Search Results
Found 15 results
510(k) Data Aggregation
(28 days)
The Compass Steerable Needle is a steerable transbronchial biopsy needle intended to be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
The Compass Steerable Needle (CSN) is sterile, single use, 22-gauge transbronchial needle with a unidirectional, steerable distal tip for the acquisition of tissue from the intrapulmonary regions. The Compass Steerable Needle include two model numbers, Model CSN1001 and Model CSN1002. Model CSN1001 includes two accessory adapters for attachment to bronchoscopes.
The subject of this 510(k) is the addition of two new adapters to Model CSN1002. Adapter SRA-1-01 connects the Model CSN1002 to the Ion™ Endoluminal System, and Adapter SRA-2-01 connects the Model CSN1002 to the Monarch™ Platform or Galaxy System™.
N/A
Ask a specific question about this device
(121 days)
The Compass Steerable Needle is a steerable biopsy needle intended be used through a compatible working channel bronchoscope or Medtronic Extended Working Channel (EWC) for the collection of tissue from the intrapulmonary regions.
The Compass Steerable Needles (CSN) are sterile, single use, 22-gauge needles with a unidirectional, steerable distal tip for the acquisition of tissue from the intrapulmonary regions. The Steerable Needle consists of a handle, shaft, and needle. The handle provides the user with control of device rotation, extension, retraction, distal tip articulation of 70°±10° unidirectionally within a plane and a sampling mechanism to extend and retract the needle out of the shaft to obtain tissue samples. A Luer connector on the proximal end of the device provides the connection for the stylet or a syringe for aspiration during sampling. There are two models of the Compass Steerable Needle. Model CSN1001 can be coupled to Olympus® 190 or Pentax® bronchoscopes with a 2.0 working channel and 600 mm working length. It is packaged with a stylet, and adapters. Model CSN1002 can be coupled to the Medtronic Illumisite™ Extended Working Channel (EWC) with a 2.0 mm working channel. It is packaged with a stylet. The Compass Steerable Needles with stylet are inserted and coupled to either a bronchoscope or a Medtronic Illumisite EWC. The translation arm advances the device into the lung. Depressing the plunger articulates the distal end of the shaft. The sampling mechanism is depressed extending the needle to obtain a sample.
The provided text is a 510(k) summary for the Serpex Medical, Inc. Compass Steerable Needle. It outlines the device's characteristics, intended use, and comparison to predicate devices, along with performance data. However, it does not describe a study that proves the device meets specific acceptance criteria in terms of AI model performance, diagnostic accuracy, or clinical effectiveness as typically seen for AI/ML-enabled devices.
Instead, the "acceptance criteria" and "study" described in the document refer to standard medical device verification and validation activities for a new physical medical device (a steerable biopsy needle), focusing on its engineering performance, safety, and functionality. This includes:
- Bench Testing: Mechanical and functional performance of the device.
- Validation Testing: Simulated use conditions, including human cadaver testing for clinical performance.
- Human Factors Usability: Assessment of user interface and potential use errors.
- Biocompatibility: Evaluation of material safety in contact with human tissue.
- Sterilization: Validation of the sterilization process.
Therefore, I cannot extract the information required for an AI/ML-driven device's acceptance criteria and study proving it meets them. The prompt's request for "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of AI (e.g., sample size for test set, data provenance, number of experts, MRMC studies, standalone performance, ground truth) is not applicable to the content of this 510(k) summary.
In summary, the provided document details the regulatory clearance process for a physical medical device (a steerable biopsy needle), not an AI/ML diagnostic or assistive device. Consequently, it does not contain the information requested regarding AI model performance, expert adjudication, or MRMC studies.
Ask a specific question about this device
(143 days)
Endoscopic biliary stent placement for biliary drainage of obstructed ducts that could be caused by common bile duct stones, malignant biliary obstruction, benign or malignant strictures or other obstructed biliary conditions requiring drainage.
The Compass BDS® Biliary Stent includes double pigtails with double radiopaque marker bands. Compass BDS® Biliary Stents are recommended for use with Cook stent introducers (PC-7, PC-7E, and FS-PC-7). The product code for Compass BDS® Biliary Stent is CBBSO-X-Y (CBBSO-7-5, CBBSO-7-10, CBBSO-7-15), where X denotes French size (Fr) and Y denotes the length in centimeters (cm). This product contains a stent and a pigtail straightener. The stent design allows the stent to be introduced on either side and the double-pigtails minimize migration, while side holes enhance biliary fluid drainage. It also has a tapered tip at both ends to facilitate smooth cannulation. The stent has two radiopaque bands on both ends for fluoroscopic visibility.
This application is for a medical device (Compass BDS Biliary Stent), not an AI/ML powered device. Therefore, the requested information regarding AI/ML powered device acceptance criteria and study details are not applicable here.
However, based on the provided document, here's what can be extracted about the device's performance data and substantial equivalence to a predicate device:
The acceptance criteria for the Compass BDS Biliary Stent are implicitly met through a comparison to a predicate device and a series of non-clinical performance tests. The study's conclusion is that the device is substantially equivalent to the predicate device and meets its design input requirements.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Tests/Characteristics | Reported Performance |
|---|---|---|
| Biocompatibility | Cytotoxicity | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. |
| Irritation | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Sensitization | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Acute Systemic Toxicity | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Systemic Toxicity (13 Week Subcutaneous Implantation) | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Muscle Implantation (4 Week) | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Muscle Implantation (13 Week) | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Mouse Lymphoma Assay | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Material Mediated USP Rabbit Pyrogen Study | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Bacterial Reverse Mutation Study | Performed; results intended to demonstrate compliance with ISO 10993-1:2018 and FDA guidance. | |
| Device Performance | Simulated Use | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. |
| Dimensional Testing | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| Visual Testing | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| Tensile Strength Testing | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| MRI Conditional Testing | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| Radiopacity | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| Flow Rate | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| Shelf-Life Testing | Testing completed to Cook Ireland's design control system. Demonstrates the device meets design input requirements. | |
| Substantial Equivalence | Comparison to Predicate Device (K172044) | The subject device shares the same intended use, technological characteristics (e.g., use in biliary duct, single use, sterile, EO sterilization, professional use only, requires 0.035" wire guide and endoscope, endoscopic placement with fluoroscopic monitoring, supplied as stent only with pigtail straightener, double pigtail, made from Ethylene-Vinyl-Acetate (EVA) co-polymer, multiple side ports with no flaps). Differences (stent marker bands, additional sideports, bidirectional stent ends) are addressed by performance testing. |
2. Sample Size Used for the Test Set and Data Provenance
This document does not specify sample sizes for the performance tests. The data provenance is Cook Ireland Ltd.'s internal design control system. It does not mention country of origin or whether the studies were retrospective or prospective, but given it's a premarket notification for a device, the tests are primarily non-clinical and conducted by the manufacturer.
3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts
This information is not applicable as the studies described are non-clinical (biocompatibility and performance testing) of a physical device, not an AI/ML algorithm requiring expert interpretation for ground truth.
4. Adjudication Method for the Test Set
This is not applicable for the non-clinical testing described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
This is not applicable as the application is for a physical medical device, not an AI/ML system, and no human reader studies are mentioned.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable as the application is for a physical medical device, not an AI/ML algorithm.
7. The Type of Ground Truth Used
For biocompatibility, the ground truth is established by adherence to recognized international standards (ISO 10993-1:2018) and FDA guidance for biological evaluation. For device performance testing, the ground truth is established by the predefined design input requirements for the device.
8. The Sample Size for the Training Set
This is not applicable as there is no mention of a training set for an AI/ML algorithm.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no mention of a training set for an AI/ML algorithm.
Ask a specific question about this device
(186 days)
The Compass Cast and MAP disposable pressure transducer system with integrated digital display and wireless transmission capability is intended for direct measurement and monitoring of physiologic pressure, including during the infusion of fluids and therapeutic and diagnostic agents.
The Compass Cast and MAP disposable pressure transducer system with integrated digital display and wireless transmission capability is intended for direct measurement and monitoring of physiologic pressure, including during the infusion of fluids and therapeutic and diagnostic agents.
The Compass CT and Compass CT Port are disposable, point-of-use pressure measurement and monitoring devices that incorporate a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The devices have a distal male luer fitting to connect to a needle or catheter, and a proximal female luer fitting that can be connected to accessory devices (e.g. syringes, caps, or infusion tubing). The devices measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD. The Compass CT Port has an additional, sealed proximal port through which commercially available guidewires can be inserted during pressure measurement.
The Compass CT Port device is being modified to add wireless Bluetooth capability to the point-of-use Compass Cast device, in order to allow for optional display of acquired pressure data and information to a separate display monitor. The modified product will consist of two components that are packaged separately, the compass Cast (Subject Device) and MAP (Monitor Accessory Plug - Accessory).
The Compass Cast (Subject Device) is a single-use, sterile pressure transducer that is physically identical to the predicate device Compass CT Port (510(k) K133624) with the exception of a slight modification to the main CT Port circuit board to accommodate the secondary Bluetooth radio board. All components and functionality of the main circuit board remain identical.
The MAP (Accessory) is a non-sterile, reusable monitor accessory plug that receives a digital pressure signal from the Compass Cast and converts the digital pressure signal to an analog output that is identical to the analog output of a traditional wired pressure transducer. There are two versions of the MAP device:
- Wireless Pressure Receiver - GE Monitor (CWMG001-5)
- Wireless Pressure Receiver Philips Monitor (CWMP001-5)
The provided text describes information about the submission of the "Compass Cast and MAP System" for FDA clearance. However, it does not contain explicit acceptance criteria and device performance data in a tabular format, nor does it detail a study that proves the device meets specific performance criteria related to its core function of physiological pressure measurement beyond stating adherence to existing standards.
The document focuses heavily on demonstrating substantial equivalence to a predicate device (Compass CT Port) by highlighting the technological characteristics and the modifications made (addition of Bluetooth capability and a reusable Monitor Accessory Plug - MAP). Performance data mentioned primarily concerns safety aspects due to these modifications, such as sterility, EO residuals, and electromagnetic compatibility.
Therefore, many of the requested items cannot be fully answered from the provided text.
Based on the available information:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state a table of new acceptance criteria established for the Compass Cast and MAP system itself in terms of pressure measurement performance. Instead, it states that the device's fundamental pressure accuracy and functional performance are "identical" to the predicate device and "Meet or exceed ANSI/AAMI BP22:1994(R)2006".
The only "acceptance criteria" and "reported performance" directly stated are for safety and compatibility tests related to the new wireless functionality and sterilization.
| Acceptance Criteria (Implied / Standard Adherence) | Reported Device Performance |
|---|---|
| Sterilization: Sterility Assurance Level (SAL) of 1 x 10^-6 | Validated (in accordance with ANSI/AAMI/ISO 11135-1:2014) |
| EO Residuals: < 4 mg/device for EO, < 9 mg/device for EC | Testing performed; compliant with ANSI/AAMI/ISO 10993-7:2008 (implied passes, as "testing was performed" on worst-case model and is "in Compliance") |
| BET Testing: Endotoxin limit: 2.15 EU/Device | Kinetic Chromogenic Test Method validated; "in Compliance" with AAMI ANSI ST72:2011 |
| EMC/Electrical Safety: Compliance with IEC 60601-1-2:2007 | "Confirmed to be compliant" and "in Compliance" |
| Wireless Communication: Compliance with relevant standards (e.g., CFR 47, Part 15, subpart B; RSS-247) | "In Compliance" |
| Pressure Accuracy and Functional Performance: Meet or exceed ANSI/AAMI BP22:1994(R)2006 | "Identical" to predicate, which meets/exceeds this standard. Explicit mention: -199 to -51 mm Hg: ±3% of reading -50 to +50 mm Hg: ±2 mm Hg 51 to 999 mm Hg: ±3% of reading |
| Zero Drift: ±1 mm Hg per 4 hours | "Identical" to predicate |
| Light Sensitivity: < 1 mm Hg | "Identical" to predicate |
2. Sample size used for the test set and the data provenance
For the safety tests (sterilization, EO residuals, BET, EMC/Wireless), specific sample sizes are mentioned:
- EO Residuals: "Three samples were processed through two routine (Cycle 20) sterilization cycles".
- Other tests: Sample sizes are not explicitly stated for EMC/Wireless, Sterilization validation, or BET, beyond stating that testing was performed.
- Pressure performance: No specific test set sample size is described for evaluating the pressure accuracy of the Compass Cast itself, as its performance is stated to be "identical" to the predicate. Therefore, any data provenance for such a study would refer to the predicate device development.
The provenance for the safety tests is internal company testing ("Centurion Protocol# 17205-16"). No country of origin is specified for these tests. The nature of these tests is prospective for this device modification.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. The regulatory submission relies on adherence to established consensus standards (e.g., ANSI/AAMI BP22, ISO 11135, IEC 60601-1-2), rather than a new study requiring expert ground truth for diagnostic accuracy.
4. Adjudication method for the test set
Not applicable, as no diagnostic accuracy study requiring expert adjudication is described in the provided text for the modified device. The compliance is against engineering and safety standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a pressure transducer system, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
A standalone performance evaluation would be the stated compliance to standards like ANSI/AAMI BP22 for pressure accuracy, and the safety/EMC standards. There is no algorithm in the AI sense for a standalone "algorithm only" performance. The device itself is the "standalone" component measuring pressure.
7. The type of ground truth used
For pressure accuracy and functional performance, the ground truth is based on established metrology reference standards and methods outlined in ANSI/AAMI BP22:1994(R)2006.
For sterilization, EO residuals, BET, and EMC, the ground truth is defined by the specific requirements and test methods detailed in the respective international and national standards (e.g., ANSI/AAMI/ISO 11135, ISO 10993-7, AAMI ANSI ST72, IEC 60601-1-2).
8. The sample size for the training set
Not applicable. This is not a machine learning/AI device, so there is no training set mentioned.
9. How the ground truth for the training set was established
Not applicable. There is no training set involved.
Ask a specific question about this device
(141 days)
The CenterVue COMPASS is intended for taking digital images of a human retina without the use of a mydriatic agent and for measuring retinal sensitivity, fixation stability and the locus of fixation. It contains a reference database that is a quantitative tool for the comparison of retinal sensitivity to a database of known normal subjects.
The CenterVue COMPASS is a scanning ophthalmoscope combined with an automatic perimeter that allows the acquisition of images of the retina, as well as the measurement of retinal threshold sensitivity and the analysis of fixation. The device works with a dedicated software application, operates as a standalone unit, integrates a dedicated tablet, a joystick, a push-button and is provided with an external power supply. COMPASS operates in non-mydriatic conditions, i.e. without the need of pharmacological dilation and is intended for prescription use only.
The Centervue COMPASS device operates on the following principles:
- An anterior segment alignment system is included, which uses two infrared LEDs with a centroid wavelength of 940 nm and two cameras, whereas the former illuminate the external eye by diffusion and the latter allow a stereoscopic reconstruction of the pupil's position, used for automated alignment purposes via pupil tracking;
- An infrared imaging system captures live monochromatic images of the central retina over a circular field of view of 60° in diameter, by an horizontal line from an infrared LED with a centroid wavelength of 850 nm and by an oscillating mirror which scans the line to uniformly illuminate the retina; such images are in turn used for auto-focusing purposes and to track eye movements, providing a measure of a patient's fixation characteristics and allowing active compensation of the position of perimetric stimuli;
- A concurrent color imaging system allows the capture of color images of the central retina over a circular field of view of 60° in diameter, using a white LED and a blue LED combined to obtain a white light illuminating the retina by the same scan mechanism;
- A fixation target projecting onto the retina a fixation target obtained from a green LED;
- A stimuli projector, projecting onto the retina white light Goldmann stimuli at variable intensity and allowing measurements of threshold sensitivity at multiple locations, according to a patient's subjective response to the light stimulus projected at a certain location.
The COMPASS device interacts with the patient by directing infrared, white, blue and green wavelength illumination into the patient's eye and by recording a patient's confirmation that a certain light stimulus has been perceived or not.
Here's an analysis of the acceptance criteria and the supporting study for the CenterVue COMPASS device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The FDA clearance letter (K150320) primarily focuses on establishing "substantial equivalence" to predicate devices, rather than explicit numerical acceptance criteria for clinical performance that might be found in a performance goal document for a novel device. However, the clinical study serves to demonstrate this equivalence. The key performance comparison is between the CenterVue COMPASS and the Humphrey HFA-II.
| Acceptance Criteria (Implied for Substantial Equivalence to HFA-II) | Reported Device Performance (CenterVue COMPASS) |
|---|---|
| Equivalence in retinal threshold sensitivity measurements for both normal and pathological subjects compared to the Humphrey HFA-II. | Mean differences in thresholds between COMPASS and HFA-II in both subject groups (normal and pathological) were found to be equivalent to those reported for the Humphrey HFA between SITA Standard and full threshold. |
| No significant adverse events during clinical testing. | No adverse event was reported during the study. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size:
- 200 normal subjects
- 120 subjects with pathology affecting the visual field (specifically glaucoma)
- Total: 320 subjects
- Data Provenance: The document does not explicitly state the country of origin. It indicates the manufacturer is in Padova, Italy, and the study was conducted to support FDA clearance in the USA, suggesting the study likely occurred in conjunction with the manufacturer's operations or clinical sites. The study is presented as prospective clinical testing ("Measurements have been obtained...").
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
The document does not specify the number of experts or their qualifications for establishing the "ground truth" for the test set (i.e., whether subjects were truly "normal" or had "glaucoma"). It only states that subjects were categorized as "normal" or with "pathology affecting the visual field (in particular glaucoma)." This implies a clinical diagnosis was used, but the specific process or number of experts for this diagnosis is not detailed.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method for the test set in terms of expert review or consensus. The study compares the performance of the COMPASS directly to the predicate device (Humphrey HFA-II) on the same subjects, rather than assessing the COMPASS's ability to classify against a pre-established ground truth determined by multiple experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This device is primarily a diagnostic instrument for measuring retinal sensitivity and imaging, not an AI-assisted diagnostic aid for interpretation by human readers. The clinical study compares the device's measurements to another device, not human performance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the clinical study presents data on the standalone performance of the CenterVue COMPASS device in measuring retinal threshold sensitivity. It directly compares the measurements obtained by the COMPASS to those obtained by the Humphrey HFA-II. The device operates as a standalone unit for acquiring images and measuring retinal sensitivity.
7. The Type of Ground Truth Used
The "ground truth" in this context is the measurement of retinal threshold sensitivity as determined by the accepted standard, the Humphrey HFA-II. The study aims to demonstrate that the COMPASS's measurements are "equivalent" to those of the HFA-II, specifically that the mean differences in thresholds are comparable to known differences within the HFA-II platform (SITA Standard vs. full threshold). The classification of subjects as "normal" or with "glaucoma" would have been based on clinical diagnosis, implicitly serving as a form of "expert consensus" or "clinical diagnosis" ground truth for subject selection, but not for the specific performance metric being evaluated (threshold sensitivity differences).
8. The Sample Size for the Training Set
The document describes a "reference database" that was developed to serve as a quantitative tool for comparison of retinal sensitivity to known normal subjects.
- Reference Database Sample Size: 200 eyes of 200 normal subjects.
- The age range of this population was 20 - 86 years (50.6 ± 15.2).
9. How the Ground Truth for the Training Set was Established
The ground truth for the "training set" (referred to as the "reference database" in the document) was established by obtaining threshold sensitivity data from 200 subjects confirmed to be "normal." The specific criteria or expert qualifications for determining "normalcy" are not detailed in this summary, but it implies a clinical assessment of individuals free from visual field pathology. The perimetric settings used to gather this data are listed (24-2 grid, 4-2 strategy, Goldmann III stimulus, etc.).
Ask a specific question about this device
(143 days)
The Compass™ CT disposable pressure transducer with integrated digital display is intended for direct measurement and monitoring of physiological pressure, including during the infusion of fluids and therapeutic and diagnostic agents.
The Compass™ CT Port disposable pressure transducer with integrated digital display is intended for direct measurement and monitoring of physiological pressure, including during the infusion of fluids and therapeutic and diagnostic agents.
The Compass CT and Compass CT Port are disposable, point-of-use pressure measurement and monitoring devices that incorporate a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The devices have a distal male luer fitting to connect to a needle or catheter, and a proximal female luer fitting that can be connected to accessory devices (e.g. syringes, caps, or infusion tubing). The devices measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD. The Compass CT Port has an additional, sealed proximal port through which commercially available guidewires can be inserted during pressure measurement.
The Mirador Compass™ CT and CT Port are disposable pressure measurement and monitoring devices with an integrated digital display. The devices are intended for direct measurement and monitoring of physiological pressure, including during the infusion of fluids and therapeutic and diagnostic agents.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. A table of acceptance criteria and the reported device performance
The provided document does not explicitly present a table of acceptance criteria with corresponding performance metrics like "Accuracy: X mmHg". Instead, it states that "The subject Compass CT and CT Port incorporate new software that allows for an extended pressure range. Therefore, all verification tests related to the software and pressure accuracy that were completed to support the substantial equivalence of the predicate Compass devices were re-executed for the subject Compass CT and CT Port. Pressure accuracy testing was completed per ANSI/AAMI BP22:1994(R)2006 and ISO 60601-2-34 310 Ed."
Without the specific results from the re-executed pressure accuracy testing, a direct table of acceptance criteria vs. reported performance cannot be created from the given text. However, the implicit acceptance criterion is adherence to the performance standards specified in ANSI/AAMI BP22:1994(R)2006 and ISO 60601-2-34 310 Ed for pressure accuracy.
The text states: "The results from this in vitro testing demonstrate that the technological and performance characteristics of the subject Compass CT and CT Port meet defined design requirements and that they can perform in a manner equivalent to devices currently on the market used for measuring physiological pressure." This implies that the device did meet the internal design requirements and the standards cited.
Implicit Acceptance Criteria & Reported Performance:
| Acceptance Criteria Category | Standard/Requirement | Reported Device Performance |
|---|---|---|
| Pressure Accuracy | Per ANSI/AAMI BP22:1994(R)2006 and ISO 60601-2-34 310 Ed. | "The results from this in vitro testing demonstrate that the technological and performance characteristics of the subject Compass CT and CT Port meet defined design requirements and that they can perform in a manner equivalent to devices currently on the market used for measuring physiological pressure." (Implying successful conformance to the standards.) |
2. Sample size used for the test set and the data provenance
The document states "Pressure accuracy testing was completed per ANSI/AAMI BP22:1994(R)2006 and ISO 60601-2-34 310 Ed." These are in vitro testing standards.
- Sample size: The specific number of devices or measurements used in the pressure accuracy testing is not provided in the given text.
- Data provenance: The testing was in vitro (laboratory-based), rather than from a clinical setting. The country of origin of the data is not explicitly stated, but the manufacturer is Mirador Biomedical, Inc. in Seattle, Washington, USA, suggesting the testing likely occurred in the US or at a certified lab. The study is prospective in the sense that these specific tests were re-executed for the new software version.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This type of information (experts, ground truth establishment) is typically not applicable or reported for in vitro device performance testing where the ground truth is established by a calibrated reference standard (e.g., a highly accurate pressure calibrator). The performance is assessed against these known, precise values, not expert interpretation.
4. Adjudication method for the test set
Not applicable for in vitro performance testing against a calibrated reference standard.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was done. This device is a direct pressure measurement device, not an AI-assisted diagnostic tool that would involve human readers interpreting images or data with and without AI.
- No AI component requiring human-in-the-loop performance measurement is described. The device incorporates an "integrated pre-programmed diagnostic computer" but this refers to its internal processing and display of pressure, not an AI for interpretation or assistance to a human reader.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the pressure accuracy testing conducted ("re-executed for the subject Compass CT and CT Port") is a form of standalone performance assessment. The performance of the device's internal pressure measurement and display system was evaluated independently (without human intervention in the measurement process itself, beyond operating the test equipment). The "algorithm" here refers to the software that processes the pressure sensor's input and drives the display.
7. The type of ground truth used
The ground truth for the pressure accuracy testing would be established by calibrated reference standards. This typically involves highly accurate, independently verified pressure transducers or calibrators against which the device's measurements are compared.
8. The sample size for the training set
The concept of a "training set" is not relevant here as this is not an AI/machine learning device that learns from data. It's a measurement device with a pre-programmed diagnostic computer.
9. How the ground truth for the training set was established
Not applicable as there is no training set mentioned or implied for this type of device.
Ask a specific question about this device
(135 days)
The Compass™ GP disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.
The Compass™ Thoracentesis disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.
The Compass™ Paracentesis disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.
The Compass™ Compartment Pressure disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.
The Compass™ Epidural Assist disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.
The Compass™ Arterial Assist disposable pressure transducer with integrated digital display is intended for direct measurement of physiological pressure.
The Compass devices are disposable, point-of-use pressure measurement and monitoring devices that incorporate a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The Compass devices are designed to attach distally to an inserted needle or catheter, measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD.
-
Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly present a table of acceptance criteria with corresponding performance results. However, it indicates that "Pressure accuracy testing was completed per ANSVAAMI BP22:1994(R)2006." The study concludes that "The results from this in vitro testing demonstrate that the technological and performance characteristics of the Compass devices meet defined design requirements and that they can perform in a manner equivalent to devices currently on the market used for measuring physiological pressure." This implies that the device met the accuracy requirements specified in the ANSVAAMI BP22:1994(R)2006 standard, which served as the acceptance criteria for pressure accuracy. Specific numerical values for the acceptance criteria and the device's performance against them are not provided in this summary.
-
Sample Size for Test Set and Data Provenance:
The document describes "in vitro testing" without specifying a particular sample size for a test set in the context of a dataset of medical images or patient records. The testing performed focused on physical device performance, specifically pressure accuracy, rather than evaluating an algorithm on a clinical dataset. The data provenance is related to the physical testing of the device.
-
Number of Experts and Qualifications for Ground Truth:
Not applicable. The study is a device performance test (pressure accuracy) and does not involve establishing ground truth by medical experts for a diagnostic or AI-driven task.
-
Adjudication Method for Test Set:
Not applicable. The study is a device performance test and does not involve adjudication of expert interpretations for a test set.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a disposable pressure transducer with an integrated digital display, not an AI system designed to assist human readers in interpreting medical images or data.
-
Standalone (Algorithm Only) Performance Study:
The device itself (Compass™) is the standalone device. The testing described is of the device's performance, specifically its pressure accuracy and software functionality. It's not an algorithm in the sense of a software-only diagnostic tool that would typically undergo a standalone performance study in the context of AI. The performance of the integrated pre-programmed diagnostic computer is assessed as part of the overall device.
-
Type of Ground Truth Used:
For the pressure accuracy testing, the "ground truth" would have been established by a reference standard pressure measurement device or system, as per the ANSVAAMI BP22:1994(R)2006 standard. This is a technical (metrological) ground truth rather than clinical ground truth (e.g., pathology, outcomes data, or expert consensus on a diagnosis).
-
Sample Size for Training Set:
Not applicable. The device is a physical pressure transducer with integrated software, not an AI system requiring a training set in the machine learning sense. The software is described as "integrated pre-programmed diagnostic computer" and "new software," suggesting it was developed through traditional software engineering processes, not by training on a dataset.
-
How Ground Truth for Training Set Was Established:
Not applicable, as there is no "training set" in the context of this device's development. The software was likely developed and verified against functional and performance requirements through established software development lifecycles and testing protocols.
Ask a specific question about this device
(30 days)
Ask a specific question about this device
(133 days)
The Compass™ Vascular Access is a disposable pressure transducer with integrated digital display intended for direct measurement and monitoring of invasive blood pressure.
The Compass™ Lumbar Puncture is a disposable pressure transducer with integrated digital display intended for direct measurement and monitoring of intracranial pressure.
The Compass is a disposable, point-of-use pressure measurement and monitoring device that incorporates a pressure transducer and an integrated pre-programmed diagnostic computer with liquid crystal display (LCD). The Compass is provided as two different models, each corresponding to a specific indication for use:
- (1) The Compass Vascular Access (VA) is for invasive blood pressure measurement and monitoring during the insertion of vascular access devices.
- (2) The Compass Lumbar Puncture (LP) is for intracranial pressure measurement and monitoring via lumbar puncture or intraventricular catheter.
Both models are designed to attach distally to an inserted needle or catheter, measure the pressure via an embedded pressure sensor, internally convert changes in pressure into electrical currents, and then display the resulting pressure via the integrated LCD.
The Compass VA and Compass LP devices are identical except for two differences: 1) the measurement units in which the pressure is displayed on the LCD (mm Hg for blood pressure measurements with the Compass VA and cm H2O for intracranial pressure measurements with the Compass LP), and 2) the Compass VA blood pressure monitoring device has a proximal port through which commercially available guidewires can be inserted during pressure measurement.
The provided 510(k) summary for the Mirador Compass™ device (K101518) describes the device, its indications for use, and the testing performed. However, it does not provide a specific table of acceptance criteria with corresponding reported device performance values in a side-by-side format as typically found in comprehensive performance studies for AI/CAD devices.
Instead, the documentation outlines various tests conducted and generally states that the device "meet defined design requirements" and can "perform in a manner equivalent to devices currently on the market."
Let's break down the information based on your requested points:
Acceptance Criteria and Reported Device Performance
The document describes several categories of testing, implying that the acceptance criteria are met if the device performs "per Mirador Biomedical's in-house test methods, protocols, and requirements or in accordance with applicable recognized standards by accredited outside laboratories," and that the results "demonstrate that the technological and performance characteristics... meet defined design requirements and that they can perform in a manner equivalent to devices currently on the market."
Due to the nature of the provided document, a specific table of acceptance criteria and reported device performance with numerical values cannot be fully constructed as the exact numerical acceptance limits and reported performance values are not explicitly stated for all tests. However, we can infer the categories of performance that were assessed.
| Acceptance Criteria Category (Inferred) | Reported Device Performance (as stated in document) |
|---|---|
| Biocompatibility | Met ISO 10993-1 standards (cytotoxicity, hemolysis, acute systemic toxicity, intracutaneous reactivity, sensitization, pyrogenicity). |
| Functional Testing (General) | Met applicable functional testing per ANSI/AAMI BP22:1994(R)2006, ISO 594-1, and ISO 594-2. |
| Luer Fittings | Met defined design requirements. |
| Mechanical Strength and Durability | Met defined design requirements. |
| Power Switch | Met defined design requirements. |
| Dimensional and Weight Verification | Met defined design requirements. |
| Battery Life | Met defined design requirements. |
| Pressure Overrange Capability | Met defined design requirements. |
| Pressure Accuracy | Met defined design requirements. (In vivo study showed accurate measurement equivalent to predicate) |
| Drift, Temperature Error Band, Temperature Error Band of Sensitivity | Met defined design requirements. |
| Light Sensitivity | Met defined design requirements. |
| Frequency Response | Met defined design requirements. |
| Guidewire Testing (Compass VA) | Met defined design requirements. |
| Spillage | Met defined design requirements. |
| Safety | Met defined design requirements. |
| Simulated Use Testing (Vascular Access & Lumbar Puncture) | Met defined design requirements. |
| In vivo Performance | Compass VA and LP devices accurately measure and monitor labeled physiologic pressures (invasive blood pressure/intracranial pressure) in a manner equivalent to the predicate device. |
Detailed Information on Studies and Ground Truth:
This device (Mirador Compass™) is a disposable pressure transducer with an integrated digital display and pre-programmed diagnostic computer. It is not an AI/CAD (Computer-Aided Detection/Diagnosis) system in the modern sense. Therefore, many of the typical questions for AI/CAD devices, such as those related to observer performance studies (MRMC), ground truth established by experts, adjudication methods, and training/test sets for AI models, are not applicable to this type of medical device.
1. Sample sized used for the test set and the data provenance:
- In vitro (Bench) Testing: The document lists several in vitro tests (Luer Fittings, Mechanical Strength, Power Switch, etc.). The sample sizes for each of these component-level or functional tests are not specified in the summary.
- In vivo Animal Study: The "in vivo animal study" used a swine model. The specific number of animals or the number of measurements taken is not provided.
- Data Provenance: The in vitro tests were conducted by "Mirador Biomedical's in-house test methods, protocols, and requirements or in accordance with applicable recognized standards by accredited outside laboratories." The animal study was conducted in a swine model. The country of origin for the animal study is not specified, but the manufacturer is US-based (Seattle, Washington).
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable for this type of device. The device's performance is measured against physical standards, calibrated equipment, and physiological measurements in an animal model, not against human expert interpretations of images or data that require adjudication for ground truth.
3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. As this is not an AI/CAD device based on expert interpretation, no adjudication method was used.
4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done. This device is a measurement tool, not an AI assistance tool for human interpretation.
5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The device itself is a "standalone" measurement device in that it directly measures and displays pressure. However, it's not an "algorithm-only" standalone performance test in the context of AI/CAD where the algorithm's output is being evaluated for accuracy without human input. Its performance was assessed through direct physical and physiological measurements.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- In vitro Tests: Ground truth would be established by reference standards, calibrated measurement equipment, and specified engineering tolerances as defined by design requirements and industry standards (e.g., ANSI/AAMI BP22, ISO 594-1/2).
- In vivo Animal Study: Ground truth for pressure measurements would be established by direct physiological measurements from the swine model, likely using a highly accurate reference pressure transducer or established invasive pressure monitoring techniques, against which the Compass device's readings are compared. The summary explicitly states that a "predicate Shenzhen DPT pressure sensor was used for relative comparison."
7. The sample size for the training set:
- Not applicable. This device is not an AI/Machine Learning diagnostic system that undergoes a "training" phase with a dataset.
8. How the ground truth for the training set was established:
- Not applicable, as there is no training set for an AI model.
In summary, the K101518 submission for the Mirador Compass™ device demonstrates substantial equivalence through extensive bench testing and an in vivo animal study against established standards and a predicate device. The information provided aligns with typical submissions for Class II measurement devices, rather than AI-driven diagnostic tools.
Ask a specific question about this device
(106 days)
COMPASS is used for quality assurance and plan verification in external beam radiation therapy for multileaf collimated fields and intensity modulated fields.
It computes dose (energy per volume deposited by ionizing radiation) three-dimensionally in a geometrical representation of a patient or a phantom. The calculation is based on read-in treatment plans, and additionally on online or offline measurements of radiation fields with radiation-transparent or non-transparent, 2 dimensional multi-element ionization chamber arrays.
The system is intended to be used with intensity modulated radiation therapy, delivered with high energy x-ray beams from an isocentric gantry linear accelerator. The modulation of these beams shall be accomplished by means of MLCs and Jaws.
The aim is to measure the fluences of the applied fields by means of electronic 2D devices, including transmission detector which can be used during patient irradiation. Alternatively the fluences can be calculated from a computation independent from the original TPS, using segmentation data from the TPS or a delivery log file.
Furthermore, the intention of the system is to recalculate the resulting dose distribution in a phantom or, using patient anatomy data, in a patient.
Moreover, the system allows performing QA tests on the radiation delivery system. When the system is used together with electronic 2D devices, preferably with transmission detector, it can also provide sensitivity calibration and performance tests on EPID systems.
Finally, the system will provide online information about the quality of the delivery, both on a delivery system level (e.g. leaf positioning accuracy, dose per segment accuracy) as on the patient level (resulting dose distribution). In case of discrepancies, corrective action (adjusted plan for further fractions) shall be proposed.
I'm sorry, but relevant information regarding the acceptance criteria, study details, and related quantitative data are missing in the provided text. The document describes the "Compass" device, its intended use, and its classification, but it does not contain the specific details required to answer your request, such as acceptance criteria, reported performance, sample sizes, ground truth establishment, or clinical study outcomes.
Ask a specific question about this device
Page 1 of 2