510(k) Data Aggregation

    K Number
    K222348
    Device Name
    JR MEDIC
    Date Cleared
    2022-10-27

    (85 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Latex Examination Powder Free Gloves are disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Latex Examination Powder Free Gloves are manufactured to meet all the current specifications listed under the ASTM Specification D3578-2019, Standard Specification for Rubber Examination Gloves. They are made from Natural Rubber Latex. These gloves are natural in color (No color is added) and are powder free and Non-Sterile.

    AI/ML Overview

    The provided document describes the FDA 510(k) premarket notification for "Latex Examination Powder Free Gloves." This is a Class I medical device, which typically does not require a complex AI/algorithm-based study with human readers, ground truth establishment by experts, or MRMC studies. The approval is based on demonstrating substantial equivalence to a predicate device through non-clinical performance data against established industry standards.

    Therefore, many of the requested elements for describing AI acceptance criteria and study methodology (like sample sizes for test/training sets, expert ground truth, adjudication, MRMC studies, standalone algorithm performance, etc.) are not applicable to this type of device and submission.

    The document explicitly states "H. SUMMARY OF CLINICAL TESTING: Not applicable." This reinforces that the submission did not rely on clinical performance data involving human readers or sophisticated AI algorithms.

    However, I can extract the acceptance criteria and performance data as presented for these gloves.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the "Latex Examination Powder Free Gloves" are based on established ASTM (American Society for Testing and Materials) standards for rubber examination gloves. The table below summarizes these criteria and the reported performance of the device (K222348) compared to the predicate device (K192329).

    1. Table of Acceptance Criteria and Reported Device Performance

    Characteristics / Test MethodStandard / PurposeAcceptance Criteria (K192329 Predicate)Reported Device Performance (K222348 - Current Device)Comparison
    Dimensions - Length (ASTM D3578-2019)To determine the length of the gloves> 230 mm (all sizes)> 230 mm (Actual values: X-Small 305 mm, Small 306 mm, Medium 307 mm, Large 308 mm, X-Large 310 mm)Similar
    Dimensions - Width (ASTM D3578-2019)To determine the width of the glovesMin 95+/-10 mm (for medium size)X-Small 76 mm, Small 85 mm, Medium 96 mm, Large 106 mm, X-Large 116 mmSimilar
    Physical Properties - Tensile Strength (ASTM D3578-2019)To Determine the physical propertiesBefore Ageing: > 18 Mpa (all sizes)After Ageing: > 14 Mpa (all sizes)Before Ageing: X-Small 22.07 Mpa, Small 22.15 Mpa, Medium 22.22 Mpa, Large 22.30 Mpa, X-Large 22.32 MpaAfter Ageing: X-Small 18.49 Mpa, Small 18.56 Mpa, Medium 18.67 Mpa, Large 18.74 Mpa, X-Large 18.76 MpaSimilar
    Physical Properties - Ultimate Elongation (ASTM D3578-2019)To Determine the physical propertiesBefore Ageing: > 650% (all sizes)After Ageing: > 500% (all sizes)Before Ageing: X-Small 858%, Small 869%, Medium 874%, Large 880%, X-Large 882%After Ageing: X-Small 841%, Small 848%, Medium 854%, Large 860%, X-Large 862%Similar
    Thickness (ASTM D3578-2019)To determine the thickness of the glovesPalm > 0.08 mm, Finger > 0.08 mm (all sizes)Palm: 0.16mm (all sizes)Finger: 0.22mm (all sizes)Similar
    Powder Free Residue (ASTM D6214)To determine the residual powder in the gloves≤ 2 mg/gloveX-Small 0.21 mg/glove, Small 0.21 mg/glove, Medium 0.22 mg/glove, Large 0.22 mg/glove, X-Large 0.22 mg/gloveSimilar
    Protein Content (ASTM D5712)To determine the extractable protein in the glovesMax 200 µg/dm²43.19 µg/dm² (reported as 43.65 µg/dm² for all sizes in detailed results)Similar
    Biocompatibility - Primary Skin Irritation (ISO 10993-10:2010)Not specified in detail, but expected to ensure safe contact with skin.Under the condition of study, not an irritantUnder the condition of study, not an irritantSame
    Biocompatibility - Dermal Sensitization (ISO 10993-10:2010)Not specified in detail, but expected to ensure safe contact with skin.Under the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizerSame
    Biocompatibility - In vitro cytotoxicity (ISO 10993-5:2009)Not specified in detail, but expected to ensure no toxic effects on cells.Under the conditions of the study, cytotoxic (This might be a typo and should be "not cytotoxic", or it implies a specific acceptable level of cytotoxicity.)Under the conditions of the study, cytotoxic (As above, likely implies within acceptable limits for the device type, or a typo).Same
    Biocompatibility - Acute Systemic Toxicity Test (ISO 10993-11:2017)Not specified in detail, but expected to ensure no systemic toxic effects.Device extracts do not pose a systemic toxicity concernDevice extracts do not pose a systemic toxicity concernSame
    Water Tightness (ASTM D5151-2019)To determine the holes in the glovesAQL 2.5Passes AQL 1.5 (AQL 1.5 is a tighter requirement than AQL 2.5, indicating better performance)Same
    Material-mediated Pyrogenicity (ISO 10993-11:2017 / USP 41<151>)To ensure the material does not cause fever.Under the conditions of the study, non-pyrogenicNo data available (for the current device in this specific row, but typically expected for such devices. May be covered by general biocompatibility; the predicate was non-pyrogenic).----
    Intended use/Indication for useJR MEDIC Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.JR MEDIC Blue Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.Latex Examination Powder Free Gloves are disposable devices intended for medical purpose that are worn on the examiner's hand to prevent contamination between patient and examiner.Similar
    Material-Natural LatexNatural LatexSame
    Color-BlueNatural (No color is added)Different
    Texture-Finger TextureFinger textureSame
    SizeASTM D3578-2019Small, Medium, Large & X LargeX Small, Small, Medium, Large, X-Large (Added X-Small size)Similar
    Single UseMedical Glove Guidance Manual - LabelingSingle UseSingle UseSame
    Sterile/non sterile-Non sterileNon sterileSame
    Powder/Powder free-Powder freePowder freeSame

    Study Details (Applicable to this Device Type)

    Given this is a 510(k) for examination gloves, the "study" is a series of non-clinical, laboratory tests performed according to specific ASTM and ISO standards, rather than a clinical trial or an AI algorithm performance study.

    2. Sample Size Used for the Test Set and Data Provenance:
    The document does not explicitly state the sample sizes for each specific test (e.g., how many gloves were tested for length, width, or tensile strength). However, it implies that standard testing methodologies as defined by ASTM and ISO were followed. These standards usually specify the number of samples required for each test. The data provenance is from JR Engineering & Medical Technologies (M) SDN.BHD., Malaysia, indicating the tests were conducted by the manufacturer or a contracted lab in Malaysia. The data is non-clinical performance data, reported as part of a premarket notification.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
    Not applicable. The "ground truth" for this device is physical and chemical adherence to specific measurable standards (e.g., a glove is X mm long, has Y tensile strength), not a diagnostic interpretation requiring expert medical opinion. The standards themselves, like ASTM D3578-2019, serve as the "ground truth" for performance.

    4. Adjudication Method for the Test Set:
    Not applicable. This is not a human-interpreted diagnostic task requiring adjudication. The tests are objective measurements against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
    No, this was not done. MRMC studies are typically for diagnostic imaging devices or algorithms where the performance of human readers (with and without AI assistance) on clinical cases is evaluated. This device is an examination glove.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    No, this is not an AI algorithm. Its performance is measured through physical and chemical properties.

    7. The Type of Ground Truth Used:
    The ground truth used is objective engineering and material science standards (e.g., ASTM D3578-2019, ASTM D5151-2019, ASTM D6124-06, ASTM D5712-95, ISO 10993 series). These are internationally recognized standards for manufacturing and safety.

    8. The Sample Size for the Training Set:
    Not applicable. This is not an AI device trained on a dataset. The "training" for such a device would be the manufacturing process development and quality control to ensure consistency.

    9. How the Ground Truth for the Training Set was Established:
    Not applicable, as it's not an AI training set. The "ground truth" for manufacturing would be the specifications and quality control procedures adherence to ensure gloves meet the physical and chemical standards.