K Number
K202990
Device Name
NinesMeasure
Manufacturer
Date Cleared
2021-02-25

(148 days)

Product Code
Regulation Number
892.2050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

NinesMeasure is a semi-automatic tool indicated for use by trained radiologists to aid in the analysis and review of adult thoracic CT images. NinesMeasure provides quantitative information about pulmonary nodule size on a single study or over the time course of several thoracic studies by providing long and short axis diameter measurements in the axial plane.

Based on analysis of DICOM images and provided input from a radiologist, indicating the location of the pulmonary nodule, the device uses artificial intelligence algorithms to automatically perform the measurements, and allows the axial measurements to be displayed and reviewed. NinesMeasure is limited for use on solid pulmonary nodules.

The device is intended to be used as a measurement tool by a trained radiologist and is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm a diagnosis. The device does not alter the original medical image.

Device Description

NinesMeasure is a semi-automatic, diagnostic patient imaging tool used to measure the size of selected pulmonary nodules in a radiological image. The software system is comprised of a set of software modules for performing image analysis at a specified image location to calculate measurements of pulmonary nodules on adult thoracic CT images. The system operates over a standard network interface and receives the DICOM images and coordinates of the pulmonary nodule to measure. The system then returns the measurements for the long and short axis diameters for review by a trained radiologist.

NinesMeasure is designed to be used with a standard PACS, where the user can indicate a location of the pulmonary nodule to measure, and then review and edit the measurements on the DICOM image.

The image analysis uses Artificial Intelligence (Al) technology to analyze chest CT images for computing the measurements. Specifically, the device utilizes a machine learning (ML) algorithm to compute segmentations of nodules, from which the long and short axis measurements are then calculated.

AI/ML Overview

The provided document describes the NinesMeasure device, a semi-automatic tool for measuring pulmonary nodule size on adult thoracic CT images. The document includes information about the device's performance testing.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document presents primary endpoints related to the normalized error in measuring nodule diameters. While explicit "acceptance criteria" (thresholds the device must meet to be considered effective) are not stated numerically, the narrative confirms that "The performance goals for the primary endpoints were met." This implies internal acceptance criteria for these normalized error values.

Primary Endpoint - All NodulesReported Device Performance [95% CI]Implied Acceptance Criterion (Met)
Normalized error on long axis diameter0.113 [Upper Bound 0.124]< 0.124 (or similar, met)
Normalized error on short axis diameter0.131 [Upper Bound 0.143]< 0.143 (or similar, met)

Stratified by Nodule Size:

Nodule SizeNumber of NodulesReported Normalized error on Long Axis Diameter [95% CI]Reported Normalized error on Short Axis Diameter [95% CI]Implied Acceptance Criterion (Met) for Long AxisImplied Acceptance Criterion (Met) for Short Axis
3-6 mm1000.104 [Upper Bound: 0.119]0.123 [Upper Bound: 0.139]< 0.119 (or similar, met)< 0.139 (or similar, met)
6-8 mm630.119 [Upper Bound: 0.138]0.143 [Upper Bound: 0.161]< 0.138 (or similar, met)< 0.161 (or similar, met)
8-10 mm460.13 [Upper Bound: 0.156]0.133 [Upper Bound: 0.166]< 0.156 (or similar, met)< 0.166 (or similar, met)

2. Sample Size and Data Provenance for the Test Set

  • Sample Size for Test Set:
    • All Nodules: The total number of nodules can be inferred by summing the stratified categories: 100 (3-6mm) + 63 (6-8mm) + 46 (8-10mm) = 209 nodules.
    • The study used cases that were "diverse, and included 3 different major scanner manufacturers, 7 different scanner models, 11 different clinical sites."
  • Data Provenance: The study was a retrospective, multi-center image comparison study. The document does not specify the country of origin for the data.

3. Number of Experts and Qualifications for Ground Truth

The document does not explicitly state the "number of experts used to establish the ground truth" or their specific qualifications (e.g., "Radiologist with 10 years of experience"). However, the device "is limited for use by trained radiologists" and relies on "provided input from a radiologist, indicating the location of the pulmonary nodule." This strongly suggests that ground truth measurements would likely be derived from expert radiologist annotations or measurements.

4. Adjudication Method for the Test Set

The document does not describe the specific adjudication method (e.g., 2+1, 3+1) used to establish the ground truth measurements for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study was not explicitly mentioned or detailed. The study described ("retrospective, multi-center image comparison study") evaluates the device's performance in measuring nodules, but it doesn't compare human readers with and without AI assistance to quantify improvement. The device is a "semi-automatic tool" and requires "provided input from a radiologist," implying a human-in-the-loop design, but the study focuses on the device's measurement accuracy after radiologist input, not the radiologist's performance change.

6. Standalone Performance (Algorithm Only)

The study describes the performance of the NinesMeasure device, which "uses artificial intelligence algorithms to automatically perform the measurements, and allows the axial measurements to be displayed and reviewed."
Given the description "semi-automatic tool" and "provided input from a radiologist, indicating the location of the pulmonary nodule," the reported performance metrics likely represent the algorithm's performance after the initial human input (locating the nodule). It does not appear to be a purely standalone (algorithm-only, end-to-end detection and measurement) performance evaluation. The algorithm performs the measurement after the radiologist pinpoints the nodule.

7. Type of Ground Truth Used

Based on the context of measuring pulmonary nodule size, the ground truth would most likely be expert consensus measurements (manual measurements performed by radiologists or other imaging experts) of the nodule diameters. Pathology or outcomes data are not typically used for direct measurement accuracy ground truth in this context.

8. Sample Size for the Training Set

The document does not specify the sample size used for the training set. It only describes the validation (test) set.

9. How the Ground Truth for the Training Set Was Established

The document does not provide details on how the ground truth for the training set was established. Given the nature of a machine learning (ML) algorithm for segmentation and measurement, it would typically involve expertly annotated images (manual segmentations or measurements) on a separate dataset from the test set.

{0}------------------------------------------------

February 25, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Nines, Inc. % John J. Smith, M.D., J.D. Regulatory Counsel Hogan Lovells US LLP 555 13th Street. NW WASHINGTON DC 20004

Re: K202990

Trade/Device Name: NinesMeasure Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 22, 2021 Received: January 22, 2021

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

  1. for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for

combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K202990

Device Name

NinesMeasure

Indications for Use (Describe)

NinesMeasure is a semi-automatic tool indicated for use by trained radiologists to aid in the analysis and review of adult thoracic CT images. NinesMeasure provides quantitative information about pulmonary nodule size on a single study or over the time course of several thoracic studies by providing long and short axis diameter measurements in the axial plane.

Based on analysis of DICOM images and provided input from a radiologist, indicating the location of the pulmonary nodule, the device uses artificial intelligence algorithms to automatically perform the measurements, and allows the axial measurements to be displayed and reviewed. NinesMeasure is limited for use on solid pulmonary nodules.

The device is intended to be used as a measurement tool by a trained radiologist and is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm a diagnosis. The device does not alter the original medical image.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

510(k) SUMMARY Nines, Inc.'s NinesMeasure K202990

Submitter:

Nines, Inc 329 Alma Street Palo Alto, CA 94301

Contact Person:

Dr. Russell Stewart Phone: 650 924 6159 russell@nines.com

Date Prepared: January 22, 2021

Name of Device: NinesMeasure

Classification Name: System, Image processing, Radiological

Regulatory Class: Class II

Product Code: LLZ

Predicate Device: Philips Medical Systems Nederland B.V.'s Lung Nodule Assessment and Comparison Option (LNA) (K162484)

Device Description

NinesMeasure is a semi-automatic, diagnostic patient imaging tool used to measure the size of selected pulmonary nodules in a radiological image. The software system is comprised of a set of software modules for performing image analysis at a specified image location to calculate measurements of pulmonary nodules on adult thoracic CT images. The system operates over a standard network interface and receives the DICOM images and coordinates of the pulmonary nodule to measure. The system then returns the measurements for the long and short axis diameters for review by a trained radiologist.

NinesMeasure is designed to be used with a standard PACS, where the user can indicate a location of the pulmonary nodule to measure, and then review and edit the measurements on the DICOM image.

The image analysis uses Artificial Intelligence (Al) technology to analyze chest CT images for computing the measurements. Specifically, the device utilizes a machine learning (ML) algorithm to compute segmentations of nodules, from which the long and short axis measurements are then calculated.

{4}------------------------------------------------

Intended Use / Indications for Use

NinesMeasure is a semi-automatic tool indicated for use by trained radiologists to aid in the analysis and review of adult thoracic CT images. NinesMeasure provides quantitative information about pulmonary nodule size on a single study or over the time course of several thoracic studies by providing long and short axis diameter measurements in the axial plane.

Based on analysis of DICOM images and provided input from a radiologist, indicating the location of the pulmonary nodule, the device uses artificial intelligence algorithms to automatically perform the measurements, and allows the axial measurements to be displayed and reviewed. NinesMeasure is limited for use on solid pulmonary nodules.

The device is intended to be used as a measurement tool by a trained radiologist and is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm a diagnosis. The device does not alter the original medical image.

Summary of Technological Characteristics

The NinesMeasure has similar technological characteristics as the predicate. Both devices utilize image processing algorithms that calculate pulmonary nodule measurements and return the measurements to the workstation. Although the predicate is cleared for multiple features in addition to pulmonary nodule measurements, these minor differences do not impact the safety of the subject device.

NinesMeasureK202990Philips Medical Systems'Lung Nodule Assessment andComparison Option(LNA)(K162484)
Device Classification NameSystem, Imageprocessing,RadiologicalSystem, Imageprocessing,Radiological
Device ClassClass IIClass II
Classification PanelRadiologyRadiology
Product CodeLLZLLZ, JAK
RegulationDescriptionRadiologicalImage ProcessingSoftwareRadiologicalImage ProcessingSoftware
RegulationNumber21 CFR 892.205021 CFR 892.205021 CFR 892.1750
Indications for UseNinesMeasure is a semi-automatic tool indicated for useby trained radiologists to aid inthe analysis and review of adultthoracic CT images.NinesMeasure providesquantitative information aboutpulmonary nodule size on aThe Lung Nodule Assessmentand Comparison Option isintended for use as a diagnosticpatient-imaging tool. It isintended for the review andanalysis of thoracic CT images,providing quantitative andcharacterizing information about
single study or over the timecourse of several thoracicstudies by providing long andshort axis diametermeasurements in the axialplane.Based on analysis of DICOMimages and provided input froma radiologist, indicating thelocation of the pulmonarynodule, the device uses artificialintelligence algorithms toautomatically perform themeasurements, and allows theaxial measurements to bedisplayed and reviewed.NinesMeasure is limited for useon solid pulmonary nodules.The device is intended to beused as a measurement tool bya trained radiologist and islimited to analysis of imagingdata and should not be used in-lieu of full patient evaluation orrelied upon to make or confirm adiagnosis. The device does notalter the original medical image.nodules in the lung in a singlestudy, or over the time course ofseveral thoracic studies.Characterizations includediameter, volume and volumeover time. The systemautomatically performs themeasurements, allowing lungnodules and measurements tobe displayed.
User PopulationRadiologistsRadiologists and Technologist
Technological CharacteristicsImage processing algorithmscomputing pulmonary nodulemeasurements and returningcomputed measurements to theworkstation.Image processing algorithmscomputing pulmonary nodulemeasurements and returningcomputed measurements to theworkstation.
ComponentsImage processing algorithms fornodule measurement-Image processing algorithms-Display, comparison, and riskcalculations
Anatomical region of interestChestChest
Features-long axis measurement-short axis measurement(perpendicular to long axis)-long axis measurement-short axis measurement(perpendicular to long axis)-Average/Max 3D/Effectivediameter (mm)-Volume (mm³)-Mean Densities (HU)-Segmentation of lung airway,lungs and lung lobes-Single click lung nodulesegmentation-Nodule Characteristics-Comparison and matching-Automatic calculation of
doubling time, percent andabsolute change of all numericalparameters-Reporting results functionsincluding dictation table, patientrelated information, LungRads,and Risk Calculator-Printing option

A table comparing the key features of the subject and predicate device is provided below.

{5}------------------------------------------------

{6}------------------------------------------------

Performance Testing

Nines performed software verification and validation testing that covers the performance of the algorithms, as well as the performance of the software and its components. In all instances, NinesMeasure functioned as intended and expected.

The algorithm performance was validated with a retrospective, multi-center image comparison study. The study was performed to evaluate the NinesMeasure device and demonstrate the product's performance as a workflow tool for pulmonary nodule measurement consistent with the proposed indications for use. The test dataset was diverse, and included 3 different major scanner manufacturers, 7 different scanner models, 11 different clinical sites.

The primary endpoints of the algorithm are listed below:

Primary Endpoint - All NodulesResult
Normalized error on long axis diameter[95% CI]0.113[Upper Bound 0.124]
Normalized error on short axis diameter[95% CI]0.131[Upper Bound 0.143]

The primary endpoint stratified by nodule size is listed below:

Nodule sizeNumber of nodulesNormalized error onlong axis diameter[95% CI]Normalized error onshort axis diameter[95% CI]
3-6 mm1000.104 [Upper Bound: 0.119]0.123 [Upper Bound: 0.139]
6-8 mm630.119 [Upper Bound: 0.138]0.143 [Upper Bound: 0.161]
8-10 mm460.13 [Upper Bound: 0.156]0.133 [Upper Bound: 0.166]

The performance goals for the primary endpoints were met.

Based on the clinical performance as documented in the clinical study, the subject software has a safety and effectiveness profile that is similar to the predicate device.

{7}------------------------------------------------

Conclusions

NinesMeasure has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the technological differences between NinesMeasure and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that NinesMeasure performs as intended. Thus, NinesMeasure is substantially equivalent.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).