Search Results
Found 2 results
510(k) Data Aggregation
(90 days)
Access EPO
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
The Access EPO Calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
The Access EPO assay is a two-site immunoenzymatic ("sandwich") assay. A sample is added to a reaction vessel along with the paramagnetic particles coated with mouse monoclonal anti-EPO, blocking reagent and the alkaline phosphatase conjugate. After incubation, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of analyte in the sample. Analyte concentration is automatically determined from a stored calibration.
The provided text describes the 510(k) premarket notification for the Beckman Coulter "Access EPO" assay when run on the Dxl 9000 Access Immunoassay Analyzer. This document outlines the device's technical characteristics, its intended use, and comparative studies against a predicate device. However, it does not describe an AI-powered diagnostic device or a study involving human readers and AI assistance. Instead, it describes an immunoassay for measuring erythropoietin levels. Therefore, many of the detailed points requested in your prompt (e.g., number of experts for ground truth, MRMC study, effect size of AI, standalone AI performance) are not applicable to this specific submission.
Despite this, I will extract and present the available information that aligns with your request regarding acceptance criteria and study details for this non-AI diagnostic device.
Here's the breakdown based on the provided document:
Device Name: Access EPO
Intended Use: The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias.
1. Table of Acceptance Criteria and Reported Device Performance
The document defines performance characteristics (acceptance criteria implicitly by demonstrating acceptable results) through various analytical studies.
Study Parameter | Acceptance Criteria (Implicit from demonstrated performance) | Reported Device Performance |
---|---|---|
Method Comparison | Slope close to 1, Intercept close to 0, High Correlation Coefficient (comparing new system to predicate) | N = 152 samples (141 native, 11 spiked) |
- Concentration Range: 0.79 - 697 mIU/mL (Access 2 values)
- Slope: 0.99 (95% CI: 0.97 – 1.00)
- Intercept: -0.040 (95% CI: -0.19 – 0.22)
- Correlation Coefficient (R): 1.00
(Demonstrates substantial equivalence to predicate) |
| Imprecision (Within-Laboratory) | ≤ 0.30 mIU/mL SD at concentrations ≤ 3.0 mIU/mL
≤ 10.0% CV at concentrations > 3.0 mIU/mL | Performed on four Dxl 9000 Access Immunoassay Analyzers, testing four native samples and one spiked sample. All reported within-laboratory %CVs are well below 10% for concentrations > 3.0 mIU/mL, and SDs are below 0.30 mIU/mL for concentrations ≤ 3.0 mIU/mL. - Sample 1 (2.0 mIU/L): Within-Laboratory SD 0.07, %CV 3.8
- Sample 2 (18 mIU/L): Within-Laboratory SD 0.5, %CV 3.0
- Sample 3 (105 mIU/L): Within-Laboratory SD 2.8, %CV 2.7
- Sample 4 (267 mIU/L): Within-Laboratory SD 6.3, %CV 2.4
- Sample 5 (548 mIU/L): Within-Laboratory SD 20.4, %CV 3.7 |
| Linearity | Assay demonstrates linearity across the measuring interval. | "The assay demonstrated linearity across the measuring interval" (Specific range is 0.6-750 mIU/L, as stated in the "Measuring Range" table). |
| Limit of Blank (LoB) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoB for Access assay is 0.6 mIU/mL. |
| Limit of Detection (LoD) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoD for Access assay is 0.6 mIU/mL. |
| Limit of Quantitation (LoQ) | Not explicitly stated as acceptance criteria, but claimed value is provided. | Claimed LoQ for Access assay is 0.6 mIU/mL. |
2. Sample Size Used for the Test Set and Data Provenance
- Method Comparison Test Set: 152 samples (141 native, 11 spiked).
- Imprecision Test Set: 5 samples (4 native, 1 spiked), tested in duplicate in 2 runs per day for 20 days on each of four analyzers (total 80 replicates per sample).
- Data Provenance: The document does not specify the country of origin for the samples or if they were retrospective or prospective. Given it's a 510(k) submission for a laboratory diagnostic, the samples are typically human biological samples collected for validation purposes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is not an image-based AI diagnostic device requiring expert adjudication of images. The "ground truth" for an immunoassay is typically established through reference methods or analysis of sample characteristics (e.g., known concentrations for spiked samples, or comparison to a cleared predicate device for native samples).
4. Adjudication Method for the Test Set
Not applicable, as it's an immunoassay for quantitative determination of erythropoietin, not a subjective interpretation task that would require adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This device is an automated immunoassay system, not an AI-assisted diagnostic tool that involves human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable in the context of an "algorithm only" AI performance. The performance described here is the "standalone" performance of the immunoassay system (Dxl 9000 Access Immunoassay Analyzer with Access EPO assay reagents).
7. The Type of Ground Truth Used
The ground truth for this device's performance evaluation is established through:
- Comparison to a Predicate Device: For the method comparison study, the predicate device (Access EPO Assay on Access 2 Immunoassay System) served as the reference for native samples.
- Known Concentrations: For imprecision and linearity studies, samples of known or targeted concentrations (including spiked samples) are used.
- Reference Standards/Materials: Implied for establishing accurate concentrations for LoB, LoD, and LoQ.
8. The Sample Size for the Training Set
Not applicable. This is an immunoassay, not a machine learning model that requires a "training set" in the conventional sense of AI development. The "training" of the instrument involves calibration using specific calibrators provided with the assay.
9. How the Ground Truth for the Training Set Was Established
Not applicable as there is no "training set" for an AI model. The "calibration" of the immunoassay is done using Access EPO Calibrators, which are intended to calibrate the assay. The ground truth for these calibrators would be established by the manufacturer through rigorous analytical methods to assign their precise concentrations.
Ask a specific question about this device
(417 days)
ACCESS EPO ASSAY
The Access EPO assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of erythropoietin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. This assay is intended as an aid in the diagnosis of anemias and polycythemias. With the advent of the administration of recombinant erythropoietin as a biologic therapy to increase red blood cell mass, an erythropoietin assay may be used also to aid in the prediction and monitoring of response to recombinant erythropoietin treatment in persons with anemias.
The Access EPO calibrators are intended to calibrate the Access EPO assay for the quantitative determination of EPO levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
The Access® EPO assay consists of the reagent pack and calibrators. Other items needed to perform the assay include the Access substrate and wash buffers.
Here's a breakdown of the acceptance criteria and study information for the Access® EPO Assay based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Performance Metric | Acceptance Criteria | Reported Device Performance |
---|---|---|
Precision | Total precision of ≤ 10% CV at EPO concentrations > 3 mIU/mL | Internal Data: |
- Precision tested at concentrations from approximately 9 to 475 mIU/mL.
- Within-run imprecision ranged from 1.8% CV to 8.7% CV.
- Total imprecision ranged from 2.6% CV to 8.7% CV.
Conclusion: Meets the criterion; total precision values are all ≤ 10% CV. |
| Analytical Sensitivity | Lowest detectable level of EPO distinguishable from zero (Access EPO Calibrator S0) is ≤0.6 mIU/mL | Internal Data: - Reported as ≤0.6 mIU/mL. |
| Dilution Recovery (Linearity) | Sample mean recovery values for all serum and plasma samples were within the range of 100 ± 15%. | Internal Data: - Dilution recovery studies performed by diluting multiple human serum and plasma (heparin) samples at various levels with Access EPO Calibrator S0.
- Sample mean recovery values were within the range of 100 ± 15%. |
| Methods Comparison | Acceptable agreement with predicate device (RDS Quantikine ELISA assays). | Internal Site Study: - Slope of 1.0511, intercept of -1.3595, and correlation coefficient (r) of 0.988.
- N=103 with EPO concentration range of approximately 3 to 182 mIU/mL.
External Site Study: - Slope of 1.1216, intercept of -2.4168, and correlation coefficient (r) of 0.995.
- N=113 with EPO concentration range of approximately 3 to 193 mIU/mL. |
| Analytical Specificity | No significant interference from therapeutic drugs, similar compounds, or potential sample contaminants (total protein, bilirubin, hemoglobin, and triglycerides). | Internal Data: - No significant interference from therapeutic drugs or similar compounds.
- No significant interference from potential sample contaminants (total protein, bilirubin, hemoglobin, and triglycerides). |
| Stability | EPO reagents stable for 28 days after opening; calibrators stable for 90 days after opening; calibration curve stable for 28 days. | Internal Data: - EPO reagents are stable for 28 days after opening.
- Calibrators are stable for 90 days after opening.
- The calibration curve is stable for 28 days. |
2. Sample Sizes Used for the Test Set and the Data Provenance:
- Precision: Not explicitly stated as a separate "test set" in the context of clinical samples, but the study was performed on samples with EPO concentrations ranging from 9 to 475 mIU/mL.
- Dilution Recovery (Linearity): "Multiple human serum and plasma (heparin) samples." The exact number is not provided.
- Methods Comparison:
- Internal Site Study: N=103
- External Site Study: N=113
- Analytical Specificity: Not explicitly stated as a numerical sample size, but indicates testing for interference from "therapeutic drugs or similar compounds" and "potential sample contaminants."
- Data Provenance: The studies are described as "Internal Site Study" and "External Site Study," suggesting the data comes from within Beckman Coulter's testing facilities and at least one other external laboratory. The samples were "human serum and plasma (heparin)." The country of origin is not explicitly stated, but the submission is to the FDA (USA). The studies appear to be prospective in nature, as they involve testing the performance of the new Access® EPO assay.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
This information is not provided in the document. For an immunoassay, the "ground truth" for the test set is typically established by comparing the device's results to a well-established, often reference, method. In this case, the predicate device (RDS Quantikine ELISA) served as the reference for method comparison. The document does not describe expert adjudication for these numerical results.
4. Adjudication Method for the Test Set:
This is not applicable in the context of an immunoassay performance study focused on quantitative measurements. Adjudication methods (like 2+1, 3+1) are typically used in imaging or diagnostic accuracy studies where expert consensus is required to establish a qualitative or subjective ground truth. Here, the comparison is against an established quantitative predicate method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. without AI Assistance:
This information is not applicable to this device. The Access® EPO Assay is an in vitro diagnostic (IVD) immunoassay, not an AI-powered diagnostic imaging device or an AI human-in-the-loop system. Therefore, MRMC studies examining human reader performance with or without AI assistance are not relevant here.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Yes, the studies presented represent the standalone performance of the Access® EPO Assay. This is an automated immunoassay system; its performance metrics (precision, sensitivity, linearity, methods comparison, specificity, stability) characterize the algorithm and instrument's ability to quantitatively determine EPO levels without direct human interpretation of the assay's core output.
7. The Type of Ground Truth Used:
For the quantitative performance studies, particularly the "Methods Comparison," the ground truth was effectively the results obtained from established predicate devices, specifically the R&D Systems Quantikine IVD Erythropoietin ELISA Kit. These predicate devices are themselves validated diagnostic assays.
8. The Sample Size for the Training Set:
This information is not provided in the document. For an immunoassay, a "training set" in the context of machine learning isn't directly applicable in the same way it would be for an AI algorithm. Instead, "training" for such a system typically involves:
- Assay development and optimization using numerous samples.
- Establishing calibration curves using specific calibrator materials.
- Method validation experiments that might involve hundreds or thousands of samples over time.
The document refers to "Summary of Performance Studies" as direct validation of the assay's performance attributes rather than detailing a distinct "training set."
9. How the Ground Truth for the Training Set Was Established:
As mentioned above, the concept of a "training set" with an explicitly established ground truth (like expert consensus or pathology for AI) is not directly applicable to this type of IVD immunoassay. The development and calibration of the assay would rely on:
- Reference materials/standards: The assay is standardized against the "WHO 2nd IRP 67/343" (International Reference Preparation). This serves as a fundamental "ground truth" for the quantitative measurement of EPO.
- Known concentrations: During assay development, samples with known EPO concentrations (often characterized by reference methods or gravimetric methods for primary standards) would be used to build and validate the assay's response curve.
- Cross-validation with existing methods: Early in development, comparison with established methods would help refine the assay's performance.
Ask a specific question about this device
Page 1 of 1