Search Results
Found 2 results
510(k) Data Aggregation
(111 days)
OxyMinder Pro (10310)
The OxyMinder Pro is an oxygen monitor with integrated pressure monitoring intended for continuous monitoring of the concentration of oxygen and pressure being delivered to patients ranging from newborns to adults. This device can be used in the hospital and subacute settings. The monitor is not intended as a life supporting device or for diagnostics.
The OxyMinder Pro is an oxygen and pressure monitor capable of measuring the oxygen concentration from 18 to 100% (cleared under K213948) and pressure from 0 to 60 cmH2O (subject of this submission) that for convenience can be mechanically mounted on to the cleared blender.
The pressure is measured via a disposable pressure tubing that connects from the monitor to an adapter placed in the patient circuit. This sampling line is identical to that cleared in predicate Maxtec K221734. We have only updated the labeling to reflect the name of the sponsor.
As indicated the oxygen monitoring portion has been previously cleared, K213948. It utilizes a cleared oxygen sensor which outputs a voltage to determine the concentration of oxygen. The OxyMinder Pro calibrates at ambient air (21%) and 100% oxygen. The OxyMinder Pro is software controlled. Again, the oxygen monitoring feature and functions are unchanged and previously cleared under reference K213948.
The new pressure monitoring feature utilizes a pressure sensor which measures the pressure within a patient circuit. There is a disposable pressure tubing that connects between the patient circuit and the pressure sensor.
This document, K251245, describes the 510(k) clearance for the OxyMinder Pro, an oxygen and pressure monitor. It uses the Maxtec MaxO2ME+p (K221734) as its predicate device for the combined oxygen and pressure monitoring features, and the Bio-Med Devices OxyMinder (K213948) as a reference for the oxygen monitoring aspects, which were previously cleared.
The core of the submission focuses on the new pressure monitoring feature, as the oxygen monitoring component is largely based on a previously cleared device (OxyMinder, K213948). The acceptance criteria and testing detailed largely pertain to the performance of the device's measurement capabilities and adherence to relevant safety and performance standards.
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes the performance specifications of the OxyMinder Pro and compares them to the predicate and reference devices. These specifications serve as the acceptance criteria for the device's functionality. The "Reported Device Performance" is implied by the similarity claims and the statement that "The test results met the applicable standards".
Acceptance Criteria (Performance Specification) | OxyMinder Pro Reported Performance | Predicate (Maxtec MaxO2ME+p K221734) | Reference (Bio-Med Devices OxyMinder K213948) |
---|---|---|---|
Oxygen Measurement Range | 18% – 100% O2 | 0.0 to 100% O2 | 18% – 100% O2 |
Pressure Measurement Range | 0 to 60 cmH2O | -15.0 to 60.0 cmH2O | N/A (Oxygen only) |
Oxygen Resolution | O2 % displayed to nearest whole integer | 0.1% O2 | O2 % displayed to nearest whole integer |
Pressure Resolution | 0.1 cmH2O | 0.5 cmH2O | N/A |
Oxygen Accuracy and Linearity | ±1% of full scale | +1% of full scale | ±1% of full scale |
Pressure Accuracy | +0.5 cmH2O | +1.0 cmH2O | N/A |
Total Oxygen Accuracy | ±2.5% Actual oxygen level over full operating temperature range | +3% Actual oxygen level over full operating temperature range | ±2.5% Actual oxygen level over full operating temperature range |
Oxygen Response Time (90% final value) | approx. 6 seconds | approx. 15 seconds at 23oC | approx. 6 seconds |
Warm-up Time | None required | None required | None required |
Operating Temperature | 0° - 50° C [32° - 122° F] | 15oC – 40oC (59oF – 104oF) | 0° - 50° C [32° - 122° F] |
Storage Temperature | 0° - 40° C [32° - 104° F] | -15oC – 50oC (5oF – 122oF) | 0° - 40° C [32° - 104° F] |
Atmospheric Pressure | 700 – 1010 mBars | 800 – 1012 mBars | 700 – 1010 mBars |
Humidity | 5 - 95% | 0-95% (non-condensing) | 5 - 95% |
Battery Life | 16 hours at 100% brightness | Approx. 5000 hours, typical use | 16 hours at 100% brightness |
Low Oxygen Alarm Range | 18% - 100% (>1% lower than high alarm) | 15% - 99% (>1% lower than high alarm) | 18% - 100% (>1% lower than high alarm) |
Low Pressure Alarm Range | Off – 55 cmH2O (> 1cmH2O lower than high pressure alarm) | Off, 1-30 cmH2O | N/A |
High Oxygen Alarm Range | 19% - 105% (>1% lower than low alarm) | 16% - 100% (>1% higher than low alarm) | N/A |
High Pressure Alarm Range | 5 – 60 cmH2O (>1 cmH2O higher than low pressure alarm) | 1-60 cmH2O, Off | N/A |
Alarm Accuracy | Exact to display alarm value | Exact to display alarm value | Exact to display alarm value |
Pressure Alarm Resolution | 1 cmH2O | 1 cmH2O | N/A |
2. Sample Size Used for the Test Set and Data Provenance
The document states that "Bench testing was performed" and lists various types of tests (Shelf-life / Aging, Software Verification and Validation, Safety and ElectroMagnetic Compatibility, etc.). It claims that "The test results met the applicable standards". However, the sample size for the test set is not explicitly stated.
The data provenance is not explicitly mentioned as country of origin, nor is it specified if the testing was retrospective or prospective. Given it's bench testing for a device clearance, it is implicitly prospective testing within a laboratory/manufacturing environment.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This type of information (number of experts, their qualifications, and their role in establishing ground truth) is not applicable or not provided for this specific 510(k) submission. Medical device performance testing, especially for devices like oxygen and pressure monitors, relies on calibrated instruments and established physical and electrical standards to determine "ground truth" (e.g., a calibrated gas mixture for oxygen concentration, a pressure calibrator for pressure measurements), not human expert consensus.
4. Adjudication Method for the Test Set
The concept of an adjudication method (like 2+1 or 3+1 used in clinical trials or image interpretation studies) is not applicable here. The "ground truth" for the device's performance (e.g., accuracy of oxygen or pressure readings) is established through comparison to validated and calibrated measurement standards, not through human interpretation that would require adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC study is not applicable and was not done for the OxyMinder Pro. MRMC studies are typically used to assess human reader performance, often in diagnostic imaging, with and without AI assistance. The OxyMinder Pro is a direct measurement device; its performance is based on its physical and electrical accuracy, not on human interpretation or an AI algorithm assisting human interpretation.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
The OxyMinder Pro itself is a standalone device. Its performance, as described by its measurement accuracy, resolution, response time, etc., is its standalone performance without a human-in-the-loop influencing the measurement itself. The device is intended to present data to a human for monitoring, but its core function (measurement) is standalone. The testing conducted, as per the "Non-Clinical Testing Summary," assesses the device's inherent performance.
7. Type of Ground Truth Used
The ground truth for the performance testing is established using calibrated instruments and reference standards. For oxygen measurements, this would involve calibrated gas mixtures with known oxygen concentrations. For pressure measurements, this would involve calibrated pressure sources or transducers with known pressure values. It is based on physical and engineering measurements against established standards, not expert consensus, pathology, or outcomes data.
8. Sample Size for the Training Set
The OxyMinder Pro uses an oxygen sensor (galvanic cell) and a pressure transducer. While it is "software controlled," it does not appear to employ machine learning or AI algorithms that would require a "training set" in the conventional sense (i.e., a dataset used to train a model). The software is likely for control, data processing, display, and alarm functions, using fixed algorithms based on physical principles, not learning from data. Therefore, a training set sample size is not applicable.
9. How the Ground Truth for the Training Set Was Established
As there is no indication of a training set in the context of machine learning, this question is not applicable. The device's operation is based on pre-programmed algorithms and calibrated sensor outputs.
Ask a specific question about this device
(90 days)
OxyMinder
The OxyMinder is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. This device can be used in the hospital and subacute settings. The monitor is not intended as a life supporting device or for diagnostics.
The OxyMinder is an air / oxygen blender mounted oxygen monitor capable of measuring the oxygen concentration from 18 to 100%. A Bio-Med specified oxygen sensor mounted to the blender outputs a voltage which is used by the OxyMinder to determine the concentration of oxygen. The OxyMinder calibrates at ambient air (21%) and 100% oxygen. The OxyMinder is software controlled. To measure the gas mixture of the blender the OxyMinder takes a sample of the gases to the sensor through a separate port and manifold which is separate from the gas pathway to the patient. This sample is then exhausted to the room. The OxyMinder sampling stream is not part of the gas pathway to the patient.
The OxyMinder is used for continuous monitoring of the concentration of oxygen delivered to patients via air oxygen blenders. The monitor provides the following features:
- Continuously displays the concentrated Oxygen level delivered to a patient.
- Accepts user inputs via touch screen or button (power button).
- Provides visual alarm messages, and audible alarms.
- Displays the current alarm setting levels (High and Low O2 alarms).
- Provides on-screen configuration tools such as O2 sensor calibration, touchscreen calibration, audio test, etc.
- Monitors and displays the battery level and power source.
- Ensures clean hospital airlines by automatically purging system periodically.
The OxyMinder is designed to be mounted to a Bio-Med Devices air / oxygen blender via a manifold that houses the oxygen sensor and a solenoid.
In addition to the primary function of monitoring oxygen concentration, the OxyMinder provides an automatic gas supply line purge function.
The provided text is related to a 510(k) premarket notification for the "OxyMinder" device, an oxygen gas analyzer. It describes the device, its intended use, and compares it to a predicate device (Maxtec MaxO2ME). The document states that "A series of non-clinical performance / bench testing was performed" and lists the types of tests, concluding that "The test results met the applicable standards and are similar to the reported performance of the predicate device." However, the document does not provide detailed acceptance criteria or the specific results of these tests. It only broadly states that the performance was similar to the predicate and met applicable standards.
Therefore, many of the requested details about the study cannot be extracted from this document.
Here's a summary of what can be inferred or explicitly stated:
Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance (OxyMinder) |
---|---|
Measurement Range: Intended for use with an air/oxygen blender which will not deliver oxygen below 18%. | 18% – 100% O2 |
Resolution: (Implied to be sufficient for reading in 10% increments given blender knob resolution) | Displayed to nearest whole integer |
Accuracy and Linearity: Similar to predicate which is ±1% of full scale at constant temperature, RH and pressure when calibrated at full scale. | ±1.0% of full scale at constant temperature and pressure |
Total Accuracy: Similar to predicate which is ±3% Actual oxygen level over full operating temperature range. | ±2.5% Actual oxygen level over the full operating temperature range |
Response Time: Similar to predicate which is 90% of final value in approx. 15 seconds at 23°C. | 90% of final value in approximately 6 seconds |
Warm-up Time: None required. | None required |
Operating Temperature: Similar to predicate which is 15°C – 40°C (59°F – 104°F). | 0° – 50°C [32° - 122° F] |
Storage Temperature: Similar to predicate which is -15°C – 50°C (5°F – 122°F). | 0° - 40° C [32° - 104° F] |
Atmospheric Pressure: Similar to predicate which is 800 - 1012 m Bars. | 345 - 2068 mBars |
Humidity: Similar to predicate which is 0-95% (non-condensing). | 5 - 95% |
Battery Life: (Implied to be sufficient for intended use with external power availability) | 16 hours at 100% brightness. |
Low Battery Indications: Similar to predicate which is "LOWBAT" icon on LCD display. | On-screen icon & audible alarm |
Expected Sensor Life: (Implied to be sufficient for intended use) | > 900,000 %O2 Hours |
Low Oxygen Alarm Range: Similar to predicate which is 15%-99% (>1% lower than high alarm). | 18%-100% (>1% lower than high alarm) |
High Oxygen Alarm Range: Similar to predicate which is 16%-99% (>1% higher than low alarm). | 19%-105% (>1% lower than low alarm) |
Alarm Accuracy: Exact to display alarm value. | Exact to display alarm value |
Compliance with Standards: Including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 80601-2-55. | Met applicable standards |
Software Verification and Validation: (Implicitly met standards) | Performed |
Shelf-life / Aging: (Implicitly met standards) | Performed |
Auto-purge functionality: (Implicitly met standards) | Performed |
Battery Performance Testing: (Implicitly met standards) | Performed |
Study Details:
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- The document mentions "a series of non-clinical performance / bench testing" but does not specify the sample size for any of these tests nor the data provenance. It implies laboratory testing rather than patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided. The testing appears to be bench/performance testing against specifications and standards, not expert-adjudicated clinical data.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not provided.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC or human-in-the-loop study was done or mentioned. This is a standalone oxygen analyzer, not an AI-assisted diagnostic tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, performance testing was done on the device (OxyMinder) as a standalone unit. The "Oxygen Accuracy with Blender" test refers to its performance when integrated with a Bio-Med Devices air/oxygen blender, but it's still about the device's accuracy rather than human interpretation.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the bench testing, the "ground truth" would be the known and controlled oxygen concentrations and environmental conditions used during calibration and testing against established performance standards and specifications.
-
The sample size for the training set:
- Not applicable. This device is not described as using machine learning or AI that would require a 'training set' in the traditional sense. It's an oxygen gas analyzer, likely with embedded software and algorithms based on known physics and engineering principles for sensor readings.
-
How the ground truth for the training set was established:
- Not applicable, as there's no mention of a training set for machine learning.
Ask a specific question about this device
Page 1 of 1