510(k) Data Aggregation
(178 days)
Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.
The provided text describes the acceptance criteria and performance data for the "Minghao Nitrile Powder Free Patient Examination Gloves, Blue Color." This is a medical device, and the information is presented in the context of a 510(k) premarket notification for demonstrating substantial equivalence to a predicate device.
Here's the breakdown of the requested information:
1. Table of acceptance criteria and reported device performance:
Characteristic | Standard / Acceptance Criteria | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 6319-10 (Length ≥ 230 mm min; Width: Small 70-90mm, Medium 85-105mm, Large 100-120mm, Xlarge 110-130mm; Thickness: Finger 0.05mm min, Palm 0.05mm min) | Meets ASTM standard D 6319-10 (Length ≥ 230 mm min for all sizes; Width as per standard; Thickness: Finger 0.05mm min, Palm 0.05mm min) |
Physical Properties | ASTM standard D 6319-10 (Before aging/after aging: Elongation ≥ 500%, Tensile Strength ≥ 14 MPa) | Meets ASTM standard D 6319-10 (Before aging/after aging: Elongation ≥ 500%, Tensile Strength ≥ 14 MPa) |
Freedom from Pinholes | 21 CFR 800.20, ASTM D6319-10, ASTM D5151-06 (Reapproved 2011) (Inspection Level I, AQL 2.5) | Meets listed standards (Inspection Level I, AQL 2.5) |
Powder Residual | ASTM standard D 6319-10 and D6124-06 (Reapproved 2011) ( |