled as "MR Conditional," meaning it can be safely used in a specified MR environment under defined conditions. | Testing supports an MR conditional label as described in FDA Guidance Document and ASTM F2503-13. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document states that "worst-case components and constructs" were evaluated for RF Heating and image artifacts. However, a specific numerical sample size for the devices tested is not provided.
- Data Provenance: The data is non-clinical, originating from laboratory testing (often referred to as ex-vivo or in-vitro when applicable to medical devices). The country of origin for the testing itself is not explicitly stated, but it follows US FDA guidance and ASTM standards. The study is retrospective in the sense that the device design already existed and was being evaluated for an additional claim.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This section is not applicable. The ground truth for MRI compatibility is established through standardized physical tests and measurements, not through expert consensus on interpretations of images or clinical outcomes. The "ground truth" is whether the device physically behaves safely and predictably in an MR environment according to predefined engineering standards.
4. Adjudication Method for the Test Set
This section is not applicable. As the evaluation involved physical measurements and adherence to engineering standards, there was no need for expert adjudication in the traditional sense of clinical trial data interpretation.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The submission is for MRI compatibility labeling based on non-clinical testing.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
A standalone performance study of an algorithm was not done. This submission pertains to the physical properties of a knee implant system and its interaction with MRI, not an algorithm.
7. Type of Ground Truth Used
The ground truth used was based on results from standardized non-clinical physical and engineering tests (magnetically induced displacement force, torque, RF heating, image artifacts) as defined by FDA guidance and ASTM standards (e.g., ASTM F2052-14, ASTM F2213-06, ASTM F2182-11a, ASTM F2219-07, ASTM F2503-13).
8. Sample Size for the Training Set
This section is not applicable. There was no "training set" as this was not a machine learning or AI-based study.
9. How the Ground Truth for the Training Set Was Established
This section is not applicable.
Ask a specific question about this device
(83 days)
SIGMA Sterilization Pouch and Roll
The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.
The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized in the Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization.
The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the steam/or EO Sterilization cycles.
The chemical indicators ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas. The color of Chemical Indicator changes from Blue to Greenish Black, when exposed to Steam. And the color changes from red to yellow, when exposed to EO gas.
The Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.
This document is a 510(k) Premarket Notification from Sigma Medical Supplies Corp. to the FDA regarding their SIGMA Sterilization Pouch and Roll. It seeks to prove substantial equivalence to a previously cleared device (K102158).
Here's an analysis of the acceptance criteria and the study proving the device meets them, based solely on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria here are derived from the comparison of the proposed device to the predicate device and the requirement to meet various ISO and ASTM standards. The study's "performance" is reported as "Passed" or "Meets requirements."
Table of Acceptance Criteria and Reported Device Performance
Feature/Test | Acceptance Criteria (Expected Outcome) | Reported Device Performance |
---|---|---|
Material Composition | Same as predicate (Medical Grade Paper, CPP, PET, PU adhesive, EO and Steam Process Indicator Print Ink) | Same as predicate (Medical Grade Paper, CPP, PET, PU adhesive, EO and Steam Process Indicator Print Ink) |
Steam Sterilization Cycle Parameters | 30 minutes at 121°C | Same (30 minutes at 121°C) |
EO Gas Sterilization Cycle Parameters | 4 hours at 55°C, 50%-85% RH, 600 mg/L sterilant concentration | Same (4 hours at 55°C, 50%-85% RH, 600 mg/L sterilant concentration) |
Pouch Types | Same as predicate (Self-sealing, Flat, Gusseted pouches; Flat, Gusseted rolls) | Same as predicate |
Sterilant Penetration (Half-Cycle Efficacy) | Meet requirement of SAL 10^-6^ | The test meets the requirement of SAL 10^-6^ |
Chemical Indicator (CI) Functionality & Endpoint | Sterilant penetrates and causes CI color change to endpoint color | The sterilant penetrated and affected the CI color change to the endpoint color |
Package Integrity (Physical Properties) | Passed (based on various ASTM/ISO standards) | Passed for all listed tests (Thickness Variations, Tensile Strength of plastic film, Tensile Strength of paper, Air Permeance of paper, Tear Resistance Test, Burst Strength, Bubble Leak Test, Seal Peel Test, Dye Penetration Test, Microbial Barrier Test) |
Toxicological Properties (Biocompatibility Test) | Passed (based on ANSI/AAMI/ISO 10993-10) | Passed. Showed "negative reaction". Meets ISO 10993-10:2010(E). EO sterilization residuals meet AAMI/ANSI/ISO 10993-7:2008 (R) 2012. |
Durability: Accelerated Aging Test | Passed (based on ASTM F 1980; ISO 11607-1) | Passed |
End Point / Post Processing Color Stability (Steam) | 6 months | 6 months |
End Point / Post Processing Color Stability (EO) | 3 Years | 3 Years |
Shelf Life (Chemical Indicator Functionality) | 3 Years | 3 Years |
Shelf Life (Accelerated aging test - Seal Strength) | 3 Years | 3 Years |
Class 1 Process Indicators for STEAM (Pouch & CI) | Remain stable before use, change color in presence of sterilant, maintain endpoint stability, meet ISO 11140-1:2014 requirements | Demonstrates ability to remain stable, change color, maintain endpoint stability. Meets ISO 11140-1:2014. |
Class 1 Process Indicators for EO Gas (Pouch & CI) | Remain stable before use, change color in presence of sterilant, maintain endpoint stability, meet ISO 11140-1:2014 requirements | Demonstrates ability to remain stable, change color, maintain endpoint stability. Meets ISO 11140-1:2014. |
Maintenance of Sterility (Steam) | Up to 6 months post Steam sterilization | Maintains up until 6 months post Steam sterilization |
Maintenance of Sterility (EO) | Up to 3 years post EO gas sterilization | Maintains up until 3 years post EO gas sterilization |
Study Details:
-
Sample Size Used for the Test Set and Data Provenance:
- The document does not specify the exact sample sizes used for each test. It states that "The testing demonstrates the ability of the Sigma Sterilization Pouch and Roll" and "The testing results demonstrate the ability of the Sigma Sterilization Pouch and Roll."
- Data Provenance: The manufacturer is Sigma Medical Supplies Corp. located in Taiwan (R.O.C.). The testing (non-clinical) would likely have been conducted by or for this company. The document does not specify if the data is retrospective or prospective, but given it's a premarket notification for a new device (albeit similar to a predicate), it would be prospective data generated specifically for this submission.
-
Number of Experts Used to Establish Ground Truth and Qualifications:
- This document describes performance testing of a medical device (sterilization pouches and rolls) against established standards (ISO, ASTM). It does not involve human interpretation of images or other data for which "ground truth" would be established by experts in a clinical sense (e.g., radiologists for diagnostic AI).
- The "ground truth" in this context is defined by the technical specifications and performance requirements of the relevant industry standards (e.g., attainment of SAL 10^-6^ for sterilization, specified physical properties, color change of chemical indicators).
- Therefore, the concept of "ground truth" established by a panel of human experts in the way it's used for AI/clinical diagnostic studies is not directly applicable here. The validity of the tests themselves is based on adherence to the referenced standards.
-
Adjudication Method for the Test Set:
- Not applicable in this context. Adjudication is typically used for reconciling disagreements among human readers or evaluators in clinical studies. Here, the tests are objective measurements against defined standards.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, an MRMC study was not done. This type of study is relevant for assessing the impact of AI on human reader performance in diagnostic tasks, which is not the purpose of this device or its reported testing.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, implicitly. All the described tests (Sterilant Penetration, Package Integrity, Biocompatibility, Chemical Indicator Efficacy, etc.) are evaluating the physical and functional properties of the device itself, independent of human interaction during its use (beyond following instructions). These are "algorithm only" in the sense that the device's performance is measured against objective criteria, not as part of a human-AI system.
-
The Type of Ground Truth Used:
- The ground truth is based on industry-recognized technical standards (e.g., ISO, ASTM) for sterilant efficacy, package integrity, biocompatibility, and chemical indicator performance. For example, "SAL 10^-6^" or physical property thresholds defined by the standards.
-
The Sample Size for the Training Set:
- This document describes premarket notification for a physical medical device, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The device is manufactured based on design specifications and then tested.
-
How the Ground Truth for the Training Set was Established:
- Not applicable, as there is no "training set" for this type of device. The design and manufacturing processes are likely informed by prior engineering knowledge and adherence to quality systems for medical device production.
Ask a specific question about this device
Page 1 of 9