Search Results
Found 4 results
510(k) Data Aggregation
(77 days)
SALTO TALARIS TOTAL ANKLE PROSTHESIS, RHS
Salto Talaris Total Ankle Prosthesis:
The Salto Talaris Total Ankle Prosthesis is indicated as a total ankle replacement in primary or revision surgery for patients with ankle joints damaged by severe rheumatoid, posttraumatic, or degenerative arthritis.
Components are intended for cemented use only.
RHS:
The RHS is intended for:
l) Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation and decreased motion at the radiohumeral and/or proximal radio-ulnar joint with:
Joint destruction and/or subluxation visible on x-ray a.
b. Resistance to conservative treatment
2) Primary replacement after fracture of the radial head
3) Symptomatic sequelae after radial head resection
4) Revision following failed radial head arthroplasty
The long stem is for single cemented use only. The short stem coated with titanium plasma-spray is for single use with or without cement.
For the Salto Talaris Total Ankle Prosthesis:
The goal of total ankle replacement is to restore function and relieve pain. The Salto Talaris Total Ankle Prosthesis is intended to accomplish these goals. The Salto Talaris Total Ankle Prosthesis is a semi-constrained anatomical design, which reproduces the kinematics of the ankle joint.
The Tornier Salto Talaris Total Ankle Prosthesis consists of two mating components: a metal tibial base in association with a conforming polyethylene articulating insert, and a metal talar resurfacing component.
For the RHS:
The RHS has been designed in order to provide surgeons and patients with a joint prosthesis to restore function and relieve pain of the radial part of the elbow joint. The RHS has an anatomical design, which reproduces the kinematics of the radial joint. The RHS consists of two components: a metal radial stem and a metal-polyethylene radial head.
The present submission corresponds to the following modification:
- Addition of a new coating subcontractor (titanium coating): Eurocoating S.p.A on cobalt chromium components.
All the prostheses of this application are strictly identical to the previously cleared devices except for the coating supplier. The indications for use of each device are not modified.
Here's an analysis of the provided text regarding the Salto Talaris Total Ankle Prosthesis and RHS, focusing on acceptance criteria and supporting studies.
Important Note: The provided text is a 510(k) summary for a modification to an existing device (specifically, a change in coating subcontractor). Therefore, the "study" described is not a clinical trial demonstrating efficacy of the entire device, but rather a non-clinical assessment to prove that the modification does not alter safety or effectiveness compared to the predicate device.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (from predicate device standards) | Reported Device Performance (with new coating subcontractor) |
---|---|
Coating specifications (e.g., thickness, pore size, pore volume) for titanium coating | Conforms to ASTM standard F1580. Coating characterization (thickness, pore size, pore volume) performed and found substantially equivalent to predicate coating. |
Shear fatigue strength | Non-clinical testing performed, results showed substantial equivalence to predicate. |
Static shear strength | Non-clinical testing performed, results showed substantial equivalence to predicate. |
Static tensile strength | Non-clinical testing performed, results showed substantial equivalence to predicate. |
Abrasion resistance | Non-clinical testing performed, results showed substantial equivalence to predicate. |
No new or higher risk introduced by the coating change | Concluded that the new subcontractor "does not induce any new or higher risk compared to the predicate coating subcontractor BioCoat Company." |
Design, materials, sizes, method of fixation, sterilization process remain identical | "The design, the material, the sizes, the method of fixation and the sterilization process are identical for both devices whatever the coating subcontractor is." |
Indications for use not modified | "The indications for use of each device are not modified." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated as a number of specific "samples" in a traditional sense. The testing was conducted on samples of the coating material and/or coated components. The extent of testing is implied by the mention of "characterization" and various mechanical tests.
- Data Provenance: The data is based on non-clinical testing of the coating and coated components. The specific origin country for the testing laboratory is not mentioned, but the submitting company (Tornier SAS) is based in France. The data is prospective in the sense that the tests were performed to evaluate the new subcontractor's coating.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not applicable in the context of this non-clinical submission. "Ground truth" for this submission would be defined by established engineering and materials standards (e.g., ISO, ASTM).
- Qualifications of Experts: Not applicable. The "experts" involved would be materials scientists and engineers performing the tests and comparing the results to established standards and the predicate device's performance.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. This was a non-clinical comparative analysis against a predicate's performance and established standards, rather than expert adjudication of clinical outcomes or images.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- Was it done?: No. This submission describes a non-clinical evaluation of a manufacturing change (coating subcontractor) for existing orthopedic implants. It is not a clinical study involving human readers or cases.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Was it done?: No. This is not an AI or algorithm-based device. It's a physical orthopedic implant.
7. The Type of Ground Truth Used
- Type of Ground Truth: The ground truth used for this submission is based on established engineering and materials standards (ISO standard 5834-2, ISO standard 5832-7 or ISO 5832-12, ASTM standard F1580), and the performance characteristics of the predicate device's coating (BioCoat Company). The goal was to prove the new coating was "substantially equivalent" and introduced no new risks.
8. The Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This is not an AI/ML device that requires a training set. The "training" for such a device would be the extensive research, development, and testing that led to the original design and material choices, which is outside the scope of this specific 510(k) modification.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Established: Not applicable, as this is not an AI/ML device.
Ask a specific question about this device
(56 days)
RHS
The RHS is intended for:
- Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepation and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with:
a. Joint destruction and/or subluxation visible on x-ray
b. Resistance to conservative treatment - Primary replacement after fracture of the radial head
- Symptomatic sequelae after radial head resection
- Revision following failed radial head arthroplasty
The long stem is for single cemented use only. The short stem coated with titanium plasma-spray is for single use with or without cement.
The RHS has been designed in order to provide surgeons and patients with a joint prosthesis to restore function and relieve pain of the radial part of the elbow joint. The RHS has an anatomical design, which reproduces the kinematics of the radial joint. The RHS consists of two components: a metal radial stern and a metal-polyethylene radial head.
This is a radial head prosthesis submission, not an AI/ML device, therefore, the requested information about acceptance criteria, study details, and AI/ML specific metrics (e.g., sample size for test/training set, number of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable.
The provided document is a 510(k) Premarket Notification for the "RHS" radial head prosthesis. The purpose of this type of submission is to demonstrate that the device is substantially equivalent to a legally marketed predicate device, not to prove performance against specific acceptance criteria in the manner of an AI/ML diagnostic device study.
The FDA's letter states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."
However, based on the general information provided for medical devices, I can extract information relevant to "claims" and "indications for use" as they relate to what a device is designed to do, which can be seen as a form of "acceptance criteria" in a broader sense for a non-AI/ML device.
Here's an interpretation based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance (as inferred from Indications for Use and Device Description)
Since this is a prosthesis and not a diagnostic device, "performance" is largely demonstrated through substantial equivalence to predicate devices and the physical properties of the materials, rather than metrics like sensitivity or specificity. "Acceptance criteria" in this context refer to the conditions under which the device is intended to be used and the clinical problems it aims to address.
Acceptance Criteria (Indications for Use) | Reported Device Performance (Implicitly met by S.E. to predicates and device design) |
---|---|
Purpose: Restore function and relieve pain of the radial part of the elbow joint. | Achieved by anatomical design and material properties. |
For replacement of radial head for degenerative or post-traumatic disabilities with: | |
- Pain, crepitation, and decreased motion at radio-humeral and/or proximal radio-ulnar joint | |
- Joint destruction and/or subluxation visible on x-ray | |
- Resistance to conservative treatment | |
Primary replacement after fracture of radial head | |
Symptomatic sequelae after radial head resection | |
Revision following failed radial head arthroplasty | |
Material Composition: Radial head: UHMWPE with CoCr shell. Stem: CoCr, some with Ti plasma-spray. | Confirmed via material specification. |
Stem Usage: Long stem for single cemented use only. Short stem (Ti plasma-spray) for single use with or without cement. | Confirmed via design specification. |
2. Sample size used for the test set and the data provenance:
- Not Applicable. This is a 510(k) for a physical implant, not an AI/ML diagnostic device requiring a "test set" in the computational sense. Substantial equivalence is typically demonstrated through comparison of design, materials, manufacturing processes, intended use, and sometimes non-clinical testing (e.g., mechanical testing, biocompatibility) to predicate devices, rather than a clinical study with a "test set" of patient data for performance evaluation against specific metrics.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not Applicable. See point 2.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. See point 2.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a physical implant, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not Applicable. This is a physical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not Applicable directly for performance evaluation in the AI/ML sense. For a prosthesis, "ground truth" might pertain to established biomechanical principles, material standards (e.g., ASTM, ISO), and historical clinical outcomes of predicate devices that establish the safety and effectiveness of the type of device. Formal "ground truth" to evaluate diagnostic accuracy is not relevant here.
8. The sample size for the training set:
- Not Applicable. See point 2.
9. How the ground truth for the training set was established:
- Not Applicable. See point 2.
Ask a specific question about this device
(74 days)
RHS ACETABULAR COMPONENT
OSTEOARTHRITIS
RHEUMATOID ARTHRITIS
TRAUMATIC ARTHRITIS
WHERE THE USE OF A MORE CONSERVATIVE PROCEDURE HAS FAILED OR IS UNACCEPTABLE
INTENDED FOR NON-CEMENTED USE SINGLE USE ONLY
The Signal Medical Corporation HHS Acetabular Component is manufactured of ASTM I-136 or F-75. The design is made available in twelve (12) sizes ranging from 48mm to 70mm, in two mm increments. The cup is coated with an F-67 sintered beaded surface for F-136 or F-75 beads for an F-75 cup to enhance bone ingrowth. The device is designed in three models, one with three circumferentially spaced pins or spikes at 120 degree intervals to provide for a primary skeletal fixation and reduce the chance of rotation, one with screw holes and one with both screw holes and spikes. The ultra high molecular weight polyethylene inserts, ASTM F648, are secured to the acetabular shell using a crown-taper fit. They are designed to accept 28mm femoral heads on their inside diameter from 48mm to 70mm and 32mm femoral heads from 54mm to 70mm. This device is intended for single use only.
This document is a 510(k) summary for a medical device (HHS Acetabular Component), not a study report for an AI/ML powered device. As such, it does not contain information about acceptance criteria, device performance, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or training set details that are typically found in such studies.
Therefore, I cannot provide a table of acceptance criteria and reported device performance or other requested information as it is not present in the provided text.
Ask a specific question about this device
(84 days)
RHS HELMET
The RHS Helmet is a cranial orthosis that applies passive pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age with moderate to severe nonsynostotic positional plagiocephaly, including infants with plagiocephalic, brachycephalic, and scaphocephalic- shaped heads.
The RHS Helmet is a cranial orthosis that applies passive pressure to prominent regions of an infant's cranium in order to improve cranial symmetry and/or shape in infants from 3 to 18 months of age with moderate to severe nonsynostotic positional plagiocephaly. It consists of a polypropylene helmet, three-eighths of an inch thick, that is vacuum formed over a plaster model of a baby's head to produce a helmet. A liner for the helmet is made of plastizote, one-fourth of an inch in thickness. Small holes are bored in the helmet for ventilation, and large holes for the child's ears.
The provided text is a 510(k) summary for the RHS Helmet, a cranial orthosis. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive performance studies against specific acceptance criteria for a novel device.
Therefore, many of the requested categories for a study proving device meets acceptance criteria are not applicable or not detailed in this document.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
This document does not specify quantitative acceptance criteria or provide a table of performance data against such criteria. The basis for clearance is "biocompatibility of the materials and the safety and effectiveness of helmet therapy," which are qualitative statements rather than specific metrics.
Acceptance Criteria | Reported Device Performance |
---|---|
Not specified in the document | Not specified in the document |
2. Sample Size Used for the Test Set and Data Provenance
No specific test set or clinical study data on device performance is provided in this 510(k) summary. The document mentions "Performance Data: Information was provided on the biocompatibility of the materials and the safety and effectiveness of helmet therapy," but does not detail any particular study results or sample sizes.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. As no specific test set or clinical study is detailed, there's no mention of experts establishing ground truth for performance.
4. Adjudication Method for the Test Set
Not applicable. No test set or clinical study is detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
Not applicable. This document does not describe an MRMC study. The focus is on demonstrating substantial equivalence to a predicate device, not on comparative effectiveness against humans or other devices.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. The RHS Helmet is an orthotic device, not an algorithm, so a standalone algorithm performance study is irrelevant.
7. The Type of Ground Truth Used
The concept of "ground truth" as typically applied to performance studies (e.g., pathology, outcomes data) is not explicitly discussed in this 510(k) summary. The device's effectiveness relies on the established principles of cranial orthosis therapy for nonsynostotic positional plagiocephaly, which is a recognized medical approach. The "effectiveness" mentioned refers to the general understanding of helmet therapy, not a specific ground truth derived from a novel study for this particular device.
8. The Sample Size for the Training Set
Not applicable. This device is a physical orthotic device, not an algorithm or AI model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. See #8.
In summary, the provided 510(k) documentation for the RHS Helmet primarily relies on the concept of "substantial equivalence" to a legally marketed predicate device (Clarren Helmet -- 510(k) # 003035). This means that the device's safety and effectiveness are supported by its similarity to an already cleared device, rather than through extensive, independent, novel performance studies with specific acceptance criteria as might be required for a completely new technology or a PMA application. The "Performance Data" mentioned likely refers to general information and existing knowledge about cranial helmet therapy and material biocompatibility, rather than data from a new, comprehensive clinical trial for the RHS Helmet itself.
Ask a specific question about this device
Page 1 of 1