K Number
K211075

Validate with FDA (Live)

Date Cleared
2021-08-05

(115 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Device Description

The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves, Non-sterile. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

AI/ML Overview

The document describes the acceptance criteria and study proving the device meets them for Disposable Medical Nitrile Examination Gloves (Non-Sterile) (K211075).

Here's the breakdown of the information requested:


1. Table of Acceptance Criteria and Reported Device Performance

Test MethodAcceptance CriteriaReported Device Performance
ASMT D6319-19 - Physical Dimensions Test (Width)Small size: 80 ± 10 mm Medium size: 95 ± 10 mm Large size: 110 ± 10 mm Extra-large size: 120 ± 10 mmSmall size: 82mm (Minimum), 82 mm (Mean), 85 mm (Maximum) Medium size: 91 mm (Minimum), 92 mm (Mean), 93 mm (Maximum) Large size: 108 mm (Minimum), 109 mm (Mean), 110 mm (Maximum) Extra-large size: 121 mm (Minimum), 122 mm (Mean), 125 mm (Maximum) All Passed
ASMT D6319-19 - Physical Dimensions Test (Length)Small size: 220 mm (Minimum) Medium size: 230 mm (Minimum) Large size: 230 mm (Minimum) Extra-large size: 230 mm (Minimum)Small size: 232 mm (Mean), 230 mm (Minimum) Medium size: 242 mm (Mean), 240 mm (Minimum) Large size: 239 mm (Mean), 237 mm (Minimum) Extra-large size: 243 mm (Mean), 240 mm (Minimum) All Passed
ASMT D6319-19 - Physical Dimensions Test (Thickness)Small size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) Medium size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) Large size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum) Extra-large size: Finger – 0.05 mm (Minimum), Palm – 0.05 mm (Minimum)Small size: Finger – 0.13 mm (Mean), 0.12 mm (Minimum); Palm – 0.08 mm (Mean), 0.08 mm (Minimum) Medium size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.08 mm (Minimum) Large size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.09 mm (Minimum) Extra-large size: Finger – 0.14 mm (Mean), 0.13 mm (Minimum); Palm – 0.09 mm (Mean), 0.08 mm (Minimum) All Passed
ASMT D6319-19 - Physical Requirements Test (Tensile Strength & Elongation)Before Aging: Tensile Strength 14 MPa (Minimum), Ultimate Elongation 500% (Minimum) After Acceleration Aging: Tensile Strength 14 MPa (Minimum), Ultimate Elongation 400% (Minimum)Before Acceleration Aging: Tensile Strength 31 MPa (Mean), 17 MPa (Minimum); Ultimate Elongation 563% (Mean), 520% (Minimum) After Acceleration Aging: Tensile Strength 41 MPa (Mean), 28 MPa (Minimum); Ultimate Elongation 555% (Mean), 524% (Minimum) Passed
ASTM D6319-19 (ASTM D5151-11) - Freedom from Holes (Water Tight)AQL 2.5Passed
ASMT D6319-19 (ASTM D6124-11) - Residual Powder< 2.0 mg/glove1.02 mg/glove Passed
ISO 10993-10 - Skin IrritationUnder the condition of study not an irritantUnder the condition of study not an irritant
ISO 10993-10 - Skin SensitizationUnder the conditions of the study not a sensitizer.Under the conditions of the study not a sensitizer.
ISO 10993-5 - In Vitro CytotoxicityUnder the conditions of study non cytotoxicUnder the condition of the study, the test article showed no cytotoxicity to L-929 cells and cytotoxic level was 0.
ISO 10993-11:2017 - Acute Systemic ToxicityUnder the conditions of the study no systemic toxicityUnder the conditions of the study no systemic toxicity

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the exact "sample size" for each test in terms of the number of individual gloves or test replicates. However, it indicates that the testing was performed according to ASTM D6319-19, which specifies sampling plans and test methods for nitrile examination gloves. For example, for "Freedom from Holes," it refers to an AQL (Acceptance Quality Limit) of 2.5, which is a statistical measure related to sampling.

  • Data Provenance: The testing was non-clinical. The specific country of origin for the data is not explicitly stated beyond the manufacturer being in China (Jiangsu Cherish Medical Technology Co., Ltd). The studies are retrospective in the sense that they were conducted on already manufactured gloves.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • This information is not applicable as the document describes non-clinical performance testing of a physical medical device (gloves), not a diagnostic algorithm requiring expert ground truth or consensus. The "ground truth" is established by the standardized test methods themselves (e.g., measuring dimensions, tensile strength, presence of holes).

4. Adjudication Method for the Test Set

  • This information is not applicable for the reasons stated above. Adjudication methods are typically relevant for human interpretation or AI model performance evaluation against expert-derived ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices often involving human readers or AI algorithms for interpretation, not for examination gloves.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Not applicable. This document evaluates a physical medical device (gloves), not a software algorithm.

7. Type of Ground Truth Used

The "ground truth" for this device's performance is established by standardized test methods and their defined specifications. These include:

  • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application)
  • ASTM D5151-11 (Standard Test Method for Detection of Holes in Medical Gloves)
  • ASTM D6124-11 (Standard Test Method for Residual Powder on Medical Gloves)
  • ISO 10993-10 (Biological Evaluation of Medical Devices – Tests for skin irritation and skin sensitization)
  • ISO 10993-5 (Biological Evaluation of Medical Devices – Tests for in vitro cytotoxicity)
  • ISO 10993-11:2017 (Biological Evaluation of Medical Devices – Tests for acute systemic toxicity)

8. Sample Size for the Training Set

  • Not applicable. This device is a physical product (examination gloves), not an AI/ML model for which "training sets" are used.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

August 5, 2021

Jiangsu Cherish Medical Technology Co., Ltd % Johnson Liu Consultant Cnmed Consultant 31 Archer St Upper MT Gravatt, QLD 4122 Australia

Re: K211075

Trade/Device Name: Disposable Medical Nitrile Examination Gloves (Non-Sterile) Regulatory Class: 21 CFR 880.6250 Product Code: LZA Dated: July 20, 2021 Received: July 26, 2021

Dear Johnson Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K211075

Device Name

Disposable Medical Nitrile Examination Gloves (Non-Sterile)

Indications for Use (Describe)

Disposable Medical Nitrile Examination Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the word "Cherish" in a blue, sans-serif font. The word is written in all lowercase letters except for the first letter, which is capitalized. There is a small "TM" symbol in the upper right corner of the word. The word is the main focus of the image and is centered.

510(K) Summary (K211075)

DISPOSABLE MEDICAL NITRILE EXAMINATION GLOVES, NON-STERILE

Preparation Date: August 5th, 2021

1. SUBMITTER

Company Name: JIANGSU CHERISH MEDICAL TECHNOLOGY CO., LTD Company Address: No. 55 Maqianan Chaoyang Road, Jiaoxi Centralized Industrial Zone Changzhou. CN 213000

Contact Person: Johnson Liu Telephone Number: +614-0158-9995 Email: Johnson@cnmed.com.au

2. NAME OF THE DEVICE

Trade Name / Proprietary Name: Cherish Nitrile Examination Gloves Device Name: Disposable Medical Nitrile Examination Gloves (Non-Sterile) Device Classification Name: Patient Examination Gloves Device Class: Class I Device Classification Number: 21 CFR 880.6250 Product Code: LZA

3. IDENTIFICATION OF THE LEGALLY MARKETED DEVICE

Predicate Device: K203191 Applicant: Nathan Trading Co., Ltd Device Name: LYDUS NITRILE EXAMINATION GLOVES, POWDER FREE Device Classification Name: Patient Examination Gloves Device Classification Number: 21 CFR 880.6250 Device Class: Class I Product Code: LZA Review Panel: General Hospital

4. DEVICE DESCRIPTION

The subject device in this 510(k) Notification is Disposable Medical Nitrile Examination Gloves, Non-sterile. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and non-sterile (Per 21 CFR 880.6250, Class I). The device meets the specifications in ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

5. Indications for Use

Disposable Medical Nitrile Examination Gloves, Non-sterile is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

{4}------------------------------------------------

6. TECHNOLOGICAL CHARACTERISTIC COMPARISON FOR THE PROPOSED AND PREDICATE DEVICES

DEVICE PERFORMANCE
CHARACTERISTICSSUBJECTPREDICATEREMARKS
510(k) NumberK211075K203191-
Device NameDisposable Medical NitrileExamination GlovesPowder Free, Non sterileLYDUS Nitrile ExaminationGloves, Powder Free-
Product CodeLZALZASame
Intended Use/Indications forUseDisposable Nitrile ExaminationGloves Non-sterile is disposabledevices intended for medical purposethat are worn on the examiner's handto prevent contamination betweenpatient and examiner. The device isfor over the-counter use.LYDUS Nitrile ExaminationGloves, Powder Free is adisposable deviceintended for medical purposesthat is worn on the examiner'shand or fingers to preventcontamination between patientand examiner.Same
Materials of Use(ASTMD6910/D6910M-19)Nitrile compoundNitrile compoundSame
ColorBlueBlueSame
TextureFinger TexturedFinger TexturedSame
Size(ASTM D6319-19)Small, Medium, Large,Extra LargeSmall, Medium, Large,Extra LargeSame
SterilizationNon-sterileNon-sterileSame
UsageSingle usageSingle usageSame
Dimensions(ASTM D6319-19)Length Min. 230 min Width Min95+/-10 mm (for medium size)Length Min. 230 min WidthMin 95+/-10 mm (for mediumsize)Same
Physical Properties(ASTM D6319-19)Before AgingTensile StrengthMin 14 Mpa Ultimate ElongationMin 500%After AgingTensile StrengthMin 14 Mpa Ultimate ElongationMin 400%Before AgingTensile StrengthMin 14 Mpa UltimateElongation Min 500%After AgingTensile StrengthMin 14 Mpa UltimateElongation Min 400%Same
Thickness(ASTM D6319-19)Palm min 0.05 mmFinger min 0.05 mmPalm min 0.05 mmFinger min 0.05 mmSame
Powder Free(ASTM D6319-19)<2 mg/glove<2 mg/gloveSame
Freedom from Holes(Water Tight -1000 ml) -ASTM D6319-19 (CrossReference D5151)PassedPassedSame
Biocompatibility -SKIN SENSITIZATION -ISO 10993-10: 2010 (E)Under the conditions of study notan irritantUnder the conditions of studynot an irritantSame
Biocompatibility -SKIN IRRITATION - ISO10993-10: 2010 (E)Under the conditions ofthe study not a sensitizerUnder the conditions ofthe study not asensitizerSame
Biocompatibility - INVITRO CYTOTOXICITY- ISO10993-5: 2009(E)Under the condition of the study,the test article showed nocytotoxicity to L-929 cells andcytotoxic level was 0.Exhibit cytotoxicreactivity at 100%extract concentration(Grade 4 with neat extract).Non-cytotoxicreactivity at 50%, 25%,12.5% and 6.25%extract concentration.Similar
Biocompatibility -ACUTE SYSTEMICTOXICITY - ISO 10993-11: 2017(E)Under the conditions ofthe study no systemic toxicityNo systemic toxicityunder the experimentalconditions employedSimilar
Manufacturer(s)Jiangsu Cherish MedicalTechnology Co., LtdNathan Trading Co., Ltd

{5}------------------------------------------------

There are no significant differences between the two products and are identical in terms of intended use, materials, design, manufacturing methods. Both devices meet the ASTM standard D6319-19.

{6}------------------------------------------------

7. NON-CLINICAL TESTING SUMMARY

PERFORMANCE DATA

Test MethodPurposeAcceptance CriteriaResult
ASMT D6319-19Standard Specification forNitrile ExaminationGloves for MedicalApplication - PhysicalDimensions TestTo determine thewidth, length, andthickness of theglovesWidth: (for Small size)$80\pm10$ mmWidth: (for small size)82mm (Minimum)82 mm (Mean)85 mm (Maximum)
Length: (for Small size)220 mm (Minimum)Length: (for small size)232 mm (Mean)230 mm (Minimum)
Thickness: (for Small size)Finger –0.05 mm (Minimum)Palm –0.05 mm (Minimum)Thickness: (for small size)Finger –0.13 mm (Mean)0.12 mm (Minimum)Palm –0.08 mm (Mean)0.08 mm (Minimum)
Passed
Width: (for Medium size)$95\pm10$ mmWidth: (for medium size)91 mm (Minimum)92 mm (Mean)93 mm (Maximum)
Length: (for Medium size)230 mm (Minimum)Length: (for medium size)242 mm (Mean)240 mm (Minimum)
Thickness: (for Medium size)Finger –0.05 mm (Minimum)Palm–0.05 mm (Minimum)Thickness: (for medium size)Finger –0.14 mm (Mean)0.13 mm (Minimum)Palm –0.09 mm (Mean)0.08 mm (Minimum)
Passed
Width: (for Large size)$110\pm10$ mmWidth: (for large size)108 mm (Minimum)109 mm (Mean)110 mm (Maximum)
Length: (for large size)230 mm (Minimum)Length: (for large size)239 mm (Mean)237 mm (Minimum)
Thickness: (for Large size)Finger–0.05 mm (Minimum)Palm0.05 mm (Minimum)Thickness: (for large size)Finger –0.14 mm (Mean)0.13 mm (Minimum)Palm –0.09 mm (Mean)0.09 mm (Minimum)
Passed
ASMT D6319-19Standard Specification forNitrile ExaminationGloves for MedicalApplication - PhysicalDimensions TestTo determine thewidth, length, andthickness of theglovesWidth: (for Extra-large size)120±10 mmLength: (for Extra-large size)230 mm (Minimum)Thickness: (for Extra-large size)Finger -0.05 mm (Minimum)Palm -0.05 mm (Minimum)Width: (for extra-large size)121 mm (Minimum)122 mm (Mean)125 mm (Maximum)Length: (for extra-large size)243 mm (Mean)240 mm (Minimum)Thickness: (for extra- large size)Finger -0.14 mm (Mean)0.13 mm (Minimum)Palm -0.09 mm (Mean)0.08 mm (Minimum)Passed
ASMT D6319-19Standard Specificationfor Nitrile ExaminationGloves for MedicalApplication - PhysicalRequirements TestTo determine thetensile strength andultimate elongationbefore and afteracceleration agingBefore Acceleration Aging:Tensile Strength (MPa):14 (Minimum)Ultimate Elongation (%):500 (Minimum)After Acceleration Aging:Tensile Strength (MPa):14 (Minimum)Ultimate Elongation (%):400 (Minimum)Before Acceleration Aging:Tensile Strength (MPa):31 (Mean)17 (Minimum)Ultimate Elongation (%):563 (Mean)520 (Minimum)After Acceleration Aging:Tensile Strength (MPa):41 (Mean)28 (Minimum)Ultimate Elongation (%):555 (Mean)524 (Minimum)Passed
ASTM D6319-19(ASTM D5151-11)Standard Test Methodfor Detection of Holes inMedical GlovesTo determine theholes in the gloves(Mean)AQL 2.5Passed
ASMT D6319-19(ASTM D6124-11)Standard Test Methodfor Residual Powder onMedical GlovesTo determine theresidual powder inthe gloves< 2.0 mg/glove1.02 mg/glove, Passed

{7}------------------------------------------------

BIO-COMPATIBILITY DATA

Test MethodPurposeAcceptanceCriteriaResult
ISO 10993-10 Biological evaluationof medical devices – Part 10: Testsfor skin irritation and skinsensitizationTo determine the potential of thematerial under test to produce skinirritation in rabbitsUnder thecondition ofstudy not anirritantUnder the condition ofstudy not an irritant

{8}------------------------------------------------

ISO 10993-10 Biological evaluationof medical devices - Part 10: Testsfor skin irritation and skinsensitizationTo determine the skinsensitization potential of thematerial both in terms ofinduction and elicitation inguinea pigs.Under theconditions of thestudy not asensitizer.Under the conditions ofthe study not a sensitizer.
ISO 10993-5 Biological evaluationof medical devices - Part 5: Testsfor in vitro cytotoxicityTo evaluate the in vitro cytotoxicpotential of the test item (bothinner and outer surface) Extractsin L-929 mouse fibroblasts cellsusing elution methodUnder theconditions ofstudy noncytotoxicUnder the condition of thestudy, the test articleshowed no cytotoxicity toL-929 cells and cytotoxiclevel was 0.
ISO 10993-11:2017 Biologicalevaluation of medical devices–Part 11: Tests for acute systemictoxicityThe test item was evaluated foracute systemic toxicity in SwissAlbino MiceUnder theconditions of thestudy no systemictoxicityUnder the conditions ofthe study no systemictoxicity

8. CLINICAL TESTING SUMMARY

Not applicable - Clinical data is not needed for gloves by the 510(k) process.

9. CONCLUSION

The conclusions drawn from the non-clinical test demonstrate that the subject device in 510(k) submission, the Disposable Medical Nitrile Examination Gloves Non sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate device K203191- LYDUS Nitrile Examination Gloves, Powder Free.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.