510(k) Data Aggregation

    K Number
    K130534
    Device Name
    DIPLOMAT
    Date Cleared
    2013-10-30

    (243 days)

    Product Code
    Regulation Number
    872.6640
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diplomat dental operative units are self-contained dental treatment units that contain vacuum system and water supply and separate compressor placed outside the dental unit. These dental operative units are designed to provide air, water, vacuum and electricity to operate various dental hand-pieces, accessories and attachments, and to serve as a base for other dental devices and accessories. Diplomat dental operative units are offered with or without attached dental chairs. When supplied with a dental chair, the chair is intended to properly position a patient to perform different dental procedures. Diplomat dental operative units are to be operated and used by dentists and other legally qualified professionals, and are intended for use to provide general dental restorative care and hygiene procedures in both traditional dental office settings applications by or under the supervision of a licensed dentist or a hygienist if permitted by applicable law.

    Device Description

    Diplomat dental operative units are modern, ergonomically designed, dental systems offered in a wide range of models depending on the specific end user requirements. Diplomat dental operative units are AC-powered devices that are intended to supply power to, and serve as a base for, other dental devices and accessories - such as hand-pieces, operating lights, curing lights, ultrasonic instruments, hygiene instruments, etc. - and deliver electric, air, water and vacuum power to them. Diplomat dental operative units are offered either with the patient dental chair or as a free-standing unit without the chair for those end-users with existing patient chairs that they want to use.

    AI/ML Overview

    This document is a 510(k) summary for Diplomat Dental Operative Units. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria for a new, innovative device feature. Therefore, many of the requested details about acceptance criteria, study design, and performance metrics are not explicitly present in the provided text.

    Based on the provided text, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    The document does not present explicit acceptance criteria in the typical sense of quantitative targets for performance metrics (e.g., sensitivity, specificity, accuracy). Instead, the "acceptance criteria" are implied by the demonstration of Substantial Equivalence to a legally marketed predicate device. This means the Diplomat Dental Operative Units are considered acceptable if they are as safe and effective as the predicate device (Sirona C8+Dental Operative Unit).

    The table provided in the document compares various features of the Diplomat Dental Operative Units to the C8+Dental Operative Unit:

    FeaturesDiplomat Dental Operative UnitsC8+Dental Operative UnitSubstantially Equivalent
    510(k)PendingK032543
    ModelsAdept Models: DA 380, DA 370, DA
    170, DA 130, DA 110;
    Consul Models: DC 350, DC 310, DC
    180, DC 170 and DC orthodontic; and
    Lux Models: DL 320, and DL 210Multiple versionsSimilar
    Product CodeEIAEIASame
    Regulation Number872.6640872.6640Same
    Regulation NameUnit, Operative, DentalUnit, Operative, DentalSame
    Place of UseDental clinicDental clinicSame
    Intended UsersDentists; assistants and hygienistsDentists; assistants and hygienistsSame
    Power Supply110V/(230 VAC)
    50/60 Hz110V/115V/127V/(230 VAC)
    50/60 HzSimilar
    Utility SupplyCompressed air and waterCompressed air and waterSame
    Protection ClassClass 1 equipmentClass 1 equipmentSame
    Degree of ProtectionType B of applied partsType B of applied partsSame
    InstallationProfessionally installed; available
    with chair and cart mounted
    options.Professionally installed; not available
    with chair and cart mounted options.Same
    Air Pressure$450 kPa/800 kPa$ (min./max.)$550 kPa/750 kPa$ (min./max.)Similar
    Water Pressure$300 kPa/600 kPa$ (min./max.)$250 kPa/600 kPa$ (min./max.)Similar
    Electrical SafetyIEC 60601-1IEC 60601-1Same
    EMCEN 60601-1-2
    ISO 7494-1IEC 60601-1-2Similar
    PerformanceDiplomat dental operative units
    are self-contained dental
    treatment units that contain
    vacuum system and water supply
    and separate compressor placed
    outside the dental unit. These
    dental operative units are designed
    to provide air, water, vacuum and
    electricity to operate various
    dental hand-pieces, accessories
    and attachments, and to serve as a
    base for other dental devices and
    accessories. Diplomat dental
    operative units are offered by with
    or without attached dental chairs.
    When supplied with a dental chair,
    the chair is intended to properly
    position a patient to perform
    different dental procedures.The C8 + Dental Operative Unit with
    accessories are intended to supply
    power to and serve as a base for dental
    devices and accessories. This product
    includes a dental chair. The unit is
    intended for use in the dental clinic
    environment and used by trained
    dentists and/or dental technicians and
    assistants. The C8 + Dental Operative
    Unit is offered wilt the optional Sivision
    3, an intraoral camera system intended
    to provide the dentist and patient with
    intraoral video images to view the
    condition of the teeth and oral cavity.Similar
    Statement of
    Intended UseDiplomat dental operative units are to be operated and used
    by dentists and other legally
    qualified professionals, and are
    intended for use to provide general
    dental restorative care and hygiene
    procedures in both traditional
    dental office settings applications
    by or under the supervision of a
    licensed dentist or a hygienist if permitted by applicable law.
    Similar

    The "reported device performance" is the conclusion that the Diplomat Dental Operative Units are "Substantially Equivalent" to the predicate device, meaning they perform similarly and do not raise new questions of safety or effectiveness. This is based on the comparison of features, intended use, and technological characteristics.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "Clinical Data: Not applicable to this 510(k) submission." This indicates that no human patient or clinical test set was used to evaluate the device's performance in a clinical setting for this submission. The evaluation was based on Non-Clinical Data - Device Bench Testing in accordance with international standards. Therefore, there is no sample size for a "test set" in the context of clinical data.

    The device testing was likely performed by the manufacturer, DIPLOMAT DENTAL, s.r.o. in Slovak Republic, Slovakia. This would be the provenance of the non-clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Since no clinical data or human test set was used, there were no experts establishing ground truth for such a set. The "ground truth" for non-clinical bench testing would be the compliance with engineering standards by the testing entity, and possibly by third-party laboratories.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for a dental operative unit (a piece of equipment), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable, as this is a dental operative unit, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical data, the ground truth would be the physical and electrical properties and performance characteristics of the device measured against the specifications outlined in the international standards (IEC 60601-1, EN 60601-1-2, ISO 7494-1). This is a form of engineering/technical ground truth.

    8. The sample size for the training set

    Not applicable, as this is not a machine learning or AI device. The "training" in this context would be the design and manufacturing processes based on engineering principles and regulatory standards.

    9. How the ground truth for the training set was established

    Not applicable, as this is related to software/AI development. For this device, the "ground truth" for its design and manufacturing (analogous to a training set for an algorithm) would be the established engineering specifications, safety standards, and performance requirements for dental operative units as defined by the relevant international standards (IEC, EN, ISO) and regulatory bodies (FDA). The manufacturer's internal quality control procedures and adherence to these standards would ensure the "ground truth" of the manufactured units.