Search Results
Found 3 results
510(k) Data Aggregation
(93 days)
Disposable Medical Nitrile Examination Gloves (Non Sterile) are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves (Non sterile). The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I) for single use only. The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.
The provided document is a 510(k) Premarket Notification for Disposable Medical Nitrile Examination Gloves (Non-Sterile). This type of document is for a medical device that does not use AI/ML, and therefore, the requested information regarding AI/ML device performance and related study details is not applicable.
However, I can extract the acceptance criteria and performance data for this device based on the non-clinical testing summary provided.
1. Table of Acceptance Criteria and Reported Device Performance
Test Method | Purpose | Acceptance Criteria | Reported Device Performance/Result |
---|---|---|---|
ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Dimensions Test | To determine the width, length, and thickness of the gloves | Width: | |
80mm ±10mm (for small size) | |||
95mm ±10mm (for medium size) | |||
110 mm ±10mm (for large size) | |||
Length: | |||
220 mm (Minimum) (for small size) | |||
230 mm (Minimum)(for medium, large size) | |||
Thickness: | |||
Finger -0.05 mm (Minimum) (for small, medium, large size) | |||
Palm -0.05 mm (Minimum)(for small, medium, large size) | Width: | ||
small size: 82 to 89mm | |||
medium size: 91 to 96mm | |||
large size: 115 to 116mm | |||
Length: | |||
small size: 230mm (minimum) | |||
medium size: 240mm (minimum) | |||
large size: 238mm (minimum) | |||
Thickness: | |||
Finger: | |||
small size: 0.12mm (minimum) | |||
medium size: 0.13mm (minimum) | |||
large size: 0.12mm (minimum) | |||
Palm: | |||
small size: 0.08 mm (minimum) | |||
medium size: 0.08mm (minimum) | |||
large size: 0.09 mm (minimum) | |||
Passed | |||
ASMT D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application - Physical Requirements Test | To determine the tensile strength and ultimate elongation before and after acceleration aging | Before Acceleration Aging: | |
Tensile Strength (MPa): 14 (Minimum) | |||
Ultimate Elongation (%): 500 (Minimum) | |||
After Acceleration Aging: | |||
Tensile Strength (MPa): 14 (Minimum) | |||
Ultimate Elongation (%): 400 (Minimum) | Before Acceleration Aging: | ||
Tensile Strength (MPa): 20 (Minimum) | |||
Ultimate Elongation (%): 521 (Minimum) | |||
After Acceleration Aging: | |||
Tensile Strength (MPa): 19 (Minimum) | |||
Ultimate Elongation (%): 505 (Minimum) | |||
Passed | |||
ASTM D6319-19 (ASTM D5151-11) Standard Test Method for Detection of Holes in Medical Gloves | To determine the holes in the gloves | (Mean) AQL 2.5 | Passed |
ASMT D6319-19 (ASTM D6124-11) Standard Test Method for Residual Powder on Medical Gloves | To determine the residual powder in the gloves |
Ask a specific question about this device
(140 days)
The Disposable Medical Nitrile Examination Glove is intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The subject device is powder free nitrile examination gloves. The subject device is blue. The subject device is non-sterile.
The provided text describes the acceptance criteria and the study conducted for a medical device: "Disposable Medical Nitrile Examination Gloves (non-sterile)." This is a Class I medical device, and the study focused on non-clinical tests to demonstrate substantial equivalence to a predicate device, rather than the types of studies typically performed for AI/ML-based diagnostic devices. Therefore, many of the requested categories are not applicable.
Here's the information extracted from the document:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ISO 10993-5 Biological evaluation of medical devices Test for in vitro cytotoxicity | To determine the potential cytotoxicity. | Non-cytotoxic | Pass |
ISO 10993-11 Biological evaluation of medical devices Part 11: Tests for Systemic Toxicity | To determine the potential systemic toxicity. | Does not cause systemic toxicity. | Pass |
ISO 10993-10 Biological evaluation on medical device Part 10: Test for Irritation and Skin Irritation | To determine the potential for irritation and skin irritation. | Non-irritant, and Non skin irritant. | Pass |
ISO 10993-10 Biological evaluation on medical device Part 10: Test Skin Sensitization | To determine the potential skin sensitization. | Does not cause skin sensitization. | Pass |
ASTM D6319 Standard Specification for Nitrile Examination Gloves for Medical Application | To test for: (1) freedom from holes (2) physical dimensions (3) Aging | (1) Shall not leak (2) For size L (mm): Width: 110 ± 10, Length: ≥ 230, Finger and palm thickness: median value ≥ 0.05 (3) After Aging: Tensile Strength: ≥ 14 MPa, Ultimate elongation: ≥400% | Pass |
2. Sample size used for the test set and the data provenance
The document does not specify a "test set" in the context of data for an AI/ML device. The tests performed are for medical gloves and involve physical and biological property assessments, not analysis of data from patients or images. The provenance of the samples (e.g., country of origin for the gloves tested) is not specified beyond the manufacturer's location (China). These were non-clinical tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The ground truth for the physical and biological properties of the gloves is established by standardized testing protocols (ISO, ASTM) and measurements, not by expert interpretation of data relevant to AI/ML.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. As noted above, the tests are for physical and biological properties of gloves, not for AI/ML performance requiring adjudication of interpretations.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a medical glove, not an AI-based diagnostic tool. No MRMC study was conducted.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a medical glove, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the non-clinical tests is based on objective measurements and established standards for material properties (e.g., tensile strength, elongation), dimensional specifications (e.g., length, width, thickness), and biocompatibility assays (e.g., cytotoxicity, irritation, sensitization, systemic toxicity).
8. The sample size for the training set
Not applicable. This device does not involve an AI/ML algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable. This device does not involve an AI/ML algorithm.
Ask a specific question about this device
(115 days)
Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves, Non-sterile. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.
The document describes the acceptance criteria and study proving the device meets them for Disposable Medical Nitrile Examination Gloves (Non-Sterile) (K211075).
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
Test Method | Acceptance Criteria | Reported Device Performance |
---|---|---|
ASMT D6319-19 - Physical Dimensions Test (Width) | Small size: 80 ± 10 mm | |
Medium size: 95 ± 10 mm | ||
Large size: 110 ± 10 mm | ||
Extra-large size: 120 ± 10 mm | Small size: 82mm (Minimum), 82 mm (Mean), 85 mm (Maximum) | |
Medium size: 91 mm (Minimum), 92 mm (Mean), 93 mm (Maximum) | ||
Large size: 108 mm (Minimum), 109 mm (Mean), 110 mm (Maximum) | ||
Extra-large size: 121 mm (Minimum), 122 mm (Mean), 125 mm (Maximum) | ||
All Passed | ||
ASMT D6319-19 - Physical Dimensions Test (Length) | Small size: 220 mm (Minimum) | |
Medium size: 230 mm (Minimum) | ||
Large size: 230 mm (Minimum) | ||
Extra-large size: 230 mm (Minimum) | Small size: 232 mm (Mean), 230 mm (Minimum) | |
Medium size: 242 mm (Mean), 240 mm (Minimum) | ||
Large size: 239 mm (Mean), 237 mm (Minimum) | ||
Extra-large size: 243 mm (Mean), 240 mm (Minimum) | ||
All Passed | ||
ASMT D6319-19 - Physical Dimensions Test (Thickness) | Small size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) | |
Medium size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) | ||
Large size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) | ||
Extra-large size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) | Small size: Finger – 0.13 mm (Mean), 0.12 mm (Minimum); Palm – 0.08 mm (Mean), 0.08 mm (Minimum) | |
Medium size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.08 mm (Minimum) | ||
Large size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.09 mm (Minimum) | ||
Extra-large size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.08 mm (Minimum) | ||
All Passed | ||
ASMT D6319-19 - Physical Requirements Test (Tensile Strength & Elongation) | Before Aging: Tensile Strength 14 MPa (Minimum), Ultimate Elongation 500% (Minimum) | |
After Acceleration Aging: Tensile Strength 14 MPa (Minimum), Ultimate Elongation 400% (Minimum) | Before Acceleration Aging: Tensile Strength 31 MPa (Mean), 17 MPa (Minimum); Ultimate Elongation 563% (Mean), 520% (Minimum) | |
After Acceleration Aging: Tensile Strength 41 MPa (Mean), 28 MPa (Minimum); Ultimate Elongation 555% (Mean), 524% (Minimum) | ||
Passed | ||
ASTM D6319-19 (ASTM D5151-11) - Freedom from Holes (Water Tight) | AQL 2.5 | Passed |
ASMT D6319-19 (ASTM D6124-11) - Residual Powder |
Ask a specific question about this device
Page 1 of 1