Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K172995
    Date Cleared
    2017-11-21

    (55 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Destination Carotid Guiding Sheath, Destination Peripheral Guiding Sheath, Destination Renal Guiding
    Sheath

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Destination® Carotid Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the carotid arteries.

    The Destination® Peripheral Guiding Sheath is designed to be used for the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to lower extremity access via a contralateral approach.

    The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the renal arteries.

    Device Description

    The Destination® Guiding Sheath is designed to perform as a guiding catheter and an introducer sheath. The Destination® Guiding Sheath is a 3-layer tubing with a Stainless-Steel coil reinforcement sandwiched between layers of PTFE Teflon and Nylon Pebax, a hemostatic valve at the proximal end, a radiopaque marker near the distal end, and a hydrophilic coating on the distal portion of the tubing of varying lengths based on the product code. The sheath is available in multiple French sizes, useable lengths and distal shape configurations. The Destination® Guiding Sheaths are packaged with the following components: a Sheath, a Dilator, a Hemostatic Valve, and a Dilator Retaining Clip (only available with codes that contain the Tuohy-borst valve).

    AI/ML Overview

    This is not an AI review. The document is for a medical device called "Destination Guiding Sheath" and pertains to its 510(k) premarket notification to the FDA. The submission seeks to establish substantial equivalence to previously marketed predicate devices, not an AI or algorithm. Therefore, the questions related to AI acceptance criteria, training sets, test sets, and expert adjudication are not applicable to this document.

    However, I can provide the available information regarding the device's performance criteria and testing based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document broadly states that "The Destination® Guiding Sheath tested met the predetermined acceptance criteria and results support a determination of substantial equivalence." Specific numerical acceptance criteria are not detailed in the summary, but the types of tests performed and the general outcome are provided.

    Test ItemAcceptance Criteria (General)Reported Device Performance (General)
    Lubricity and Durability/AdhesionMet in-house standardMet the predetermined acceptance criteria and results support a determination of substantial equivalence.
    Simulated Use and ParticulateMet FDA Guidance Doc 1608Met the predetermined acceptance criteria and results support a determination of substantial equivalence.
    Simulated Use Coating Integrity EvaluationMet FDA Guidance Doc 1608Met the predetermined acceptance criteria and results support a determination of substantial equivalence.
    BiocompatibilityConforms to FDA Guidance / EN ISO 10993-1Tested and results demonstrate biocompatibility of the finished device.
    SterilizationAchieves SAL 10-6 (ISO 11135:2014)Sterilization process validated, achieved SAL 10-6.
    Residual EO/ECHDo not exceed 4 mg (EO) and 9 mg (ECH) dailyLevel of residual EO and ECH do not exceed an average daily dose of 4 mg and 9 mg respectively.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified in the provided text for non-clinical performance tests.
    • Data Provenance: The tests were conducted internally ("In-house standard") or against FDA guidance documents. The document doesn't specify country of origin for the test data, but the manufacturer is Terumo Medical Corporation with locations in Somerset, NJ and Elkton, MD, suggesting U.S.-based testing or oversight. The studies were non-clinical (laboratory/bench testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This question is not applicable as the document describes a traditional medical device (catheter introducer), not an AI/ML device that requires expert ground truth for image or data interpretation. Performance was evaluated against engineering standards and regulatory guidelines.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable, as this refers to adjudication for expert-labeled ground truth in AI/ML studies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable, as this is not an AI/ML device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable, as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the non-clinical tests was established by predefined engineering specifications, applicable FDA guidance documents, and international standards (e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization).

    8. The sample size for the training set:

    Not applicable, as this is not an AI/ML device and thus does not involve a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as this is not an AI/ML device and thus does not involve a training set or its ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K081045
    Date Cleared
    2008-05-28

    (44 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    DESTINATION RENAL GUIDING SHEATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries.

    Device Description

    The Destination® Renal Guiding Sheath is designed to perform as a guiding catheter and an introducer sheath. The Guiding Sheath is packaged with the following components: a Sheath, a Dilator, a Hemostatic Valve, and a Dilator Retaining Clip.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "Destination® Renal Guiding Sheath." This is a medical device, and the evaluation focuses on its substantial equivalence to a predicate device, not on complex AI performance metrics. Therefore, many of the requested fields are not applicable in this context.

    Here's the breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    For a traditional medical device like a guiding sheath, "acceptance criteria" are typically related to meeting established performance standards (e.g., sterilization, biocompatibility, mechanical integrity) and demonstrating equivalence to a predicate device. Performance is shown through bench and vessel model testing, not statistical metrics like sensitivity/specificity for disease detection.

    Acceptance Criteria / Performance CategoryDevice Performance (Reported)
    Intended UseIntroduction of interventional and diagnostic devices into the human vasculature, including but not limited to the Renal Arteries. (Substantially equivalent to predicate)
    Design / MaterialsUses similar materials to the predicate device. Differences do not raise new safety/effectiveness issues. (Substantially equivalent to predicate)
    Principle of Operation / TechnologyOperated manually or by a manual process. (Substantially equivalent to predicate)
    SterilizationValidated in accordance with EN ISO 11135-1 to provide a Sterility Assurance Level of 10-6.
    BiocompatibilityBlood contacting materials tested in accordance with FDA G95-1 (ISO 10993). Found to be biocompatible for "Externally Communicating Device, Circulating Blood, Limited Contact (≤24hrs)".
    Mechanical PerformanceDemonstrated through bench and vessel model testing. (Substantially equivalent to predicate)
    Expiration Dating30 months.
    Specifications (Physical)Sheath Size: 5Fr., Sheath Length: 45-55 cm, Hydrophilic Coating: Distal 5 cm, Distal Shape Configurations: Straight, HS, RDC, LIMA, MP. (Implied to meet design specifications)

    Studies that Prove the Device Meets Acceptance Criteria:

    The document states:

    • "The performance of the Destination® Renal Guiding Sheath is substantially equivalent to the performance of the unmodified Renal Guiding Sheath. The equivalence was shown through bench and vessel model testing."
    • "Sterilization conditions have been validated in accordance with EN ISO 11135-1..."
    • "Blood contacting materials were tested in accordance with the test recommendations in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993..."

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the summary. For a medical device 510(k), especially one comparing physical characteristics to a predicate, specific "test set" sample sizes or data provenance in the way one would describe for an AI/ML study are not typically detailed in this type of summary document. The "bench and vessel model testing" implies laboratory testing rather than human subject data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is not applicable and not provided. The ground truth for device performance in this context would be engineering specifications, physical measurements, and biological compatibility standards, not expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This is not applicable and not provided. Adjudication methods are relevant for studies where human interpretation (e.g., of medical images) generates a "ground truth" that needs to be resolved by multiple experts. For physical device performance, this concept doesn't apply.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This device is a physical medical instrument (a guiding sheath), not an AI algorithm or a diagnostic tool that involves human readers interpreting data assisted by AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is based on established engineering and biocompatibility standards, and measurable physical properties. For example:

    • Sterility: Achieves a measured Sterility Assurance Level (SAL) of 10^-6, verified by laboratory testing according to EN ISO 11135-1.
    • Biocompatibility: Meets the requirements outlined in ISO 10993 for limited contact with circulating blood, verified by laboratory testing.
    • Physical Performance (e.g., torqueability, pushability, trackability): Assessed through "bench and vessel model testing" against a predicate device. The ground truth here would be engineering metrics and comparison to the predicate's established performance.

    8. The sample size for the training set

    This is not applicable and not provided. This device does not involve a "training set" in the context of machine learning.

    9. How the ground truth for the training set was established

    This is not applicable and not provided. This device does not involve machine learning or a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K012854
    Date Cleared
    2001-11-21

    (89 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    RENAL GUIDING SHEATH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the renal arteries.

    Device Description

    The Renal Guiding Sheath is designed to perform as a guiding catheter and an introducer sheath. The sheath is coil reinforced, has a radiopaque marker and is hydrophilically coated on the distal 5 cm. It comes packaged with a dilator, a rotating hemostatic valve (or a detachable hemostatic valve), and a dilator retaining clip.

    AI/ML Overview

    This document describes a medical device, the "Renal Guiding Sheath," and its submission for FDA clearance. It focuses on demonstrating substantial equivalence to a predicate device, not on AI/algorithm performance. Therefore, many of the requested fields related to AI, ground truth, and expert evaluation are not applicable.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Tests)Reported Device Performance (Reference to Predicate)
    PenetrationSubstantially equivalent to Super Arrow-Flex Percutaneous Sheath Introducer Set
    Kink ResistanceSubstantially equivalent to Super Arrow-Flex Percutaneous Sheath Introducer Set
    LeakageSubstantially equivalent to Super Arrow-Flex Percutaneous Sheath Introducer Set
    Tensile StrengthSubstantially equivalent to Super Arrow-Flex Percutaneous Sheath Introducer Set
    Tortuous Path / DurabilitySubstantially equivalent to Super Arrow-Flex Percutaneous Sheath Introducer Set
    Sterility Assurance Level (SAL)10^-6 (validated according to ANSI / AAMI / ISO 11135-1994)
    BiocompatibilityBlood contacting materials found to be biocompatible (tested in accordance with FDA General Program Memorandum #G95-1 / ISO 10993)
    Expiration Dating30 months

    2. Sample Size Used for the Test Set and Data Provenance

    This document does not specify sample sizes for each performance test. The data provenance is not mentioned beyond the fact that tests were performed to demonstrate substantial equivalence. It is likely internal testing without country-of-origin specification, and it is implied to be prospective testing (performance evaluations of the new device).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is a physical medical instrument, not an AI or diagnostic algorithm that requires expert-established ground truth for its performance evaluation in this context. The performance tests are engineering and biological evaluations.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for expert review of images or clinical cases to establish ground truth for AI algorithms. For physical device performance tests, standards and objective measurements are used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes a physical medical device, not an AI system. Therefore, no MRMC study or AI assistance effect size is relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm, so standalone performance is not relevant.

    7. The Type of Ground Truth Used

    For the performance tests:

    • Engineering Standards / Objective Measurements: For tests like Penetration, Kink Resistance, Leakage, Tensile Strength, and Tortuous Path / Durability, the "ground truth" would be the successful adherence to pre-defined engineering specifications and measurable performance within acceptable limits, often compared directly to the predicate device's measured performance.
    • Validated Standards: For sterilization, the ground truth is established by validation against ANSI / AAMI / ISO 11135-1994 to achieve a specified SAL.
    • Biocompatibility Standards: For biocompatibility, the ground truth is conformity with test recommendations in FDA General Program Memorandum #G95-1 (ISO 10993).

    8. The Sample Size for the Training Set

    Not applicable. This is a physical device, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. This is a physical device, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1