Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K112565
    Manufacturer
    Date Cleared
    2011-12-06

    (95 days)

    Product Code
    Regulation Number
    870.4210
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NovaPort® One Vascular Access cannula with its associated Introducer Kit is a Novalung coated percutaneous cannula intended to be used to cannulate blood vessels in patients requiring extracorporeal support such as cardiopulmonary bypass. The application of NovaPort® One cannula is limited to < 6 hours.

    Device Description

    The NovaPort® One Vascular Access cannula with Novalung coating are a wire reinforced, thin walled femoral cannula made of polyurethane. A cone without wire reinforcement allows clamping and serves as transition to a connector. The beveled tip is not a separate piece and is part of the whole cannula body. The NovaPort® one cannulas are inserted percutaneously into vessels. Insertion depth marks aid in positioning the cannula. They have a beveled tip that is made as part of the PVC body. The Cannula is intended to be used to cannulate vessels, perfuse vessels or organs and/or connect with accessory extracorporeal equipment. The cannula introducer is intended to facilitate proper insertion and placement of the appropriate sized cannula within the vessel for cardiopulmonary bypass. All surfaces of the NovaPort® One which come into contact with blood are protected with the Novalung Coating. This coating is a bioactive, stable, biocompatible and non-thrombogenic surface. This coating consists of high molecular weight heparin from pig mucosa bound covalently and ionically to immobilized polypeptides. NovaPort® one cannula is available in insertion lengths of 90 mm and with outer diameters of 13 F, 17 F, 19 F and 21 F. All are supplied sterile with Novalung coated, non-pyrogenic, and are single use.

    AI/ML Overview

    The provided 510(k) summary for the Novalung® GmbH NovaPort® One Vascular Access device primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing specific performance criteria through a detailed clinical or standalone performance study as would be seen for AI/ML devices.

    Here's an analysis based on the information provided, emphasizing what is present and what is absent from the request's specific questions regarding acceptance criteria and study details:


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for a medical device (a vascular access cannula) and not an AI/ML diagnostic tool, there are no specific diagnostic "performance metrics" like sensitivity, specificity, or AUC. Instead, the acceptance is based on demonstrating equivalence to predicate devices across various attributes. The "acceptance criteria" here implicitly refer to meeting the expectations for functional and safety performance that allow for declaration of substantial equivalence.

    Acceptance Criteria Category (Implied)Reported Device Performance (Summary in 510(k))
    Intended UseIdentical to predicate devices
    Indications for UseIdentical to predicate devices, limited to < 6 hours. Novalung coating indications same as Bioline coating (K102532) and Novalung sLA (K072362).
    Fundamental TechnologyIdentical to predicate devices
    Design CharacteristicsMinor differences from predicate devices, raising no new safety/effectiveness issues. Wire-reinforced, thin-walled femoral cannula made of polyurethane, with Novalung coating. Beveled tip part of cannula body.
    Generic Materials of ConstructionIdentical to predicate devices (polyurethane). Novalung coating made of high molecular weight heparin from pig mucosa bound to immobilized polypeptides.
    Operational CharacteristicsIdentical to predicate devices
    Pressure/Burst ResistanceTested and found acceptable
    Simulated UseTested and found acceptable
    Kink ResistanceTested and found acceptable
    Tensile StrengthTested and found acceptable
    Flow CharacteristicsTested and found acceptable
    HemolysisTested and found acceptable
    BiocompatibilityTested and found acceptable
    Coating ValidationTested and found acceptable
    SterilitySupplied sterile
    Non-pyrogenicSupplied non-pyrogenic
    Single UseDesigned for single use

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "The NovaPort® One Vascular Access was subjected to numerous tests and comparisons to the predicate device: testing included pressure/burst, simulated use, kink resistance, tensile strength, flow characteristics, hemolysis, biocompatibility, and validation testing to coating."

    • Sample Size: The specific sample sizes for these various engineering and bench tests are not provided in the 510(k) summary. These would typically be detailed in the full test reports submitted to the FDA, not in the summary.
    • Data Provenance: The tests are described as "numerous tests," indicating they were likely benchtop/laboratory studies conducted by Novalung GmbH (likely in Germany, given the company's location) rather than human clinical trials. There is no mention of country of origin for any human data (as no human data appears to be used for acceptance) or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This question is not applicable in the context of this 510(k) summary. "Ground truth" established by experts (e.g., radiologists) is relevant for diagnostic devices where human expert interpretation is the gold standard for comparison (e.g., in AI/ML tasks). For this vascular access cannula, the "truth" is established through physical and material property testing, and comparison to the known performance of predicate devices. There's no mention of expert review panels for determining ground truth in the way described.


    4. Adjudication Method for the Test Set

    This question is not applicable for the reasons stated above. Adjudication methods (like 2+1 or 3+1) are used to resolve disagreements among human experts in diagnostic ground truth establishment, which isn't part of this device's acceptance testing.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or imaging devices to assess how a device (often with AI assistance) impacts human reader performance. The NovaPort® One is a physical medical device (cannula), not a diagnostic tool requiring human interpretation.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm-only) performance study was not done. This concept is entirely specific to AI/ML or autonomous diagnostic algorithms. The acceptance testing for the NovaPort® One involves physical performance evaluations and material characterization, not algorithmic performance.


    7. The Type of Ground Truth Used

    For the NovaPort® One Vascular Access, the "ground truth" for its acceptance testing is based on:

    • Engineering specifications and regulatory standards: For tests like pressure/burst, tensile strength, kink resistance, and flow characteristics.
    • Established biocompatibility standards: For biocompatibility testing.
    • Predicate device performance: The core of a 510(k) relies on demonstrating that the new device performs as safely and effectively as a legally marketed predicate device. This "ground truth" is therefore derived from the established safety and efficacy profiles of the predicate devices (Bio-Medicus Multi-Stage Venous Femoral Percutaneous Kit K052524 and One Piece Pediatric Arterial Cannula 7000 Models K024069).

    8. The Sample Size for the Training Set

    Not applicable. A "training set" refers to data used to train an AI/ML algorithm. This device is a physical medical instrument, not an AI/ML algorithm, so there is no training set.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI/ML algorithm, this question is not relevant to the NovaPort® One Vascular Access device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1