(170 days)
Nitrile powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Nitrile powder-free patient examination, that meets all of the requirements of ASTM standard D 6319-00ae3.
The provided document describes the safety and effectiveness information for the LEGEND™ Nitrile powder-free patient examination glove. The study demonstrates the device meets acceptance criteria by comparing its performance to established ASTM standards and FDA regulations.
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard/Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Dimension | ASTM standard D 6319-00ae3 | Meets |
| Palm Width (Small) | 80 +/- 10 mm | 80-90 mm |
| Palm Width (Medium) | 95 +/- 10 mm | 90-100 mm |
| Palm Width (Large) | 110 +/- 10 mm | 100-110 mm |
| Palm Width (X-Large) | 120 +/- 10 mm | 110-120 mm |
| Length (Small) | 220 mm min | 240-250 mm (all sizes) |
| Length (Medium) | 230 mm min | 240-250 mm (all sizes) |
| Length (Large) | 230 mm min | 240-250 mm (all sizes) |
| Length (X-Large) | 230 mm min | 240-250 mm (all sizes) |
| Thickness (Finger) | 0.05 mm min | 0.12 +/- 0.01 mm |
| Thickness (Palm) | 0.05 mm min | 0.12 +/- 0.01 mm |
| Physical Properties | ASTM standard D 6319-00ae3 | Meets |
| Tensile strength (Before aging) | 14 MPa min | 19.0 MPa min |
| Ultimate elongation (Before aging) | 500% min | 630% min |
| Tensile strength (After aging, 22 hours @ 100°C) | 14 MPa min | 19.0 MPa min |
| Ultimate elongation (After aging, 22 hours @ 100°C) | 400% min | 530% min |
| Freedom from pinholes | 21 CFR 800.20 | Meets (Water Tight Test performed to manufacturer's AQL under 21 CFR 820.181) |
| Powder Residual | ASTM standard D 6124-01 | Meets (<2mg/glove) |
| Biocompatibility | Primary Skin Irritation (Rabbits) | Passes (Not a Primary Skin Irritation) |
| Dermal sensitization (Guinea Pig) | Passes (Not a Dermal sensitization) |
2. Sample Size and Data Provenance
The document does not explicitly state the exact sample sizes used for each specific test (e.g., dimension, physical properties, pinholes, biocompatibility). However, it indicates that testing was conducted per recognized standards:
- Dimensions & Physical Properties: ASTM standard D 6319-00ae3.
- Freedom from Pinholes: 21 CFR 800.20 and the manufacturer's quality system acceptance criteria in 21 CFR 820.181.
- Powder Residual: ASTM standard D 6124-01.
- Biocompatibility: ISO10993-10 (specifically Primary Skin Irritation and Dermal Sensitization).
The data provenance is not explicitly stated in terms of country of origin but is implied to be from the manufacturer's internal testing as part of their premarket notification for a U.S. FDA submission. The studies were non-clinical (nonclinical submitted, reference, or relied on). The nature of the tests indicates prospective testing performed on samples of the manufactured gloves.
3. Number of Experts and Qualifications for Ground Truth
This information is not applicable. The device is a medical glove, and its performance is evaluated against objective, measurable physical, chemical, and biological standards, rather than requiring expert interpretation (e.g., radiologists, clinicians) for ground truth establishment.
4. Adjudication Method
This information is not applicable for this type of device and testing. Performance is measured against quantitative standards rather than subjective expert consensus.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic or AI-assisted interpretation devices where human performance is measured and compared with and without AI assistance. This device is a medical glove, and its effectiveness is determined by its physical and biocompatibility properties.
6. Standalone (Algorithm Only) Performance Study
No, a standalone (algorithm only) performance study was not done. This concept is relevant for software or AI-based devices. The LEGEND™ Nitrile powder-free patient examination glove is a physical medical device.
7. Type of Ground Truth Used
The ground truth used for evaluating the device's performance is based on established industry standards and regulatory requirements. Specifically, these include:
- ASTM standard D 6319-00ae3 (for dimensions and physical properties)
- 21 CFR 800.20 (for freedom from pinholes/water impermeability)
- ASTM standard D 6124-01 (for powder residual)
- ISO 10993-10 (for biocompatibility - primary skin irritation and dermal sensitization)
8. Sample Size for the Training Set
This information is not applicable. The device is a physical product, not a machine learning model, so there is no training set in the AI sense. The manufacturer would have internal quality control processes and statistical sampling plans for their manufacturing, but these are not disclosed as a "training set" for an algorithm.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable, as there is no "training set" in the context of an AI/machine learning device. The "ground truth" for the device's performance (i.e., the specifications that the device must meet) is established by the referenced national and international standards (ASTM, FDA regulations, ISO).
{0}------------------------------------------------
AUG = 8 2003
C Summary
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 "
applicant leave blank) "The assigned 510(k) number is:
Premarket Notification [510(k)] Summary
((a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | JDA (TIANJIN) PLASTIC RUBBER CO., LTD. |
|---|---|
| Submitter's address : | 17 HAIBIN 7 RD. TIANJIN TREE TRADE ZONE,TIANJIN ,300456,P.R.China |
| Phone number : | (86)22-25760468 |
| Fax number : | (86)22-25760466 |
| Name of contact person: | Wan Dong Hua, Technology Department Manager |
| Date the summary was prepared: | 23 May 2003 |
[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name | Nitrile powder-free patient examination glove |
|---|---|
| Proprietary/Trade name: | LEGEND™ Nitrile powder-free patient examination glove |
| Common Name: | Other clients private labelingPatient examination glove |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital (80) |
| Product Code: | LZA |
(a)(3)). An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* Nitrile powder-free patient examination, that meets all of the requirements of ASTM standard D 6319-00ae3.
Predicate device : Nitrile powder-free patient examination glove, JDA International Inc. K993247 .
|(a)(4)| A description of the device
{1}------------------------------------------------
Device Description : Nitrile powder-free patient examination, that meets all of the requirements of ASTM standard D 6319-00ae3.
[(a)(5)] The summary describes the intended use of the device
Device Intended Use: Nitrile powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The nitrile powder-free patient examination glove, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|---|---|---|
| Dimension | ASTM standard D 6319-00ae3. | Meets |
| Physical Properties | ASTM standard D 6319-00ae3. | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Residual | ASTM standard D 6124-01 | Meets<2mg/glove |
| Biocompatability | Primary Skin Irritation in rabbits | PassesNot a Primary Skin Irritation |
| Dermal sensitization in the guineapig | PassesNot a Dermal sensitization |
((b)(1)) A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Nitrile powder-free patient examination glove meet requirements per ASTM D6319-00ac3, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.
[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
[{b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the LEGEND™ Nitrile powder-free patient examination gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .
{2}------------------------------------------------
Appendix 2.0 Product specification
The Nitrile powder-free patient examination glove meet ALL the current specifications listed under the ASTM Specification D 6319-00ae3.
-
- Dimension
| (mm) | ||
|---|---|---|
| DEMENTION | ASTM D6319-00ae3 | LEGEND'S |
| Palm Width | ||
| Small | $80+/-10$ | 80-90 |
| Medium | $95 +/-10$ | 90-100 |
| Large | $110 +/-10$ | 100-110 |
| X-Large | $120 +/-10$ | 110-120 |
| Length | ||
| Small | 220 min | 240-250 |
| Medium | 230 min | for all sizes |
| Large | 230 min | for all sizes |
| X-Large | 230 min | for all sizes |
| Thickness | ||
| finger | 0.05 min | $0.12+/-0.01$ |
| palm | 0.05 min | $0.12+/-0.01$ |
2. Physical properties
| (mm) | ||||
|---|---|---|---|---|
| Tensile strength (MPa) | Ultimate elongation(%) | |||
| ASTMD6319-00ae3 | LEGEND'S | ASTM D6319-00ae3 | LEGEND'S | |
| Before agingSmallMediumLargex-large | 14 min | 19.0 min | 500 min | 630min |
| After aging22hours@100°CSmallMediumLargex-large | 14 min | 19.0 min | 400 min | 530min |
3. Water Tight Test
The pinhole or leak testing, the sampling and testing conform to the test methods and AQL established by the manufacturer under their quality system acceptance criteria in 21 CFR 820.181.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol featuring three wavy lines emanating from a central point, resembling a human figure.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 8 2003
JDA Tianjin Plastic Rubber Company, Limited C/O Mr. Chen Yuhong TUV Rheinland Beijing Office Rm 707 Avic Bldg. 2 Dong San Huan Nan Road Chaoyang District BEIJING 100022, P.R. CHINA
Re: K030529
Trade/Device Name: Legend (Brand) Nitrile Powder-Free Patient Examination Glove Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: April 23, 2003 Received: June 2, 2003
Dear Mr. Yuhong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Yuhong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Suser Runse
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
1K030524
3.0 Indications for Use Statement: Include the following or equivalent Indications for Use page. The information, data and labeling claims in the entire 510(k) submission must support and agree with the Indications for Use statement.
INDICATIONS FOR USE
Applicant:_ JDA (Tianjin) Plastic Rubber Co., Ltd.
510(k) Number (if known):_*
Device Name: __ Nitrile powder-free patient examination glove _
Indications For Use:
Nitrile powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Clim S. Lin
510(k) Number: K030529
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.