510(k) Data Aggregation
(52 days)
The Encompass™ SRS Headframe provides noninvasive stereotactic head and neck immobilization by using a patient specific thermoplastic mask that conforms to the patient's features to provide accurate, reproducible positioning. repositioning, and immobilization. The Encompass™ SRS Headframe allows the patient to undergo diagnostic imaging in the same position as that of the treatment position enabling more accurate radiation therapy.
Qfix, has made a modification to its previously cleared Encompass™ SRS Immobilization System (cleared in K152321). In this modified variant of the Encompass™ SRS Immobilization System, the essential geometry of the head portion of the Encompass™ SRS Standalone Device is isolated to create a separate attachment which can be affixed to patient positioning devices that accept S-type thermoplastic masks, such as Qfix's Portrait™ Intracranial, Head and Neck Device (a device for use with S-type thermoplastics). Ofix intends to initially market the subject device for use with the Encompass™ SRS Fibreplast® System with optional IntegraBite™ (also cleared in K152321) and its devices which share a hole pattern and geometry with its Portrait™ Intracranial, Head and Neck Device. Examples of such devices include the kVue™ Portrait™ Intracranial. Head and Neck Insert and the Symphony® Portrait™ Transfer Device (cleared in K160627). This creates a flexible solution for a variety of imaging and radiotherapy treatment applications using stereotactic radiosurgery, such as radiation oncology and the treatment of noncancerous conditions and malformations such as trigeminal neuralgia.
The provided text is a 510(k) summary for the Qfix Encompass™ SRS Headframe, a medical device for patient immobilization during radiation therapy. It claims substantial equivalence to a previously cleared predicate device rather than presenting a novel AI/software-as-a-medical-device (SaMD) study. Therefore, the information typically requested about AI/SaMD acceptance criteria, test set details, expert ground truth, MRMC studies, and training set information is not present in this document.
However, I can extract information related to the device's performance claims and general testing.
Here's a breakdown of what can be extracted based on the provided text, and what cannot be extracted because it's not an AI/SaMD submission:
Acceptance Criteria and Device Performance (General Device Performance):
The document does not provide a table of explicit, quantifiable acceptance criteria with corresponding performance data in the typical sense of a SaMD validation study. Instead, it makes a general claim about performance relative to its predicate device.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Accuracy/Immobilization | Provides sub-millimetric immobilization for imaging and radiotherapy, including SRS. |
Limiting motion to less than 1 mm in most clinical applications using stereotactic radiosurgery. | |
Compatibility | Compatible with radiation therapy environments, including stereotactic radiosurgery. |
Accepts the same Encompass™ SRS Fibreplast® System patient immobilization masks. | |
Intended for use with patient positioning devices that accept S-type thermoplastic masks (e.g., Qfix's Portrait™ Intracranial, Head and Neck Device). | |
MR compatible variants (Encompass™ SRS MR Headframe is MR Safe per ASTM F2503-13). | |
Safety & Effectiveness | No new questions of safety or effectiveness. |
Risk management in compliance with ISO 14971:2007. | |
Adheres to recognized and established industry practices and standards. | |
Non-Clinical Bench Testing & Validation | Met all acceptance criteria for testing conducted and was appropriately validated per its intended use based on: |
- Verification of hardware specifications
- Accuracy validation via motion studies
- Ease of use/ergonomics assessments
- Verification of aluminum equivalence
- Verification of compatibility with optical tracking systems |
Information NOT available in the provided text (typical for AI/SaMD studies):
- Sample size used for the test set and the data provenance: Not applicable to this device type and submission. The document mentions "testing using healthy volunteers" for non-clinical bench testing but does not specify a sample size for these volunteers.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical immobilization device is typically established through direct measurement (e.g., motion studies, calibration) rather than expert interpretation of images or data.
- Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI-assisted device for diagnostic interpretation.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is a physical medical device, not an algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For this physical device, ground truth for accuracy claims would be based on physical measurements (e.g., displacement measurements during motion studies) rather than expert interpretation or pathological findings.
- The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
- How the ground truth for the training set was established: Not applicable, as this is not an AI/machine learning device.
Summary of Device and Study Type:
This 510(k) submission is for a physical medical device (Encompass™ SRS Headframe) used for patient immobilization during radiation therapy. The submission relies on demonstrating substantial equivalence to a previously cleared predicate device (Encompass™ SRS Immobilization System, K152321). Therefore, the "study" described is primarily non-clinical bench testing and testing with healthy volunteers to ensure the modified device maintains the performance characteristics (e.g., sub-millimetric accuracy, compatibility) and safety profile of the predicate device, rather than a clinical trial or AI/SaMD validation study with diagnostic endpoints.