    Device Name :

    IMMUNICON CELLTRACKS ANALYZER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use for the Immunicon CellTracks™ System is as a general-purpose laboratory bench top scanning cytometer designed to enumerate fluorescently labeled leukocyte subsets that are immuno-magnetically selected and aligned. The system is for in vitro diagnostic use.

    To enumerate fluorescently labeled leukocyte subsets that are immuno-magnetically selected and aligned.

    Device Description

    The CellTracks Analyzer is used in conjunction with reagents containing ferrofluids and fluorescent conjugates. Depending on the application, samples may be prepared manually or using the CellTracks AutoPrep Sample Preparation System. Ferrofluid consists of a magnetic core surrounded by a polymeric layer coated with antibodies for capturing cells. Ferrofluid particles are colloidal (25-100 nanometers in radius) and when mixed with a sample containing target cells, the antibodies bound to the ferrofluid bind the associated antigens to the target cells. Fluorescent reagents are added for the identification and the enumeration of the target cells. The ferrofluid/sample mixture is placed in a magnetic field (Magnest), to select and align the labeled target cells. The Magnest, containing the fluorescence-labeled magnetically aligned target cells, is placed on the CellTracks analyzer and is ready for analysis.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance Criteria (Implied)Reported Device Performance
    CD4 Slope (vs. predicate)Close to 1.00.95
    CD4 R² (vs. predicate)High, close to 1.00.9802
    CD3 Slope (vs. predicate)Close to 1.00.77
    CD3 R² (vs. predicate)High, close to 1.00.9375
    LinearityNot explicitly stated but expected to be broad0 to 2,000 cells/ul
    CD3 Normal Control Within-run CVNot explicitly stated but expected to be low3.68% to 4.34%
    CD3 Low Control Within-run CVNot explicitly stated but expected to be low3.89% to 6.60%
    Interference from LipidsNo adverse affectNo adverse affect up to 5,000 mg/dL
    Interference from HematocritsNo adverse affectNo adverse affect from 43.3% to 76.2%
    Interference from Platelet CountsNo adverse affectNo adverse affect from 210 x 10² to 2,040 x 10²
    Interference from WBC CountsNo adverse affectNo adverse affect from 4.1 x 10³ to 25.0 x 10³
    Interference from Free HemoglobinNo adverse affectNo adverse affect from 10.3 to 12.4 g/dL
    Electrical Safety (IEC 61326-1998, IEC 601010-2-101)Met requirementsMet requirements
    Laser Safety (21 CFR Part 1040.10)Met requirementsMet requirements