Search Results
Found 2 results
510(k) Data Aggregation
(123 days)
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Black
), Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Black), Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Orange) are made from synthetic rubber latex. It is single use and powder-free. The subject device is a patient examination glove made from nitrile latex compound, Black and Orange color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. The available sizes are Small, Medium, Large ,X-Large and XX-Large.
Here's a breakdown of the acceptance criteria and the study information based on the provided text, formatted as requested:
1. A table of acceptance criteria and the reported device performance
The provided document lists various performance criteria for the Nitrile Powder Free Examination Gloves (Black and Orange) and reports the results against these criteria.
Characteristic / Test | Acceptance Criteria | Reported Device Performance (Black Gloves) | Reported Device Performance (Orange Gloves) |
---|---|---|---|
Freedom From Holes | Meet requirement inspection level G-1, AQL 2.5 Sampling size 125pcs (Ac: 7, Re: 8) | Pass (≤3pcs) | Pass (≤4pcs) |
Dimension: Length | 220 mm min (small) | ||
230 mm min (other sizes) | 241 mm min | 240 mm min | |
Dimension: Width (S) | 80 ± 10 mm | average 85.5mm | average 85.8mm |
Dimension: Width (M) | 95 ± 10 mm | average 96.3mm | average 96.1mm |
Dimension: Width (L) | 110 ± 10 mm | average 106.4mm | average 106.8mm |
Dimension: Width (XL) | 120 ± 10 mm | average 115.5mm | average 115.9mm |
Dimension: Width (XXL) | 130 ± 10 mm | average 125.1mm | average 125.6mm |
Dimension: Thickness (Palm) | Minimum 0.05 mm | 0.178mm min. | 0.181mm min. |
Dimension: Thickness (Finger) | Minimum 0.05 mm | 0.199mm min. | 0.212mm min. |
Physical Properties (Before Aging): Tensile Strength | 14 MPa, min. | 15.1 MPa, min. | 15.3 MPa, min. |
Physical Properties (Before Aging): Elongation | 500%, min. | 518%, min. | 502%, min. |
Physical Properties (After Aging): Tensile Strength | 14 MPa, min. | 15.0 MPa, min. | 14.1 MPa, min. |
Physical Properties (After Aging): Elongation | 400%, min. | 460%, min. | 462%, min. |
Residual Powder Content | Not more than 2 mg per glove | average 0.32mg | average 0.30mg |
Biocompatibility: Primary Skin Irritation | Not a primary skin irritant | Pass (Conclusion: The test result showed that the response of the test article was categorized as negligible under the test condition) | Pass (Conclusion: The test result showed that the response of the test article was categorized as negligible under the test condition - implied similar from the text for both colors) |
Biocompatibility: Skin Sensitization | Not a contact sensitizer | Pass (Conclusion: Under the conditions of this study, the test article showed no significant evidence of causing skin sensitization in the guinea pig) | Pass (Conclusion: Under the conditions of this study, the test article showed no significant evidence of causing skin sensitization in the guinea pig - implied similar from the text for both colors) |
Biocompatibility: Acute Systemic Toxicity | No systemic toxicity | Pass (Conclusion: Under the conditions of this study, there was no evidence of acute systemic toxicity from the extract. The test article extract met the requirements of the study.) | Pass (Conclusion: Under the conditions of this study, there was no evidence of acute systemic toxicity from the extract. The test article extract met the requirements of the study. - implied similar from the text for both colors) |
Chemotherapy Drugs & Fentanyl Citrate Permeation (Minimum Breakthrough Detection Time in minutes) | As per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. (Specific breakthrough times vary by drug) | ||
Carmustine (BCNU) | Not explicitly stated as a general minimum, but the document highlights extremely low permeation times for this drug and warns against its use. | 21.6 minutes | 22.7 minutes |
Cyclophosphamide | >240 min. | >240 min. | >240 min. |
Doxorubicin HCl | >240 min. | >240 min. | >240 min. |
Etoposide | >240 min. | >240 min. | >240 min. |
Fluorouracil | >240 min. | >240 min. | >240 min. |
Mitomycin C | >240 min. | >240 min. | >240 min. |
Oxaliplatin | >240 min. | >240 min. | >240 min. |
Paclitaxel | >240 min. | >240 min. | >240 min. |
Thiotepa | Not explicitly stated as a general minimum, but the document highlights extremely low permeation times for this drug and warns against its use. | 28 minutes | 37.1 minutes |
Fentanyl Citrate Injection | >240 min. | >240 min. | >240 min. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Physical/Dimensional/Chemical Tests: For "Freedom From Holes," the sampling size is stated as 125 pieces (Ac: 7, Re: 8). The specific sample size for other physical/dimensional tests (length, width, thickness, tensile strength, elongation, residual powder content) is not explicitly stated, but the tests are reported as meeting ASTM D6319-19 and ASTM D6124-06 requirements. For Biocompatibility tests, no specific sample size is given, but the studies involve guinea pigs and rats, implying animal testing was conducted.
- Sample Size for Chemotherapy Drug Permeation Tests: Not explicitly stated within the document. The tests are reported as per ASTM D6978-05 (Reapproved 2019).
- Data Provenance: The manufacturing company is Dehai (Shandong) Medical Gloves, Co., Ltd. located in Zibo City, Shandong Province, China. The testing would presumably have been conducted in a laboratory, potentially in China or a third-party testing facility. The document doesn't specify if the data is retrospective or prospective, but testing of device performance characteristics is generally prospective for regulatory submissions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a medical glove, and its performance is evaluated against established physical, chemical, and biological standards (ASTM, ISO). There is no "ground truth" established by human experts in the way an AI diagnostic algorithm would require. The "ground truth" is defined by the objective metrics and thresholds specified in the referenced ASTM and ISO standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus ground truth in studies involving human interpretation (e.g., radiology reads). This document pertains to physical and chemical testing of a medical device against predefined standards, which does not involve such adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for a medical glove, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance evaluation was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a medical glove, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is based on established industry standards and objective material properties. Specifically:
- ASTM D6319-19: Standard specification for Nitrile Examination Gloves.
- ASTM D5151-19: Standard test methods for detection of holes in medical gloves.
- ASTM D6124-06: Standard test method for residual powder on medical gloves.
- ASTM D6978-05 (Reapproved 2019): Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
- ISO 10993-23, ISO 10993-10, ISO 10993-11: Biocompatibility standards for medical devices (Primary Skin Irritation, Skin Sensitization, Acute Systemic Toxicity).
These standards define the acceptable range or threshold for each characteristic (e.g., minimum tensile strength, maximum powder residue, minimum breakthrough time for drugs).
8. The sample size for the training set
Not applicable. This device is a medical glove, not a machine learning model, so there is no training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(49 days)
BLACK NITRILE POWDER-FREE EXAMINATION GLOVES
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free black nitrile patient examination glove, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
Here's an analysis of the provided information, structured according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Standard | Acceptance Criteria (from Standard) | Reported Device Performance |
---|---|---|---|
Dimension | ASTM D6319-10 | (Implicitly, compliance with all dimensional specifications in ASTM D6319-10) | Meets (meaning it complies with ASTM D6319-10) |
Physical Properties | ASTM D6319-10 | (Implicitly, compliance with all physical property specifications in ASTM D6319-10) | Meets (meaning it complies with ASTM D6319-10) |
Freedom from Pinholes | ASTM D5151-06 & ASTM D6319-10 | (Implicitly, compliance with pinhole requirements in ASTM D5151-06 and D6319-10) | Meets (meaning it complies with both ASTM D5151-06 and D6319-10) |
Powder Residual | ASTM D6124-06 | (Implicitly, compliance with powder residual limits in ASTM D6124-06, specifically |
Ask a specific question about this device
Page 1 of 1