Search Results
Found 2 results
510(k) Data Aggregation
(188 days)
MultiStage Balloon Dilatation Catheter
The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.
The MultiStage Balloon Dilatation Catheters are indicated for use to endoscopically dilate strictures of the alimentary tract and the Sphincter of Oddi with or without prior sphincterotomy. It consists of an inflatable balloon on a catheter shaft with multiple lumens for inflation and passage of a guidewire. The MultiStage balloon is designed with three-in-one technology and provides successive, gradual dilation of strictures. The balloon is made of Pebax material. The MultiStage Balloon Dilatation is designed to pass over a 0.035in (0.89mm) guidewire through its guidewire lumen or through an minimum working channel of a 2.8mm alimentary tract, or a 3.7mm biliary tract. The hub is made of polycarbonate (PC). Two radiopaque tantalum marker bands are positioned within the balloon shoulders to provide visual reference points fluoroscopically for balloon positioning within the stricture. Accordingly, product models, are divided into those with a wireguided (MSW) and non-wireguided (MSO) component. The MSW model allows the guidewire and catheter to be introduced into the endoscope together. The MSO model as a traditional operation method, the guidewire is first inserted into the endoscope for positioning and then the catheter is inserted through the guidewire for good passability. The physician can choose the model as needed.
The provided text is a 510(k) summary for a medical device called the "MultiStage Balloon Dilatation Catheter." This document describes the device, its intended use, and its comparison to a predicate device to establish substantial equivalence for regulatory clearance. It does not contain information about an AI/ML algorithm or a study proving that an AI/ML device meets acceptance criteria.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance for an AI/ML device.
- Sample sizes or data provenance for an AI/ML test set.
- Number of experts or their qualifications for establishing ground truth for an AI/ML test set.
- Adjudication method for an AI/ML test set.
- MRMC comparative effectiveness study results for an AI/ML device.
- Standalone performance of an AI/ML algorithm.
- Type of ground truth used for an AI/ML study.
- Sample size for the training set of an AI/ML algorithm.
- How the ground truth for an AI/ML training set was established.
The document focuses on the non-clinical testing of a physical medical device (balloon catheter), which includes:
- Biocompatibility testing (Cytotoxicity, Sensitization, Intracutaneous reactivity test)
- Shelf life testing
- Sterilization validation
- Performance characteristics (Visual Inspection, Bacteriostasis performance of pouch, Dye penetration, Tension test, Guidewire compatibility, Channel compatibility, Product Dimensional Inspection, Compliance, Balloon fatigue, Balloon rated burst pressure (RBP), Peak tensile, Kink resistance, Leakage Test, Hydratability, Hub Corrosion resistance, X-ray detectability, Chemical properties).
All of these non-clinical tests were reported to have "passed." The document explicitly states: "No clinical study is included in this submission."
For an AI/ML-based medical device, the information you requested would typically be found in sections detailing performance studies, often referred to as "clinical performance" or "algorithm performance" studies, which are absent here.
Ask a specific question about this device
(222 days)
Multistage Balloon Dilatation Catheter
MultiStage balloon dilatation catheter is intended to be used to dilate strictures located in the airway tree.
The MultiStage Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree. Balloons designed with three-in-one technology and provides successive, gradual dilation of strictures. The balloon material is made of Pebax material. The MultiStage Balloon Dilatation Catheter designed to pass over a 0.035in (0.89mm) guidewire through its guidewire lumen or through a minimum 2.8mm working channel bronchoscope. The Catheter hub is made of polycarbonate (PC). Two radiopaque tantalum marker bands are positioned within the balloon shoulders to provide visual reference points fluoroscopically for balloon positioning within the stricture.
This document is a 510(k) Summary for a medical device called the "MultiStage Balloon Dilatation Catheter." It is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. As such, it does not describe specific acceptance criteria and a study proving the device meets those criteria in the way one might expect for a novel AI/software device.
Instead, the document focuses on demonstrating that the proposed device is "substantially equivalent" to a predicate device. This is achieved by showing similar intended use, technological characteristics, and principles of operation, and by performing non-clinical tests to ensure safety and effectiveness.
Here's an attempt to extract the requested information, understanding that it will be framed within the context of substantial equivalence rather than a direct performance study against defined acceptance criteria for, for example, a diagnostic accuracy claim:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" in terms of specific quantitative thresholds to be met for diagnostic performance (as would be typical for an AI/software device making a diagnostic claim). Instead, the "acceptance criteria" are implied by the standards and performance characteristics tested for a physical medical device, and the "reported device performance" is that the device passed these tests and is considered "safe and effective."
Acceptance Criteria (Implied by Standards & Tests) | Reported Device Performance |
---|---|
Biocompatibility (ISO 10993-1) | All tests passed: No cytotoxicity, not an irritant, not a sensitizer. |
Shelf Life (ASTM F1980) | All tests passed. |
Sterilization (ISO 11607-1:2019, ISO 11135:2014) | Sterile with SAL of 10⁻⁶; all tests passed. |
Visual Inspection | (Implied: Passed standard visual inspection) |
Bacteriostasis performance of pouch | (Implied: Passed) |
Dye penetration | (Implied: Passed) |
Pouch | (Implied: Passed) |
Tension test | (Implied: Passed) |
Guidewire compatibility test | (Implied: Passed) |
Channel compatibility | (Implied: Passed) |
Product Dimensional Inspection | (Implied: Passed) |
Balloon fatigue; No leakage and damage when inflation | (Implied: Passed) |
Balloon inflation and deflation time | (Implied: Passed) |
Balloon rated burst pressure (RBP) | (Implied: Passed) |
Peak tensile | (Implied: Passed) |
Leakage Test | (Implied: Passed) |
Hydratability | (Implied: Passed) |
Corrosion resistance | (Implied: Passed) |
X-ray detectability | (Implied: Passed; marker bands provide visual reference) |
Torsion test | (Implied: Passed) |
Overall Safety and Effectiveness | Considered safe and effective for its intended use. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists "non-clinical tests" which include physical, chemical, and biological performance testing of the device and its components. The sample sizes for these specific engineering and biocompatibility tests are not disclosed in this summary. The data provenance is primarily from laboratory testing of the manufactured device. There is no mention of patient data or its origin.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable to this type of submission. The device is a physical medical instrument, not a diagnostic AI/software that requires expert-established ground truth from clinical data. The "ground truth" for the non-clinical tests is established by adherence to recognized international standards (e.g., ISO, ASTM) and engineering principles.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human readers/interpreters to establish a consensus ground truth for diagnostic decisions. This submission focuses on non-clinical performance and substantial equivalence to a predicate device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic devices, not for a physical balloon dilatation catheter.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to AI/software performance. The device is a physical catheter, so a "standalone" algorithm performance study is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical tests, the "ground truth" is defined by the specifications, requirements, and methods outlined in the referenced international standards (e.g., ISO 10993-1 for biocompatibility, ASTM F1980 for shelf life, ISO 11135 and ISO 11607-1 for sterilization). There is no clinical ground truth involving patient outcomes or expert consensus for this type of submission.
8. The sample size for the training set
This is not applicable. The device is a physical medical instrument, not an AI/software device that requires a training set for model development.
9. How the ground truth for the training set was established
This is not applicable as there is no training set for a physical medical device.
Ask a specific question about this device
Page 1 of 1