Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K202990
    Device Name
    NinesMeasure
    Manufacturer
    Date Cleared
    2021-02-25

    (148 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    NinesMeasure

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NinesMeasure is a semi-automatic tool indicated for use by trained radiologists to aid in the analysis and review of adult thoracic CT images. NinesMeasure provides quantitative information about pulmonary nodule size on a single study or over the time course of several thoracic studies by providing long and short axis diameter measurements in the axial plane.

    Based on analysis of DICOM images and provided input from a radiologist, indicating the location of the pulmonary nodule, the device uses artificial intelligence algorithms to automatically perform the measurements, and allows the axial measurements to be displayed and reviewed. NinesMeasure is limited for use on solid pulmonary nodules.

    The device is intended to be used as a measurement tool by a trained radiologist and is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm a diagnosis. The device does not alter the original medical image.

    Device Description

    NinesMeasure is a semi-automatic, diagnostic patient imaging tool used to measure the size of selected pulmonary nodules in a radiological image. The software system is comprised of a set of software modules for performing image analysis at a specified image location to calculate measurements of pulmonary nodules on adult thoracic CT images. The system operates over a standard network interface and receives the DICOM images and coordinates of the pulmonary nodule to measure. The system then returns the measurements for the long and short axis diameters for review by a trained radiologist.

    NinesMeasure is designed to be used with a standard PACS, where the user can indicate a location of the pulmonary nodule to measure, and then review and edit the measurements on the DICOM image.

    The image analysis uses Artificial Intelligence (Al) technology to analyze chest CT images for computing the measurements. Specifically, the device utilizes a machine learning (ML) algorithm to compute segmentations of nodules, from which the long and short axis measurements are then calculated.

    AI/ML Overview

    The provided document describes the NinesMeasure device, a semi-automatic tool for measuring pulmonary nodule size on adult thoracic CT images. The document includes information about the device's performance testing.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents primary endpoints related to the normalized error in measuring nodule diameters. While explicit "acceptance criteria" (thresholds the device must meet to be considered effective) are not stated numerically, the narrative confirms that "The performance goals for the primary endpoints were met." This implies internal acceptance criteria for these normalized error values.

    Primary Endpoint - All NodulesReported Device Performance [95% CI]Implied Acceptance Criterion (Met)
    Normalized error on long axis diameter0.113 [Upper Bound 0.124]
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1