Search Results
Found 5 results
510(k) Data Aggregation
(59 days)
Bovie Bantam/PRO Electrosurgical Generator, Bovie DERM 941 Electrosurgical Generator, Bovie DERM 942
The intended use for the DERM 942 is for removal and destruction of skin lesions and the coagulation of tissue.
The intended use for the Bantam/PRO is for removal and destruction of skin lesions and for electrosurgical cutting, blending, coagulation, fulguration, and bipolar procedures.
The intended use for the DERM 941 and DERM 942 is for removal and destruction of skin lesions and the coagulation of tissue.
The intended use for the Bantam/PRO is for removal and destruction of skin lesions and for electrosurgical cutting, blending, coagulation, fulguration, and bipolar procedures.
The Bantam/PRO performs the following functions:
Cut mode: Allows the user to utilize electrosurgical current to vaporize or cut tissue.
Blend mode: Combines cutting with hemostasis, which achieves a bloodless cut.
Coagulation mode: Is used for the destruction of tissue and hemostasis.
Fulguration mode: Allows the user to coagulate over a broad area with less tissue penetration.
Bipolar mode: Allows coagulation of tissue using forceps.
The DERM 942 has the following functions shared with the Bantam/PRO: Fulguration mode: Allows the user to coagulate over a broad area with less tissue penetration.
Bipolar mode: Allows coagulation of tissue using forceps.
The DERM 941 has only the basic Fulguration mode function.
The DERM 942 and 941 are termed High Frequency Dessicators since they perform treatment of skin lesions. The fulguration and bipolar are termed the dessication modes. The Bantam/PRO performs treatment of skin lesions but also has the basic cut and coagulation functions of a typical electrosurgical generator. All these units are considered low wattage (
This document is a 510(k) premarket notification for electrosurgical generators and does not describe a study involving an AI device or its acceptance criteria. Therefore, most of the requested information regarding AI device performance, sample sizes, ground truth establishment, or expert evaluation methods is not applicable.
The document focuses on demonstrating substantial equivalence of the new Bovie Bantam/PRO, Bovie Derm 941, and Bovie Derm 942 Electrosurgical Generators to legally marketed predicate devices.
Here's the information that is available from the provided text, adapted to the closest relevant categories:
1. Table of Acceptance Criteria and Reported Device Performance
Test Category | Purpose | Verification Performed |
---|---|---|
Electrical Verification | To verify electrical product and performance specification requirements where the test method is other than inspection or proof by design evidence. | Perform testing required to verify the electrical functionality of the generator. |
FPGA Validation for Main Board | To specify the FPGA validation procedure for the logic design used in the generator. | Validate the operation of the programmable systems on the main board of the generator to ensure FPGA design meets functional and timing requirements. |
Mechanical Verification | To verify mechanical product and performance specification requirements. | Perform testing required to verify the mechanical functionality of the generator and to ensure generator has adequate mechanical strength and resistance to heat. |
Usability Validation | Validate the user interface of the Electrosurgical Generator. | Validate that the device fulfills the user needs and intended uses. |
No specific quantitative acceptance criteria or performance metrics (like sensitivity, specificity, accuracy, etc., for an AI device) are reported in this document beyond verification that the devices pass these tests and meet relevant standards. The conclusion states that "Performance testing demonstrates equivalence between the proposed and predicate devices," implying successful completion of these tests.
2. Sample Size Used for the Test Set and Data Provenance
Not applicable. This is not a study assessing an AI device on a test set of data. The "test set" here refers to the physical devices being tested.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. Ground truth in the context of an AI device is not relevant here. The "ground truth" for these electrosurgical generators would be their functional specifications and safety standards.
4. Adjudication Method for the Test Set
Not applicable, as this is not a study with expert adjudication of data.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document does not describe an AI device or an MRMC study.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the device's performance is established by adherence to recognized international standards and specifications for electrosurgical equipment. The document explicitly lists:
- IEC-60601-1, Edition 3.1: Medical Electrical Equipment - Part 1: General Requirements For Safety
- IEC 60601-1-2:2007: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC-60601-2-2 : 2009: Particular requirements for the safety of high frequency surgical equipment
8. The Sample Size for the Training Set
Not applicable. This is not an AI device trained on a dataset.
9. How the Ground Truth for the Training Set Was Established
Not applicable.
Ask a specific question about this device
(243 days)
EASYSTAND BANTAM MEDIUM
The Altimate Medical EasyStand Bantam Medium is intended to assist persons who have difficulty rising from a seated position to a standing position and is indicated for persons weighing up to 200lbs.
The EasyStand Bantam Medium is a standing frame for indoor use that allows users with various degrees of physical disability to be supported in a standing, weight-bearing position. The device is a sitto-stand stander with the option of supine positioning. The optional Shadow Tray supports the user from sitting to standing and can be used as a desk/workstation in the seated and /or standing position. The device offers two options to raise the seat, the standard nanual hydraulic lift or the optional Pow'r Up lift. The Bantam Medium accommodates most individuals within the height range of 48"-66" and up to 200 lbs.
This document is a 510(k) summary for Altimate Medical's EasyStand Bantam Medium, a medical device. The document describes the device, its intended use, and its substantial equivalence to a predicate device. It also outlines the performance standards it meets.
Here's the information requested, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Performance Standards) | Reported Device Performance |
---|---|
IEC 60601-1 (3rd Edition) and CSA C22.2#60601-1 (3rd Edition) for basic safety and performance, including risk management | The EasyStand Bantam Medium meets these standards. |
IEC 60601-1 (2nd Edition) to address medical electrical equipment general requirements for safety, including mechanical strength testing for claimed weight capacities | The EasyStand Bantam Medium and predicate devices all meet this standard, including mechanical strength testing for the claimed weight capacities. |
IEC 60601-1-11 to address specific requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment | The EasyStand Bantam Medium meets this standard. |
IEC 60601-1-2 to address Class B emissions and immunity for non-life-supporting equipment | The EasyStand Bantam Medium meets this standard. |
BS EN 12182:1999 to address safety requirements for medical electrical systems. (Included in this test report, load testing results support EasyStand Bantam Medium substantial equivalency.) | Load testing results supporting substantial equivalency were included in the test report against this standard. |
2. Sample size used for the test set and the data provenance:
The document does not specify a "test set" in the context of patient data or a clinical study. Instead, it refers to the device being tested against various international performance standards. The testing described is likely engineering and safety testing on the device itself, not a study involving human subjects or data from a specific population.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable. The "ground truth" for the device's performance is established by its compliance with recognized international safety and performance standards for medical electrical equipment, not by expert consensus on clinical data. The experts would be the engineers and testing personnel who performed the evaluations against the standards.
4. Adjudication method for the test set:
This information is not applicable as the "test set" refers to compliance with engineering and safety standards, not a clinical trial requiring adjudication of patient outcomes or imaging data.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC comparative effectiveness study was not done. The device is a physical medical device (standing frame), not an AI diagnostic or assistive technology for human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
No, a standalone (algorithm only) performance study was not done. This device is a mechanical and electrical medical device, not an algorithm.
7. The type of ground truth used:
The "ground truth" for the device's acceptance is its compliance with recognized international performance standards (IEC 60601-1, CSA C22.2#60601-1, IEC 60601-1-11, IEC 60601-1-2, BS EN 12182:1999). This would involve engineering measurements, stress tests, electrical safety checks, and other objective evaluations against the quantitative and qualitative requirements of these standards.
8. The sample size for the training set:
This information is not applicable. The device is not an AI or machine learning model that requires a training set.
9. How the ground truth for the training set was established:
This information is not applicable for the same reason as above.
Ask a specific question about this device
(61 days)
BANTAM CATHETER (280MM BALLOON LENGTHS)
Balloon dilatation of the femoral, popliteal and infra-popliteal arteries. These catheters are not designed to be used in the coronary arteries.
The Bantam (280mm balloon lengths) catheter is a standard over-the-wire PTA catheter. The co-axial catheter has a balloon located near the distal tip. One lumen is used for inflation of the balloon, while the internal lumen allows access to the distal tip of the catheter for guidewire insertion (max 0.018"). The balloon expands to a known diameter at specific pressure.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Bantam OTW PTA Catheter:
Executive Summary:
The Bantam OTW PTA Catheter (280mm balloon lengths) was cleared through a Special 510(k) pathway, demonstrating substantial equivalence to a predicate device (Bantam K093139). The clearance relies entirely on in vitro testing to show that its technological characteristics and performance criteria are comparable to the predicate. No clinical studies, human-in-the-loop performance, or AI-assisted studies were conducted or required for this type of device clearance.
1. Acceptance Criteria and Reported Device Performance
The document does not explicitly list "acceptance criteria" as pass/fail thresholds with specific numerical values for each test. Instead, it describes "performance criteria" that were evaluated through in vitro testing to demonstrate comparability to the predicate device. The reported device performance is implicitly understood to have met these criteria because the device was granted substantial equivalence.
Acceptance Criteria Category (Evaluated Performance Criteria) | Reported Device Performance (as demonstrated by in vitro testing) |
---|---|
Visual and functional testing | Comparable to predicate device |
Catheter Body Diameters | Comparable to predicate device |
Inflation/Deflation time | Comparable to predicate device |
Introducer sheath withdrawal | Comparable to predicate device |
Leak and rated burst pressure testing | Comparable to predicate device |
Guidewire lumen stability test | Comparable to predicate device |
Tensile testing - Hub bond | Comparable to predicate device |
Tensile testing - proximal bond | Comparable to predicate device |
Working surface | Comparable to predicate device |
Balloon compliance | Comparable to predicate device |
Average burst | Comparable to predicate device |
Note: The document states that the results "demonstrate that the technological characteristics and performance criteria... are comparable to the predicate device." This implies that the Bantam (280mm) performed within acceptable ranges relative to the Bantam K093139 for all tested parameters. Specific numerical values for the predicate or the new device are not provided in this summary.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample size for each in vitro test. It generally states that "in vitro tests were performed."
- Data Provenance: The data is from in vitro testing conducted by Clearstream Technologies Ltd. The country of origin for the data is implicitly Ireland, where Clearstream Technologies Ltd. is located. This is a prospective study in the sense that the testing was performed specifically to support this 510(k) submission, but it is not a clinical (human) prospective study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This was an in vitro engineering performance study, not a clinical study requiring expert ground truth or assessment of cases. The "ground truth" for these tests would be the established engineering standards and methodologies (e.g., ISO 10555-1) and the performance of the predicate device under these same tests.
4. Adjudication Method for the Test Set
Not applicable. As this was an in vitro engineering performance study, there was no human reader or panel involved in adjudicating the "correctness" of the findings in the way one would for a clinical study with subjective interpretations. The tests yield objective measurements against predefined standards or comparisons.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable.
- This submission is for a medical device (PTA catheter), not an AI algorithm.
- No MRMC or human-in-the-loop studies were conducted or required.
- The concept of human readers improving with AI assistance is not relevant to this device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device. There is no algorithm component to evaluate in a standalone manner.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this submission is based on:
- Engineering standards: Compliance with ISO 10555-1.
- Predicate device performance: The performance of the legally marketed Bantam (K093139) under the same in vitro tests, as the new device is compared to it for "substantial equivalence."
- FDA guidance: Adherence to FDA guidance on non-clinical testing of medical devices.
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device, not a machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set for this type of device.
Ask a specific question about this device
(86 days)
BANTAM,48XXXXXX, 50XXXXXX
Balloon dilatation of the femoral, popliteal and infra-popliteal arteries. These catheters are not designed to be used in the coronary arteries.
The Bantam and Bantam & PTA Catheters are standard over-the-wire PTA catheters. The co-axial catheter has a balloon located near the distal tip. One lumen is used for inflation of the balloon, while the internal lumen allows access to the distal tip of the catheter for guidewire insertion (max 0.018" and 0.014"). The balloon expands to a known diameter at specific pressure.
The provided document is a 510(k) summary for a medical device (PTA Catheters) and does not describe acceptance criteria, a study proving device meeting acceptance criteria, or any of the specific details requested in the prompt (e.g., sample sizes, ground truth establishment, expert qualifications, MRMC studies, standalone performance).
Instead, it focuses on:
- Applicant Information: ClearStream Technologies Ltd, contact details.
- Device Information: Trade name (Bantam and Bantam α PTA Catheter), common name, classification, product code.
- Predicate Device: Savvy Long PTA Catheter (K072947).
- Device Description: Standard over-the-wire PTA catheters with a balloon for inflation and a lumen for guidewire insertion.
- Intended Use: Balloon dilatation of femoral, popliteal, and infra-popliteal arteries (excluding coronary arteries).
- Comparison to Predicate Device: Stated as "substantially equivalent."
- Test Data: Mentions safety and effectiveness demonstrated through "non-clinical design verification and design validation tests and analyses," but provides no details on these tests, their results, or acceptance criteria.
- FDA Correspondence: Official letter from FDA granting marketing clearance based on substantial equivalence to the predicate device.
- Indications for Use Form: Reiterates the intended use.
Therefore, I cannot provide the requested information based on the input document. The document primarily serves as a regulatory submission and clearance notice, not a detailed scientific study report.
Ask a specific question about this device
(69 days)
BANTAM
The Altimate Medical EasyStand Bantam is intended to assist persons who have difficulty rising from a seated position to a standing position and is indicated for persons weighing between 50 - 100 lbs., including pediatrics.
The EasyStand Bantam is a standing frame for indov use that allows users with various degrees of pirysical disability to ob supported in a suitioning, weight-bearing position. The optional Shadow Tray device is a sit-io stands standing and can be used as a desk/workstation in the seated and/or standing position. The device comes in two il (accommodating individuals ranging in height from 28" – 40" (71cm – 102 cm) and so books (23kg) and small (accommodating individuals ranging in height from 20" – 10 – 10 – 10 – 137 cm - 137 cm) and up to 100lbs (23kg).
The provided document is a 510(k) summary for the EasyStand Bantam, an electric lift chair. It primarily focuses on demonstrating substantial equivalence to a predicate device for regulatory approval. It does not contain information about acceptance criteria, device performance metrics, or a study design to measure such performance, as would be typical for a clinical or performance study of an AI/ML device.
Therefore, I cannot provide the requested information in the format specified. The document does not describe:
- A table of acceptance criteria and reported device performance.
- Sample sizes for test sets, data provenance (country, retrospective/prospective).
- Number and qualifications of experts for ground truth.
- Adjudication method.
- MRMC comparative effectiveness study or human reader improvement.
- Standalone algorithm performance.
- Type of ground truth used (pathology, outcomes data).
- Sample size for the training set.
- How ground truth for the training set was established.
Instead, the document focuses on:
- Device identification: Trade name, common name, classification, manufacturer.
- Intended Use: To assist persons who have difficulty rising from a seated position to a standing position and is indicated for persons weighing between 50 - 100 lbs., including pediatrics.
- Substantial Equivalence: Comparing the EasyStand Bantam to a legally marketed predicate device (Altimate Medical EasyStand Evolv; K062402) based on intended use, construction, and functionality. Key differences noted are user population (pediatric vs. larger persons) and additional features like a gas cylinder for seat elevation and a supine position option.
- Regulatory Status: 510(k) clearance by the FDA (K091242) for marketing the device.
- Performance Standards: States that "Although no performance standards or special controls have been developed to test positioning chairs, Altimate Medical has chosen to test the EasyStand Bantam against the standards as referenced in this submission." (The specific standards referenced are not detailed in the provided pages).
This document serves as regulatory submission for device marketing authorization, not a clinical or performance study with the detailed metrics you've requested for an AI/ML device.
Ask a specific question about this device
Page 1 of 1