Search Results
Found 4 results
510(k) Data Aggregation
(267 days)
Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable or for single patient use and designed for use with Mindray monitor Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.
These devices are used by qualified medical professional only.
The subject devices are used for patient temperature measurement. The probes are reusable or disposable depending on models. These probes consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. The subject devices are designed to be used in healthcare facilities like hospital and compatible with a monitor of Mindray Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.
The six models have two types of structure designs corresponding to reusable and disposable use which consists of different materials. The NTC of the six models are identical. Reusable models T1306, T2306, T3306, T4306 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials. Disposable models T5106 and T6106 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials.
Model: T1306, Description: Skin contact Temperature Probe, adult, reusable
Model: T2306, Description: Body cavity Temperature Probe, Esophageal/Rectal, adult, reusable
Model: T3306, Description: Skin contact Temperature Probe, pediatric, reusable
Model: T4306, Description: Body cavity Temperature Probe, Esophageal/Rectal, pediatric, reusable
Model: T5106, Description: Skin contact Temperature Probe, adult/ pediatric, disposable
Model: T6106, Description: Body cavity Temperature Probe, Esophageal/Rectal, adult/ pediatric, disposable
The provided FDA 510(k) clearance letter describes a medical device, the Reusable and Disposable Temperature Probes, but does not include information about AI/ML performance. Therefore, I will respond to the prompt by extracting the acceptance criteria and study information pertinent to this medical device, which focuses on traditional medical device performance rather than AI/ML.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance | Study Supporting Performance |
|---|---|---|
| Accuracy | ±0.1℃ | Bench Testing (ISO 80601-2-56) |
| Measurement Range | 25-45℃ | Bench Testing (ISO 80601-2-56) |
| Electrical Safety | Complies with IEC 60601-1 | Bench Testing (IEC 60601-1) |
| Electromagnetic Compatibility (EMC) | Complies with IEC 60601-1-2 | Bench Testing (IEC 60601-1-2) |
| Biocompatibility | Complies with ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23 | Biocompatibility testing |
| Operating Environment | +5 to +40°C, ≤80% humidity (non-condensing), 86kPa~106kPa | Bench Testing (IEC 60601-1 and ISO 80601-2-56) |
| Storage Environment | -20℃ to 55℃, ≤93% humidity, 86kPa~106kPa | Bench Testing (IEC 60601-1 and ISO 80601-2-56) |
| Compatibility with Monitors | Verifies compatibility with Mindray Model PM-8000 and other YSI 400 series compatible monitors | Bench Testing |
Note: The document presents "Accuracy" and "Measurement Range" as inherent characteristics of the device and states that bench testing was conducted to verify that design specifications were met, which implies these values are the acceptance criteria.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for the test sets (e.g., number of units tested, number of temperature measurements, or specific test configurations) for the bench testing or biocompatibility testing.
The document also does not provide information about the provenance of data in terms of country of origin or whether studies were retrospective or prospective. The testing described appears to be laboratory-based verification and validation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable and not provided in the document. The device is a clinical electronic thermometer, and its performance is assessed against technical specifications and international standards, not against human expert interpretation of medical images or data. Ground truth for temperature measurement is typically established by reference standards or calibrated equipment.
4. Adjudication Method for the Test Set
This information is not applicable and not provided in the document. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective assessment (e.g., image interpretation by multiple readers), which is not relevant for the objective performance testing of a temperature probe.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. An MRMC comparative effectiveness study was not conducted as this is a medical device for objective temperature measurement, not an AI-assisted diagnostic tool requiring human-in-the-loop performance evaluation. The document does not mention any AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
No. This device is an electronic temperature probe, not an algorithm or software. Its performance is inherent to its physical and electrical properties, evaluated through bench testing against established standards.
7. Type of Ground Truth Used
The ground truth for the performance evaluations (accuracy, measurement range, electrical safety, etc.) would be established by:
- Reference Standards/Calibrated Equipment: For accuracy and measurement range, the device's readings would be compared against highly accurate and calibrated reference thermometers in controlled temperature environments.
- International Standards: Compliance with electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and thermometer-specific performance (ISO 80601-2-56) serves as the ground truth for safety and performance.
- Laboratory Analysis: For biocompatibility, laboratory tests (cytotoxicity, sensitization, irritation) are conducted to assess the biological response to the device materials according to ISO 10993 standards.
8. Sample Size for the Training Set
This information is not applicable and not provided. This device is a hardware medical device with no mention of machine learning or algorithms that would require a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable. As there is no training set for an AI/ML algorithm involved, no ground truth was established for a training set.
Ask a specific question about this device
(248 days)
Reusable Temperature Probes:
Reusable Temperature Probes are intended to be used for monitoring temperature for multi- patient use. The temperature probes are reusable and designed for use with monitor of GE-Marquette Model DASH3000. These devices are used by qualified medical professionals only.
Disposable Temperature Probes:
Disposable Temperature Probes are intended to be used for monitoring temperature for single patient use. The temperature probes are non-sterile and designed for use with monitors of GE- Marquette Model DASH3000. These devices are used by qualified medical professionals only.
The proposed devices are used for patient temperature measurement. The probes are reusable or disposable depending on the model. These probes consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increasing with temperature decrease, and the linear changes to the characteristics of the temperature measurement. The proposed devices are designed to be used in a healthcare facility and compatible with a monitor, GE-Marquette Model DASH3000.
The six models have two types of structure designs corresponding to reusable and disposable use which consists of different materials. The Negative Temperature Coefficient (NTC) of the six models are identical. Reusable models ST1304, ST2304, ST3304, ST4304 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials. Disposable models ST5105A and ST6105A have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials.
The provided text describes the regulatory clearance of a medical device (Reusable and Disposable Temperature Probes) and its substantial equivalence to a predicate device, rather than a study designed to prove the device meets acceptance criteria through detailed performance metrics and a comparison to a defined ground truth.
Therefore, many of the requested details about acceptance criteria, specific study design elements (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, effect sizes, standalone performance, training set details) are not present in the provided document.
However, I can extract the following information that is available:
1. A table of acceptance criteria and the reported device performance:
The document mentions compliance with standards which inherently include performance criteria. Specifically, the "Accuracy" is listed.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Accuracy | ±0.1°C |
| Measurement Range | 25-45°C |
| Measurement Time | 60s |
| Biocompatibility | Complied with ISO 10993-1, ISO 10993-5, and ISO 10993-10 |
| Electrical Safety | Complied with IEC 60601-1 |
| Performance Standard | Complied with ISO 80601-2-56 |
| Electromagnetic Compatibility | Complied with IEC 60601-1-2 |
2. Sample size used for the test set and the data provenance:
- Sample Size for Test Set: Not specified. The document states "Non-clinical testing has been conducted to verify that the subject devices meet all design specifications." The exact number of units or measurements in these tests is not provided.
- Data Provenance: The tests are non-clinical, conducted by the manufacturer, Shenzhen SINO-K Medical Technology Co., Ltd., which is based in Shenzhen, Guangdong, China. The testing confirms compliance with international standards (IEC, ISO, ASTM).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: Not applicable/not specified. The testing described is non-clinical performance testing against established international standards, not expert evaluation of medical images or diagnostic outputs.
4. Adjudication method for the test set:
- Adjudication Method: Not applicable. This document describes objective performance testing against engineering standards, not a diagnostic study requiring expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This device is a temperature probe, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Standalone Performance: Yes, in the sense that the device's inherent performance (e.g., accuracy, measurement range) was demonstrated through non-clinical testing against standards (e.g., ISO 80601-2-56) without human interpretation of results beyond reading the device's output.
7. The type of ground truth used:
- Ground Truth Type: International standards (ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2), and established biocompatibility standards (ISO 10993 series). These specify acceptable ranges and methodologies for determining performance, forming the "ground truth" for compliance.
8. The sample size for the training set:
- Sample Size for Training Set: Not applicable. This device is a hardware temperature probe, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- Ground Truth for Training Set: Not applicable, as there is no training set for this device.
Ask a specific question about this device
(343 days)
The Reusable Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable and designed for use with monitors of Nihon Kohden model BSM-5135A. These devices are indicated for used by qualified medical personnel only.
Reusable temperature probes are used during patient temperature measurement for multi-patient use. These probes consist of a phone plug connector on the monitor end and a thermistor on the patient end. Temperature probes measure temperature by a resistor that is sensitive to temperature changes. These probes are connected to the patient monitor either directly by using a phono plug or by an interconnect cable. These probes have a skin or core contact with a patient.
These temperature probes are typically used with legacy Nihon Kohden monitors BSM-5135A.
Products are packed individually into a plastic bag in non-sterile condition. Package label describes product LOT codes, CE-mark, legal entity information and a caution "Rx Only ".
The provided text describes the performance data for the Reusable Temperature Probe (Models: TRAG-2252, TRAS-2252, TRCG-2252 and TRCS-2252) in support of its substantial equivalence determination.
Here's a breakdown of the requested information based on the text:
1. A table of acceptance criteria and the reported device performance
| Criteria | Acceptance Criteria (Standard / Description) | Reported Device Performance |
|---|---|---|
| Material Accuracy | 25-45°C | 25-45°C |
| Accuracy | ±0.2°C | ±0.2°C |
| Laboratory Accuracy | Not greater than 0.3 °C for a continuous clinical thermometer that is not an adjusted mode clinical thermometer (per ISO 80601-2-56) | Pass (within ±0.2°C) |
| Time Response | Heating transient time < 150s | < 150s |
| Electrical Safety | Meets requirements of IEC 60601-1 | Pass |
| Electromagnetic Compatibility (EMC) | Meets requirements of IEC 60601-1-2 and ISO 80601-2-56 | Pass |
| Disinfection Validation | Low-level disinfection validation; High-level disinfection validation | Pass |
| Chemical Disinfection Agent Residue | Non-Cytotoxicity according to ISO 10993-5 (after disinfection) | Pass |
| Biocompatibility | Complies with ISO 10993-1 (Cytotoxicity, Skin Sensitization, Skin Irritation, Rectal Irritation) | Pass (no cytotoxicity, negligible irritation, no sensitization) |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the sample size used for performance testing (e.g., how many probes were tested) nor does it specify the provenance (e.g., country of origin) of the data. It indicates that "All subject devices were conducted performance testing according to ISO 80601-2-56" and references "Laboratory accuracy" and "Time response" results as well as biocompatibility testing. The nature of these tests (e.g. for electrical safety and accuracy) suggests they are conducted in a laboratory setting, not on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable as the device is a temperature probe, and the performance testing described is objective measurement against standards and physical characteristics, not expert interpretation of diagnostic images or clinical scenarios requiring ground truth establishment by experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable for the reasons stated above. Adjudication methods are typically relevant for studies involving human interpretation (e.g., radiologists reading images) where disagreement among experts needs resolution.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. The device is a Reusable Temperature Probe, which is a physical measurement device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is not applicable, as the device is a temperature probe and does not involve an algorithm for diagnostic interpretation. Its performance is inherent in its design and manufacturing.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the performance metrics (accuracy, time response, safety, EMC) of the temperature probe, the "ground truth" is established by objective measurement against the specified international standards (e.g., ISO 80601-2-56 for accuracy and time response, IEC 60601-1 for electrical safety, ISO 10993 series for biocompatibility). There is no "expert consensus" or "pathology" involved in determining if a physical temperature measurement falls within a specified range.
8. The sample size for the training set
This section is not applicable. The device is a Reusable Temperature Probe, which is a hardware medical device, not an AI algorithm that requires a training set.
9. How the ground truth for the training set was established
This section is not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(26 days)
The reusable temperature probes are intended to be used for monitoring temperature. The temperature probes are designed for use with Datex-Ohmeda and GE Medical Systems IT monitoring systems and other monitors compatible with 400 Series temperature probes.
These devices are indicated for use by qualified medical personnel only.
Reusable temperature probes (M1024254 Skin Temperature probe, reusable; M1024247, GP Temperature Probe, Adult, reusable; M1024251 GP Temperature Probe, Pediatric, reusable) are used during patient temperature measurement. These probes consist of a phone plug connector on the monitor end and a thermistor on the patient end. These probes are to be used with 400-series compatible temperature measurement systems only.
Temperature probes measure temperature by a resistor that is sensitive to temperature changes. The probe is connected to the patient monitor either directly by using a phono plug or by an interconnect cable. These probes have a skin or core contact with a patient.
These temperature probes are typically used with legacy GE Medical System monitors like Dash 3000/4000 (K033304), Solar (K012467), TRAM (K900540) and also with the new GE Healthcare S/5 modules like M-PRESTN (K041772) and also legacy Datex-Ohmeda patient monitors and modules like Cardiocap 5 (K992323), Light (K981378) or M-ESTPR (K953175).
Products are packed individually into a plastic bag in non-sterile condition. Package label describes product REF codes, manufacturing date, CE-mark, legal entity information and a caution "Rx Only (USA), U.S. Federal law restricts this device to sale by or on the order of a physician."
This is an interesting case where the provided document is a 510(k) summary for a medical device that isn't AI-powered. Therefore, much of the requested information about AI model performance, ground truth, and expert evaluation is not applicable.
However, I can extract the relevant information regarding the acceptance criteria and the study proving the non-AI device meets those criteria from the provided text.
Here's a breakdown based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding performance data in the traditional sense of a clinical study or AI performance metrics. Instead, it refers to compliance with established international and national standards for medical devices, particularly regarding safety and performance.
| Acceptance Criteria (Standard) | Reported Device Performance (Compliance Claim) |
|---|---|
| EN 60601-1:1990 +A1:1993 +A2:1995 +A13:1996 Part 1: General requirements for safety | Assessed against, thoroughly tested through validation and verification of specifications. |
| IEC 60601-2-49:2001 (Part 2:-49: Particular requirements for the safety of multifunction patient monitoring equipment) | Assessed against, thoroughly tested through validation and verification of specifications. |
| 21 CFR Part 898 (FDA Radiation Control Regulations for electronic products) | Assessed against, thoroughly tested through validation and verification of specifications. |
| EN 1041:1998 Information supplied by the manufacturer with medical devices | Assessed against, thoroughly tested through validation and verification of specifications. |
| EN 980:2003 Graphical symbols for use in the labeling of medical devices | Assessed against, thoroughly tested through validation and verification of specifications. |
| ISO 10993-5:1995, ISO 10993-10:1996 Biological evaluation of medical devices | Assessed against, thoroughly tested through validation and verification of specifications. |
| ISO 14971:2000 Medical devices Application of risk management to medical devices | Assessed against, thoroughly tested through validation and verification of specifications. |
| EN12470-4:2000 Performance of electrical thermometers for continuous measurement | Assessed against, thoroughly tested through validation and verification of specifications. |
| Accuracy in 32 to 42° C range (Claim of similarity to predicate device) | The proposed reusable temperature probes have "accuracy in 32 to 42° C range" similar to the predicate device, implying this range is met. |
Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the acceptance criteria is referred to as "validation and verification of specifications" against the listed standards. This is not a clinical study with human subjects, but rather a series of engineering and quality assurance tests. The document states: "The reusable temperature probes... have been assessed against the standards below. The device has been thoroughly tested through validation and verification of specifications."
2. Sample Size Used for the Test Set and Data Provenance
This is not applicable as the device is a physical temperature probe and not an AI algorithm processing data. The "test set" refers to the specific probes manufactured and subjected to testing to ensure compliance with the standards. There is no mention of a fixed "sample size" for a data test set in the context of AI.
The data provenance, again, is not applicable in the AI sense. The "data" here would be the measurements and test results from the physical probes themselves during their validation and verification.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This is not applicable. Ground truth, in the context of AI, refers to annotated data. For a physical device like a temperature probe, the "ground truth" for accuracy is typically established by comparing the probe's readings to highly accurate reference thermometers traceable to national standards, carried out by laboratory technicians or engineers, not medical experts establishing a diagnosis.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (like 2+1 or 3+1 for resolving disagreements among human annotators) are relevant for establishing ground truth in AI training or evaluation data. For physical device testing, the methods involve standardized test procedures and measurement comparison.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
This is not applicable as the device is a physical temperature probe, not an AI algorithm assisting human readers. Therefore, no MRMC study looking at human reader improvement with AI assistance would have been conducted or is relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable for the same reason as above. This is a physical device, not an algorithm.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance of these temperature probes (specifically accuracy) would be established by comparison to reference standards (e.g., highly accurate, calibrated thermometers in a controlled environment), not expert consensus, pathology, or outcomes data. The document explicitly mentions "accuracy in 32 to 42° C range" among its similarities to the predicate device, suggesting this was a key performance metric evaluated against an objective standard.
8. The Sample Size for the Training Set
This is not applicable as this is a physical medical device, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as this is a physical medical device, not an AI model.
Ask a specific question about this device
Page 1 of 1