K Number
K151704

Validate with FDA (Live)

Date Cleared
2016-04-13

(294 days)

Product Code
Regulation Number
888.3070
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DIPLOMAT® Spinal System, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

  • Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
  • Spondylolisthesis
  • Spinal stenosis
  • Fracture
  • Dislocation
  • Scoliosis
  • Kyphosis
  • Spinal tumor
  • Pseudoarthrosis
  • Failed previous fusion

In addition, the DIPLOMAT® Spinal System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

The DIPLOMAT® Spinal System is intended to be used with autograft and/or allograft.

Device Description

The DIPLOMAT® Spinal System is a multiple component posterior spinal fixation system which consists of pedicle screws, rods, and connectors. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are manufactured from Ti-6AL-4V ELI per ASTM F136 and cobalt chrome per ASTM F75. The safety and effectiveness of this device has not been established when used in conjunction with bone cement or for use with poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye

AI/ML Overview

This document is a 510(k) Premarket Notification for the DIPLOMAT® Spinal System, a medical device for spinal fixation. It describes the device's indications for use, technological characteristics, and performance data to demonstrate substantial equivalence to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Test Mode per ASTM F1717-14)Reported Device Performance
Static Compression-Bending TestStrength sufficient for intended use and substantially equivalent to legally marketed predicate devices.
Static Torsion TestStrength sufficient for intended use and substantially equivalent to legally marketed predicate devices.
Dynamic Compression-Bending TestStrength sufficient for intended use and substantially equivalent to legally marketed predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the context of human patient data or a clinical study. The performance evaluation is based on non-clinical mechanical testing. Therefore, information regarding human sample size and data provenance (country of origin, retrospective/prospective) is not applicable or provided.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable. The study is a non-clinical mechanical testing study, not a study involving expert assessment of patient data.

4. Adjudication Method

This information is not applicable for a non-clinical mechanical testing study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This document describes mechanical testing of a spinal implant, not a study evaluating human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

No, a standalone (algorithm only) performance study was not done. This document focuses on the mechanical performance of a physical medical device, not an algorithm or AI.

7. Type of Ground Truth Used

The "ground truth" for this study is established by engineering standards and specifications as defined in the ASTM F1717-14 test methodology. The performance of the DIPLOMAT® Spinal System is compared against the expected mechanical properties for devices of its type and against predicate devices.

8. Sample Size for the Training Set

This information is not applicable. There is no concept of a "training set" in a non-clinical mechanical testing study.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set mentioned in the context of this non-clinical mechanical testing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 14, 2016

SIGNUS Medizintechnik GmbH % Mr. Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K151704

Trade/Device Name: DIPLOMAT® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: April 13, 2016 Received: April 13, 2016

Dear Mr. Maxwell:

This letter corrects our substantially equivalent letter of April 13, 2016.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Caroline Rhim -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K151704

Device Name

DIPLOMAT® Spinal System

Indications for Use (Describe)

The DIPLOMAT® Spinal System, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

  • Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)

  • Spondylolisthesis

  • Spinal stenosis

  • Fracture

  • Dislocation

  • Scoliosis

  • Kyphosis

  • Spinal tumor

  • Pseudoarthrosis

  • Failed previous fusion

In addition, the DIPLOMAT® Spinal System is intended for treatment of severe

spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

The DIPLOMAT® Spinal System is intended to be used with autograft and/or allograft.

Type of Use (Select one or both, as applicable)

Z Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995. DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

5.510(K) SUMMARY

Submitter's Name:SIGNUS Medizintechnik GmbH
Submitter's Address:Industriestr. 263755 Alzenau / Germany
Submitter's Telephone:+49 6023 9166-213
Contact Person:Kenneth C. MaxwellEmpirical Testing Corp.719.291.6874
Date Summary was Prepared:29 January 2016
Trade or Proprietary Name:DIPLOMAT® Spinal System
Common or Usual Name:Orthosis, Spinal Pedicle FixationOrthosis, Spondylolisthesis Spinal FixationOrthosis, Spinal Pedicle Fixation, For Degenerative DiscDiseaseAppliance, Fixation, Spinal Interlaminal
Classification:Class III per 21 CFR §888.3070 Pedicle screw spinalsystem
Product Code:NKB, MNI, MNH
Classification Panel:87 Orthopedics Panel

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The DIPLOMAT® Spinal System is a multiple component posterior spinal fixation system which consists of pedicle screws, rods, and connectors. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are manufactured from Ti-6AL-4V ELI per ASTM F136 and cobalt chrome per ASTM F75. The safety and effectiveness of this device has not been established when used in conjunction with bone cement or for use with poor bone quality (e.g., osteoporosis, osteopenia). This device is intended only to be used with saline or radiopaque dye

INDICATIONS FOR USE

The DIPLOMAT® Spinal System, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

  • Degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)

  • Spondylolisthesis

  • Spinal stenosis

  • Fracture

  • Dislocation

  • Scoliosis

  • Kyphosis

{5}------------------------------------------------

  • Spinal tumor
  • Pseudoarthrosis
  • Failed previous fusion

In addition, the DIPLOMAT® Spinal System is intended for treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft, having implants attached to the lumbosacral spine and/or ilium with removal of the implants after attainment of a solid fusion. Levels of pedicle screw fixation for these patients are L3-sacrum/ilium.

The DIPLOMAT® Spinal System is intended to be used with autograft and/or allograft.

The indications for use for the Diplomat™ system is similar to that of the predicate device noted below in Table 5-1.

TECHNOLOGICAL CHARACTERISTICS

The SIGNUS Medizintechnik GmbH DIPLOMAT® is made from Ti-6AL-4V ELI per ASTM F136. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:

  • Indications for Use
  • Materials of manufacture ●
  • Structural support mechanism .
  • Sterilization ●
  • . Principles of Operation
510k NumberTrade or Proprietary orModel NameManufacturerType
K000236Synergy VLS OpenInterporePrimary
K131820, K103490, K033901,K955348Moss Miami TitaniumDePuy Spine Inc.Additional
K081080TSRHMedtronicAdditional
K020279, K051971, K024096OPTIMA™U&I CorporationAdditional
K102870Spine Proliant ScrewSystemExactechAdditional
K120838Matrix SystemSynthesAdditional
K110280RevlokGlobusAdditional
K123717UCentumUlrichAdditional
Table 5-1 Predicate Devices

{6}------------------------------------------------

PERFORMANCE DATA

The DIPLOMAT® has been tested in the following test modes:

  • Static Compression-Bending Test per ASTM 1717-14 .
  • . Static Torsion Test per ASTM 1717-14
  • Dynamic Compression-Bending Test per ASTM F1717-14 .

The results of this non-clinical testing show that the strength of the Diplomat is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the DIPLOMAT® Spinal System is substantially equivalent to the predicate device.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.