(29 days)
The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.
The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.
The provided text describes the 510(k) premarket notification for the Mirage™ and X-pedion™ Hydrophilic Guidewires, asserting their substantial equivalence to a predicate device. The information primarily focuses on bench testing rather than AI/ML algorithm performance. Therefore, many of the requested details, such as those related to AI model training, expert ground truth, and human reader studies, are not applicable or cannot be extracted from this document.
Here's the relevant information based on the provided text:
1. A table of acceptance criteria and the reported device performance
| Test | Acceptance Criteria (Implicit: "met the acceptance criteria") | Reported Device Performance |
|---|---|---|
| Performance | ||
| Visual Inspection and Dimensional Verification | Met the acceptance criteria for visual inspection and dimensional verification. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for visual inspection and dimensional verification. |
| Friction Force | Met the acceptance criteria for friction force. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for friction force. |
| Flexing | Met the acceptance criteria for flexing. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for flexing. |
| Particulate | Met the acceptance criteria for particulate. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for particulate. |
| Tensile Strength | Met the acceptance criteria for tensile strength. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for tensile strength. |
| Corrosion Resistance | Met the acceptance criteria for corrosion resistance. | The Mirage™ and X-pedion™ Hydrophilic Guidewires met the acceptance criteria for corrosion resistance. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes (number of guidewires) used for each bench test. The data provenance is from non-clinical bench testing conducted by Micro Therapeutics, Inc. d/b/a ev3 Neurovascular (the applicant).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device clearance based on substantial equivalence demonstrated through non-clinical bench testing, not an AI/ML algorithm requiring expert-established ground truth. The "ground truth" for these tests would be the physical properties and performance measurements of the guidewires against engineering specifications.
4. Adjudication method for the test set
Not applicable, as this is bench testing of physical device properties, not a study involving human interpretation of data where adjudication would be necessary.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for a physical medical guidewire, not an AI-powered diagnostic or assistive tool. No MRMC study was conducted or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is for a physical medical guidewire.
7. The type of ground truth used
The "ground truth" for the bench tests would be the physical and mechanical specifications for guidewires, as referenced by standards such as ISO 11070 and USP <788>, as well as internal quality procedures (QP50324, TM0047). The tests evaluated observed performance against these predefined criteria.
8. The sample size for the training set
Not applicable. There is no training set for a physical medical device.
9. How the ground truth for the training set was established
Not applicable. There is no training set for a physical medical device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 21, 2020
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Prerana Gurubasavaraj Regulatory Affairs Specialist 9775 Toledo Way Irvine, CA 92618
Re: K201690
Trade/Device Name: Mirage Hydrophilic Guidewire; X-pedion Hydrophilic Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX, MOF Dated: June 18, 2020 Received: June 22, 2020
Dear Prerana Gurubasavaraj:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201690
Device Name
Mirage™ Hydrophilic Guidewire; X-pedion™ Hydrophilic Guidewire
Indications for Use (Describe)
The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular | ||
|---|---|---|---|
| 9775 Toledo Way | |||
| Irvine, CA 92618 | |||
| Establishment Registration: 2029214 | |||
| Contact Person: | Prerana Gurubasavaraj | ||
| Specialist, Regulatory Affairs | |||
| Telephone: (949) 297-5804 | |||
| Email: prerana.gurubasavaraj@medtronic.com | |||
| Date SummaryPrepared: | June 18 ,2020 | ||
| Trade Name ofDevice: | Mirage™ Hydrophilic GuidewireX-pedion™ Hydrophilic Guidewire | ||
| Device ClassificationName: | Guide, Wire, Catheter, Neurovasculature | ||
| Regulation MedicalSpecialty: | Cardiovascular | ||
| 510(k) Review Panel: | Neurology | ||
| Classification ProductCode: | DQX | ||
| Subsequent ProductCode: | MOF | ||
| Regulation Number: | 870.1330 | ||
| RegulationDescription: | Catheter Guide Wire | ||
| Device Classification: | Class II | ||
| Predicate Device: | Mirage™ and X-pedion™ Hydrophilic Guidewires510(k): K193548 | ||
| Reference Device: | React™ 68 Catheter510(k): K180715 |
510(k) Summary K201690
Device Description:
The Hydrophilic Guidewire is a stainless-steel guidewire with a radiopaque, platinum distal coil. The guidewire has a hydrophilic coating that spans the distal 170 cm. Included within the sterile pouch is a pin vise to assist in guidewire manipulation and an introducer needle to ease the introduction of the guidewire into the catheter hub and/or hemostasis valve.
Indication for Use Statement:
The Hydrophilic Guidewire is indicated for general intravascular use to aid in the selective placement of catheters in the peripheral, visceral and cerebral vasculature during diagnostic and/or therapeutic procedures.
{4}------------------------------------------------
| Legally Marketed Predicate Device(K193548) | Mirage™ and X-pedion™ Hydrophilic Guidewires | |
|---|---|---|
| Indication for Use Statement | The Hydrophilic Guidewire isindicated for general intravascular useto aid in the selective placement ofcatheters in the peripheral, visceraland cerebral vasculature duringdiagnostic and/or therapeuticprocedures. | Same |
| Dimensions | ||
| Nominal WireDiameter | Mirage™: 0.008"X-pedion™: 0.010” | Same |
| Device Length | $200 cm \pm 2.5 cm$ . | Same |
| Tip Length | 10 cm | Same |
| Tip Type and Shape | Shapeable to 90° | Same |
| Coating Length | 170 cm | Same |
| Material | ||
| Coil | Platinum | Same |
| Wire | Stainless-Steel | Same |
| Coating | Hydrophilic | Same |
| Packaging | ||
| Pouch | Polyester/Tyvek | Same |
| Carton | 0.020" Solid Bleached Sulfate | Same |
| Accessories | ||
| Pin Vise | Yes | Same |
| Introducer Needle | Yes | Same |
| Sterilization | ||
| Method | Ethylene Oxide (EO) | Same |
| Stability | ||
| Shelf Life | 12 Months | 36 Months |
Device Comparison:
Performance Data – Bench:
The following non-clinical bench testing was performed to evaluate the performance of the Mirage™ and X-pedion™ Hydrophilic Guidewires. The successful results of the testing demonstrated that the changes do not raise questions of safety and effectiveness, supporting the substantial equivalence to the predicate device.
| Test | Test Method Summary | Results |
|---|---|---|
| Performance | ||
| Visual Inspection andDimensional Verification | The MirageTM and X-pedionTMHydrophilic Guidewires wereevaluated per QP50324. | The MirageTM and X-pedionTMHydrophilic Guidewires met theacceptance criteria for visualinspection and dimensionalverification. |
| Friction Force | The MirageTM and X-pedionTMHydrophilic Guidewires wereevaluated per TM0047. | The MirageTM and X-pedionTMHydrophilic Guidewires met theacceptance criteria for frictionforce. |
{5}------------------------------------------------
| Test | Test Method Summary | Results |
|---|---|---|
| Flexing | The MirageTM and X-pedionTMHydrophilic Guidewires wereevaluated per ISO 11070. | The MirageTM and X-pedionTMHydrophilic Guidewires met theacceptance criteria for flexing. |
| Particulate | The MirageTM and X-pedionTMHydrophilic Guidewires wereevaluated per USP <788>. | The MirageTM and X-pedionTMHydrophilic Guidewires met theacceptance criteria forparticulate. |
| Tensile Strength | The MirageTM and X-pedionTMHydrophilic Guidewires wereevaluated per ISO 11070. | The MirageTM and X-pedionTMHydrophilic Guidewires met theacceptance criteria for tensilestrength. |
| Corrosion Resistance | The MirageTM and X-pedionTMHydrophilic Guidewires wereevaluated per ISO 11070. | The MirageTM and X-pedionTMHydrophilic Guidewires met theacceptance criteria for corrosionresistance. |
Performance Data - Animal:
There is no change to the Indication for Use (IFU) Statement for the Mirage™ and X-pedion™ Hydrophilic Guidewires in comparison to the legally marketed predicate device. The differences in technological characteristics do not raise questions of safety and effectiveness as demonstrated through non-clinical bench testing using well-established acceptable scientific methods.
Performance Data - Clinical:
There is no change to the Indication for Use (IFU) Statement for the Mirage™ and X-pedion™ Hydrophilic Guidewires in comparison to the legally marketed predicate device. The differences in technological characteristics do not raise questions of safety and effectiveness as demonstrated through non-clinical bench testing using well-established acceptable scientific methods.
Conclusion:
There is no change to the Indication for Use (IFU) Statement for the Mirage™ and X-pedion™ Hydrophilic Guidewires in comparison to the legally marketed predicate device. In addition, the differences in technological characteristics do not raise questions of safety and effectiveness as demonstrated through non-clinical bench testing using well-established acceptable scientific methods. The information provided in this submission supports a determination of substantial equivalence for the Mirage™ and X-pedion™ Hydrophilic Guidewires.
§ 870.1330 Catheter guide wire.
(a)
Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.(b)
Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.