K Number
K033304
Date Cleared
2003-11-12

(29 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dash 3000/4000 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash is designed as a bedside, portable, and transport monitor that can operate in all professional medical facilities and medical transport modes including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, freestanding surgical centers, and other alternate care facilities, intra-hospital patient transport, inter-hospital patient transport via ground vehicles (i.e., ambulance, etc.) and fixed and rotary winged aircraft, and pre-hospital emergency response.

Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, non-invasive blood pressure, pulse. temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, and anesthetic agents as summarized in the operator's manual.

The Dash 3000/4000 Patient Monitor is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital data at the point-of-care.

This information can be displayed, trended, stored, and printed.

The Dash 3000/4000 Patient Monitor was developed to interface with non-proprietary third party peripheral devices that support serial data outputs.

Device Description

The Dash 3000/4000 Patient Monitor is a device that is designed to be used to monitor, display, and print a patient's basic physiological parameters including: electrocardiography (ECG), invasive blood pressure, non-invasive blood pressure, oxygen saturation, temperature, impedance respiration, end-tidal carbon dioxide, oxygen, nitrous oxide and anesthetic agents. Other features include arrhythmia, cardiac output, cardiac and pulmonary calculations, dose calculations, PA wedge, ST analysis, and interpretive 12 lead ECG analysis (12SL). Additionally, the network interface allows for the display and transfer of network available patient data.

AI/ML Overview

The GE Medical Systems Information Technologies Dash 3000/4000 Patient Monitor is a physiological patient monitor designed to monitor, display, and print a patient's basic physiological parameters. These parameters include electrocardiography (ECG), invasive blood pressure, non-invasive blood pressure, oxygen saturation, temperature, impedance respiration, end-tidal carbon dioxide, oxygen, nitrous oxide, and anesthetic agents. The device also offers features like arrhythmia detection, cardiac output, cardiac and pulmonary calculations, dose calculations, PA wedge, ST analysis, and interpretive 12-lead ECG analysis (12SL).

This submission describes the acceptance criteria and study used to demonstrate the substantial equivalence of the Dash 3000/4000 Patient Monitor to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly list specific numerical acceptance criteria for each physiological parameter monitored by the Dash 3000/4000 Patient Monitor, nor does it provide detailed reported device performance values in a quantitative table.

Instead, the submission outlines a general approach to compliance and testing. The "Test Summary" section states:
"The Dash 3000/4000 Patient Monitor complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the system:

  • Requirements specification review
  • Code inspections
  • Software and hardware testing
  • Safety testing
  • Environmental testing
  • Final validation"

The "Conclusion" then states:
"The results of these measurements demonstrated that the Dash 3000/4000 Patient Monitor is as safe, as effective, and performs as well as the predicate devices."

This implies that the acceptance criteria were based on adherence to relevant voluntary standards and that the device's performance was found to be comparable to that of the predicate devices for all specified physiological parameters. Without "Section 9" or further detailed test reports, specific numerical performance figures and acceptance thresholds cannot be extracted from this summary.

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not specify the sample size used for the test set. It mentions "software and hardware testing," "safety testing," and "environmental testing," but does not provide details about the number of patients, data points, or scenarios included in these tests.

The data provenance (e.g., country of origin, retrospective or prospective) is also not mentioned in the summary. The studies appear to be internal verification and validation tests rather than clinical trials with external data sources.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The provided summary does not mention the use of experts to establish ground truth for any specific physiological parameter. The testing appears to be focused on technical performance against established standards and predicate device comparisons rather than requiring expert adjudication of clinical outcomes or interpretations.

4. Adjudication Method for the Test Set

Since the summary does not indicate the involvement of experts for establishing ground truth, there is no adjudication method mentioned (e.g., 2+1, 3+1). The testing described likely relied on instrumental measurements and technical specifications.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The summary does not mention a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. The device is a physiological patient monitor, and the described testing focuses on its technical performance and compliance with standards, not on human reader interpretation accuracy or the effect of AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

The summary does not explicitly describe a standalone performance study in terms of "algorithm only without human-in-the-loop performance." However, the "Test Summary" detailing "software and hardware testing" and "final validation" implies that the device's functionality was evaluated in a standalone manner to ensure it meets requirements and performs as intended. The listed physiological parameters are measured and displayed by the device itself, suggesting an inherent standalone performance evaluation against established reference methods or predicate devices.

7. Type of Ground Truth Used

The type of ground truth used for the testing highlighted in the summary would be based on:

  • Voluntary standards: Compliance with recognized industry and regulatory standards (as mentioned in Section 9, although not provided).
  • Predicate device comparison: Performance being "as safe, as effective, and performs as well as the predicate devices," indicating that the predicate device's performance serves as a benchmark for comparison.
  • Technical specifications and requirements: The device's performance against its own "requirements specification."
  • Instrumental measurements: For parameters like ECG, blood pressure, temperature, etc., the "ground truth" would be derived from calibrated reference instruments or established measurement methodologies.

8. Sample Size for the Training Set

The Dash 3000/4000 Patient Monitor is presented as a physiological monitoring device, not a machine learning or AI-driven diagnostic algorithm that typically relies on a distinct "training set." Therefore, the concept of a "training set sample size" as understood in AI/ML contexts does not apply to the information provided in this 510(k) summary. The device's functionality is based on established signal processing and measurement techniques, not on a data-driven learning model requiring traditional training data.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" is not applicable to the type of device and information provided in this 510(k) summary. Therefore, how ground truth for a training set was established is not relevant or described.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

Date:October 9, 2003
Submitter:GE Medical Systems Information Technologies8200 West Tower AvenueMilwaukee, WI 53223 USA
Contact Person:Melissa RobinsonRegulatory Affairs SpecialistGE Medical Systems Information Technologies8200 West Tower AvenueMilwaukee, WI 53223 USAPhone: 813-887-2133Fax: 813-887-2545
Device:Trade Name:Dash 3000/4000 Patient Monitor
Common/Usual Name:Physiological Patient Monitor
Classification Names:21 CFR 868.1400 Analyzer, Gas, Carbon Dioxide, Gaseous-Phase21 CFR 868.2375 Breathing Frequency Monitor21 CFR 870.1025 Detector and Alarm, Arrhythmia21 CFR 870.1100 Monitor, Blood Pressure, Indwelling21 CFR 870.1130 Noninvasive Blood Pressure Measurement System21 CFR 870.1100 Blood Pressure Alarm21 CFR 870.1425 Programmable Diagnostic Computer21 CFR 870.2340 Electrocardiograph21 CFR 870.1435 Monitor, Cardiac Output, Thermal (Balloon Type Catheter)21 CFR 880.2910 Monitor, Temperature (with probe)21 CFR 870.2300 Monitor, Cardiac (Incl. Cardiotachometer & rate alarm)21 CFR 870.2700 Oximeter, PulseCCKFLSDSICAABXDDSJDQKFYWKFNBWXDRTDQAMHX
21 CFR 870.1025 Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)21 CFR 870.2770 Plethysmograph, ImpedanceDSB
Predicate Device:K030431 Dash 3000/4000 Patient Monitor

The Dash 3000/4000 Patient Monitor is a device that is designed to be Device Description: used to monitor, display, and print a patient's basic physiological parameters including: electrocardiography (ECG), invasive blood pressure, non-invasive blood pressure, oxygen saturation, temperature, impedance respiration, end-tidal carbon dioxide, oxygen, nitrous oxide and anesthetic agents. Other features include arrhythmia, cardiac output, cardiac and pulmonary calculations, dose calculations, PA wedge, ST analysis, and interpretive 12 lead ECG analysis (12SL). Additionally, the network interface allows for the display and transfer of network available patient data.

Intended Use:

The Dash 3000/4000 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash is designed as a bedside, portable, and transport monitor that can operate in all professional medical facilities and medical transport modes including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, freestanding

GE Medical Systems Information Technologies Confidential

{1}------------------------------------------------

surgical centers, and other alternate care facilities, intra-hospital patient transport, inter-hospital patient transport via ground vehicles (i.e., ambulance, etc.) and fixed and rotary winged aircraft, and pre-hospital emergency response.

Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, noninvasive blood pressure, pulse. temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, and anesthetic agents as summarized in the operator's manual.

The Dash 3000/4000 Patient Monitor is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital data at the point-of-care.

This information can be displayed, trended, stored, and printed.

The Dash 3000/4000 Patient Monitor was developed to interface with non-proprietary third party peripheral devices that support serial data outputs.

The Dash 3000/4000 Patient Monitor employs the same functional Technology: technology as the predicate devices.

  • Test Summary: The Dash 3000/4000 Patient Monitor complies with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the system:
    • Requirements specification review .
    • Code inspections .
    • Software and hardware testing .
    • Safety testing .
    • Environmental testing .
    • . Final validation

Conclusion:

The results of these measurements demonstrated that the Dash 3000/4000 Patient Monitor is as safe, as effective, and performs as well as the predicate devices.

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 2 2003

GE Medical Systems Information Technologies c/o Ms. Melissa Robinson Regulatory Affairs Specialist 8200 West Tower Avenue Milwaukee, WI 53223

Re: K033304

Trade Name: Dash 3000/4000 Patient Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: MHX Dated: October 9, 2003 Received: October 14, 2003

Dear Ms. Robinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Melissa Robinson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Aepasmar

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

Device Name: Dash 3000/4000 Patient Monitor

Indications For Use:

The Dash 3000/4000 Patient Monitor is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use of the system is to monitor physiologic parameter data on adult, pediatric and neonatal patients. The Dash is designed as a bedside, portable, and transport monitor that can operate in all professional medical, facilities and medical transport modes including but not limited to: emergency department, operating room, post anesthesia recovery, critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care areas located in hospitals, outpatient clinics, freestanding surgical centers, and other alternate care facilities, intra-hospital patient transport, interhospital patient transport via ground vehicles (i.e., ambulance, etc.) and fixed and rotary winged aircraft, and pre-hospital emergency response.

Physiologic data includes but is not restricted to: electrocardiogram, invasive blood pressure, noninyasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, carbon dioxide, oxygen, and anesthetic agents as summarized in the operator's manual.

The Dash 3000/4000 Patient Monitor is also intended to provide physiologic data over the Unity network to clinical information systems and allow the user to access hospital data at the point-of-care.

This information can be displayed, trended, stored, and printed.

The Dash 3000/4000 Patient Monitor was developed to interface with non-proprietary third party peripheral devices that support serial data outputs.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Cmman frBDt

(Division Sign-Off)
Division of Cardiovascular Devices

K033304

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.