K Number
K213948

Validate with FDA (Live)

Device Name
OxyMinder
Date Cleared
2022-03-17

(90 days)

Product Code
Regulation Number
868.1720
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OxyMinder is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. This device can be used in the hospital and subacute settings. The monitor is not intended as a life supporting device or for diagnostics.

Device Description

The OxyMinder is an air / oxygen blender mounted oxygen monitor capable of measuring the oxygen concentration from 18 to 100%. A Bio-Med specified oxygen sensor mounted to the blender outputs a voltage which is used by the OxyMinder to determine the concentration of oxygen. The OxyMinder calibrates at ambient air (21%) and 100% oxygen. The OxyMinder is software controlled. To measure the gas mixture of the blender the OxyMinder takes a sample of the gases to the sensor through a separate port and manifold which is separate from the gas pathway to the patient. This sample is then exhausted to the room. The OxyMinder sampling stream is not part of the gas pathway to the patient.

The OxyMinder is used for continuous monitoring of the concentration of oxygen delivered to patients via air oxygen blenders. The monitor provides the following features:

  • Continuously displays the concentrated Oxygen level delivered to a patient.
  • Accepts user inputs via touch screen or button (power button).
  • Provides visual alarm messages, and audible alarms.
  • Displays the current alarm setting levels (High and Low O2 alarms).
  • Provides on-screen configuration tools such as O2 sensor calibration, touchscreen calibration, audio test, etc.
  • Monitors and displays the battery level and power source.
  • Ensures clean hospital airlines by automatically purging system periodically.

The OxyMinder is designed to be mounted to a Bio-Med Devices air / oxygen blender via a manifold that houses the oxygen sensor and a solenoid.

In addition to the primary function of monitoring oxygen concentration, the OxyMinder provides an automatic gas supply line purge function.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for the "OxyMinder" device, an oxygen gas analyzer. It describes the device, its intended use, and compares it to a predicate device (Maxtec MaxO2ME). The document states that "A series of non-clinical performance / bench testing was performed" and lists the types of tests, concluding that "The test results met the applicable standards and are similar to the reported performance of the predicate device." However, the document does not provide detailed acceptance criteria or the specific results of these tests. It only broadly states that the performance was similar to the predicate and met applicable standards.

Therefore, many of the requested details about the study cannot be extracted from this document.

Here's a summary of what can be inferred or explicitly stated:

Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance (OxyMinder)
Measurement Range: Intended for use with an air/oxygen blender which will not deliver oxygen below 18%.18% – 100% O2
Resolution: (Implied to be sufficient for reading in 10% increments given blender knob resolution)Displayed to nearest whole integer
Accuracy and Linearity: Similar to predicate which is ±1% of full scale at constant temperature, RH and pressure when calibrated at full scale.±1.0% of full scale at constant temperature and pressure
Total Accuracy: Similar to predicate which is ±3% Actual oxygen level over full operating temperature range.±2.5% Actual oxygen level over the full operating temperature range
Response Time: Similar to predicate which is 90% of final value in approx. 15 seconds at 23°C.90% of final value in approximately 6 seconds
Warm-up Time: None required.None required
Operating Temperature: Similar to predicate which is 15°C – 40°C (59°F – 104°F).0° – 50°C [32° - 122° F]
Storage Temperature: Similar to predicate which is -15°C – 50°C (5°F – 122°F).0° - 40° C [32° - 104° F]
Atmospheric Pressure: Similar to predicate which is 800 - 1012 m Bars.345 - 2068 mBars
Humidity: Similar to predicate which is 0-95% (non-condensing).5 - 95%
Battery Life: (Implied to be sufficient for intended use with external power availability)16 hours at 100% brightness.
Low Battery Indications: Similar to predicate which is "LOWBAT" icon on LCD display.On-screen icon & audible alarm
Expected Sensor Life: (Implied to be sufficient for intended use)> 900,000 %O2 Hours
Low Oxygen Alarm Range: Similar to predicate which is 15%-99% (>1% lower than high alarm).18%-100% (>1% lower than high alarm)
High Oxygen Alarm Range: Similar to predicate which is 16%-99% (>1% higher than low alarm).19%-105% (>1% lower than low alarm)
Alarm Accuracy: Exact to display alarm value.Exact to display alarm value
Compliance with Standards: Including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 80601-2-55.Met applicable standards
Software Verification and Validation: (Implicitly met standards)Performed
Shelf-life / Aging: (Implicitly met standards)Performed
Auto-purge functionality: (Implicitly met standards)Performed
Battery Performance Testing: (Implicitly met standards)Performed

Study Details:

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • The document mentions "a series of non-clinical performance / bench testing" but does not specify the sample size for any of these tests nor the data provenance. It implies laboratory testing rather than patient data.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. The testing appears to be bench/performance testing against specifications and standards, not expert-adjudicated clinical data.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not provided.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC or human-in-the-loop study was done or mentioned. This is a standalone oxygen analyzer, not an AI-assisted diagnostic tool.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, performance testing was done on the device (OxyMinder) as a standalone unit. The "Oxygen Accuracy with Blender" test refers to its performance when integrated with a Bio-Med Devices air/oxygen blender, but it's still about the device's accuracy rather than human interpretation.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For the bench testing, the "ground truth" would be the known and controlled oxygen concentrations and environmental conditions used during calibration and testing against established performance standards and specifications.
  7. The sample size for the training set:

    • Not applicable. This device is not described as using machine learning or AI that would require a 'training set' in the traditional sense. It's an oxygen gas analyzer, likely with embedded software and algorithms based on known physics and engineering principles for sensor readings.
  8. How the ground truth for the training set was established:

    • Not applicable, as there's no mention of a training set for machine learning.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

March 17, 2022

Bio-Med Devices, Inc. % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704

Re: K213948

Trade/Device Name: OxyMinder Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: Class II Product Code: CCL Dated: February 25, 2022 Received: February 28, 2022

Dear Paul Dryden:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K213948

Device Name OxyMinder

Indications for Use (Describe)

The OxyMinder is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. This device can be used in the hospital and subacute settings. The monitor is not intended as a life supporting device or for diagnostics.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Date Prepared:

Bio-Med Devices, Inc. 61 Soundview Rd Guilford, CT 06437 USA Tel - 203-458-0202

Official Contact:Ken K Close - Regulatory Affairs Manager
Submission Correspondent:Paul Dryden
ProMedic, LLC
131 Bay Point Dr NE
St. Petersburg, FL 33704
Proprietary or Trade Name:OxyMinder
Common/Usual Name:Oxygen gas analyzer
Classification CFR:21 CFR 868.1720
Classification Code:CCL
Classification Name:Analyzer, Gas, Oxygen, Gaseous-Phase
Class:Class II
Predicate Device:Maxtec MaxO2ME (K153659)
Common/Usual Name:Oxygen gas analyzer
Classification CFR:21 CFR 868.1720
Classification Code:CCL
Classification Name:Analyzer, Gas, Oxygen, Gaseous-Phase
Class:Class II

Device Description:

The OxyMinder is an air / oxygen blender mounted oxygen monitor capable of measuring the oxygen concentration from 18 to 100%. A Bio-Med specified oxygen sensor mounted to the blender outputs a voltage which is used by the OxyMinder to determine the concentration of oxygen. The OxyMinder calibrates at ambient air (21%) and 100% oxygen. The OxyMinder is software controlled. To measure the gas mixture of the blender the OxyMinder takes a sample of the gases to the sensor through a separate port and manifold which is separate from the gas pathway to the patient. This sample is then exhausted to the room. The OxyMinder sampling stream is not part of the gas pathway to the patient.

The OxyMinder is used for continuous monitoring of the concentration of oxygen delivered to patients via air oxygen blenders. The monitor provides the following features:

  • Continuously displays the concentrated Oxygen level delivered to a patient. O
  • Accepts user inputs via touch screen or button (power button). O
  • Provides visual alarm messages, and audible alarms. O
  • Displays the current alarm setting levels (High and Low O2 alarms). O
  • Provides on-screen configuration tools such as O2 sensor calibration, touchscreen calibration, O audio test, etc.
  • Monitors and displays the battery level and power source. O
  • Ensures clean hospital airlines by automatically purging system periodically. o

{4}------------------------------------------------

510(k) Summary Page 2 of 7

The OxyMinder is designed to be mounted to a Bio-Med Devices air / oxygen blender via a manifold that houses the oxygen sensor and a solenoid.

In addition to the primary function of monitoring oxygen concentration, the OxyMinder provides an automatic gas supply line purge function.

Indications for Use:

The OxyMinder is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. This device can be used in the hospital and subacute settings. The monitor is not intended as a life supporting device or for diagnostics.

Patient Population:

The monitor is not population dependent. As it fits on a cleared Bio-Med blender has established the population.

Environments of use:

Hospital and subacute settings.

We present the proposed device vs. the predicate in the Table below.

{5}------------------------------------------------

510(k) Summary Page 104 of 7

AttributesPredicateSubjectDifferences
Maxtec MaxO2ME (K153659)Bio-Med Devices OxyMinder
Indications for UseThe MaxO2ME oxygen monitor isintended for continuous monitoring ofthe concentration of oxygen beingdelivered to patients ranging fromnewborns to adults. It can be used inthe pre-hospital, hospital and sub-acute settings. The MaxO2ME is notintended as a life supporting device.The OxyMinder is intended forcontinuous monitoring of theconcentration of oxygen beingdelivered to patients ranging fromnewborns to adults. This device canbe used in the hospital and subacutesettings. The monitor is notintended as a life supporting deviceor for diagnostics.Similar
Environments ofUsePre-hospital, hospital and sub-acutesettingsHospital and subacute settingsSimilar Subject device does not include pre-hospital which is not in its intended useBoth devices are MRI Unsafe
Patient Populationnewborns to adultsnewborns to adultsSimilar (based on indications)
MeasurementRange0.0 to 100%18% – 100% O2The subject device is only intended for use withan air/oxygen blender which will not deliveroxygen below 18%, thus this is the minimumrange for the OxyMinder.
Resolution0.1%Displayed to nearest whole integerThe OxyMinder verifies the proper oxygenconcentration from the a ir/oxygen blender. Theresolution of the a ir/oxygen blender knob is in10% increments. The predicate can be used as astandalone oxygen monitor where the resolutionof 0.1% is beyond the accuracy of the samplingcell.
Accuracy andLinearity±1% of full scale at constanttemperature, RH and pressure whencalibrated at full scale±1.0% of full scale at constanttemperature and pressureSimilar
Total Accuracy±3% Actual oxygen level over fulloperating temperature range±2.5% Actual oxygen level over thefull operating temperature rangeSimilar
Response Time90% of final value in approx. 15seconds at 23°C90% of final value in approximately6 secondsSimilar
Warm-up TimeNone requiredNone requiredSimilar
OperatingTemperature15°C – 40°C (59°F – 104°F)0° – 50°C [32° - 122° F]Similar
AttributesPredicateMaxtec MaxO2ME (K153659)SubjectBio-Med Devices OxyMinderDifferences
StorageTemperature-15°C – 50°C (5°F – 122°F)0° - 40° C [32° - 104° F]Similar
AtmosphericPressure800 - 1012 m Bars345 - 2068 mBarsSimilar
Humidity0-95% (non-condensing)5 - 95%Similar
Power requirements4 - AA Alkaline batteriesExternal DC and internalrechargeable lithium-ion batterySubject device includes external power andrechargeable, appropriate for its intended use.
Battery LifeApprox. 5000 hours, typical use16 hours at 100% brightness.Subject device is primarily used stationary(connected to air / oxygen blender) whereexternal power is available.
Low BatteryIndications"LOWBAT" icon on LCD displayOn-screen icon & audible alarmSimilar
Sensor TypeMaxtec MAX-550E galvanic fuel cellAnalytical IndustriesPSR-11-917-J10 (510(k K952736)Similar
Expected SensorLife> 1,500,000%O2 Hours, over 2 yearstypical application> 900,000%O2 HoursThe sensor manufacturer provides the followingspecifications.902,880 %O2 Hours = 60 months in thefollowing conditions: In air (20.9% O2) at 25°Cand 1 atm.Changes in Oxygen levels, temperature andpressure produce a proportional change in outputand an inversely proportional change inexpected life.
Alarm SystemsHigh/Low alarms, flashing yellowLEDsNominal975 Hz audio buzzer (IEC60601-1-8)Connect AC power (external powerdisconnected when battery is low)SimilarMore alarm functionality for the subject device
Low O2 alarm
High O2 alarm
Recalibrate O2 sensorReplace O2 sensorLow battery
AttributesPredicateMaxtec MaxO2ME (K153659)SubjectBio-Med Devices OxyMinderDifferences
Low Oxygen Alarm Range15%-99% (>1% lower than high alarm)18%-100% (>1% lower than high alarm)Similar
High Oxygen Alarm Range16%-99% (>1% higher than low alarm)19%-105% (>1% lower than low alarm)Similar
Alarm AccuracyExact to display alarm valueExact to display alarm valueSimilar
Dimensions3.6"(W) x 5.8"(H)x1.2”(D)7.86" W x 8.12" H x 4.39" DSimilar
WeightApprox. 0.89 lbs.1.5 lbs.Similar
Materials in Gas PathwayThe oxygen sensorNo materials in the patient gas pathwayThe OxyMinder sampling stream is sidestream off the patient gas pathway through a separate port and manifold than that which is used to deliver the gas to the patient.Similar except the oxygen sampling gas stream is not in the patient gas pathway.
AccessoriesDiverterTee adapter (15 mmm x 22 mm fittings)Galvanic cellMounting bracketsDC power adapterGalvanic cellPower AdaptorMounting Bracket
AutoPurgeNot available but has a manual purge when the blender is on.The OxyMinder has a scheduled purge feature to allow an automated purge of the gas sources connected to the blender when the device is not in use. This allows purging of equal amounts of flow of oxygen and air from the gas sources, ensuring that any contamination to the air and oxygen sources caused by backflow through the blender will be purgedThis feature is not part of the patient performance or risk analysis.It is used when the blender is left connected to the gas sources and not in use.

{6}------------------------------------------------

510(k) Summary Page 105 of 7

{7}------------------------------------------------

510(k) Summary Page 106 of 7

{8}------------------------------------------------

Substantial Equivalence Discussion and Rationale

The table above compares the key features of the proposed device with the identified predicate -Maxtec MaxO2ME (K153659). The comparison demonstrates that the proposed devices can be found to be substantially equivalent.

Indications for Use -

The indications for use are similar for the proposed device when compared to the predicate device.

Discussion - Both devices are indicated for use for monitoring oxygen levels. Both devices have equivalent range and accuracy specification. Both devices have similar alarm functionality. Minor difference does not raise different concerns of risk than the predicate.

Technology and construction -

The technology is identical to the predicate device using a galvanic fuel cell for measurement technology.

Discussion - The differences in how the gas is measured and additional features do not raise different concerns of safety or effectiveness compared to the predicate.

Environment of Use -

The environments of use are similar to predicate which are clinical settings. Discussion - The subject device does not include pre-hospital which is in keeping with its use with the compatible Bio-Med Devices air / oxygen blender.

Patient Population -

The patient population of the predicate device is newborns to adults as is the subject device. Discussion - The subject and predicate have identical patient populations

Non-Clinical Testing Summary -

A series of non-clinical performance / bench testing was performed that includes:

  • Shelf-life / Aging ●
  • Software Verification and Validation ●
  • IEC 60601-1 ●
  • IEC 60601-1-2
  • IEC 60601-1-8 ●
  • IEC 80601-2-55
  • Auto-purge functionality ●
  • Oxygen Accuracy with Blender ●
  • Battery Performance Testing ●

Discussion - The test results met the applicable standards and are similar to the reported performance of the predicate device.

Biocompatibility -

The predicate device has materials in the gas pathway, the subject device does not. Discussion - As the subject device has not patient contacting materials (indirect or gas pathway) biocompatibility testing is not applicable

{9}------------------------------------------------

Discussion of Differences –

There are no significant differences in critical function between the proposed device and the predicate device. The subject device is intended to specifically be used with only a Bio-Med Devices cleared air / oxygen blender. A secondary feature of the OxyMinder is the auto-purge feature, which mitigates gas line contamination in the hospital medical gas pipelines due to minimal backflow when air/oxygen blender is left connected and unused for extended periods of time. The predicate device is not limited to use with only an air / oxygen blender, has a manual purge feature when it is in use, and can be used in the pre-hospital environment.

The performance testing has demonstrated that the subject device met the applicable standard performance requirements. The above table plus the risk analysis do not identify any new or different risks compared to the predicate.

Substantial Equivalence Conclusion

The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.

§ 868.1720 Oxygen gas analyzer.

(a)
Identification. An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.(b)
Classification. Class II (performance standards).