Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This glove has been tested for use with specific chemotherapy drugs listed below.

Device Description

Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim, meets all of the requirements of ASTM D 6319-10.

AI/ML Overview

This document is a 510(k) Premarket Notification for Powder Free Nitrile Patient Examination Gloves, Green, Non-Sterile, with Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim. The submission seeks to demonstrate substantial equivalence to a predicate device.

Here's the breakdown of the acceptance criteria and the studies that prove the device meets them:


1. Table of Acceptance Criteria and the Reported Device Performance

CharacteristicStandard (Acceptance Criteria)Device Performance
DimensionsASTM D 6319-10Meets
Physical PropertiesASTM D 6319-10Meets
Freedom from pin-holesASTM D 5151-11; ASTM D 6319-10Meets
Powder Free ResidueASTM D 6124-11; ASTM D 6319-10Meets
Low Dermatitis Potential (Biocompatibility)Modified Draize-95 Test (as per FDA's guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510(k)] Submissions for Testing for Skin Sensitization to Chemicals in Natural Rubber Products, 1999")No clinical evidence of residual chemical additives that may induce Type IV allergy in human subjects. Results of the human repeat insult patch test (HRIPT) on 200 subjects were negative, indicating no potential for inducing dermal irritation or sensitization in the general user population.
Dermal Sensitization (Biocompatibility)ISO 10993-10:2010Not a contact skin sensitizer
Primary Skin Irritation Test (Biocompatibility)16 CFR Part 1500.41Not a primary skin irritant
Chemotherapy Drugs Permeation TestASTM D6978-05Dacarbazine (DTIC) (10.0 mg/ml): > 240 minutes Carmustine (BCNU) (3.3 mg/ml): 15.3 minutes Cyclophosphamide (Cytoxan) (20.0 mg/ml): > 240 minutes Doxorubicin Hydrochloride (2.0 mg/ml): > 240 minutes Fluorouracil (50.0 mg/ml): > 240 minutes Cisplatin (1.0 mg/ml): > 240 minutes Etoposide (20.0 mg/ml): > 240 minutes Paclitaxel (Taxol) (6.0 mg/ml): > 240 minutes Thiotepa (10.0 mg/ml): 195.8 minutes

2. Sample size used for the test set and the data provenance

  • Low Dermatitis Potential (Modified Draize-95 Test):
    • Sample Size: A total of 200 human subjects.
    • Data Provenance: Prospective, clinical study. The document states, "The study was conducted in two stages. In the first stage, a population of 50 human subjects was tested... The second stage was initiated on a further number of subjects to a total of a minimum of 200 individuals after the first stage..."
    • Country of Origin: Not explicitly stated, though the submitter is based in Malaysia. The study aimed to reflect "the general user population in the US."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number or qualifications of experts used to establish ground truth for the Modified Draize-95 Test or the Chemotherapy Drugs Permeation Test. These tests are standardized laboratory and human subject tests, respectively, where the "ground truth" is determined by the specific criteria and readings of the test methods rather than expert consensus on subjective findings.


4. Adjudication method for the test set

The document does not describe an adjudication method for the tests performed. For the human subject testing (Modified Draize-95), the results were simply "all negative," indicating no ambiguous findings requiring adjudication. For the other performance tests (e.g., dimensions, physical properties, pinholes, powder residue, chemotherapy permeation), the results are quantitative or pass/fail against established standards.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a medical glove, and the nature of its evaluation does not involve human readers interpreting data or AI assistance.


6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This is not applicable as the device is a medical glove, not an algorithm or AI system.


7. The type of ground truth used

  • For physical properties, pinholes, and powder free residue: Ground truth was established by adherence to recognized ASTM (American Society for Testing and Materials) standards (ASTM D6319-10, ASTM D5151-11, ASTM D6124-11), which define objective test methods and acceptance criteria.
  • For biocompatibility (dermal sensitization and primary skin irritation): Ground truth was established by adherence to ISO 10993-10:2010 and 16 CFR Part 1500.41, which are standardized testing protocols for biological evaluation of medical devices.
  • For Low Dermatitis Potential: Ground truth was established through a human clinical study (Modified Draize-95 Test) observing the absence of allergic or irritant reactions in human subjects.
  • For Chemotherapy Drugs Permeation Test: Ground truth was established by adherence to the ASTM D6978-05 standard, which outlines specific methods for measuring drug permeation through protective materials.

8. The sample size for the training set

This is not applicable as the device is a medical glove and does not involve machine learning or AI models with training sets.


9. How the ground truth for the training set was established

This is not applicable as the device is a medical glove and does not involve machine learning or AI models with training sets.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Kossan Latex Industries. The logo includes the letters "KL1" in a bold, black font on the left side of the image. To the right of the letters, the text "KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)" is displayed in a smaller, serif font.

FDA 510(k), Premarket Notification: 510(k) Summary of Safety and Effectiveness Information

1.0 Submitter:

Kossan Latex Industries (M) Sdn Bhd Lot 6129, Jalan Haji Abdul Manan, Batu 5 1/4, Jalan Meru, Klang, 41050 Selangor, Malaysia

Telephone No .: +603 3392 3088 Fax No .: +603 3291 0584

2.0 Contact Person:

Contact:Ms S F Cho
Telephone No.:+603 3392 3088
Fax No.:+603 3291 0584

3.0 Preparation Date: 29 November 2012

4.0 Name of Device:

Trade Name: Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim.

Powder-Free Nitrile Patient Examination Glove Common Name: Classification Name: Patient Examination Glove (21 CFR Part 880.6250) Regulatory Class: Class I Product Code: 80 LZC, 80 LZA.

5.0 Identification of the Legally Marketed Device:

Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim; Class I Patient Examination Gloves, Nitrile-80LZA, 80LZC, meets all of the requirements of ASTM D 6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application.

Predicate Device: K120066, Powder Free Nitrile Patient Examination Glove, Blue, White, Green. Non-Sterile. Low Dermatitis Potential Labeling Claim.

Image /page/0/Picture/14 description: The image shows the text "Page 1 of 6". This text indicates that the image is the first page of a document that has a total of six pages. The text is written in a simple, sans-serif font and is positioned at the top of the image.

{1}------------------------------------------------

K L 1 KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)

6.0 Description of Device:

Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim, meets all of the requirements of ASTM D 6319-10.

7.0 Intended Use of the Device:

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

This glove has been tested for use with specific chemotherapy drugs listed below.

Chemotherapy Drug Permeation (Minimum Breakthrough Detection Time in Minutes)

Dacarbazine (DTIC) (10.0 mg/ml)> 240
Carmustine (BCNU) (3.3 mg/ml)15.3
Cyclophosphamide (Cytoxan) (20.0 mg/ml)> 240
Doxorubicin Hydrochloride (2.0 mg/ml)> 240
Fluorouracil (50.0 mg/ml)> 240
Cisplatin (1.0 mg/ml)> 240
Etoposide (20.0 mg/ml)> 240
Paclitaxel (Taxol) (6.0 mg/ml)> 240
Thiotepa (10.0 mg/ml)195.8

Warning: Do not use with Carmustine.

8.0 Summary of the Technological Characteristics of the Device:

Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim possess the following technological characteristic (as compared to ASTM or equivalent standards):

CharacteristicStandardsDevice Performance
DimensionsASTM D 6319-10Meets
Physical PropertiesASTM D 6319-10Meets
Freedom from pin-holesASTM D 5151-11Meets
ASTM D 6319-10Meets
Powder Free ResidueASTM D 6124-11Meets
ASTM D 6319-10Meets
BiocompatibilityDermal Sensitization(as per ISO 10993-10:2010)Not a contact skin sensitizer
Primary SkinIrritation Test(as per 16 CFR Part1500.41)Not a primary skin irritant

{2}------------------------------------------------

KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)

CharacteristicStandardsDevice Performance
Low DermatitisPotentialModified Draize-95 TestNo clinical evidence presence of residualchemical additives that may induce TypeIV allergy in human subjects.
Chemotherapy DrugsPermeation TestASTM D6978-05Chemotherapy Drug Permeation(Minimum Breakthrough Detection Timein Minutes)
Dacarbazine (DTIC) (10.0 mg/ml) and the corrected html is: ```html Characteristic Standards Device Performance Low DermatitisPotential Modified Draize-95 Test No clinical evidence presence of residualchemical additives that may induce TypeIV allergy in human subjects. Chemotherapy DrugsPermeation Test ASTM D6978-05 Chemotherapy Drug Permeation(Minimum Breakthrough Detection Timein Minutes) Dacarbazine (DTIC) (10.0 mg/ml) Carmustine (BCNU) (3.3 mg/ml) Cyclophosphamide (Cytoxan)(20.0 mg/ml) Doxorubicin Hydrochloride(2.0 mg/ml) Fluorouracil (50.0 mg/ml) Cisplatin (1.0 mg/ml) Etoposide (20.0 mg/ml) Paclitaxel (Taxol) (6.0 mg/ml) Thiotepa (10.0 mg/ml) > 240 15.3 > 240 > 240 > 240 > 240 > 240 > 240 195.8

9.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data

Testing was performed per ASTM D6319-10, ASTM D5151-11, ASTM D6124-11, ISO 10993-10:2010, and 16 CFR Part 1500.41. The gloves meet standard requirements referenced in Section 8.0 above. Biocompatibility test indicates the gloves are not a contact skin sensitizer and not a primary skin irritant.

Substantial Equivalent Based on Assessment of Clinical Performance Data 10.0

Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim, were tested in accordance with Modified Draize-95 Test, per FDA's guidance document "Guidance for Industry and FDA Reviewers/Staff: Premarket Notification [510(k)] Submissions for Testing for Skin Sensitization to Chemicals in Natural Rubber Products, 1999".

The study was conducted in two stages. In the first stage, a population of 50 human subjects was tested to evaluate the product for the potential to cause irritation or sensitization. The second stage was initiated on a further number of subjects to a total of a minimum of 200 individuals after the first stage has shown that the test product does not indicate a potential for inducing dermal irritation and does not shown sensitization capability.

Image /page/2/Picture/8 description: The image shows the text "Page 3 of 6". This indicates that it is a page number within a document. The document has at least 6 pages, and this image represents the third page.

{3}------------------------------------------------

K L 1 KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K)

The study completed on 200 non-sensitized adult human subjects, who reasonably reflect the general user population in the US, gave all negative results. There was no clinical evidence of the presence of residual chemical additives at the level that may induce Type IV allergy in the un-sensitized general user population in the tested article.

11.0 Conclusion

It can be concluded that Powder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim is safe and effective for use, and perform according to the glove performance standards referenced in Section 8.0 above, thereby meeting ASTM standards, FDA requirements, and the labeling claims for the product.

This device is identical and substantially equivalent to currently marketed devices, per Substantial Equivalence Comparison Table below:-

{4}------------------------------------------------

KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K) KLI

Substantial Equivalence Comparison Table

Featur(120066, Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterent Colored, Non-SterenPredicate DevicePowder Free Nitrile Patient Examination Glovential Chemotherapy Drugs Protection Labeling Cla
tended UProposed Devicedentica
ications for Usntended for medical purposes that is worn on thepatient examination glove is a disposable device ended for medical purposes that is worn on theent and examiner. I glove has been tested for use with specifichemotherapy drugs listed below
なお気になる8 556 Obizisaearbazine (DZTIC) ( 0.0 mg/mL) velophosphamide ( ( 10.0 mg/mL) velophosphamide ( ( 10.0 mg/mL) velophosphamide ( ( 10.0 mg/mL) velophosille ( 1) video ( ( 12.0 mg/mL) 1nemotherapy Drug Permeation (Minimutes
arning: Do not use with Carmu

{5}------------------------------------------------

KOSSAN LATEX INDUSTRIES (M) SDN. BHD. (169832-K) KLI

FeatureK120066, Powder Free Nitrile Patient Examination Glove, Blue, White and Green Colored, Non-Sterile (Low Dermatitis Potential Labeling Claim) of Kossan Latex Industries Sdn BhdPredicate DevicePowder Free Nitrile Patient Examination Glove, Green, Non-Sterile, Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim
MaterialNitrileBlue, White and GreenProposed Device Identical
ColorYesGreenIdentical
Single UseYesYesIdentical
SterilizationNot Applicable; Non-SterileNot Applicable; Non-SterileIdentical
DimensionsMeets ASTM D6319Meets ASTM D6319Identical
Physical PropertiesMeets ASTM D6319Meets ASTM D6319Identical
Freedom from PinholesMeets ASTM D5151 and ASTM D6319Meets ASTM D5151 and ASTM D6319Identical
Residue PowderMeets ASTM D6124 and ASTM D6319Meets ASTM D6124 and ASTM D6319Identical
Biocompatibility TestPass Dermal Sensitization Test Pass Primary Skin Irritation TestPass Dermal Sensitization Test Pass Primary Skin Irritation TestIdentical
Human Draize TestPassPassIdentical
Chemotherapy Drugs Permeation TestNot Applicable; Not TestedNot Applicable; Not TestedMeets ASTM D6978-05
Labeling ClaimLow Dermatitis Potential Labeling ClaimLow Dermatitis Potential Labeling ClaimProposed device With Low Dermatitis Potential and Chemotherapy Drugs Protection Labeling Claim

Page 6 of 6

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a stylized graphic on the right and a circular arrangement of text on the left. The graphic resembles a stylized caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular fashion around the graphic.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 1, 2013

Ms. Cho Sow Fong Manager, Regulatory Affairs Kossan Latex Industries (M) Sdn Bhd Lot 16632, 51/4 Miles, Jalan Meru Klang Selangor, Malaysia 41050

Re: K123749

Trade/Device Name: Powder Free Nitrile Examination Gloves, Green, Non Sterile. Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs Labeling Claim.

Regulation Number: 21 CFR 880.6250

Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA. LZC Dated: January 21, 2013 Received: January 25, 2013

Dear Ms. Fong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Ms. Fong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/7/Picture/7 description: The image shows a stylized signature or logo. It features a combination of abstract shapes and what appears to be stylized text. The design includes looping lines, geometric shapes resembling a mountain or pyramid, and illegible script-like elements, all rendered in black against a white background.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use

510(k) Number (if known): K123749

Device Name:

Powder Free Nitrile Patient Examination Gloves, Green, Non-Sterile. Low Dermatitis Potential, and Tested for Use with Chemotherapy Drugs Labeling Claim.

Indications for Use:

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This glove has been tested for use with specific chemotherapy drugs listed below,

Chemotherapy Drug Permeation (Minimum Breakthrough Detection Time in Minutes)

Dacarbazine (DTIC) (10.0 mg/ml)> 240
Carmustine (BCNU) (3.3 mg/ml)15.3
Cyclophosphamide (Cytoxan) (20.0 mg/ml)> 240
Doxorubicin Hydrochloride (2.0 mg/ml)> 240
Fluorouracil (50.0 mg/ml)> 240
Cisplatin (1.0 mg/ml)> 240
Etoposide (20.0 mg/ml)> 240
Paclitaxel (Taxol) (6.0 mg/ml)> 240
Thiotepa (10.0 mg/ml)195.8

Please note that Carmustine has extremely low permeation times of 15.3 minutes.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C) x

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office Of Device Evaluation (ODE)

Elizabeth-F.Claverie
2013.02.27 17:10:04 -05'00'

(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division of Anesu.co.org/w/infection Control

510(k) Number,


Page 1 of _

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.