Search Filters

Search Results

Found 262 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Unimed Reusable SpO2 Sensors (-117 and -121 Series) (U403-117); Unimed Reusable SpO2 Sensors (-117 and
    -121 Series) (U103-117); Unimed Reusable SpO2 Sensors (-117 and -121 Series) (U410-121); Unimed Reusable

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Unimed Reusable SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only.

    Device Description

    The subject devices are Unimed Reusable SpO2 Sensors intended for non-invasive measurement of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) in clinical settings. The sensors are designed for compatibility with GE B40i, and are supplied non-sterile.

    Each sensor consists of a connector, a cable, and a reusable patient-contacting sensor element incorporating a light-emitting diode (LED) and photodetector (PD). The sensors are available in multiple configurations, including finger clip, wrap, and soft-tip types, to accommodate various patient needs and anatomical sites.

    The subject devices operate on the same principle and share similar design features, materials, and performance characteristics as the predicate device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Unimed Reusable SpO2 Sensors, based on the provided FDA 510(k) clearance letter:

    Acceptance Criteria and Device Performance

    CriteriaAcceptance CriteriaReported Device Performance
    Saturation Accuracy, No Motion (70-100%)±3%±3%
    Pulse Rate Accuracy, No Motion±2 bpm (30-250 bpm)±2 bpm (30-250 bpm)
    BiocompatibilityPass ISO 10993 cytotoxicity, skin irritation, and skin sensitivity testsPass ISO 10993 cytotoxicity, skin irritation, and skin sensitivity tests

    Study Information

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Twelve human adult volunteers were enrolled for the clinical study. The study contains more than the minimum 200 data points.
      • Data Provenance: The study was conducted on human adult volunteers and includes sufficient darkly pigmented subjects (three dark subjects with Fitzpatrick Type 5-6). It is a prospective clinical study. The country of origin is not explicitly stated but implies testing in a controlled clinical environment, likely linked to the manufacturer's location or a designated clinical trial site.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document implies that arterial oxygen saturation (SaO2) as determined by co-oximetry was used as the ground truth. This is a direct measurement from blood samples. Therefore, typical "experts" in the sense of human readers adjudicating images are not applicable here. The accuracy of co-oximetry itself is the standard.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. The ground truth (SaO2 by co-oximetry) is a direct, objective measurement, not subject to human interpretation or adjudication in the same way as an imaging study.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This device is a sensor (hardware) for SpO2 and pulse rate measurement, not an AI-assisted diagnostic tool or an imaging system that would involve human reader interpretation. No MRMC study was conducted.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. The device is a sensor that measures physiological parameters. Its performance is inherent to its design and function, not an algorithm's performance. The "clinical test data" section describes the validation of the sensor's accuracy in vivo.
    6. The type of ground truth used:

      • Arterial oxygen saturation (SaO2) as determined by co-oximetry. This is an objective "gold standard" for blood oxygen measurement.
    7. The sample size for the training set:

      • Not applicable. This device is a hardware sensor, not a machine learning model that requires a training set. The clinical study described is for validation/testing, not training.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243000
    Date Cleared
    2025-06-20

    (267 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Reusable Temperature Probe (T1306, T2306, T3306, T4306); Disposable Temperature Probe (T5106, T6106)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Temperature Probes are intended to be used for monitoring temperature. The temperature probes are reusable or for single patient use and designed for use with Mindray monitor Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.
    These devices are used by qualified medical professional only.

    Device Description

    The subject devices are used for patient temperature measurement. The probes are reusable or disposable depending on models. These probes consist of a connector on the monitor end and a thermistor on the patient end. The working principle is resistance based on the metal conductor increases with temperature decrease, and the linear changes to the characteristics of the temperature measurement. The subject devices are designed to be used in healthcare facilities like hospital and compatible with a monitor of Mindray Model PM-8000 and other monitors compatible with YSI 400 series temperature probes.

    The six models have two types of structure designs corresponding to reusable and disposable use which consists of different materials. The NTC of the six models are identical. Reusable models T1306, T2306, T3306, T4306 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials. Disposable models T5106 and T6106 have a similar structure design with two different sensor shapes for different measure sites and consist of the same materials.

    Model: T1306, Description: Skin contact Temperature Probe, adult, reusable
    Model: T2306, Description: Body cavity Temperature Probe, Esophageal/Rectal, adult, reusable
    Model: T3306, Description: Skin contact Temperature Probe, pediatric, reusable
    Model: T4306, Description: Body cavity Temperature Probe, Esophageal/Rectal, pediatric, reusable
    Model: T5106, Description: Skin contact Temperature Probe, adult/ pediatric, disposable
    Model: T6106, Description: Body cavity Temperature Probe, Esophageal/Rectal, adult/ pediatric, disposable

    AI/ML Overview

    The provided FDA 510(k) clearance letter describes a medical device, the Reusable and Disposable Temperature Probes, but does not include information about AI/ML performance. Therefore, I will respond to the prompt by extracting the acceptance criteria and study information pertinent to this medical device, which focuses on traditional medical device performance rather than AI/ML.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device PerformanceStudy Supporting Performance
    Accuracy±0.1℃Bench Testing (ISO 80601-2-56)
    Measurement Range25-45℃Bench Testing (ISO 80601-2-56)
    Electrical SafetyComplies with IEC 60601-1Bench Testing (IEC 60601-1)
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2Bench Testing (IEC 60601-1-2)
    BiocompatibilityComplies with ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23Biocompatibility testing
    Operating Environment+5 to +40°C, ≤80% humidity (non-condensing), 86kPa~106kPaBench Testing (IEC 60601-1 and ISO 80601-2-56)
    Storage Environment-20℃ to 55℃, ≤93% humidity, 86kPa~106kPaBench Testing (IEC 60601-1 and ISO 80601-2-56)
    Compatibility with MonitorsVerifies compatibility with Mindray Model PM-8000 and other YSI 400 series compatible monitorsBench Testing

    Note: The document presents "Accuracy" and "Measurement Range" as inherent characteristics of the device and states that bench testing was conducted to verify that design specifications were met, which implies these values are the acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for the test sets (e.g., number of units tested, number of temperature measurements, or specific test configurations) for the bench testing or biocompatibility testing.

    The document also does not provide information about the provenance of data in terms of country of origin or whether studies were retrospective or prospective. The testing described appears to be laboratory-based verification and validation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable and not provided in the document. The device is a clinical electronic thermometer, and its performance is assessed against technical specifications and international standards, not against human expert interpretation of medical images or data. Ground truth for temperature measurement is typically established by reference standards or calibrated equipment.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided in the document. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective assessment (e.g., image interpretation by multiple readers), which is not relevant for the objective performance testing of a temperature probe.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. An MRMC comparative effectiveness study was not conducted as this is a medical device for objective temperature measurement, not an AI-assisted diagnostic tool requiring human-in-the-loop performance evaluation. The document does not mention any AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No. This device is an electronic temperature probe, not an algorithm or software. Its performance is inherent to its physical and electrical properties, evaluated through bench testing against established standards.

    7. Type of Ground Truth Used

    The ground truth for the performance evaluations (accuracy, measurement range, electrical safety, etc.) would be established by:

    • Reference Standards/Calibrated Equipment: For accuracy and measurement range, the device's readings would be compared against highly accurate and calibrated reference thermometers in controlled temperature environments.
    • International Standards: Compliance with electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and thermometer-specific performance (ISO 80601-2-56) serves as the ground truth for safety and performance.
    • Laboratory Analysis: For biocompatibility, laboratory tests (cytotoxicity, sensitization, irritation) are conducted to assess the biological response to the device materials according to ISO 10993 standards.

    8. Sample Size for the Training Set

    This information is not applicable and not provided. This device is a hardware medical device with no mention of machine learning or algorithms that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As there is no training set for an AI/ML algorithm involved, no ground truth was established for a training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Unimed Reusable Finger Clip SpO2 Sensors (UXXX-91 Series) (U403-91); Unimed Reusable Finger Clip SpO2

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Models: U403-91, U410-91 Unimed Reusable Finger Clip SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg. These devices are for prescription use only.

    Device Description

    The subject devices, Unimed Reusable Finger Clip SpO2 Sensors (UXXX-91 Series), are fully compatible reusable sensors for use with Philips Itellivue MP30. These sensors are supplied non-sterile. The subject sensors (Models: U403-91 and U410-91) consist of a plug/connector (Philips D8 sensor connector), a cable, and a patient-contacting (finger clip) where light-emitting diode (LED) and photodetector (PD) are located. The subject sensors share the same principle of operation as the predicate device for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) measurements. Patient-contacting materials of the subject sensors include ABS, silicone, and TPU, all of which have been qualified against biocompatibility requirements in ISO 10993-5, ISO 10993-10, and ISO 10993-23.

    AI/ML Overview

    The Unimed Reusable Finger Clip SpO2 Sensors (UXXX-91 Series) are subject to the following acceptance criteria and the performance study described was conducted to prove the device meets these criteria. It's important to note that the provided document is a 510(k) summary, which is a premarket notification to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily a full clinical trial report for novel device approval.

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature / Acceptance CriteriaSubject Device (Model: U410-91, U403-91) Reported PerformancePredicate Device (K062605) Performance
    Performance (Arms)
    Saturation Accuracy (SpO2), No Motion (70-100%)±3% (70-100%)±3% (70-100%)
    Pulse Rate Accuracy (PR), No Motion (30-250 bpm)±3 bpm (30-250 bpm)±2% or 1 bpm, whichever is greater (30 to 300 bpm)
    Low Perfusion Accuracy (SpO2, 70-100%)SpO2 ±3% (70-100%)SpO2 ±3% (70-100%)
    Low Perfusion Accuracy (Pulse, 30-250 bpm)Pulse ±3 bpm (30-250 bpm)Pulse ±2% or 1 bpm, whichever is greater (30 to 300 bpm)

    Note¹: The document states that according to ISO 80601-2-61:2017, both Subject Device and Predicate Device have validated accuracy over the claimed pulse range and are therefore considered substantially equivalent, despite the difference in stated accuracy metrics for pulse rate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size:
      • Clinical Study Test Set: 12 human adult volunteers.
      • Data Points: The clinical study contains "more than the minimum 200 data points."
    • Data Provenance:
      • The document implies the clinical study was conducted by or on behalf of Unimed Medical Supplies Inc., which is based in China. The data provenance regarding the country of origin of the study conduct is therefore likely China or an authorized testing facility.
      • The study involved clinical hypoxia tests, which are prospective in nature, as they involve actively inducing hypoxia in volunteers to collect data on sensor performance.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not explicitly state the number of experts or their specific qualifications for establishing the ground truth. However, for pulse oximetry accuracy studies involving induced hypoxia, the ground truth (arterial oxygen saturation, SaO2) is typically established by co-oximetry readings from arterial blood samples, which are analyzed by laboratory equipment. This process relies on the accuracy of the co-oximeter and the proper drawing and handling of blood samples by trained medical personnel. It doesn't involve subjective expert consensus in the same way an imaging study would.

    4. Adjudication Method for the Test Set

    Not applicable. For pulse oximetry accuracy studies using induced hypoxia and co-oximetry, the ground truth is an objective physiological measurement from a laboratory instrument (co-oximeter) from arterial blood samples. There is no human adjudication process involved in establishing this ground truth, unlike with subjective visual assessments in imaging.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. The provided document details a clinical study to validate the device's standalone accuracy against established physiological measurements (co-oximetry) in human volunteers. It is not an MRMC study comparing human reader performance with and without AI assistance, as the device is a sensor for measuring physiological parameters, not an AI-powered diagnostic imaging tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Yes. The clinical hypoxia test described is a standalone performance test for the Unimed Reusable Finger Clip SpO2 Sensors. The study directly assesses the device's ability to accurately measure SpO2 against the ground truth (SaO2 from co-oximetry) without human interpretation or intervention in the measurement process itself.

    7. The Type of Ground Truth Used

    The ground truth used was outcomes data / physiological measurement, specifically:

    • Arterial oxygen saturation (SaO2) as determined by co-oximetry. This is considered the gold standard for measuring oxygen saturation in blood for pulse oximeter accuracy studies.

    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" sample size. For medical device premarket notifications like this, the focus is on a verification/validation study (the "clinical study" mentioned here) rather than an AI/machine learning model where distinct training and test sets are common. The device is a sensor, not a machine learning algorithm that undergoes a training phase.

    9. How the Ground Truth for the Training Set was Established

    As this is a sensor device and not an AI/ML model, there isn't a "training set" in the typical sense for which ground truth would be established for model training. The "ground truth" (SaO2 via co-oximetry) was established for the 12-subject clinical validation study as described above.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Unimed Reusable Finger Clip SpO2 Sensors (U103-125); Unimed Reusable Finger Clip SpO2 Sensors (U403-254
    ); Unimed Reusable Finger Clip SpO2 Sensors (U103-254); Unimed Reusable Finger Clip SpO2 Sensors (U403

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Unimed Reusable Finger Clip SpO2 Sensors are indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for adult patients weighing greater than 30 kg and pediatric patients weighing 10-50 kg. These devices are for prescription use only.

    Device Description

    The subject devices, Unimed Reusable Finger Clip SpO2 Sensors, are fully compatible reusable sensors for use with monitors that incorporate Masimo technology. These sensors are supplied non-sterile. The subject sensors consist of a plug/connector (U403-125 and U103-125 with M8 sensor connector, and U403-254 and U103-254 with M10 sensor connector), a cable, and a patient-contacting (finger clip) where light-emitting diode (LED) and photodetector (PD) are located. The subject sensors share the same principle of operation as the predicate device for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) measurements. Also, the subject sensors have identical material composition and performance characteristics to the predicate device.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Unimed Reusable Finger Clip SpO2 Sensors, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (Predicate Device)Reported Device Performance (Subject Device)Comparison
    Saturation Accuracy, No Motion (70-100%)± 3%± 3%Identical
    Pulse Rate Accuracy, No Motion (30-250 bpm)± 3 bpm (30-250 bpm)± 3 bpm (30-250 bpm)Identical

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document states that the clinical study was conducted with "human adult volunteers," including "sufficient darkly pigmented subjects." However, the exact number of subjects (the sample size) is not specified.
    • Data Provenance: The data is prospective as it was collected in a "clinical study... under an approved protocol with subject informed consent." The country of origin is not explicitly stated, but the submission is from "Unimed Medical Supplies Inc." located in "Shenzhen, China." It's reasonable to infer the study was conducted there or arranged by their organization.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document mentions that "arterial oxygen saturation (SaO2) as determined by co-oximetry" was used to establish the ground truth. This is a direct physiological measurement and does not involve human expert interpretation for ground truth establishment. Therefore, information about the number or qualifications of experts for ground truth is not applicable in this context.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the ground truth was established by co-oximetry, which is an objective measurement, not a subjective interpretation requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is an oximeter sensor, which directly measures physiological parameters (SpO2 and pulse rate). Its performance is evaluated against a gold standard method (co-oximetry), not by comparing human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the clinical study performed appears to be a standalone (algorithm only) performance evaluation. The device, an SpO2 sensor, directly provides measurements. The study evaluated the accuracy of these measurements against a reference method (co-oximetry) without integrating a human in the loop for interpreting the device's output.

    7. Type of Ground Truth Used

    The ground truth used was outcomes data (physiological measurements), specifically "arterial oxygen saturation (SaO2) as determined by co-oximetry." Co-oximetry is considered a clinical gold standard for measuring blood oxygen levels.

    8. Sample Size for the Training Set

    Not applicable. The document describes a medical device (SpO2 sensor), not an AI/machine learning algorithm that requires a training set. The device operates based on physical principles (two-wavelength relative optical absorption) and is validated through clinical performance against a reference standard.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As the device is not an AI/machine learning algorithm, there is no training set and therefore no ground truth established for a training set.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    Philips Reusable ECG Lead Sets and Trunk Cables : Model Number,Model Name: 989803170171, OR 3-Lead ECG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Philips IntelliVue reusable ECG lead sets/trunk cables and Philips OR ECG trunk cables are indicated for continuous monitoring of cardiac signals for both diagnostic and monitoring purposes. These devices are limited by the indications for use of the connected monitoring and diagnostic equipment in healthcare facilities. These devices are intended to interact with patient intact skin only.

    Device Description

    Philips Reusable ECG Lead Sets and Trunk Cables are non-invasive reusable medical devices designed to transfer ECG signals from the patient to the monitor used in the healthcare facilities, which includes intensive care unit (ICU) for both adults and neonates and the operating room (OR). The ECG cables and lead sets are used to forward an electrical cardiac signal via electrical wires from the electrode attached on patient skin to the ECG measurement hardware in the patient monitor and to support other outputs as needed. The portfolio consists of two types of trunk cables and two types of lead sets: ICU and OR, including various electrode connectors with AAMI and IEC coloring. The ECG Trunk cables and lead sets are suitable for multiple patient use. The Philips Reusable ECG Lead Sets and Trunk Cables have a service life of 18 months. The Philips Reusable ECG Lead Sets and Trunk Cables can be used with any Philips monitors including Philips monitor/defibrillator product lines and Philips monitoring or diagnostic equipment which claim compatibility to Philips Reusable ECG Lead Sets and Trunk Cables.

    AI/ML Overview

    This document is a 510(k) premarket notification for Philips Reusable ECG Lead Sets and Trunk Cables. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a standalone study with clinical accuracy metrics. Therefore, many of the requested details about acceptance criteria, clinical study design, and ground truth are not applicable or explicitly stated in this type of submission.

    However, I can extract the information that is present regarding performance testing and standards.

    1. Table of Acceptance Criteria and Reported Device Performance

    Instead of specific quantitative clinical acceptance criteria and device performance metrics, this submission relies on compliance with recognized harmonized standards. The "performance" is demonstrated through successful completion of tests outlined in these standards.

    Acceptance Criteria (Compliance with Standard)Reported Device Performance
    IEC 60601-1:2005+A1:2012+A2:2020 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    ANSI AAMI EC53:2013/(R)2020 (ECG trunk cables and patient lead wires)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    IEC 60601-2-25:2011/(R)2016 (Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    IEC 60601-2-27:2011(R)2016 (Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic equipment)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    IEC 60601-1-12 Edition 1.1 2020-07 CONSOLIDATED VERSION (Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard. (Specifically, ICU Lead Sets and Trunk cables comply with this additional standard).
    IEC 62366-1:2015+AMD1:2020: (Part 1: Application of usability engineering to medical devices)The subject devices have passed all safety tests for demonstrated compliance with this recognized standard.
    Reprocessing Standards: ISO 17664-2:2021, AAMI TIR12 (2020)/(R)2023, and ANSI AAMI ST98:2022, and FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling"The subject device adheres to these standards, and its proposed reprocessing instructions are as effective as the predicate device, meeting regulatory requirements for reusable devices.
    Biocompatibility: (ISO 10993-1, as inferred from comparison to predicate)Both the subject and predicate devices meet standards for biocompatibility.
    Electrical Conductivity, Mechanical Strength: (as inferred from comparison to predicate)Both the subject and predicate devices meet standards for electrical conductivity and mechanical strength.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
    This document explicitly states: "The subject devices, like the primary predicate devices, did not require clinical trials." Therefore, there is no "test set" in the context of a clinical study for external validation of performance. Performance is primarily demonstrated through compliance with recognized standards via non-clinical bench testing. The data provenance would be from internal lab testing, not patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
    Not applicable. No clinical test set requiring expert ground truth was used as per the statement above.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
    Not applicable. No clinical test set requiring expert adjudication was used.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This device is an ECG cable and lead set, a hardware accessory. It is not an AI algorithm for interpreting medical images or signals, nor does it involve "human readers" or "AI assistance" in that sense.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
    Not applicable. This is a hardware accessory, not an algorithm. Performance is assessed through its physical and electrical properties, not algorithmic output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
    Not applicable in the sense of clinical ground truth. For non-clinical bench testing, the "ground truth" would be established by the specifications and measurement techniques defined within the harmonized standards (e.g., electrical resistance, mechanical durability measured against stated tolerances, biocompatibility testing results).

    8. The sample size for the training set
    Not applicable. This is a hardware accessory, not a machine learning model, so there is no training set.

    9. How the ground truth for the training set was established
    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242623
    Date Cleared
    2024-10-31

    (58 days)

    Product Code
    Regulation Number
    870.1120
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Blood Pressure cuff; Reusable Blood Pressure cuff

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

    The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate size. The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. It is available in neonatal, infant, child and adult sizes. The subject device includes disposable blood pressure cuff and reusable blood pressure cuff and contains Cuff with/without bladder and single/dual tubes Air Hose.

    AI/ML Overview

    The provided document is a 510(k) summary for a blood pressure cuff and does not contain detailed information about acceptance criteria, reported device performance for specific metrics, or any study proving such performance. The document explicitly states:

    "No clinical study is included in this submission"

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets them because such information is not present in the provided text. The document focuses on demonstrating substantial equivalence to a predicate device primarily through non-clinical data and compliance with voluntary standards for design specifications.

    However, I can extract the information that is present:

    8. Non-clinical data

    • Performance testing: Non-clinical tests were conducted to verify that the proposed devices met all design specifications.
    • Compliance with standards: The test results demonstrated that the proposed device complies with ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007.

    Based on this, here's what can be stated about the available information:

    1. A table of acceptance criteria and the reported device performance: Not provided. The document mentions compliance with ISO 81060-1, which would contain performance requirements for non-automated sphygmomanometers (blood pressure cuffs). However, the specific acceptance criteria used for the device and the reported device performance against those criteria are not detailed.
    2. Sample sized used for the test set and the data provenance: Not provided. The document only mentions "non-clinical tests" and "performance testing" but does not give details about sample sizes or data provenance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical study or expert-based ground truth establishment is mentioned. The testing focused on technical specifications and standards compliance.
    4. Adjudication method for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. The document explicitly states "No clinical study is included in this submission."
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a blood pressure cuff, which is a physical accessory, not an algorithm.
    7. The type of ground truth used: Not applicable for performance evaluation as no clinical data or comparative performance study is provided. For demonstrating substantial equivalence, the ground truth indirectly relies on the established performance and safety of the predicate device and the standards.
    8. The sample size for the training set: Not applicable. This is a non-AI/ML device.
    9. How the ground truth for the training set was established: Not applicable. This is a non-AI/ML device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K234132
    Date Cleared
    2024-09-17

    (264 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Medline Reusable Sterilization Wrappers

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Medline Reusable Sterilization Wrappers are intended to be used to enclose another medical device or component that is to be sterilized by health care professionals. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used. The Medline Reusable Sterilization Wrap is a non-sterile, reusable device, with a wash/dry/autoclave life cycle as indicated on the Wrap.

    Medline Reusable Sterilization Wrappers are available in 3 different fabrics that are either single or double ply bonded sheets of cotton and polyester blends. Model Angelstat is reusable up to 50 times; Models Resistat and Ripstop are reusable up to 75 times . The Medline Reusable Sterilization Wrappers are available in Misty Green, Jade Green or Ceil Blue with configurations 12"x12" through 72"x72" and 60"x90". The Medline Reusable Sterilization Wrappers are intended to be laundered then sterilized prior to use, and examined prior to use for defects and debris.

    A prevacuum steam sterilization cycle at 270 F for 4 minutes has been validated for use on textile packs, with a minimum weight of 25 pounds and 30 minute dry time.

    Medline recommends following the below standards for pack configuration and assembly in sterilization facilities. · AAMI ANSI ST79 • AAMI ANSI ST65

    Device Description

    The Medline Reusable Sterilization Wrappers are offered in bulk, non-sterile packaging. The Medline Reusable Sterilization Wrappers are available in 3 different fabrics that are either single or double ply bonded sheets of cotton and polyester blends. The Medline Reusable Sterilization Wrappers are available in Misty Green, Jade Green or Ceil Blue with configurations 12"x12" through 72"x72" and 60"x90". The Medline Reusable Sterilization Wrappers are intended to be sterilized and laundered per the instructions.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of non-clinical testing for the Medline Reusable Sterilization Wrappers.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test PerformedTest MethodAcceptance CriteriaReported Device Performance
    Resistance to Water PenetrationAATCC 127-2018Test and Report - Average hydrostatic pressurePass
    Tensile Strength and ElongationASTM D5034-21Test and Report - Average load at break and elongation at break for machine and cross directionPass
    Particle Generation (Linting)ISO 9073-10:2004Test and Report - Average lint counts of particles and coefficient of lintingPass
    CytotoxicityISO 10993-5Non-Cytotoxicity - device did not show cytotoxicity potentialPass
    SensitizationISO 10993-10Non-sensitizing - device showed no significant evidence of causing delayed dermal contact sensitizationPass
    IrritationISO 10993-10Non-irritating - the irritation response category of the device was classified as NegligiblePass
    Acute Systemic ToxicityISO 10993-11Non-systemic toxicity – device did not show systemic toxicity potentialPass
    Sterilization ValidationST79:2017All biological indicators shall be negative, positive controls shall be positive, chemical integrators shall demonstrate adequate steam penetration, negative and environmental controls shall be negativePass
    Dry Time ValidationST79:2017Average pre and post sterilization wrought difference of less than 0.2% within 5 minutes of cycle completion, no visible moisture following 30 minute cooling period and all integrators shall demonstrate steam penetrationPass
    Cleaning ValidationAAMI TIR30:2011Each test sample shall show a TOC level of less than 12µg/cm2 and no visible soil after cleaningPass

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the specific sample sizes for each test listed. It mentions "Each test sample" for Cleaning Validation, implying multiple samples were tested. For other tests like Resistance to Water Penetration, Tensile Strength, and Particle Generation, it refers to "Test and Report" without specifying quantity.
    The data provenance is not specified. It can be inferred that these tests were conducted by the manufacturer, Medline Industries, LP, to support their 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The listed tests are non-clinical bench tests and sterilization validations, which typically rely on standardized methods and laboratory analysis rather than expert human interpretation for "ground truth" establishment in the same way, for example, medical image analysis would.

    4. Adjudication method for the test set

    Not applicable. The tests are objective, laboratory-based physical, chemical, and biological evaluations, not requiring human adjudication of results in the traditional sense.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical accessory (sterilization wrapper) and does not involve AI or human "readers" for interpretation. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable, as this device does not involve an algorithm or AI component.

    7. The type of ground truth used

    The ground truth for these tests is established by:

    • Standardized Test Methods and Criteria: For physical and material properties (e.g., AATCC 127-2018 for water penetration, ASTM D5034-21 for tensile strength, ISO 9073-10:2004 for particle generation).
    • Biological Test Outcomes: For biocompatibility tests (Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity), the "ground truth" is determined by the biological response observed according to the protocols of ISO 10993 series.
    • Sterilization Efficacy Indicators: For Sterilization Validation and Dry Time Validation, the "ground truth" is determined by the results of biological indicators, chemical integrators, and physical observations (e.g., moisture, weight difference) as per ST79:2017.
    • Chemical Analysis: For Cleaning Validation, the "ground truth" is based on measurable chemical residuals (TOC level) and visual inspection for soil, guided by AAMI TIR30:2011.

    8. The sample size for the training set

    Not applicable. This device does not use an AI/ML component, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242061
    Manufacturer
    Date Cleared
    2024-08-15

    (31 days)

    Product Code
    Regulation Number
    876.4300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Reusable Interface Cable 1.5 m (PRD-IFC-002)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Interface Cable is for connection only of a compatible Creo Medical Instrument (Surgical Accessory) to the CROMA Electrosurgical Generator to deliver Radiofrequency (RF) and/or Microwave energy for the cutting, coagulation and ablation of gastrointestinal tissue.

    Device Description

    The Reusable Interface Cable comprises a flexible, low-loss coaxial cable intended for high-power microwave operation and has latching connectors at each end for connection to the CROMA Electrosurgical Generator and compatible Creo Medical Instruments (Surgical Accessories). The established lifetime of the Reusable Interface Cable is for 20 uses in total.

    AI/ML Overview

    The provided text is related to a 510(k) premarket notification for a medical device (Reusable Interface Cable 1.5 m), which is an electrosurgical accessory. The purpose of this notification is to demonstrate substantial equivalence to a legally marketed predicate device, not necessarily to provide detailed acceptance criteria and study results for clinical efficacy or sophisticated AI performance.

    Therefore, the document does not contain information relevant to:

    • A table of acceptance criteria and reported device performance related to a diagnostic or AI-driven outcome.
    • Sample size used for a test set in the context of an accuracy study or provenance of data.
    • Number of experts for ground truth establishment or their qualifications.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies or effect sizes of AI assistance.
    • Standalone algorithm performance.
    • Type of ground truth (expert consensus, pathology, outcomes data).
    • Sample size for training set.
    • How ground truth for the training set was established.

    Instead, the document focuses on:

    • Device Description: A flexible, low-loss coaxial cable connecting a Creo Medical Instrument to a CROMA Electrosurgical Generator for delivering RF and/or Microwave energy. It specifies a lifetime of 20 uses.
    • Intended Use/Indications for Use: Connection of compatible Creo Medical Instrument to the CROMA Electrosurgical Generator to deliver RF/Microwave energy for cutting, coagulation, and ablation of gastrointestinal tissue. This indication is stated to be the same as the predicate device, with reworded readability.
    • Technological Comparison: The subject device is a reusable version of the predicate device. The conductor cable was modified for durability while maintaining identical electrical properties. Principles of operation, electrical properties, energy source, materials, dimensions, and cleaning methods are identical to the predicate.
    • Non-Clinical Tests:
      • Electrical safety testing: Conducted in compliance with FDA-recognized versions of IEC 60601-2-2 and IEC 60601-2-6 standards.
      • Performance bench testing: Conducted to establish device durability over its established lifetime for:
        • Cleaning between uses (reprocessing)
        • Rotation load
        • Cyclic load
        • Connector connection and disconnection forces.

    Conclusion:

    This submission is for an accessory device (cable) that facilitates the use of an electrosurgical unit. The "acceptance criteria" and "study" described are primarily related to electrical safety and physical durability/reusability of the cable, ensuring it functions as intended for its specified lifespan and remains substantially equivalent to the predicate device. It does not involve complex performance metrics or clinical studies in the context of diagnostic accuracy or AI-driven decision support.

    Since the provided text does not contain the requested information about diagnostic performance, AI, or specific clinical study details as implied by the questions, I cannot populate the table or answer those specific points. The information available is about engineering and regulatory compliance testing for a reusable medical cable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240898
    Manufacturer
    Date Cleared
    2024-07-31

    (121 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    SteriTite rigid reusable sterilization container with MediTray Products

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K232060
    Manufacturer
    Date Cleared
    2024-04-12

    (276 days)

    Product Code
    Regulation Number
    884.5470
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    TOSAMA Reusable Applicator

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TOSAMA Reusable Applicator is intended to be used to insert a digital (no applicator) menstrual tampon into the vagina.

    Device Description

    The TOSAMA Reusable Applicator is a non-sterile, single user, reusable medical device. It is intended to be sold in three different packaging configurations: Applicator with prepackaged tampons (regular, super and super plus absorbency versions) and Applicator with loose tampons of super and regular absorbency. The device is comprised of a tampon holder and pusher made of thermoplastic elastomer and a lipstick-style cap to protect the applicator, which is made of polystyrene. The tampon holder and pusher are connected at the hinged base of the holder, which opens to load a tampon and closes to allow the pusher to eject a tampon. The holder and pusher are designed to stay connected and not separate during use, storage, or cleaning. The TOSAMA Reusable Applicator requires the user to load the applicator with a legally marketed menstrual tampon. The use-life of the subject TOSAMA Reusable Applicator is 2 years. The device requires the user to clean (holder and pusher) and disinfect (lipstick cap) before initial use, before long term storage and at the end of the menstrual cycle, as well as to clean the device after each use. The applicator is available in one size for all sizes of tampon absorbencies (Light, Regular, Super and Super Plus) and accommodates a maximal tampon diameter of 15 mm. The holder inner diameter is 8.2 + 0.25/- 0.10 mm, pusher length is 83.7 ±0.5 mm and total length and diameter of the applicator (without the lipstick cap) is 70.8 ± 1.0 mm and 19.6 ±0.05 mm respectively.

    AI/ML Overview

    The provided text describes the regulatory clearance of the "TOSAMA Reusable Applicator" (K232060) and includes some information about its testing. However, it does not contain a detailed study report that proves the device meets specific acceptance criteria in the format requested.

    The text generally states that "The results of the non-clinical performance data were acceptable and met the established acceptance criteria." but does not provide the specific "acceptance criteria" themselves or the detailed "reported device performance" against those criteria in a table. It also lacks significant details regarding the study design, sample sizes for test sets, data provenance, ground truth establishment, or multi-reader studies as typically found in comprehensive medical device performance reports.

    Based on the provided text, here's what can be extracted and what information is missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Inferred/General)Reported Device Performance (General Statement)
    Non-cytotoxicThe subject device was non-cytotoxic.
    Non-sensitizingThe subject device was a non-sensitizer.
    Non-irritating (vaginal)The subject device was non-irritating.
    Met applicable design and performance requirements for device weightAcceptable and met established acceptance criteria.
    Met applicable design and performance requirements for physical dimensionsAcceptable and met established acceptance criteria.
    Met applicable design and performance requirements for functional evaluation (e.g., ejection force, plunger pull force, pusher stability, hinge durability, lipstick cap compatibility)Acceptable and met established acceptance criteria.
    Met applicable design and performance requirements for use-life testingAcceptable and met established acceptance criteria.
    Met reprocessing, cleaning, and disinfecting requirementsCleaning instructions developed per FDA guidance; validated processes for washing, boiling, and alcohol disinfection.

    Missing specific acceptance values and corresponding performance results.


    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified for any of the tests (biocompatibility, bench testing, reprocessing).
    • Data Provenance: Not specified. It's a regulatory submission from a Slovenian company (Tosama, d.o.o., Domžale, Slovenia). The testing would likely have been conducted by or for this company. Whether the data is retrospective or prospective is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This type of study (e.g., biocompatibility and functional bench testing for a reusable tampon applicator) would not typically involve human experts establishing "ground truth" in the way an AI diagnostic device would. Instead, the ground truth is based on standardized biochemical assays (for biocompatibility) or engineering measurements against design specifications (for bench testing). Therefore, this question is not directly applicable in the context of this device's testing according to the provided information.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable for the types of tests described (biocompatibility, bench testing). Adjudication methods are typically relevant for studies involving human interpretation or clinical endpoints, not for objective laboratory or engineering tests.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant for AI-powered diagnostic devices where human readers (e.g., radiologists) use AI as an aid. The TOSAMA Reusable Applicator is a physical medical device (tampon applicator), not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm performance study was not done. This device is not an algorithm or AI. The testing described focuses on the physical and biological safety and performance of the applicator itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For Biocompatibility: The ground truth is based on established biological and toxicological standards and test methods (e.g., ISO 10993 series for cytotoxicity, sensitization, irritation) and their defined pass/fail criteria.
    • For Bench Testing: The ground truth is established by engineering design specifications and performance requirements (e.g., specific force values for ejection, dimensional tolerances, durability cycles).
    • For Reprocessing: The ground truth is based on micro-biological testing (e.g., reduction of bioburden) and visual inspection to ensure cleanliness and disinfection efficacy, validated against pre-defined criteria.

    8. The sample size for the training set

    • Not applicable. This device is not an AI model, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 27