Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123749
    Date Cleared
    2013-03-01

    (85 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    POWDER FREE NITRILE PATIENT EXAMINATION GLOVES, GREEN, NON-STERILE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This glove has been tested for use with specific chemotherapy drugs listed below.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim, meets all of the requirements of ASTM D 6319-10.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for Powder Free Nitrile Patient Examination Gloves, Green, Non-Sterile, with Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim. The submission seeks to demonstrate substantial equivalence to a predicate device.

    Here's the breakdown of the acceptance criteria and the studies that prove the device meets them:


    1. Table of Acceptance Criteria and the Reported Device Performance

    CharacteristicStandard (Acceptance Criteria)Device Performance
    DimensionsASTM D 6319-10Meets
    Physical PropertiesASTM D 6319-10Meets
    Freedom from pin-holesASTM D 5151-11; ASTM D 6319-10Meets
    Powder Free ResidueASTM D 6124-11; ASTM D 6319-10Meets
    Low Dermatitis Potential (Biocompatibility)Modified Draize-95 Test (as per FDA's guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510(k)] Submissions for Testing for Skin Sensitization to Chemicals in Natural Rubber Products, 1999")No clinical evidence of residual chemical additives that may induce Type IV allergy in human subjects. Results of the human repeat insult patch test (HRIPT) on 200 subjects were negative, indicating no potential for inducing dermal irritation or sensitization in the general user population.
    Dermal Sensitization (Biocompatibility)ISO 10993-10:2010Not a contact skin sensitizer
    Primary Skin Irritation Test (Biocompatibility)16 CFR Part 1500.41Not a primary skin irritant
    Chemotherapy Drugs Permeation TestASTM D6978-05Dacarbazine (DTIC) (10.0 mg/ml): > 240 minutes
    Carmustine (BCNU) (3.3 mg/ml): 15.3 minutes
    Cyclophosphamide (Cytoxan) (20.0 mg/ml): > 240 minutes
    Doxorubicin Hydrochloride (2.0 mg/ml): > 240 minutes
    Fluorouracil (50.0 mg/ml): > 240 minutes
    Cisplatin (1.0 mg/ml): > 240 minutes
    Etoposide (20.0 mg/ml): > 240 minutes
    Paclitaxel (Taxol) (6.0 mg/ml): > 240 minutes
    Thiotepa (10.0 mg/ml): 195.8 minutes

    2. Sample size used for the test set and the data provenance

    • Low Dermatitis Potential (Modified Draize-95 Test):
      • Sample Size: A total of 200 human subjects.
      • Data Provenance: Prospective, clinical study. The document states, "The study was conducted in two stages. In the first stage, a population of 50 human subjects was tested... The second stage was initiated on a further number of subjects to a total of a minimum of 200 individuals after the first stage..."
      • Country of Origin: Not explicitly stated, though the submitter is based in Malaysia. The study aimed to reflect "the general user population in the US."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish ground truth for the Modified Draize-95 Test or the Chemotherapy Drugs Permeation Test. These tests are standardized laboratory and human subject tests, respectively, where the "ground truth" is determined by the specific criteria and readings of the test methods rather than expert consensus on subjective findings.


    4. Adjudication method for the test set

    The document does not describe an adjudication method for the tests performed. For the human subject testing (Modified Draize-95), the results were simply "all negative," indicating no ambiguous findings requiring adjudication. For the other performance tests (e.g., dimensions, physical properties, pinholes, powder residue, chemotherapy permeation), the results are quantitative or pass/fail against established standards.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a medical glove, and the nature of its evaluation does not involve human readers interpreting data or AI assistance.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable as the device is a medical glove, not an algorithm or AI system.


    7. The type of ground truth used

    • For physical properties, pinholes, and powder free residue: Ground truth was established by adherence to recognized ASTM (American Society for Testing and Materials) standards (ASTM D6319-10, ASTM D5151-11, ASTM D6124-11), which define objective test methods and acceptance criteria.
    • For biocompatibility (dermal sensitization and primary skin irritation): Ground truth was established by adherence to ISO 10993-10:2010 and 16 CFR Part 1500.41, which are standardized testing protocols for biological evaluation of medical devices.
    • For Low Dermatitis Potential: Ground truth was established through a human clinical study (Modified Draize-95 Test) observing the absence of allergic or irritant reactions in human subjects.
    • For Chemotherapy Drugs Permeation Test: Ground truth was established by adherence to the ASTM D6978-05 standard, which outlines specific methods for measuring drug permeation through protective materials.

    8. The sample size for the training set

    This is not applicable as the device is a medical glove and does not involve machine learning or AI models with training sets.


    9. How the ground truth for the training set was established

    This is not applicable as the device is a medical glove and does not involve machine learning or AI models with training sets.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1