    Device Name :

    Blue Latex Examination Powder Free Gloves

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Blue Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D 3578-05 (Reapproved 2015), Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are blue in color and are powder free.

    AI/ML Overview

    The document provided describes the acceptance criteria and the study results for the Blue Latex Examination Powder Free Gloves (K192329). This is a medical device, specifically examination gloves, and the studies evaluate physical properties and biocompatibility.

    Here's the requested information:

    1. A table of acceptance criteria and the reported device performance

    Characteristic / Test MethodAcceptance CriteriaReported Device Performance (K192329)Comparison with Predicate (K141042)
    Dimensions - Length (ASTM D3578-05)Min 230 mm for all sizesSmall: 304 mm, Medium: 304 mm, Large: 305 mm, X-Large: 305 mm, XX-Large: 305 mmSimilar
    Dimensions - Width (ASTM D3578-05)Small: 80+/-6 mm, Medium: 95+/-6mm, Large: 105+/-6 mm, X-Large: 114+/-6 mm, XX-Large: 120 +/-6 mmSmall: 84 mm, Medium: 94 mm, Large: 105 mm, X-Large: 114 mm, XX-Large: 123 mmSimilar
    Thickness (ASTM D3578-05)Palm: 0.08 mm min, Finger: 0.08 mm min for all sizesPalm: 0.31mm (all sizes), Finger: 0.38mm (all sizes)Similar
    Physical Properties - Tensile Strength (ASTM D3578-05)Before Ageing: 18 Mpa Min, After Ageing: 14 Mpa MinBefore Ageing: Small: 33.0 Mpa, Medium: 32.9 Mpa, Large: 32.2 Mpa, X-Large: 31.9 Mpa, XX-Large: 31.1 Mpa. After Ageing: Small: 30.0 Mpa, Medium: 30.6 Mpa, Large: 29.9 Mpa, X-Large: 29.7 Mpa, XX-Large: 28.2 MpaSimilar
    Physical Properties - Ultimate Elongation (ASTM D3578-05)Before Ageing: 650% Min, After Ageing: 500% MinBefore Ageing: Small: 1322%, Medium: 1250%, Large: 1392%, X-Large: 1130%, XX-Large: 1149%. After Ageing: Small: 1046%, Medium: 1122%, Large: 1257%, X-Large: 1011%, XX-Large: 1110%Similar
    Physical Properties - Stress at 500% Elongation (ASTM D3578-05)Before Ageing: 5.5 Mpa MaxSmall: 5.1 Mpa, Medium: 5.2 Mpa, Large: 5.2 Mpa, X-Large: 5.1 Mpa, XX-Large: 5.2 MpaNot directly compared to predicate (N/A in predicate column) but meets the criteria
    Water Tight (Holes) (ASTM D5151-06)AQL 2.5Passes AQL 1.5Same
    Powder Free Residue (ASTM D6124-06)≤ 2 mg/glove MaxSmall: 0.20 mg/glove, Medium: 0.21 mg/glove, Large: 0.22 mg/glove, X-Large: 0.22 mg/glove, XX-Large: 0.23 mg/gloveSimilar
    Extractable Protein (ASTM D 5712-95)200 µg/dm² MaxSmall: 45 µg/dm², Medium: 44 µg/dm², Large: 44 µg/dm², X-Large: 43 µg/dm², XX-Large: 45 µg/dm²Not specified for predicate
    Biocompatibility - Primary Skin Irritation (ISO 10993-10:2010(E))Not an irritantNot an irritantSame
    Biocompatibility - Dermal Sensitization (ISO 10993-10:2010(E))Not a sensitizerNot a sensitizerSame
    Biocompatibility - In vitro cytotoxicity (ISO10993-5:2009(E))Not explicitly stated in predicate, but implied (safe)Cytotoxic, which is to be expected as latex is the positive control for this test.Different (explained as expected)
    Biocompatibility - Material mediated Pyrogenicity (ISO 10993-11:2017(E) / USP)Data not available for predicate, implied non-pyrogenic if not listed.Non-pyrogenicNot available for predicate
    Intended UseFor medical purposes to prevent contamination between patient and examiner.Disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.Same
    MaterialNatural LatexNatural LatexIdentical
    ColorBlueBlueIdentical
    TextureFinger TextureFinger textureIdentical
    Size (ASTM D3578-05)Small, Medium, Large, X Large & XX LargeSmall, Medium, Large, X Large & XX LargeSame
    Single Use (Medical Glove Guidance Manual - Labeling)Single UseSingle UseSame
    Sterile/non sterileNon sterileNon sterileSame
    Powder/Powder freePowder freePowder freeSame
    Label and LabelingMeets FDA's label and labeling requirementsMeets FDA's label and labeling requirementsSame

    2. Sample size used for the test set and the data provenance

    The document provides specific values for various measurements (e.g., length, width, thickness, tensile strength, ultimate elongation, residual powder content, extractable protein content) across different glove sizes (Small, Medium, Large, X-Large, XX-Large). For tests like Water Tightness (AQL 1.5 achieved) and Biocompatibility testing (Primary Skin Irritation, Dermal Sensitization, In vitro cytotoxicity, Material mediated Pyrogenicity), the methodology implies testing on a statistically significant sample, but the exact sample sizes are not explicitly stated in the provided text for each test.

    The data provenance is from non-clinical tests conducted by JR Engineering & Medical Technologies (M) SDN.BHD. in Malaysia. The study is a prospective assessment of the device's adherence to established ASTM standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the document describes non-clinical testing of physical and chemical properties of a medical device (gloves), not studies requiring expert interpretation of medical images or conditions to establish a ground truth. The "ground truth" here is defined by established engineering and medical device standards (e.g., ASTM, ISO).

    4. Adjudication method for the test set

    This is not applicable for the type of testing described (physical and chemical properties testing against standards). Adjudication methods like 2+1 or 3+1 are typically used in clinical studies for medical image interpretation or diagnostic accuracy assessments where human expert consensus is required.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The document describes the performance of an examination glove, not an AI-powered diagnostic or assistive technology.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This document evaluates a physical medical device (gloves), not a software algorithm.

    7. The type of ground truth used

    The ground truth used for these tests is based on established industry standards and specifications:

    • ASTM D3578-05 (Reapproved 2015) Standard Specification for Rubber Examination Gloves
    • ASTM D5151-06 (Reapproved 2015) Standard Test Method for Detection of Holes in Medical Gloves
    • ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
    • ASTM D 5712-95 (Reapproved 2010) Standard Test Method for the Analysis of Protein in Natural Rubber
    • ISO 10993-10:2010(E) for Primary Skin Irritation and Dermal Sensitization
    • ISO 10993-5:2009(E) for In vitro cytotoxicity
    • ISO 10993-11:2017(E) / USP for Material mediated Pyrogenicity

    8. The sample size for the training set

    This is not applicable as this pertains to a physical product's performance validation, not a machine learning model.

    9. How the ground truth for the training set was established

    This is not applicable (see #8).