Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182099
    Device Name
    Delta TT Pro
    Date Cleared
    2018-12-26

    (145 days)

    Product Code
    Regulation Number
    888.3358
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Delta TT Pro

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

    • · Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
    • Rheumatoid arthritis;
    • · Post-traumatic arthritis,
    • · Correction of functional deformity;
    • · Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
    • · Revisions in cases of good remaining bone stock.

    The Delta Acetabular System is intended for cementless use.

    Device Description

    The Delta TT Pro is an acetabular system, intended for total hip replacement in cementless applications. The system consists of a Delta TT Pro acetabular cup (Ti6Al4V) and LimaVit liners (cross-linked UHMWPE with Vitamin E), available in three different design: neutral, protruded and high-wall. The LimaVit liners articulate with the Biolox®Delta and CoCrMo femoral heads, which are coupled to LimaCorporate femoral stems. Bone screws may be used to give further stability to the cup.

    AI/ML Overview

    This FDA 510(k) summary is for the Delta TT Pro, a total hip prosthesis. As such, it describes a physical medical device and not a software algorithm or AI model. Therefore, many of the requested categories related to acceptance criteria for AI/software, expert review, and ground truth are not applicable.

    Here's the information that can be extracted or deduced from the provided text, primarily focusing on the physical device's performance and regulatory pathway rather than AI-specific criteria.

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific numerical acceptance criteria or performance metrics for the device. Instead, it states that:

    Acceptance Criteria CategoryReported Device Performance
    Overall Performance"All verification and validation testing confirms that the product specifications are met, in support of the substantial equivalence of the intended use and technological characteristic as the predicate devices."
    Stress Testing"The Delta TT Pro has been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices."
    Mechanical TestsVarious tests performed to confirm substantial equivalence, including "Push out, lever out and torque out disassembly test," "Stiffness test," "Impingement test," and "Wear test."

    2. Sample size used for the test set and the data provenance

    • Sample size: Not explicitly stated. The document refers to "extensive verification and validation testing" but does not quantify the number of devices or components tested for each type of test.
    • Data provenance: Not explicitly stated, but as a physical device, the data would originate from laboratory testing of the manufactured components. The manufacturer, Limacorporate S.p.A., is located in Italy, so the testing would likely have occurred there or at a certified lab.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This concept is not applicable as this is a physical medical device, not an AI/software device that requires ground truth established by experts for performance evaluation. The "ground truth" for mechanical testing would be the physical properties and performance characteristics measured against established engineering standards.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human readers and AI for agreement on findings, which is not relevant for the mechanical testing of a hip prosthesis.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical device, and MRMC studies are not relevant. This type of study focuses on the diagnostic performance of human readers, typically aided by AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The Delta TT Pro is a physical hip prosthesis, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    As it is a physical device, the "ground truth" for its performance is based on established engineering and biomechanical principles and standards. The tests mentioned (push out, lever out, torque out, stiffness, impingement, wear) would have their results compared against these industry-accepted benchmarks and the performance of the predicate devices. The document states that the testing "confirms that the product specifications are met" and that the device operates "in a manner substantially equivalent to the predicate devices."

    8. The sample size for the training set

    Not applicable. There is no "training set" for a physical device in the context of machine learning or AI.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a physical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1