Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K212288
    Manufacturer
    Date Cleared
    2022-06-02

    (316 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Minjie Catheter System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Minjie Catheter System is indicated for the introduction of interventional devices into the peripheral and neurovasculature.

    Device Description

    The Minjie Catheter System consists of the Minjie Catheter, Introducer Peel Away Tool and packaging hoop with Elbow Flush Luer. The Minjie Catheter is a single lumen. flexible, variable stiffness composite catheter with a Nitinol structure. A radiopaque marker band on the distal tip of the catheter is used for visualization under fluoroscopy. The distal section of the catheter is coated with a hydrophilic coating, to reduce friction during intravascular use. The Minjie Catheter System dimensions are included in the individual device label. The devices are supplied sterile and are intended for single use only.

    AI/ML Overview

    The provided FDA 510(k) summary for the Minjie Catheter System does not contain acceptance criteria or reported device performance in the format requested for AI/algorithm-based devices. This document describes a medical device (a catheter system) and its performance through non-clinical (bench and animal) testing to demonstrate substantial equivalence to a predicate device, rather than the performance of an AI algorithm.

    Therefore, many of the requested fields related to AI algorithm evaluation (sample sizes for test/training sets, data provenance, number/qualification of experts, adjudication methods, MRMC studies, standalone performance, ground truth types) are not applicable to the information provided in this document.

    However, I can extract information related to the device's non-clinical testing and its conclusions, which serve as the "proof" that the device meets some form of acceptance criteria for medical devices of this type.

    Here's a summary based on the provided document, adapting the requested structure where possible for a non-AI device:

    Acceptance Criteria and Study for the Minjie Catheter System

    The Minjie Catheter System is a physical medical device (catheter) and its performance is evaluated against engineering specifications, biocompatibility standards, and functional capabilities rather than AI algorithm metrics. The studies performed are non-clinical, including bench testing and an animal study, to demonstrate substantial equivalence and safety/effectiveness for its intended use.

    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document states that "All devices met acceptance criteria" for each bench test, but it does not explicitly define the numeric acceptance criteria themselves. The reported performance is therefore a qualitative statement of compliance.

    Test NameTest Method SummaryAcceptance Criteria (Implied)Reported Device Performance
    Catheter Visual InspectionInspection for dents, kinks, cracks, damage, or anomalies.No damage or anomalies affecting function.All devices met acceptance criteria.
    Coating IntegrityInspection under magnification for coating defects before and after simulated use and particulate testing.No coating defects.All devices met acceptance criteria.
    Particulate TestingEvaluation for particulate generation under simulated use in a tortuous anatomical model.Meets particulate generation limits.All devices met acceptance criteria.
    Simulated UseRepeated navigation through a tortuous benchtop model to assess compatibility with accessories, stability, and navigation to M1 and M2 segments of MCA.Successful navigation, compatibility, and stability.All devices met acceptance criteria.
    Coating Frictional Forces / DurabilityEvaluation of frictional forces and durability via repeated navigation through simulated use test model.Acceptable frictional forces and durability.All devices met acceptance criteria.
    Distal Tip BucklingTesting distal tip under compressive loads at 5mm, 10mm, and 20mm to evaluate stiffness.Meets specified stiffness requirements.All devices met acceptance criteria.
    Kink ResistanceWrapping device around mandrels of clinically relevant diameters and inspecting for kinks.No kinks.All devices met acceptance criteria.
    Liquid Leakage under PressureTested per ISO 10555-1, Annex C.No liquid leakage.All devices met acceptance criteria.
    Hub Air Aspiration LeakTested per ISO 10555-1, Annex D.No hub air aspiration leak.All devices met acceptance criteria.
    Torque to FailureTested in a simulated use model to determine number of rotations to failure.Withstands specified torque without failure.All devices met acceptance criteria.
    Manual Injection / Peak PressureTested with manual syringe injection of worst-case contrast media after simulated use.Withstands generated pressures.All devices met acceptance criteria.
    Static / Dynamic BurstTested under full-length static conditions to burst per ISO 10555-1, Annex F.Meets burst pressure requirements.All devices met acceptance criteria.
    Luer Hub CompatibilityTested per ISO 80369-7 and ISO 80369-20.Compatible with luer hubs.All devices met acceptance criteria.
    Dimensional InspectionMeasurement of usable length, proximal/distal inner/outer diameters.Meets specified dimensional tolerances.All devices met acceptance criteria.
    Lumen PatencyMandrel of required size must pass from proximal hub to distal tip.Mandrel passes freely.All devices met acceptance criteria.
    Shaft Peak Tensile ForceTested to failure at distal tip section and each joint per ISO 10555-1, Annex B.Withstands specified tensile forces.All devices met acceptance criteria.
    Hub Peak Tensile ForceTested to failure per ISO 10555-1, Annex B.Withstands specified tensile forces.All devices met acceptance criteria.
    Physician Usability TestingNavigated through a tortuous benchtop model to assess compatibility, stability, injection ability, and navigation to M1/M2 segments of MCA.Usable and effective for intended functions.All devices met acceptance criteria.

    Biocompatibility Test NameTest Results Conclusion
    CytotoxicityNon-cytotoxic
    SensitizationNon-sensitizer
    Intracutaneous IrritationNon-irritant
    Acute Systemic ToxicityNo acute systemic toxicity
    Material Mediated PyrogenNon-pyrogenic
    Hemocompatibility (Hemolysis)Non-hemolytic
    Hemocompatibility (Complement Activation)Non-activator of the complement system
    Hemocompatibility (PTT)Non-activator of coagulation
    Hemocompatibility (Platelet & Leukocyte Counts)Non-activator of platelet and leukocyte

    Sterilization/Shelf-Life TestAcceptance Criteria (Implied)Reported Performance
    EO Sterilization ValidationSAL of 10^-6 per ISO 11135:2014.Achieved SAL of 10^-6.
    Endotoxin TestingMeets FDA guidance for pyrogen and endotoxins.Compliant.
    Packaging IntegrityMaintains sterility and integrity for 24 months.Met all acceptance criteria.
    Product Stability (Aging)Device remains functional for 24 months.Met all acceptance criteria.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Bench Testing: The document does not specify the exact sample size for each bench test (e.g., "All devices met acceptance criteria" implies multiple devices were tested). There is no explicit mention of data provenance (e.g., country of origin). This is non-clinical, prospective testing.
    • Biocompatibility Testing: Not specified for individual tests.
    • Sterilization and Shelf-Life: Not specified, but validation studies would involve multiple units.
    • Animal Testing: The document states "An animal study was performed," but does not specify the number of animals used. Data provenance is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as this is a physical device and not an AI algorithm requiring expert ground truth for image interpretation or diagnosis. Bench testing results are typically objective measurements against engineering specifications. Animal study evaluations would be performed by qualified veterinary and pathology staff, but specific numbers and qualifications are not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for physical device testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic or therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this physical device:

    • Bench Testing: Engineering specifications, industry standards (e.g., ISO), and defined performance parameters serve as the "ground truth" or acceptance criteria.
    • Biocompatibility Testing: Established biological response parameters outlined in ISO 10993 series and FDA guidance.
    • Animal Testing: Angiographic and histological evaluations, which are considered objective medical findings.

    8. The sample size for the training set

    Not applicable. This is not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1