K Number
K182099

Validate with FDA (Live)

Device Name
Delta TT Pro
Date Cleared
2018-12-26

(145 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • · Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
  • Rheumatoid arthritis;
  • · Post-traumatic arthritis,
  • · Correction of functional deformity;
  • · Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
  • · Revisions in cases of good remaining bone stock.

The Delta Acetabular System is intended for cementless use.

Device Description

The Delta TT Pro is an acetabular system, intended for total hip replacement in cementless applications. The system consists of a Delta TT Pro acetabular cup (Ti6Al4V) and LimaVit liners (cross-linked UHMWPE with Vitamin E), available in three different design: neutral, protruded and high-wall. The LimaVit liners articulate with the Biolox®Delta and CoCrMo femoral heads, which are coupled to LimaCorporate femoral stems. Bone screws may be used to give further stability to the cup.

AI/ML Overview

This FDA 510(k) summary is for the Delta TT Pro, a total hip prosthesis. As such, it describes a physical medical device and not a software algorithm or AI model. Therefore, many of the requested categories related to acceptance criteria for AI/software, expert review, and ground truth are not applicable.

Here's the information that can be extracted or deduced from the provided text, primarily focusing on the physical device's performance and regulatory pathway rather than AI-specific criteria.

1. A table of acceptance criteria and the reported device performance

The document does not provide specific numerical acceptance criteria or performance metrics for the device. Instead, it states that:

Acceptance Criteria CategoryReported Device Performance
Overall Performance"All verification and validation testing confirms that the product specifications are met, in support of the substantial equivalence of the intended use and technological characteristic as the predicate devices."
Stress Testing"The Delta TT Pro has been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices."
Mechanical TestsVarious tests performed to confirm substantial equivalence, including "Push out, lever out and torque out disassembly test," "Stiffness test," "Impingement test," and "Wear test."

2. Sample size used for the test set and the data provenance

  • Sample size: Not explicitly stated. The document refers to "extensive verification and validation testing" but does not quantify the number of devices or components tested for each type of test.
  • Data provenance: Not explicitly stated, but as a physical device, the data would originate from laboratory testing of the manufactured components. The manufacturer, Limacorporate S.p.A., is located in Italy, so the testing would likely have occurred there or at a certified lab.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This concept is not applicable as this is a physical medical device, not an AI/software device that requires ground truth established by experts for performance evaluation. The "ground truth" for mechanical testing would be the physical properties and performance characteristics measured against established engineering standards.

4. Adjudication method for the test set

Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human readers and AI for agreement on findings, which is not relevant for the mechanical testing of a hip prosthesis.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical device, and MRMC studies are not relevant. This type of study focuses on the diagnostic performance of human readers, typically aided by AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The Delta TT Pro is a physical hip prosthesis, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

As it is a physical device, the "ground truth" for its performance is based on established engineering and biomechanical principles and standards. The tests mentioned (push out, lever out, torque out, stiffness, impingement, wear) would have their results compared against these industry-accepted benchmarks and the performance of the predicate devices. The document states that the testing "confirms that the product specifications are met" and that the device operates "in a manner substantially equivalent to the predicate devices."

8. The sample size for the training set

Not applicable. There is no "training set" for a physical device in the context of machine learning or AI.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for a physical device.

{0}------------------------------------------------

December 26, 2018

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Limacorporate S.p.A. Roberto Gabetta Regulatory Manager via Nazionale 52 Villanova di San Daniele, 33038 It

Re: K182099

Trade/Device Name: Delta TT Pro Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, MBL Dated: November 23, 2018 Received: November 30, 2018

Dear Roberto Gabetta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone

(1-800-638-2041 or 301-796-7100).

Sincerely, Daniel S. Ramsey -S 2018.12.26 15:34:29 -05'00'

FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K182099

Device Name Delta TT Pro

Indications for Use (Describe)

The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • · Non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
  • Rheumatoid arthritis;
  • · Post-traumatic arthritis,
  • · Correction of functional deformity;
  • · Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
  • · Revisions in cases of good remaining bone stock.

The Delta Acetabular System is intended for cementless use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------------------------------

|X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Lima Corporate. The logo consists of a red abstract symbol on the left, followed by the words "Lima" in bold black letters and "Corporate" in light gray letters. Below "Lima Corporate" is the text "Orthopaedic" in light gray, followed by a red diamond shape and the word "motion" in black.

Section 7 - 510(k) Summary

(in accordance with 21 CFR 807.92)

510(k) Number K

I. Applicant Information

Applicant:Limacorporate S.p.A.
Via Nazionale, 52
Villanova di San Daniele
Udine 33038
Italy
Contact Person:Roberto Gabetta
Regulatory Affairs Manager
Tel: +39 0432945511
Fax: +39 0432945512
e-mail: roberto.gabetta@limacorporate.com
Application Correspondent:Same as Applicant.

Date Prepared:

August 3, 2018

II. Device Identification

Proprietary Name:Delta TT Pro
Common/Usual Name:Total Hip Prosthesis
Classification Name:Hip joint metal/polymer/metal semi-constrained
porous-coated uncemented prosthesis
Regulation Number:21 CFR 888.3358
Product Codes:LPH, MBL
Classification:Class II
Classification Panel:Orthopedic Devices

III. Predicate Devices

The Delta TT Pro device is substantially equivalent to the following FDA cleared predicate devices with regard to indications for use, performance and technological characteristics:

510(k) Number: K112898, K141395

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for LimaCorporate. The logo consists of a red symbol on the left, followed by the word "Lima" in black and the word "Corporate" in gray. Below the word "Corporate" is the phrase "Orthopaedic motion" in gray, with the word "motion" in red.

Trade Name:Delta TT Acetabular System
Manufacturer:Limacorporate S.p.A.
Classification Name:Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Common/Usual Name:Total Hip Prosthesis
Regulation Number:21 CFR 888.3358
Product Code:LPH, MBL, JDI, LZO
Classification:Class II
510(k) Number:K140669
Trade Name:G7 Acetabular System
Manufacturer:Biomet Manufacturing Corp.
Classification Name:Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis
Common/Usual Name:Total Hip Prosthesis
Regulation Number:21 CFR 888.3358
Product Code:LPH, LZO, OQG, KWZ, JDI, OQH, OQI, PBI
Classification:Class II

IV. Device Description

The Delta TT Pro is an acetabular system, intended for total hip replacement in cementless applications. The system consists of a Delta TT Pro acetabular cup (Ti6Al4V) and LimaVit liners (cross-linked UHMWPE with Vitamin E), available in three different design: neutral, protruded and high-wall. The LimaVit liners articulate with the Biolox®Delta and CoCrMo femoral heads, which are coupled to LimaCorporate femoral stems. Bone screws may be used to give further stability to the cup.

Indications for Use Statement V.

The Delta Acetabular System is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • Non-inflammatory degenerative joint disease such as osteoarthritis, ● avascular necrosis and hip dysplasia;
  • Rheumatoid arthritis;
  • Post-traumatic arthritis, ●
  • Correction of functional deformity;
  • Fractures, dislocation of the hip and unsuccessful cup arthroplasty;
  • Revisions in cases of good remaining bone stock. ●

The Delta Acetabular System is intended for cementless use.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Lima Corporate Orthopaedic. The logo features a red abstract symbol resembling a person in motion, followed by the words "Lima" in bold black and "Corporate" in gray. Below "Lima Corporate" is the word "Orthopaedic" in gray, followed by a red diamond shape and the word "motion" in black.

VI. Summary of the Technical Characteristics

The intended use, design, and materials of the Delta TT Pro are substantially equivalent to the intended use, design, and materials of the predicate devices. All required design control activities have been completed and the results indicated that the subject device is substantially equivalent to the predicate devices, in terms of safety and effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Lima Corporate Orthopaedic Motion. The words "Lima Corporate" are stacked vertically in a bold, sans-serif font. To the right of "Corporate" are the words "Orthopaedic" and "motion" stacked on top of each other in a smaller font. A red symbol is located below the word "Lima".

Predicate Device Comparison Table

Device NameSubject Device:Delta TT ProPredicate Device:(K112898,K141395)Delta TT Acetabular SystemPredicate Device:(K140669)G7 Acetabular System(OsseoTi acetabular shell withEl acetabular liner)Significant Differences
Product CodeLPH, MBLLPH, MBL, JDI, LZOLPH, LZO, OQG, KWZ, JDI,OQH, OQI, PBINo significant differences.The subject device isclassified as Product CodeLPH and MBL underregulation 888.3358, whereasthe predicate devices areclassified under regulation888.3358 with Product CodesLPH, MBL, JDI, LZO (DeltaTT) and LPH, LZO, OQG,KWZ, JDI, OQH, OQI, PBI(G7 Acetabular System).However, the three productspresent substantiallyequivalent features as well assubstantially equivalentsafety and effectivenesscharacteristics.
Regulation #888.3358888.3358888.3358
ClassIIIIIINo difference.
Device NameSignificant Differences
Subject Device:Delta TT ProPredicate Device:(K112898,K141395)Delta TT Acetabular SystemPredicate Device:(K140669)G7 Acetabular System(OsseoTi acetabular shell withE1 acetabular liner)
Indications forUse• Non-inflammatorydegenerative joint diseasesuch as osteoarthritis,avascular necrosis and hipdysplasia;• Rheumatoid arthritis;• Post-traumatic arthritis;• Correction of functionaldeformity;• Fractures, dislocation of thehip and unsuccessful cuparthroplasty;• Revisions in cases of goodremaining bone stock.Intended for cementless use.• Non-inflammatorydegenerative joint diseasesuch as osteoarthritis,avascular necrosis and hipdysplasia;• Rheumatoid arthritis;• Post-traumatic arthritis;• Correction of functionaldeformity;• Fractures, dislocation of thehip and unsuccessful cuparthroplasty;• Revisions in cases of goodremaining bone stock.Intended for cementless use.• Non-inflammatorydegenerative joint diseaseincluding osteoarthritis andavascular necrosis.• Rheumatoid arthritis.• Correction of functionaldeformity.• Treatment of non-union,femoral neck fracture, andtrochanteric fractures of theproximal femur with headinvolvement, unmanageableusing other techniques.• Revision of previously failedtotal hip arthroplasty.Porous acetabular shells andfemoral stems are indicated foruncemented biological fixation.Non coated or polyethyleneNo significant differences.The subject and predicatedevices have essentiallyidentical Indications for Use,with the minor exception thatthe G7 Acetabular System isalso indicated in cases of non-union and trochantericfractures (see parts inyellow).
Subject Device:Predicate Device:(K112898,K141395)Predicate Device:(K140669)Significant Differences
Device NameDelta TT ProDelta TT Acetabular SystemG7 Acetabular System(OsseoTi acetabular shell withE1 acetabular liner)
mating components that areindicated for either cemented oruncemented use.
Environment of UseClinical/Hospital Environment.Clinical/Hospital Environment.Clinical/Hospital Environment.No difference.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Lima Corporate. The text "Lima Corporate" is stacked vertically in black font. To the right of "Corporate" is the text "Orthopaedic motion" in a smaller, gray font, with a red arrow pointing to the left.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Lima Corporate Orthopaedic Motion. The logo is vertically oriented, with the Lima Corporate text stacked on top of each other. Below the text is a red symbol, and to the right of the text is the word Orthopaedic and the word motion with a red symbol.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Lima Corporate. The logo consists of a red abstract symbol resembling a person with outstretched arms, followed by the words "Lima" in black and "Corporate" in gray. Below "Lima Corporate" is the word "Orthopaedic" in gray, followed by a red symbol and the word "motion" in gray.

VII. Summary of Performance Data

Limacorporate has conducted extensive verification and validation testing of the Delta TT Pro as a Hip Prosthesis intended for use in total hip arthroplasty. All verification and validation testing confirms that the product specifications are met, in support of the substantial equivalence of the intended use and technological characteristic as the predicate devices.

The Delta TT Pro has been stress tested to ensure that the system as a whole provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate devices.

The main groups of tests performed include:

  • Push out, lever out and torque out disassembly test
  • . Stiffness test
  • Impingement test ●
  • o Wear test

Substantial Equivalence Conclusions VIII.

Based on the comparison of intended use, principles of operation and technological characteristics, the Delta TT Pro device is deemed to be substantially equivalent to the Delta TT Acetabular System, manufactured by Limacorporate S.p.A. (K112898, K141395) and the G7 Acetabular System, manufactured by Biomet Manufacturing Corp. (K140669).

The minor differences identified between the Delta TT Pro and its predicate devices raise no new issues of safety or effectiveness. The performance and validation data demonstrate that the Delta TT Pro is as safe and effective as the predicate devices.

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.