Search Results
Found 3 results
510(k) Data Aggregation
(29 days)
Mirage Hydrophilic Guidewire; X-pedion Hydrophilic Guidewire
The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.
The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.
The provided text describes the 510(k) premarket notification for the Mirage™ and X-pedion™ Hydrophilic Guidewires, asserting their substantial equivalence to a predicate device. The information primarily focuses on bench testing rather than AI/ML algorithm performance. Therefore, many of the requested details, such as those related to AI model training, expert ground truth, and human reader studies, are not applicable or cannot be extracted from this document.
Here's the relevant information based on the provided text:
1. A table of acceptance criteria and the reported device performance
Test | Acceptance Criteria (Implicit: "met the acceptance criteria") | Reported Device Performance |
---|---|---|
Performance | ||
Visual Inspection and Dimensional Verification | Met the acceptance criteria for visual inspection and dimensional verification. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for visual inspection and dimensional verification. |
Friction Force | Met the acceptance criteria for friction force. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for friction force. |
Flexing | Met the acceptance criteria for flexing. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for flexing. |
Particulate | Met the acceptance criteria for particulate. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for particulate. |
Tensile Strength | Met the acceptance criteria for tensile strength. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for tensile strength. |
Corrosion Resistance | Met the acceptance criteria for corrosion resistance. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for corrosion resistance. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes (number of guidewires) used for each bench test. The data provenance is from non-clinical bench testing conducted by Micro Therapeutics, Inc. d/b/a ev3 Neurovascular (the applicant).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device clearance based on substantial equivalence demonstrated through non-clinical bench testing, not an AI/ML algorithm requiring expert-established ground truth. The "ground truth" for these tests would be the physical properties and performance measurements of the guidewires against engineering specifications.
4. Adjudication method for the test set
Not applicable, as this is bench testing of physical device properties, not a study involving human interpretation of data where adjudication would be necessary.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for a physical medical guidewire, not an AI-powered diagnostic or assistive tool. No MRMC study was conducted or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for a physical medical guidewire.
7. The type of ground truth used
The "ground truth" for the bench tests would be the physical and mechanical specifications for guidewires, as referenced by standards such as ISO 11070 and USP , as well as internal quality procedures (QP50324, TM0047). The tests evaluated observed performance against these predefined criteria.
8. The sample size for the training set
Not applicable. There is no training set for a physical medical device.
9. How the ground truth for the training set was established
Not applicable. There is no training set for a physical medical device.
Ask a specific question about this device
(29 days)
Mirage Hydrophilic Guidewire, X-pedion Hydrophilic Guidewire
The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.
The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.
The provided text describes the 510(k) premarket notification for the Mirage™ Hydrophilic Guidewire and X-pedion™ Hydrophilic Guidewire. This document primarily focuses on demonstrating substantial equivalence to a predicate device through non-clinical (bench and biocompatibility) testing, rather than a study on an AI/ML-based medical device.
Therefore, the information required to answer most of your questions about acceptance criteria for an AI/ML device, sample sizes for AI test sets, expert ground truthing, MRMC studies, standalone performance, training sets, and ground truth establishment for AI/ML models is not present in this document.
However, I can extract information related to the acceptance criteria and performance data for the physical medical device as described:
Acceptance Criteria and Device Performance (Based on Non-Clinical Bench Testing):
The document details numerous non-clinical bench tests performed to demonstrate the safety and performance of the guidewires. The results consistently state that "The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for [Test Name]."
Table of Acceptance Criteria and Reported Device Performance (Non-Clinical):
Test Category | Specific Test / Parameter | Acceptance Criteria (Implied by positive outcome) | Reported Device Performance |
---|---|---|---|
Biocompatibility | |||
Chemical Characterization (Extractables/Leachables) | No risk to the patient from Chemicals of Potential Concern (COPC). | The extractables/leachables found are acceptable. | |
Cytotoxicity | No evidence of cell lysis or toxicity; grade , and USP . | Met the acceptance criteria for bacterial endotoxin. | |
Packaging | |||
Terminally Sterilized Medical Devices | Met criteria per ISO 11607. | Met the acceptance criteria for packaging terminally sterilized medical devices. | |
Performance (Bench) | |||
Device Compatibility/Distal Access | Compatible with ancillary devices, navigates tortuous vessels. | Met the acceptance criteria for device compatibility/distal access. | |
Distal Flexibility | Navigates through tortuous vessels. | Met the acceptance criteria for distal flexibility. | |
Visual Fracture | Met criteria per ISO 10555-1 and ISO 11070. | Met the acceptance criteria for visual fracture. | |
Radiopacity | Clearly visible during use. | Met the acceptance criteria for radiopacity. | |
Tip Buckling | Withstands forces typical of clinical use. | Met the acceptance criteria for tip buckling. | |
Tip Retention | Satisfactory tip retention. | Met the acceptance criteria for tip retention. | |
Tip Shapeability | Satisfactory tip shapeability. | Met the acceptance criteria for tip shapeability. | |
Torque Response | Distal tip responds to proximal manipulations. | Met the acceptance criteria for torque response. | |
Turns to Failure | Withstands torsional forces typical of clinical use. | Met the acceptance criteria for turns to failure. | |
Visual Inspection and Dimensional Verification | Meets specified dimensions. | Met the acceptance criteria for visual inspection and dimensional verification. | |
Particulate | Meets criteria per USP for particulate counts and sizes. | Met the acceptance criteria for particulate counts and sizes. | |
Friction Force | Satisfactory friction force. | Met the acceptance criteria for friction force. | |
Flexing | Met criteria per ISO 11070. | Met the acceptance criteria for flexing. | |
Corrosion Resistance | Met criteria per ISO 11070. | Met the acceptance criteria for corrosion resistance. | |
Tensile Strength | Met criteria per ISO 11070. | Met the acceptance criteria for tensile strength. |
Since this is a 510(k) submission for a non-AI/ML medical device, the following information is not applicable or not provided in the document:
- Sample sizes used for the test set and the data provenance: Not an AI device, so no "test set" in the AI sense. Bench tests follow specific standards, implying sample sizes per those standards, but not explicitly stated here.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable; ground truth for a physical device is established via standardized physical and chemical tests, not expert human interpretation of data for AI.
- Adjudication method for the test set: Not applicable for non-AI bench testing.
- If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used: For a physical guidewire, "ground truth" is typically defined by adherence to engineering specifications, material properties, and performance under simulated clinical conditions, verified through empirical testing following established ISO and USP standards.
- The sample size for the training set: Not applicable; this is not an AI/ML device.
- How the ground truth for the training set was established: Not applicable; this is not an AI/ML device.
In summary, this document is a regulatory submission for a physical medical device demonstrating substantial equivalence, not an AI/ML device requiring clinical validation for diagnostic or prognostic purposes.
Ask a specific question about this device
(13 days)
MIRAGE HYDROPHILIC GUIDEWIRE, MODEL 103-0608
Ask a specific question about this device
Page 1 of 1