Search Results
Found 3 results
510(k) Data Aggregation
(366 days)
The Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim (50 micrograms or less total water extractable protein per gram of glove), is a disposable device intended for Medical Purposes that is worn on the examiners hand to prevent contamination between patient and examiner.
Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim
Here's a breakdown of the acceptance criteria and the study information for the "Non-Sterile, Powder-Free, Latex Examination Gloves with Lanolin and Vitamin E Coating, Green and with Protein Labeling Claim," based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Criteria | Acceptance Criteria (Minimum/Maximum) | Reported Device Performance |
|---|---|---|
| Material | Latex | Latex |
| Cuff | Beaded | Beaded |
| Powder Residue | 2mg/glove maximum | N/A (implied compliance) |
| Inspection Level: Dimensions | S-2, AQL 4.0 | N/A (implied compliance) |
| Inspection Level: Physical Properties | S-2, AQL 4.0 | N/A (implied compliance) |
| Inspection Level: Water Tight Test 1000ml | G-1, AQL 1.5 | N/A (implied compliance) |
| Inspection Level: Visual Major Defects | G-1, AQL 1.5 | N/A (implied compliance) |
| Inspection Level: Visual Minor Defects | G-1, AQL 2.5 | N/A (implied compliance) |
| Overall Length | 240 mm minimum | 240 mm minimum |
| Width (medium glove) | 95 mm minimum | 95 mm minimum |
| Thickness | 0.08 mm minimum | 0.08 mm minimum |
| Tensile Strength (Before Aging) | 18.0 Mpa minimum | 18.0 Mpa minimum |
| Ultimate Elongation (Before Aging) | 650% minimum | 650% minimum |
| Tensile Strength (After Aging) | 14.0 Mpa minimum | 14.0 Mpa minimum |
| Ultimate Elongation (After Aging) | 500% minimum | 500% minimum |
| Special Properties | Lanolin and Vitamin E inner coating | Lanolin and Vitamin E inner coating |
| Protein Labeling Claim Level | < 50ug/g | < 50ug/g |
Note: The reported device performance often states "minimum" or "maximum" values that align with the acceptance criteria, suggesting the device met or exceeded these thresholds. For inspection parameters, the text states "In compliance with ASTM D3578-10.ISO 2859-1:1999" and lists the AQLs, implying the device met these standards through the specified inspection levels.
The provided text describes a 510(k) premarket notification for a medical device (examination gloves). This type of submission focuses on demonstrating substantial equivalence to a previously legally marketed device (predicate device), rather than a detailed clinical study proving efficacy in the same way a new drug or novel high-risk device might. Therefore, many of the typical questions for a "study that proves the device meets acceptance criteria" are not directly applicable or are answered implicitly through the regulatory framework.
Here's a breakdown of the other requested information based on the provided text:
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: The document does not specify a distinct "test set" sample size in terms of the number of gloves tested for each criterion. Instead, it refers to industry standards (ASTM D3578-10, ISO 2859-1:1999) and inspection levels (S-2, G-1) with Acceptance Quality Limits (AQLs). These standards inherently define the sampling plans used for quality control testing.
- Data Provenance: The manufacturer is "UG Global Resources Sdn Bhd 181 Floor No. 18 Jalan Dato' Abd Rahman 70000 Seremban Malaysia." This indicates the testing and data likely originated from Malaysia. The study is prospective in nature, as it involves testing the manufactured gloves to ensure they meet the defined specifications before market release.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- This question is not applicable to this type of device and submission. The "ground truth" for examination gloves is defined by adherence to established engineering and material science standards (ASTM, ISO), manufacturing quality control processes, and biocompatibility testing, not by expert consensus on clinical interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This question is not applicable. Adjudication methods like 2+1 or 3+1 are used for interpreting subjective data, typically in clinical trials involving image analysis or symptom assessment. The tests performed on these gloves are objective physical and chemical measurements.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This question is not applicable. MRMC studies are used for evaluating diagnostic devices, particularly those involving human interpretation (e.g., radiology AI). This device is an examination glove, which does not involve "readers" or "AI assistance."
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- This question is not applicable. This device is a physical product, not an algorithm, and does not involve AI or standalone algorithmic performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The "ground truth" for the performance of these gloves is based on:
- Established Industry Standards: Primarily ASTM D3578-10 for physical properties and ISO 2859-1:1999 for quality assurance.
- FDA 1000 ml water test: A specific regulatory test for watertight integrity.
- Biocompatibility Test Results: To ensure safety in contact with skin.
- Residual Powder Contents tests: To verify "powder-free" claim.
- Protein Labeling Claim Level: Chemical testing to quantify protein levels.
8. The sample size for the training set
- This question is not applicable. "Training sets" are relevant for machine learning algorithms. For manufactured goods like gloves, there isn't a "training set" in this context. The manufacturing process is designed and monitored to consistently produce gloves meeting specifications, and ongoing quality control (using the stated inspection levels and AQLs) ensures this.
9. How the ground truth for the training set was established
- This question is not applicable as there is no "training set" in the context of this device. The "ground truth" for acceptable glove performance is established by the specifications themselves, derived from regulatory requirements and industry standards.
Ask a specific question about this device
(354 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner.
Non-Sterile, Powder-Free, Nitrile Examination Gloves with Lanolin and Vitamin E Coating, Blue
1. Table of acceptance criteria and the reported device performance:
| Characteristics | Acceptance Criteria (Standards) | Device Performance |
|---|---|---|
| Dimensions | ASTM D 6319-10 | Meets |
| Physical Properties | ASTM D 6319-10 | Meets |
| Freedom from Pinholes | ASTM D 6319-10 and ASTM D 5151-06 (1000ml water test both before and after aging) | Meets |
| Powder-Free Residue | ASTM D 6319-10 and ASTM D 6124-06 (Maximum 2mg/glove) | Meets |
| Biocompatibility | Dermal Sensitization (as per ISO 10993-10:2002) | Not a contact skin sensitizer |
| Primary Skin Irritation Test | Not a primary skin irritant |
2. Sample size used for the test set and the data provenance:
The document does not explicitly state the specific sample sizes used for each test (e.g., separate sample sizes for physical properties, pinholes, etc.). However, it refers to compliance with ASTM standards, which would dictate the sample sizes and testing methodologies. The data provenance is from Ug Global Resources Sdn Bhd, based in Malaysia, indicating the tests were conducted by the manufacturer for their 510(k) submission. The type of study is retrospective, as it describes tests performed on existing glove samples to demonstrate compliance with standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to this type of device and study. The "ground truth" for glove performance is established by standardized, objective tests outlined in ASTM and ISO standards, not by expert consensus or interpretation of subjective data.
4. Adjudication method for the test set:
This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in medical imaging or clinical trial settings where there's subjective interpretation by experts. For glove testing, the results are quantitative measurements against predefined thresholds in the standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. This is a submission for a medical device (examination gloves), not an AI algorithm or diagnostic tool. Therefore, MRMC studies and AI assistance are not relevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This information is not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used:
The ground truth for this device is based on objective measurements and performance criteria defined by international and national standards, specifically:
- ASTM D 6319-10: Standard Specification for Nitrile Examination Gloves for Medical Application
- ASTM D 5151-06: Standard Test Method for Detection of Holes in Medical Gloves
- ASTM D 6124-06: Standard Test Method for Residual Powder on Medical Gloves
- ISO 10993-10:2002: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
These standards provide the methodology and acceptable limits for various properties like dimensions, physical properties, freedom from pinholes, powder residue, and biocompatibility.
8. The sample size for the training set:
This information is not applicable. This is not an AI/machine learning study that involves training and test sets. The glove's performance is evaluated against established standards through direct testing.
9. How the ground truth for the training set was established:
This information is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(15 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between the patient and the examiner.
Lano-E Powder-Free Nitrile Examination Glove. Class I Powder-Free Nitrile examination glove 80LZA that meets a all of the requirements of ASTM Standard D 6319 - 00 Rev A. Powder Residue: Maximum 2mg/glove. Material: Nitrile. Cuff: Beaded. Processed with pharmaceutical quality lanolin as the emollient and conditioning agent. Also contains Vitamin E which complies with the current USP, Ph.Eur., DAB and BP monographs.
This looks like a 510(k) submission for examination gloves, not an AI/ML device. Therefore, the requested information about acceptance criteria, study details, ground truth, and AI-specific metrics (like MRMC studies or human reader improvement with AI) is not applicable to this document.
The document describes the device, its intended use, and its compliance with ASTM standards for physical properties. The "study" here is essentially the testing performed to ensure the gloves meet these established standards.
Here's the relevant information adapted to the context of this device:
Acceptance Criteria and Device Performance (for Lano-E Powder-Free Nitrile Examination Glove)
The Lano-E Powder-Free Nitrile Examination Glove meets the physical property requirements of ASTM D 6319-00 Rev A and the FDA 1000 ml water test both before and after aging.
1. Table of Acceptance Criteria and Reported Device Performance
Inspection Parameters (Acceptance by Inspection Level and AQL)
| Criteria | Inspection Level | AQL (Acceptance Quality Limit) |
|---|---|---|
| Dimensions | S-2 | 4.0 |
| Physical Properties | S-2 | 4.0 |
| Water Tight Test 1000ml | G-1 | 1.5 |
| Visual Major Defects | G-1 | 1.5 |
| Visual Minor Defects | G-1 | 2.5 |
Physical Properties (Before Aging)
| Dimensions/Property | Acceptance Criteria | Reported Device Performance** |
|---|---|---|
| Overall Length | 240 mm minimum | Met |
| Width (medium glove) | 95 mm minimum | Met |
| Palm Thickness | 0.10 to 0.15 mm | Met |
| Finger Thickness | 0.10 to 0.17 mm | Met |
| Cuff Thickness | 0.10 to 0.15 mm | Met |
| Tensile Strength | 14.0 Mpa minimum | Met |
| Ultimate Elongation | 700% minimum | Met |
| Pinhole AQL | 1.5 minimum | Met |
Physical Properties (After Aging)
| Property | Acceptance Criteria | Reported Device Performance** |
|---|---|---|
| Tensile Strength | 14.0 Mpa minimum | Met |
| Ultimate Elongation | 500% minimum | Met |
| Pinhole AQL | 1.5 minimum | Met |
Note: The document explicitly states: "The Lano-E Powder-Free Nitrile Examination Glove meets the physical property requirements of ASTM D 6319-00 Rev A and the FDA 1000 ml water test both before and after aging." This implies that the device performance met all listed acceptance criteria.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a precise sample size for the testing of each individual criterion. However, the testing was conducted "in compliance with ASTM D6319-00 Rev A, EN 455-3 : 2000, EN 455-2 : 2000, EN 455-1 : 2000, ISO 2859-1:1999". These standards typically define the sampling plans (e.g., AQL tables) for quality control and testing.
- Data Provenance: The device manufacturer, N.S. Uni-Gloves Sdn. Bhd., is based in Malaysia. The testing data would have been generated internally by the manufacturer or by a contracted laboratory following the specified international standards. The data is retrospective in the sense that it was collected as part of the product development and quality assurance process prior to the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Not Applicable: For this type of device (examination gloves), "ground truth" is established by adherence to measurable physical and chemical specifications outlined in established industry standards (like ASTM and EN standards). There are no "experts" in the sense of clinical reviewers establishing a subjective truth. The "truth" is objective measurement against a standard.
4. Adjudication Method for the Test Set
- Not Applicable: Adjudication methods like 2+1 or 3+1 are typically used for subjective assessments (e.g., image interpretation). For examination gloves, the testing involves objective measurements (e.g., dimension, tensile strength, water tightness), which do not require human adjudication in the traditional sense. The pass/fail is determined by whether the measured value falls within the specified range or meets the AQL.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable: This is an examination glove, not an AI/ML diagnostic device. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable: This is an examination glove. There is no algorithm or AI component involved.
7. The type of ground truth used
- Objective Standards and Measurements: The "ground truth" is defined by the objective, quantifiable specifications outlined in recognized international standards (ASTM D 6319-00 Rev A, EN 455 series, ISO 2859-1:1999) and the FDA 1000ml water tight test. This includes material properties, physical dimensions, and defect rates.
8. The sample size for the training set
- Not Applicable: There is no "training set" in the context of examination gloves. The manufacturing process is controlled to produce gloves that consistently meet the specifications. Continuous quality monitoring, potentially using statistical process control, ensures ongoing adherence.
9. How the ground truth for the training set was established
- Not Applicable: As there is no training set, this question is not applicable. The design and manufacturing process for the gloves are based on achieving the objective criteria defined by the standards.
Ask a specific question about this device
Page 1 of 1