Search Results
Found 6 results
510(k) Data Aggregation
(108 days)
The MANTIS Clip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- Endoscopic marking
- Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm
- Bleeding ulcers
- Arteries < 2 mm
- Polyps < 1.5 cm in diameter
- Diverticula in the colon
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
- Anchoring to affix jejunal feeding tubes to the wall of the small bowel; and Anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus
- As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
The MANTIS Clip is a sterile device consisting of a pre-loaded, radiopaque, singleuse, endoscopic clipping device consisting of two main components: the delivery system and the clip.
The delivery system consists of a handle assembly and delivery catheter. The deliverv system is constructed using stainless steel, and polyester materials. The delivery system will allow for the device to rotate at the distal end. The MANTIS Clip delivery system is offered in a 235cm working length. The clip consists of a stainlesssteel capsule and clip arms, a cobalt chrome yoke, and a styrene tension breaker. The clip is deployed from the delivery system during use. The MANTIS Clip jaws are engineered such that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening, closing, and rotation capability may be limited by clinical circumstances and patient anatomy. There are no associated accessories included with this device. The clip jaw teeth are designed with smaller angle between the teeth and the clip arm body.
The MANTIS Clip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of: endoscopic marking; hemostasis for mucosal/sub-mucosal defects < 3 cm, bleeding ulcers, arteries < 2 mm, polyps < 1.5 cm in diameter, diverticula in the colon, and prophylactic clipping to reduce the risk of delayed bleeding post lesion resection; anchoring to affix jejunal feeding tubes to the wall of the small bowel; and anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus; and as a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively.
Here's the breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
| Component | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Clip | 4.1 Clip Assembly Repeated Open/Close Function | PASS |
| Clip | 4.2 Clip Opening Gap | PASS |
| Clip | 4.4 Deployed Clip: Retention Force | PASS |
| Clip | 4.5 Clip Approach: Cantilever Force | PASS |
| Clip | 4.6 Clip Approach: Vertical Oblique | PASS |
| Clip + Delivery System | 4.7 Torque – Full Device | PASS |
| Clip | 4.9.1 Clip Opening Force | PASS |
| Clip | 4.9.2 Clip Close Force | PASS |
| Clip | 4.14 Jaw Deflection | PASS |
| Clip | 4.15 Grasping Capability | PASS |
| Clip + Delivery System | 7.1 Scope Compatibility / Usability | PASS |
| Clip + Delivery System | 7.3 Endoscope Damage – Clip Passability | PASS |
| Clip + Delivery System | 7.6 Biopsy Valve Compatibility | PASS |
| Device | MR Conditionality (consistent with FDA Guidance for MR Environment) | MR Conditional |
| Device | Biological Evaluation of Medical Devices (ISO 10993-1, ISO 10993-7) | Meets requirements |
| Device | Sterilization of Health Care products (ISO 11135-1) | Meets requirements |
2. Sample Size and Data Provenance for Test Set
The document does not specify a separate "test set" in the context of a clinical study with human patients. The performance data presented refers to bench testing and in-vitro assessments. The sample sizes for each specific bench test item (e.g., number of clips tested for "Repeated Open/Close Function") are not provided in this summary.
Data Provenance: The data provenance is from bench testing conducted by Boston Scientific Corporation. This is not human data, so country of origin and retrospective/prospective distinctions are not applicable.
3. Number of Experts and Qualifications for Ground Truth
Not applicable. The ground truth for this submission is established through pre-defined engineering specifications and bench test results, not through expert consensus on medical images or clinical outcomes.
4. Adjudication Method
Not applicable. As the validation relies on bench testing against engineering specifications, there is no need for expert adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not conducted or mentioned in this summary. The device is a medical instrument (clip), not an AI algorithm assisting human readers.
6. Standalone Performance Study (Algorithm Only)
No, a standalone (algorithm only) performance study was not conducted. This device is a physical medical instrument, not a software algorithm.
7. Type of Ground Truth Used
The ground truth used for proving the device meets its acceptance criteria is based on pre-defined engineering specifications and performance metrics derived from bench testing. Additionally, the device's biological compatibility and sterility were assessed against relevant ISO standards.
8. Sample Size for the Training Set
Not applicable. This device is a physical medical instrument, not a machine learning model, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for a physical device.
Ask a specific question about this device
(61 days)
The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthrosis; and Failed Previous Fusion.
The Radius® Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation system, the Radius® Spinal system is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Radius® Spinal System can also be linked to the XIA® Titanium Spinal System via the Ø5.5mm to Ø6.0mm Radius® rod-to-rod connector.
The Stryker Spine TRIO® Plate System is intended for posterior, noncervical (T10-S1) pedical and nonpedical fixation of the spine for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.
The Stryker Spine TRIO® Spinal Fixation System is intended for posterior, noncervical pedicle and non-pedicle fixation of the spine. The Stryker Spine TRIO® Spinal Fixation System is indicated for: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® Spinal Fixation Sytem is intended to be used in conjunction with the OSS Diapason Rods, Opus Spinal System Rods, and the Multi-Axis Cross Connectors.
The Stryker Spine TRIO® Spinal System is intended for posterior, noncervical pedicle and nonpedicle fixation of the spine to provide immobilization and stabilization of spinal segements in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The TRIO® + Spinal System is intended to be used in conjunction with the OSS/Diapason Rods or Opus Rods, XIA® Pre-bent Rods, and the Multi-Axis Cross Connectors.
The Stryker Spine TRIO® TRAUMA Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion.
The XIA® Spinal System and XIA® 4.5 Spinal System are intended for anterior/anterolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DDD) (Defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e., fracture or dislocation); Spinal Stenosis; Curvatures (i.e., Scoliosis, Kyphosis, and/or Lordosis); Tumor; Pseudoarthrosis and; Failed previous fusion. The 6mm diameter rods from the DIAPASON® Spinal System and OPUS® Spinal System are intended to be used with the other components of the XIA® Titanium Spinal System. The Titanium Multi-Axial Cross Connector are intended to be used with the other components of the XIA® Titanium Spinal System.
The XIA® 3 Spinal System is intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, noncervical pedicle fixation system, the XIA® 3 Spinal System is intended to provide additional support during autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities: Degenerative Disc Disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor, Pseudoarthorisis; and Failed Pervious Fusion. The Ø5.5mm rods from the Stryker Spine Radius® Spinal System and the Ø6.0mm Vitallium Rods from the XIA® Spinal System are intended to be used with the other components of the XIA® 3 Spinal System. When used for posterior noncervical pedicle screw fixation in pediatric patients, the XIA® Spinal System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. The XIA® 3 Spinal System for pediatic use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The XIA® 4.5 Spinal System is intended for anterior/anteriolateral and posterior, noncervical pedicle and non-pedicle fixation for the following indications: Degenerative Disc Disease (DD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); Spondylolisthesis; Trauma (i.e. fracture or dislocation); Spinal Stenosis; Curvature (i.e. scoliosis, kyphosis, and/or lordosis); Tumor; Pseudoarthorisis; and Failed Pervious Fusion. The Stryker Spine DIAPASON® Spinal System, OPUS® Spinal System and XIA® 4.5 Spinal System can be linked to the XIA® 4.5 Spinal System via the rod-to-rod connector when used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. Except for the staples, the XIA® 4.5 Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical pedicle screw fixation in pediatric patients. The XIA® 4.5 Spinal System for pediatric use in intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The STRYKER Spine thoraco-lumbar spinal fixation systems, subject of this 510(k), are non-cervical, pedicle and non-pedicle fixation systems comprised of screws, rods, plates, hooks, connectors, washers and staples. The components are manufactured from either Titanium (Titanium Alloy and CP Titanium), Stainless Steel or Cobalt-Chromium-Molybdenum Alloy (Vitallium®). This Special 510(k) submission seeks clearance for sterile labeling of the listed STRYKER Spine thoraco-lumbar spinal fixation systems. All of the components of the subject devices will be sterilized by gamma radiation, a traditional sterilization method as per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA.
This is a 510(k) Premarket Notification for sterile packaging of a range of spinal systems. The key claim is substantial equivalence to previously cleared non-sterile versions of the same devices. Therefore, the "device" in question for this submission is specifically the sterile packaging and the sterilization process, not the spinal implant itself as a medical device performing a physiological function. The "performance" being evaluated is the ability to maintain sterility and the structural integrity of the device within its packaging over its shelf life.
Here's the breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Sterilization Effectiveness | Demonstrated by gamma radiation, a traditional sterilization method per FDA guidance document, Updated 510(k) Sterility Review Guidance K90-1; Guidance for Industry and FDA. |
| Sterile Barrier Packaging System | Demonstrated effective in maintaining sterility for the recommended 5-year shelf-life. |
| Impact on Device Characteristics | No design modifications were incorporated to facilitate sterile packaging. The sterile-packed implant components maintain the same technological characteristics (design, technical requirements, materials of construction, indications/intended use) as the non-sterile predicate devices. |
| Shelf Life | Demonstrated a 5-year shelf life in sterile packaging. |
| Safety and Effectiveness | Concluded to be safe and effective, retaining the same intended use, indications, technological characteristics, and mode of operation as the predicate non-sterile devices. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily discusses the sterilization process and packaging rather than clinical performance of the implant itself. Therefore, typical "test set" in the context of clinical or diagnostic AI/software devices (e.g., patient data) is not applicable here.
- Sterilization and Packaging Validation Data: The document mentions "accelerated aging data" for validating the 5-year shelf life. It does not specify the sample size for this testing, nor the country of origin or whether it was retrospective or prospective. However, these types of tests are typically prospective laboratory studies conducted in controlled environments.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This submission is about the sterility and packaging of medical implants, not a diagnostic or clinical performance study that would require expert ground truth labeling. The "ground truth" here is the scientific validation of sterilization methods and shelf life according to recognized standards and FDA guidance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods are typically used in clinical studies for establishing ground truth, which is not relevant to this submission on sterile packaging.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/software device and no MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this submission relies on:
- Scientific Validation of Sterilization: Adherence to established scientific principles and FDA guidance for sterilization (gamma radiation), demonstrating a Sterility Assurance Level (SAL).
- Accelerated Aging Data: Laboratory testing to simulate real-time aging and demonstrate that the sterile barrier packaging system maintains sterility for the stated shelf life.
- Material and Design Specification Conformity: Confirmation that the components are manufactured from specified materials and that their design, technical requirements, and intended use are identical to the non-sterile predicate devices.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device that requires a training set.
9. How the ground truth for the training set was established
Not applicable. As this is not a machine learning or AI device, there is no training set and therefore no ground truth establishment for a training set.
Ask a specific question about this device
(15 days)
The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications:
- Degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);
- Spondylolisthesis;
- Trauma (i.e., fracture or dislocation);
- Spinal stenosis;
- Curvatures (i.e., scoliosis, kyphosis, and/or lordosis);
- Tumor:
- Pseudoarthrosis; and
- Failed previous fusion.
The Titanium and Vitallium® rods from the Stryker Spine RADIUS® Spinal System are intended to be used with the other components of MANTIS® Spinal System and MANTIS® Redux Spinal System.
This 510(k) is intended to introduce a line extension to the existing MANTIS® Redux Spinal System. The line extension consists of additional sizes of the MANTIS® Redux Long Arm Polyaxial Screws. The additional sizes include the 7.5 mm and 8.5 screw diameter, ranging 30 - 90 mm in length for the MANTIS Redux Spinal System. No modifications were made to the MANTIS Spinal System.
The provided text is a 510(k) summary for a medical device (Stryker Spine MANTIS® Spinal System and MANTIS® Redux Spinal System). It focuses on demonstrating substantial equivalence to predicate devices for a line extension. It does not contain information on acceptance criteria, a study proving device performance against acceptance criteria, or any of the detailed study parameters you've requested (sample sizes, expert qualifications, adjudication, MRMC, standalone performance, ground truth types, or training set details).
This type of submission typically relies on non-clinical (bench) testing to show that the new components meet established performance standards and are comparable to existing, legally marketed devices. It does not involve clinical studies with human participants or the assessment of AI algorithm performance as described in your prompt.
Therefore, I cannot populate the table or answer the specific questions about acceptance criteria and a study proving device performance because that information is not present in the provided document.
The document discusses:
- Device Modification: Additional sizes (7.5 mm and 8.5 mm diameter, 30-90 mm length) for MANTIS® Redux Long Arm Polyaxial Screws.
- Intended Use: Percutaneous, posterior, non-cervical pedicle and non-pedicle fixation of the spine for immobilization and stabilization as an adjunct to fusion in skeletally mature patients for various indications (e.g., DDD, spondylolisthesis, trauma).
- Technological Characteristics: Substantial equivalence in terms of material, design, and indications for use to predicate devices.
- Testing: Static Compression Bending testing, Static Torsion testing, and Fatigue Compression Bending testing per ASTM F1717 were conducted on the subject components and compared to a predicate system to demonstrate substantial equivalence.
In summary, the provided text describes a regulatory submission for a medical device line extension, not a study evaluating AI performance or clinical efficacy against specific acceptance criteria for diagnostic or therapeutic accuracy.
Ask a specific question about this device
(30 days)
The MANTIS® Spinal System and MANTIS® Redux Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondvlolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.
The Titanium and Vitallium rods from the Stryker Spine RADIUS® Spinal System are also intended to be used with other components of MANTIS® Spinal System and MANTIS® Redux Spinal System.
This 510(k) is intended to introduce an extension to the existing MANTIS® Spinal System. The proposed line extension includes the addition of titanium Reduction Screws, and use of Stryker Spine Radius titanium and Vitallium® rods with other components of the MANTIS® and MANTIS® Redux Spinal Systems.
This is a 510(k) premarket notification for a medical device (MANTIS® Spinal System and MANTIS® Redux Spinal System) and as such, it focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study to prove meeting those criteria in the way a clinical trial for a new drug or a novel AI diagnostic might.
Therefore, many of the requested fields cannot be directly extracted from this document as they are not relevant to a 510(k) for a spinal fixation system line extension. The information provided is primarily about mechanical testing and conformity to existing standards, showing the new components perform similarly to the predicate devices.
Here's an attempt to answer the questions based on the provided text, with explanations where information is not applicable or available:
1. A table of acceptance criteria and the reported device performance
The document states that "Engineering analysis and testing to demonstrate compliance with FDA's Guidance for Spinal System 510(k)'s May 3, 2004 were completed for the systems." This implies that the acceptance criteria are generally established by this FDA guidance and relevant consensus standards (e.g., relating to mechanical strength, fatigue, etc. for spinal fixation devices). However, specific quantitative acceptance criteria and reported performance values are not detailed in this summary.
| Acceptance Criteria (Implied by FDA Guidance/Standards) | Reported Device Performance (Implied by Submission) |
|---|---|
| Mechanical Strength | Demonstrated compliance through engineering analysis and testing |
| Fatigue Life | Demonstrated compliance through engineering analysis and testing |
| Biocompatibility (via materials) | Titanium and Vitallium® rods are known biocompatible materials used in predicate devices; assumed to meet standards. |
| Substantial Equivalence | Device is substantially equivalent to predicate devices in terms of material, design, and indications for use. |
2. Sample size used for the test set and the data provenance
The document mentions "Engineering analysis and testing." For mechanical devices, this typically involves testing a sample number of components and assemblies. The specific sample size for the mechanical testing is not provided in this summary.
- Test Set Sample Size: Not specified.
- Data Provenance: The testing would have been conducted by the manufacturer (Stryker Spine) or a contracted lab, likely in the US (Allendale, NJ is mentioned). It is a prospective generation of data for regulatory submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to a 510(k) for a spinal fixation system line extension. "Ground truth" in the context of clinical expert review is relevant for diagnostic devices or AI algorithms evaluating patient data. For a mechanical device, performance is established through physical and engineering tests, not expert consensus on clinical cases.
4. Adjudication method for the test set
Not applicable. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in expert opinions, typically in clinical studies or AI algorithm validation. Mechanical tests have objective outputs.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a mechanical device, not an AI diagnostic or assistance system. MRMC studies are for evaluating the performance of human readers, sometimes aided by AI, on clinical cases.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used
For mechanical testing, the "ground truth" is typically the physical properties and performance characteristics of the materials and device components, measured against established engineering standards (e.g., tensile strength, fatigue limits, compression strength, etc.).
8. The sample size for the training set
Not applicable. This concept applies to AI/ML models. For mechanical testing, there is no "training set" in this sense.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(28 days)
The MANTIS® Spinal System is intended for posterior, non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.
This Special 510(k) submission is intended to introduce a line extension to the MANTIS Spinal System, which entails additional diameter size cannulated polyaxial screw components as well as additional straight rod components in a variety of lengths. Modifications were made to the existing MANTIS Spinal System (K061813) to render the new components. The MANTIS Spinal System will include cannulated polyaxial screw, pre-bent rod and straight rod components that can be used via either a percutaneous surgical approach or a standard open approach.
The provided text does not describe the acceptance criteria or a study that proves a device meets such criteria. Instead, it is a 510(k) summary for a spinal fixation system, focusing on its substantial equivalence to predicate devices and its intended use.
Therefore, I cannot provide the requested information in the format specified because the necessary details are not present in the given input.
The document discusses:
- Proprietary Name: MANTIS® Spinal System
- Common Name: Spinal Fixation Appliances
- Regulatory Class: Class III
- Device Product Code: 87 MNH, 87 MNI, 87 NKB
- Predicate Devices: Stryker Spine MANTIS Spinal System (K061813), Stryker Spine Xia® Titanium Spinal System (K013823, K043473, K002858, K060361), and Stryker Spine Osteonics Spinal System (K951725).
- Description of Device Modification: Line extension including additional diameter size cannulated polyaxial screw components and additional straight rod components.
- Intended Use: Posterior, non-cervical pedicle fixation for indications such as degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, and failed previous fusion.
- Technological Comparison: States that the additional screw and rod components have equivalent mechanical properties to the predicate devices and are made from the same materials.
Missing Information:
The document does not contain any information regarding:
- A table of acceptance criteria and reported device performance.
- Sample size used for a test set or data provenance.
- Number of experts or their qualifications for ground truth establishment.
- Adjudication method for a test set.
- Multi-reader, multi-case (MRMC) comparative effectiveness study or effect size.
- Standalone (algorithm only) performance study.
- Type of ground truth used.
- Sample size for a training set.
- How ground truth for a training set was established.
This is a regulatory submission aimed at demonstrating substantial equivalence based on material and mechanical properties compared to already cleared devices, not a study evaluating performance against specific acceptance criteria for a new feature or algorithm.
Ask a specific question about this device
(84 days)
The MANTIS™ Spinal System is intended for percutaneous, posterior, non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.
The MANTIS™ Spinal System includes cannulated polyaxial screw and straight or pre-bent rod components that can be used via a percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are fabricated from titanium alloy. The implants will be provided non-sterile.
The provided text is a 510(k) Summary for the MANTIS™ Spinal System, a medical device for spinal fixation. It details the device's description, intended use, and substantial equivalence to predicate devices. However, this document does not contain any information regarding acceptance criteria or a study proving device performance against such criteria in the context of the requested AI/algorithm evaluation.
The 510(k) process for this type of medical device (spinal fixation system) typically involves demonstrating substantial equivalence to a legally marketed predicate device through engineering tests (e.g., mechanical strength, fatigue, biocompatibility) rather than clinical studies of "acceptance criteria" or "device performance" as understood in algorithm evaluation (like sensitivity, specificity, or reader studies).
Therefore, I cannot provide the requested information from the provided document.
To be explicit, the following information is not available in the provided text:
- A table of acceptance criteria and the reported device performance: Not present. The 510(k) focuses on substantial equivalence to predicate devices, not specific performance metrics against pre-defined acceptance criteria, especially not in an AI/algorithm context.
- Sample sized used for the test set and the data provenance: Not applicable/not present. No "test set" for an algorithm is described.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/not present.
- Adjudication method: Not applicable/not present.
- If a multi reader multi case (MRMC) comparative effectiveness study was done: No. This document describes a physical medical device, not an AI algorithm.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No.
- The type of ground truth used: Not applicable/not present.
- The sample size for the training set: Not applicable/not present.
- How the ground truth for the training set was established: Not applicable/not present.
Ask a specific question about this device
Page 1 of 1