Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K091954
    Device Name
    MICRO-STICK SET
    Manufacturer
    Date Cleared
    2010-01-08

    (191 days)

    Product Code
    Regulation Number
    870.1340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MICRO-STICK SET

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Micro-Stick® Set is indicated for percutaneous introduction of up to a 0.038 inch guidewire or catheter into the vascular system following a small 21 gauge needle stick. The Micro-Stick® Set is not intended for use in the coronary or cerebral vasculature.

    Device Description

    The Micro-Stick® Set is designed to provide dilation of the initial entry point and provide a channel for attaining vascular access. The coaxial introducer consists of an inner dilator within a slightly shorter outer sheath.

    The kit contains a 21 gauge introducer needle, a .018" diameter guidewire, and a 4F or 5F coaxial introducer. The coaxial introducer is offered in a standard and stiff version.

    The Micro-Stick® Set is packaged sterile.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and study proving the device meets them:

    No AI/ML device is described in the provided text. The document describes a medical device called the "Micro-Stick® Set," which is an introducer set used for vascular access. The testing performed is in-vitro bench testing to ensure reliable design and performance, not clinical studies involving human readers or AI algorithms.

    Therefore, many of the requested sections related to AI/ML device studies (sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, standalone performance, etc.) are not applicable to this document.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes in-vitro bench testing for the Micro-Stick® Set. It states that the device was tested "to assure reliable design and performance in accordance with ISO standards and/or internal procedures." However, specific numerical acceptance criteria or detailed performance results (e.g., exact force at break in Newtons, precise leakage rates) are not provided in this summary. It only lists the types of tests performed and a general conclusion of conformance.

    Acceptance Criterion (Type of Test)Reported Device Performance
    Guidewire PassageMet requirements (implied)
    Air LeakageMet requirements (implied)
    Liquid LeakageMet requirements (implied)
    Force at BreakMet requirements (implied)
    Simulated UseMet requirements (implied)
    Biocompatibility (ISO 10993)Met requirements

    Note: The specific numerical or descriptive acceptance levels for these tests are not detailed in the summary. The summary only confirms that the "devices meet the performance criteria of design verification as specified by ISO standards and test protocols."


    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified in the document. The testing was in-vitro bench testing, so there isn't a "test set" in the context of patient data.
    • Data Provenance: The testing was "in-vitro" and conducted as "bench performance data." This implies laboratory testing of the physical device components, not involving patient data from a specific country or retrospective/prospective collection.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is not an AI/ML device, and the testing was in-vitro bench testing. No human experts were used to establish ground truth for this type of performance evaluation.


    4. Adjudication method for the test set

    Not applicable. This is not an AI/ML device, and no human readers/experts were involved in assessing the 'ground truth' of the in-vitro performance.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document does not describe an AI/ML device, nor does it conduct MRMC studies.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document does not describe an AI/ML algorithm.


    7. The type of ground truth used

    The "ground truth" for the in-vitro tests would be the established ISO standards and internal procedures/specifications for physical device performance (e.g., a specific leakage rate is acceptable, a certain force is required for breakage, a guidewire must pass smoothly). These are engineering and material science metrics, not clinical ground truth like pathology or expert consensus.


    8. The sample size for the training set

    Not applicable. This document does not describe an AI/ML device, and therefore there is no training set mentioned.


    9. How the ground truth for the training set was established

    Not applicable. There is no training set for an AI/ML device described here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1