K Number
K222663
Date Cleared
2023-03-09

(188 days)

Product Code
Regulation Number
876.5010
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.

Device Description

The MultiStage Balloon Dilatation Catheters are indicated for use to endoscopically dilate strictures of the alimentary tract and the Sphincter of Oddi with or without prior sphincterotomy. It consists of an inflatable balloon on a catheter shaft with multiple lumens for inflation and passage of a guidewire. The MultiStage balloon is designed with three-in-one technology and provides successive, gradual dilation of strictures. The balloon is made of Pebax material. The MultiStage Balloon Dilatation is designed to pass over a 0.035in (0.89mm) guidewire through its guidewire lumen or through an minimum working channel of a 2.8mm alimentary tract, or a 3.7mm biliary tract. The hub is made of polycarbonate (PC). Two radiopaque tantalum marker bands are positioned within the balloon shoulders to provide visual reference points fluoroscopically for balloon positioning within the stricture. Accordingly, product models, are divided into those with a wireguided (MSW) and non-wireguided (MSO) component. The MSW model allows the guidewire and catheter to be introduced into the endoscope together. The MSO model as a traditional operation method, the guidewire is first inserted into the endoscope for positioning and then the catheter is inserted through the guidewire for good passability. The physician can choose the model as needed.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called the "MultiStage Balloon Dilatation Catheter." This document describes the device, its intended use, and its comparison to a predicate device to establish substantial equivalence for regulatory clearance. It does not contain information about an AI/ML algorithm or a study proving that an AI/ML device meets acceptance criteria.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and reported device performance for an AI/ML device.
  2. Sample sizes or data provenance for an AI/ML test set.
  3. Number of experts or their qualifications for establishing ground truth for an AI/ML test set.
  4. Adjudication method for an AI/ML test set.
  5. MRMC comparative effectiveness study results for an AI/ML device.
  6. Standalone performance of an AI/ML algorithm.
  7. Type of ground truth used for an AI/ML study.
  8. Sample size for the training set of an AI/ML algorithm.
  9. How the ground truth for an AI/ML training set was established.

The document focuses on the non-clinical testing of a physical medical device (balloon catheter), which includes:

  • Biocompatibility testing (Cytotoxicity, Sensitization, Intracutaneous reactivity test)
  • Shelf life testing
  • Sterilization validation
  • Performance characteristics (Visual Inspection, Bacteriostasis performance of pouch, Dye penetration, Tension test, Guidewire compatibility, Channel compatibility, Product Dimensional Inspection, Compliance, Balloon fatigue, Balloon rated burst pressure (RBP), Peak tensile, Kink resistance, Leakage Test, Hydratability, Hub Corrosion resistance, X-ray detectability, Chemical properties).

All of these non-clinical tests were reported to have "passed." The document explicitly states: "No clinical study is included in this submission."

For an AI/ML-based medical device, the information you requested would typically be found in sections detailing performance studies, often referred to as "clinical performance" or "algorithm performance" studies, which are absent here.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 9, 2023

Dongguan TT Medical, Inc. % Mingzi Hussey Principal Regulatory Consultant Zi-medical. Inc. 93 Springs Rd Bedford, MA 01730

Re: K222663

Trade/Device Name: MultiStage Balloon Dilatation Catheter Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: Class II Product Code: FGE Dated: February 10, 2023 Received: February 10, 2023

Dear Mingzi Hussey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any

{1}------------------------------------------------

Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image contains the text "Je An -S" in a large, sans-serif font. The letters are black and bolded, and they are arranged horizontally. The background is white, with a faint watermark-like design in a light blue color.

Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K222663

Device Name MultiStage Balloon Dilatation Catheter

Indications for Use (Describe)

The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.

Type of Use (Select one or both, as applicable)
Repackaging (Part 201.22(a) Subject Shipment) Own-Use Controlled (201.22(a) Subject Shipment) Repackaging (Part 201.22(a) Subject Shipment) Own-Use Controlled (201.22(a) Subject Shipment)
Repackaging (Part 201.22(a) Subject Shipment)
Own-Use Controlled (201.22(a) Subject Shipment)

Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary

[As required by 21 CFR 807.92]

1. Submission Sponsor

Dongguan TT Medical, Inc. Address: BLD#1, 1 Taoyuan Road, Songshan Lake, Dongguan, Guangdong, 523808 China Phone: +86 15553341910 Contact: Yuying Bi, Regulatory Affairs Specialist Email: ybi @ttmedicalinc.com

2. Submission Correspondent

Mingzi Hussey Zi-medical, Inc. Address: 93 Springs Rd, Bedford, MA 01730 US Phone: 206-981-0675 Email: mingzi@zi-medical.com

3. Date Prepared

Aug 10th, 2022

4. Device Identification

Trade Name: MultiStage Balloon Dilatation Catheter Regulation name: Biliary Catheter And Accessories Product Code: FGE Regulation Number: 21 CFR 876.5010 Regulation Class: Class II Review Panel: Gastroenterology/Urology

5. Predicate Devices

The proposed device is substantially equivalent to the following predicate device:

ApplicantDevice name510(k) NumberProduct code
Boston ScientificCorpCRETM Wireguided BalloonDilatation CatheterK112994FGE

6. Device Description

The MultiStage Balloon Dilatation Catheters are indicated for use to endoscopically dilate strictures of the alimentary tract and the Sphincter of Oddi with or without prior sphincterotomy. It consists of an inflatable balloon on a catheter shaft with multiple lumens for inflation and passage of a guidewire.

{4}------------------------------------------------

The MultiStage balloon is designed with three-in-one technology and provides successive, gradual dilation of strictures. The balloon is made of Pebax material. The MultiStage Balloon Dilatation is designed to pass over a 0.035in (0.89mm) guidewire through its guidewire lumen or through an minimum working channel of a 2.8mm alimentary tract, or a 3.7mm biliary tract. The hub is made of polycarbonate (PC). Two radiopaque tantalum marker bands are positioned within the balloon shoulders to provide visual reference points fluoroscopically for balloon positioning within the stricture. Accordingly, product models, are divided into those with a wireguided (MSW) and non-wireguided (MSO) component.

The MSW model allows the guidewire and catheter to be introduced into the endoscope together. The MSO model as a traditional operation method, the guidewire is first inserted into the endoscope for positioning and then the catheter is inserted through the guidewire for good passability. The physician can choose the model as needed.

7. Indication for Use Statement

The MultiStage Balloon Dilatation Catheters are indicated for use in adult and adolescent populations to endoscopically dilate strictures of the alimentary tract. Also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy.

8. Comparison of Technological Characteristics

The proposed device is compared with the predicate devices in terms of intended use, technological characteristics and principles of operation. The proposed device shares the similar indications for use, similar materials, same device operation, and same overall technical capabilities as the predicate device. The MSW model has same components with the predicate device while the MSO model does not contain a guidewire for the users to install a guidewire by their choice. The proposed device also has the same standards and requirements as the predicate device, and has been verified for its safety and effectiveness to establish substantially equivalence.

Nominal Balloon Diameter
NominalLength6mm8mm10mm12mm15mm18mm
30mmMSO063008MSO083008MSO103008MSO123008MSO153008MSO183008
30mmMSO063017MSO083017MSO103017MSO123017MSO153017MSO183017
30mmMSO063023MSO083023MSO103023MSO123023MSO153023MSO183023
55mmMSO065508MSO085508MSO105508MSO125508MSO155508MSO185508
55mmMSO065517MSO085517MSO105517MSO125517MSO155517MSO185517
55mmMSO065523MSO085523MSO105523MSO125523MSO155523MSO185523
30mmMSW063008MSW083008MSW103008MSW123008MSW153008MSW183008
30mmMSW063017MSW083017MSW103017MSW123017MSW153017MSW183017

MultiStage Balloon Dilatation Catheter has 72 models:

{5}------------------------------------------------

30mmMSW063023MSW083023MSW103023MSW123023MSW153023MSW183023
55mmMSW065508MSW085508MSW105508MSW125508MSW155508MSW185508
55mmMSW065517MSW085517MSW105517MSW125517MSW155517MSW185517
55mmMSW065523MSW085523MSW105523MSW125523MSW155523MSW185523

9. Description of Non-clinical Testing

The non-clinical tests of the subject device and predicate device are in compliance with the following standards and guidance.

The biocompatibility evaluation for MultiStage Balloon Dilatation Catheter was conducted in accordance with ISO 10993-1.

  • · Cytotoxicity (ISO 10993-5)
  • · Sensitization (ISO 10993-10)
  • · Intracutaneous reactivity test (ISO 10993-10)

Result: All tests were passed.

Shelf life for MultiStage Balloon Dilatation Catheter test is conducted based on ASTM F1980: · Shelf life test report

  • Shell life test report

Result: All tests were passed.

Sterile barrier systems for MultiStage Balloon Dilatation Catheter were evaluated in accordance with ISO 11607-1:2019.

Sterilization Process has been validated accordance with ISO 11135:2014.

Result: The MultiStage Balloon Dilatation Catheter is sterile with a validation of

of the sterilization process that demonstrates achievement of the required SAL of 10% and is in accordance with the half cycle overkill approach.

Technological characteristics for MultiStage Balloon Dilatation Catheter have been tested for its functions as intended including verification of performance characteristics per performances characteristics relevant to functions as intended:

  • · Visual Inspection
  • · Bacteriostasis performance of pouch
  • · Dye penetration
  • Pouch
  • · Tension test
  • · Guidewire compatibility test
  • · Channel compatibility

{6}------------------------------------------------

  • Product Dimensional Inspection
  • . Compliance (deployed balloon dimensional analysis)
  • Balloon fatigue; No leakage and damage when inflation .
  • · Balloon rated burst pressure (RBP)
  • · Peak tensile
  • · Kink resistance
  • Leakage Test
  • · Hydratability
  • Hub
  • · Corrosion resistance
  • · X-ray detectability
  • · Chemical properties

Result: All tests were passed.

The results of Non-Clinical Performance testing demonstrate that the MultiStage Balloon Dilatation Catheter is considered as safe and effective as the predicate.

10. Performance Data-Clinical

No clinical study is included in this submission.

11. Conclusion

The differences in technological characteristics between the proposed device and the predicate device do not raise any different questions of safety or effectiveness. Performance testing and compliance with voluntary standards demonstrate the subject device is substantially equivalent to the relevant aspects of the predicate device in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.

Therefore, the proposed devices are determined to be substantially equivalent to the referenced predicate device.

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.