Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K983281
    Device Name
    APOB
    Date Cleared
    1999-01-26

    (130 days)

    Product Code
    Regulation Number
    866.5600
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apolipoprotein B assay is used for the quantitation of apolipoprotein B in human serum or plasma. Apolipoprotein B meaurements are useful in the diagnosis of premature coronary artery disease, hyper-b-lipoproteinemia, and hypo-b-lipoproteinemia.

    Device Description

    Apolipoprotein B is an in vitro diagnostic assay for the quantitative determination of apolipoprotein B in human serum or plasma. Antibodies to apolipoprotein B combine with apolipoprotein B in the sample to form insoluble immune complexes. The immune complexes cause an increase in light scattering (turbidity). The resulting increase in sample turbidity, measured at 604 nm, is directly proportional to the concentration of apolipoprotein B in the sample.

    AI/ML Overview

    Here's an analysis of the provided text, focusing on the acceptance criteria and the study used to demonstrate equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/MetricPredicate Device (K-ASSAY® Apo B on Hitachi® 717 Analyzer)Submitted Device (ApoB on AEROSET™ System)Result/PerformanceAcceptance Status
    Method ComparisonCorrelation CoefficientN/A (Comparator)N/A (Submitted Device)0.9793Met (Excellent)
    SlopeN/A (Comparator)N/A (Submitted Device)1.064Met (Close to 1)
    Y-interceptN/A (Comparator)N/A (Submitted Device)-2.920 mg/dLMet (Small value)
    PrecisionTotal %CV (Level 1)N/A (No specific value given for predicate)N/A (Submitted Device)2.5%Met (Good)
    Total %CV (Level 2)N/A (No specific value given for predicate)N/A (Submitted Device)1.8%Met (Good)
    Assay RangeQuantitative RangeN/A (A difference noted, but no specific value for predicate)N/A (Submitted Device)1.55 to 285.31 mg/dLMet
    SensitivityLimit of QuantitationN/A (No specific value given for predicate)N/A (Submitted Device)1.55 mg/dLMet
    Qualitative EquivalenceSimilar clinical resultsYesYesAchievedMet
    In Vitro ImmunoassayYesYesAchievedMet
    Quantitative DeterminationYesYesAchievedMet
    Immune Complex FormationYesYesAchievedMet

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for the comparative performance studies. It mentions "Comparative performance studies were conducted," but details on the number of samples are absent.

    Regarding data provenance: The studies were conducted by Abbott Laboratories, in support of their submission to the FDA. The specific country of origin for the samples is not mentioned, nor is it explicitly stated whether the study was retrospective or prospective. Given the nature of a 510(k) submission for an in vitro diagnostic, these are typically laboratory-based studies using patient samples, but the exact details are not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to an in vitro diagnostic assay like ApoB. "Ground truth" in this context refers to the established true value of apolipoprotein B in a sample. For diagnostic assays, the "ground truth" for the test set is typically established by using a reference method or a legally marketed predicate device, rather than expert consensus on images or clinical cases. In this case, the predicate device (K-ASSAY® Apo B on the Hitachi® 717 Analyzer) served as the defacto "ground truth" or comparator for the new device's performance.

    4. Adjudication Method for the Test Set

    Not applicable. As this is an in vitro diagnostic assay comparing performance against a predicate, there is no need for adjudication by human experts in the way it would be applied to, for example, image interpretation. The comparison is objective and quantitative.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for devices that involve human interpretation, such as diagnostic imaging AI, where the impact of AI assistance on human reader performance is evaluated. The ApoB assay is an automated in vitro diagnostic test.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the study described is a standalone performance study of the ApoB assay. It assesses the assay's performance characteristics (correlation, precision, range, sensitivity) as an isolated device on the AEROSET™ System, comparing its results to those obtained from a predicate device. There is no human-in-the-loop component in the operation or result generation for this type of assay.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is the results obtained from the legally marketed predicate device, the K-ASSAY® Apo B on the Hitachi® 717 Analyzer. The new device's measurements are compared directly to the measurements of this established and accepted method.

    8. The Sample Size for the Training Set

    No information regarding a "training set" or its sample size is provided. For an in vitro diagnostic assay like this, the development typically involves analytical validation studies rather than machine learning model training. The assay's parameters (e.g., reagent concentrations, reaction times) would be optimized during development, but this is not typically referred to as a "training set" in the context of AI/ML.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a "training set" in the context of machine learning. The "ground truth" for developing such an assay would be based on established analytical chemistry principles, reference materials, and clinical validation against known patient samples, but these details are not provided in this summary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1