Search Results
Found 1 results
510(k) Data Aggregation
K Number
K030529Manufacturer
Date Cleared
2003-08-08
(170 days)
Product Code
Regulation Number
880.6250Type
TraditionalPanel
General HospitalReference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
Nitrile powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Description
Nitrile powder-free patient examination, that meets all of the requirements of ASTM standard D 6319-00ae3.
AI/ML Overview
The provided document describes the safety and effectiveness information for the LEGEND™ Nitrile powder-free patient examination glove. The study demonstrates the device meets acceptance criteria by comparing its performance to established ASTM standards and FDA regulations.
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Standard/Acceptance Criteria | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 6319-00ae3 | Meets |
Palm Width (Small) | 80 +/- 10 mm | 80-90 mm |
Palm Width (Medium) | 95 +/- 10 mm | 90-100 mm |
Palm Width (Large) | 110 +/- 10 mm | 100-110 mm |
Palm Width (X-Large) | 120 +/- 10 mm | 110-120 mm |
Length (Small) | 220 mm min | 240-250 mm (all sizes) |
Length (Medium) | 230 mm min | 240-250 mm (all sizes) |
Length (Large) | 230 mm min | 240-250 mm (all sizes) |
Length (X-Large) | 230 mm min | 240-250 mm (all sizes) |
Thickness (Finger) | 0.05 mm min | 0.12 +/- 0.01 mm |
Thickness (Palm) | 0.05 mm min | 0.12 +/- 0.01 mm |
Physical Properties | ASTM standard D 6319-00ae3 | Meets |
Tensile strength (Before aging) | 14 MPa min | 19.0 MPa min |
Ultimate elongation (Before aging) | 500% min | 630% min |
Tensile strength (After aging, 22 hours @ 100°C) | 14 MPa min | 19.0 MPa min |
Ultimate elongation (After aging, 22 hours @ 100°C) | 400% min | 530% min |
Freedom from pinholes | 21 CFR 800.20 | Meets (Water Tight Test performed to manufacturer's AQL under 21 CFR 820.181) |
Powder Residual | ASTM standard D 6124-01 | Meets ( |