    Device Name :

    Modified Bencox Hip System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Modified Bencox Hip System is intended for use in total or partial hip arthroplasty in primary or revision surgery for the following conditions:

    a. Non-inflammatory degenerative joint disease, such as avascular necrosis, osteoarthritis, traumatic arthritis

    b. Inflammatory degenerative joint disease, such as rheumatoid arthritis

    c. Treatment of non-union, femoral neck fracture and trochantric fractures of the proximal femur with head involvement, unmanageable using other techniques

    d. Patients with failed previous surgery where pain, deformity, or dysfunction persists

    e. Revision of previously failed total hip arthroplasty

    Device Description

    The Modified Bencox Mirabo Hip System consists of the following components,

    • Acetabular Insert Bencox Mirabo PE Liner (Std. & Elevated), .
    • Femoral Stem Bencox ID Stem (Offset), .
    • Femoral Head Bencox Delta Head (XL), ●
    • Instrumentation Bencox Total Hip System Instrumentation. .

    The components are modification of Acetabular Insert cleared in K120924 (& K103431), Femoral Stem cleared in K112019 & specification inclusion of Femoral Heads cleared in K121665. The modified components are cementless, metal-onpolyethylene hip system for hip arthroplasty similar to devices cleared in respective mentioned 510(k)'s.

    Modified Bencox Mirabo PE Liner is manufactured from similar material cleared under K120924 & K103431, conforming to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants, Type 2 (GUR 1050), and irradiated with average dose of 10.0 Mrad of gamma radiation similar to predicate device Bencox Mirabo PE Liner and other predicates such as Stryker Crossfire.

    Modified Bencox ID Stem (Offset) is similar to the predicate device Bencox ID Stem cleared in K112019, except for the slightly thicker neck geometry and horizontal offset.

    Bencox Delta Head XL specification inclusion consists of diameters 32, 36 & 40 mm design is similar to the 510k cleared devices, Bencox Delta Head (K121665). The specification is also similar with other predicate devices and has the same supplier, CeramTec, AG.

    AI/ML Overview

    This document describes the Modified Bencox Hip System, a hip implant. The information provided outlines the performance data rather than the acceptance criteria for an AI/ML device. Therefore, it is important to note that the following analysis is based on the provided text, which details the performance testing conducted for a medical device (hip implant components), not an AI/ML algorithm.

    Given the context of a hip implant system, the "acceptance criteria" are typically related to the physical and mechanical properties of the device, ensuring it meets established international standards for safety and performance in the human body. The "study" refers to the mechanical and material testing performed on the device to demonstrate its substantial equivalence to previously cleared predicate devices.

    Here's an interpretation based on the provided document:

    Acceptance Criteria and Device Performance (Applied to a Medical Device)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Bencox Mirabo PE Insert:
    Conformity to ISO 14242 (Wear)Performed comparable to predicate devices.
    Conformity to ASTM F1820 (Liner Torsion & Lever Out)Performed comparable to predicate devices.
    Conformity to ASTM F2582 (Push Out and Impingement)Performed comparable to predicate devices.
    Material Characterization (ASTM 2565)Obtained values of various parameters were comparable to predicate devices.
    Bencox ID Stem (Offset):
    Conformity to ISO 7204-4 (Fatigue)Satisfied the standards and performed similar to predicate devices.
    Conformity to ISO 7204-6 (Range of Motion)Satisfied the standards and performed similar to predicate devices.
    Bencox Delta Heads XL:
    Reference to predicate device submission, K121665 and CeramTec Device Master File (Implies meeting the established standards for these components)Performance refers to predicate device submission K121665 and CeramTec Device Master File.
    Overall: No new issues of safety or efficacy from technological differencesDifferences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy.

    Regarding the specific questions intended for AI/ML device evaluation, the provided document does not contain information for most of these points, as it describes a physical medical device. However, I will address them to the best of my ability based on the information not present or by inferring the closest equivalent for a physical device.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated in terms of "test set" samples as one would for an AI/ML model. For physical devices, "sample size" typically refers to the number of units tested. The document mentions "worst case combination" for the Bencox ID Stem, implying specific device configurations were selected for testing. The number of individual components or assemblies tested to meet standard requirements is not detailed.
    • Data Provenance: Not applicable in the context of an AI/ML model's data. The testing was physical and performed to international standards. The manufacturer is Corentec Co., Ltd. from South Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. "Ground truth" in this context would be the objective criteria defined by the ISO and ASTM standards and the measurements taken during mechanical and material testing. No human experts are used to "establish ground truth" in the way they would for labeling in an AI/ML diagnostic task.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not applicable. This concept pertains to resolving discrepancies in expert labels for AI/ML training/testing data. For medical device testing, adjudication is typically not performed in this manner; results are objectively measured against established standard thresholds.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is a framework for evaluating diagnostic AI/ML systems. No human readers or AI assistance are involved in the performance testing of a hip implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This refers to the performance of an AI algorithm alone. The "performance" described here is for a physical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For a physical medical device, the "ground truth" is defined by the objective, measurable criteria within the referenced ISO and ASTM international standards. These standards specify test methods (e.g., how to measure wear, fatigue strength, material properties) and acceptable performance limits. The device's performance is then compared directly to these limits.

    8. The sample size for the training set

    • Not applicable. There is no "training set" in the context of developing and testing this type of physical medical device. The device itself is manufactured, and specific samples are then tested for performance.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no "training set" for an AI/ML model, there is no ground truth established for one. The "ground truth" for the device's performance is the adherence to the published standards as mentioned in point 7.