(49 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free black nitrile patient examination glove, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
Here's an analysis of the provided information, structured according to your request:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard | Acceptance Criteria (from Standard) | Reported Device Performance |
|---|---|---|---|
| Dimension | ASTM D6319-10 | (Implicitly, compliance with all dimensional specifications in ASTM D6319-10) | Meets (meaning it complies with ASTM D6319-10) |
| Physical Properties | ASTM D6319-10 | (Implicitly, compliance with all physical property specifications in ASTM D6319-10) | Meets (meaning it complies with ASTM D6319-10) |
| Freedom from Pinholes | ASTM D5151-06 & ASTM D6319-10 | (Implicitly, compliance with pinhole requirements in ASTM D5151-06 and D6319-10) | Meets (meaning it complies with both ASTM D5151-06 and D6319-10) |
| Powder Residual | ASTM D6124-06 | (Implicitly, compliance with powder residual limits in ASTM D6124-06, specifically < 2mg) | Meets (Results generated values below 2mg of residual powder) |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10:2010) | Non-irritating | Gloves are non-irritating |
| Biocompatibility | Dermal Sensitization in the guinea pig (ISO 10993-10:2010) | No potential for sensitization | Gloves do not display any potential for sensitization |
2. Sample Size Used for the Test Set and Data Provenance
For the physical and chemical tests (Dimension, Physical Properties, Freedom from Pinholes, Powder Residual), the sample sizes are not explicitly stated in this summary. However, these standards (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06) typically define the sampling plans and quantities required for compliance testing. The data provenance is not explicitly stated in terms of country of origin but is implied to be from testing conducted by or for the device manufacturer to demonstrate compliance with international standards. These tests are generally prospective, in that they are performed specifically to assess the device's adherence to the standard.
For Biocompatibility, the sample sizes are also not explicitly stated in this summary, but ISO 10993-10:2010 would specify the minimum number of animals (rabbits for irritation, guinea pigs for sensitization) for testing. This data is prospectively generated for device testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of device (patient examination glove) does not typically involve expert review for "ground truth" in the way a diagnostic imaging device would. The "ground truth" for the performance criteria is established by the specifications defined within the referenced ASTM and ISO consensus standards. The "experts" would be the scientists and engineers who developed and validated these standards, and the laboratory personnel performing the tests according to standardized protocols. Qualifications of these personnel are not provided, but they would be expected to be trained and certified in laboratory testing procedures relevant to their specific roles.
4. Adjudication Method for the Test Set
Not applicable for this type of device and testing. The tests are objective and quantitative, based on standardized measurement protocols and predefined acceptance criteria in consensus standards. There is no subjective interpretation requiring adjudication among human reviewers.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is a medical device (glove) that does not involve "readers" or artificial intelligence. It's a physical barrier device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is a medical device (glove) that does not involve algorithms.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
The "ground truth" for the device's performance is the compliance with established, quantitative consensus standards (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, ISO 10993-10:2010). These standards define objective chemical and physical properties as well as biological responses.
8. The Sample Size for the Training Set
Not applicable. This device does not use machine learning or AI models, so there is no training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
{0}------------------------------------------------
1 Traditional 510(k) Premarket Notification ' ' Black Powder Free Nitrile Patient Exam Glove
510(k) Summary Page 1
K 121529
ﺮ ﺗﺮ
510(k) SUMMARY
February 17, 2012 DATE:
OWNER:
4
Northstar Healthcare Holdings 70 Sir John Rogerson's Quay Dublin 2, Ireland
JUL 1 2 2012
OFFICIAL CORRESPONDENT: Alex Nagy Manager, Quality Systems and Regulatory Affairs Cypress Medical Products LLC 1200 South Route 31 McHenry, IL 60050 Telephone: 815-385-0100 Fax: 815-385-0114 DEVICE NAME: Trade Name: Black Powder Free Nitrile Patient Exam Glove Common Name: Patient Examination Gloves Classification: Patient Examination Gloves Class: Class I
Product Code:
PREDICATE DEVICE(S):
| Predicate510(k) | Device Name | Indication | ClearanceDate | Company |
|---|---|---|---|---|
| K060724 | Nitrile PowderFree ExaminationGloves (Black andGray) | The examination glove is a disposabledevice intended for medical purposesthat is worn on the examiner's hand toprevent contamination between patientand examiner. | 17 May 2006 | SmartGlove Corp.SDN BHN.SelangorDarulEhsan,Malaysia |
DEVICE DESCRIPTION: Powder free black nitrile patient examination glove, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
LZA
{1}------------------------------------------------
Traditional 510(k) Premarket Notification Black Powder Free Nitrile Patient Exam Glove
K121529
51,0(k) Summary Page 2
، ﺗ
STATEMENT OF INTENDED USE:
The nitrile examination glove is a disposable device intended for medical purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.
TECHNOLOGICAL The Black Powder Free Nitrile Patient Exam Glove CHARACTERISTICS: is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with Nitrile using similar manufacturing processes.
ASSESSMENT OF NONCLINICAL DATA:
| Characteristic | Standard | Device Performance |
|---|---|---|
| Dimension | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from Pinholes | ASTM D5151-06 & ASTMD6319-10 | Meets |
| Powder Residual | ASTM D6124-06 | MeetsResults generatedvalues below 2mg ofresidual powder |
| Biocompatibility | Primary Skin Irritation in rabbits(ISO 10993-10:2010)Dermal Sensitization in the guineapig (ISO 10993-10:2010) | Gloves are non-irritatingGloves do not displayany potential forsensitization |
CONCLUSIONS: The Black Powder Free Nitrile Patient Exam Glove meets the requirements of established standards ASTM D6319-10, ASTM D6124-06, ASTM D5151-06, and ISO 10993-10:2010.
Based on the comparison of intended use, design, materials and performance, the Black Powder Free Nitrile Patient Exam Glove is substantially equivalent to the predicate device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Northstar Healthcare Holdings C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
JUL 12 2012
Re: K121529
Trade/Device Name: Black Powder Free Nitrile Patient Exam Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: June 27, 2012 Received: June 28, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1_of 1
INDICATIONS FOR USE
510(k) Number (if known): K121529
Black Powder Free Nitrile Patient Exam Glove Device Name:
Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
Page1 of1
510(k) Number: K121529
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.