Search Results
Found 5 results
510(k) Data Aggregation
(232 days)
AlertWatch:AC is intended for use by physicians for secondary monitoring of ICU patients. AlertWatch:AC is also intended for use by physicians providing supplemental remote support to bedside care teams in the management and care of ICU patients. AlertWatch:AC is not intended for use in monitoring pediatric or neonatal patients. AlertWatch:AC is a software system that combines data from the electronic medical record, networked physiologic monitors, and ancillary systems, and displays them on a dashboard view of the unit and patient. The clinical decision support is generated to aid in understanding the patient's current condition and changes over time. Once alerted by AlertWatch:AC, the physician must refer to the primary monitor, device or data source before making a clinical decision.
AlertWatch:AC is a secondary monitoring system used by physicians to monitor adult patients in an ICU environment. The purpose of the device is to synthesize a wide range of patient data and inform physicians of potential problems. Once alerted, the physician is instructed to refer to the primary monitoring device or EMR before making a clinical decision. The software design includes a default set of rules and alerts that can be configured by the hospital during the installation process. AlertWatch:AC is intended to supplement, not replace, a hospital's primary EMR. The device retrieves data from the electronic medical record (EMR) system and networked physiologic monitors, integrates this data, and performs a series of calculations to assess potential clinical issues. The information is conveyed both via organ colors and messages in the alert panel. Any alert can also be configured to send pages to physicians assigned to the patient.
The provided text describes the 510(k) clearance for AlertWatch:AC, a secondary monitoring system for ICU patients. However, it does not contain information about acceptance criteria or a specific study that proves the device meets such criteria in terms of the accuracy or performance of its clinical decision support algorithms.
The document focuses on regulatory compliance, outlining the device's intended use, technological comparison to a predicate device, and various verification and validation activities (software V&V, human factors study, default limits review, and wireless co-existence testing).
Therefore, I cannot provide the requested information regarding a table of acceptance criteria and reported device performance using figures like sensitivity, specificity, or AUC, nor can I detail the sample size, ground truth establishment, or expert qualifications for such a study, because this information is not present in the provided text.
Based on the document, here's what can be inferred or explicitly stated about the device's validation:
-
A table of acceptance criteria and the reported device performance: Not available in the provided text. The document refers to "software verification and validation testing" and "performance testing" but does not provide specific quantitative acceptance criteria or results for the clinical decision support functionality (e.g., accuracy of alerts). It states that "the results of performance testing demonstrate that the subject device performs in accordance with specifications and meets user needs and intended uses," but no specifics are given.
-
Sample sizes used for the test set and the data provenance:
- Software Verification and Validation Testing: Performed with "both constructed data and data from the EMR." No specific sample size for the "data from the EMR" portion is provided.
- Human Factors Study:
- Summative usability study: 18 users.
- Summative usability study on verbal alarm signals: 15 users.
- Data Provenance: Not explicitly stated beyond "data from the EMR." No geographical origin (e.g., country) or retrospective/prospective nature is specified.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Expert Committee for Default Limits and Thresholds: "Acute care physicians at the University of Michigan Health System." The exact number of physicians is not given, but it implies multiple experts. Their specific qualifications (e.g., years of experience, board certifications) are not detailed beyond "acute care physicians."
-
Adjudication method for the test set: Not explicitly mentioned for any testing related to the clinical decision support's accuracy. For the "Expert Committee" review of default limits, it states clinicians "reviewed the limits, provided feedback, and reviewed the final results," implying a consensus-based approach without detailing a specific adjudication method like 2+1 or 3+1.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No, the document explicitly states "Clinical Data: Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." This indicates that no MRMC study comparing human readers with and without AI assistance was performed or presented.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The document mentions "Software verification and validation testing... ensure that the product works as designed, and was tested with both constructed data and data from the EMR." While this implies algorithm-only testing as part of V&V, it doesn't present it as a separate performance study with metrics suitable for standalone performance (e.g., sensitivity, specificity for specific conditions detected by the algorithm). The device is positioned as clinical decision support where the physician "must refer to the primary monitor, device or data source before making a clinical decision," suggesting it's not intended for standalone diagnostic use.
-
The type of ground truth used:
- For the "Default Limits and Thresholds": Ground truth was established by "Review of References" (published studies) and "Expert Committee" (consensus/feedback from acute care physicians). This suggests a form of expert consensus and literature-based validation for the rule-based alerts.
- For "Software Verification and Validation Testing" using EMR data: The method for establishing ground truth for this EMR data is not described.
-
The sample size for the training set: Not applicable. The AlertWatch:AC is described as a "software system that combines data... and performs a series of calculations to assess potential clinical issues." It uses "a default set of rules and alerts" and "established patient risk and acuity calculations (SOFA and SIRS)." This indicates a rule-based or calculational system rather than a machine learning model that would typically have a "training set." Therefore, no training set size is mentioned.
-
How the ground truth for the training set was established: Not applicable, as it's not a machine learning model with a distinct training set. The "default limits and thresholds" and "established patient risk and acuity calculations" are based on literature review and expert consensus rather than labelled training data for an AI model.
Ask a specific question about this device
(140 days)
AlertWatch:OB is intended for use by clinicians for secondary monitoring of maternal patients in the labor and delivery unit. AlertWatch:OB is a maternal surveillance system that combines data from validated electronic medical record systems, and displays them in one place. Once alerted by AlertWatch:OB, the clinician must refer to the primary monitor, device, or data source before making a clinical decision.
AlertWatch:OB is a secondary monitoring system used by OB nurses, obstetricians, and OB anesthesiologists to monitor women in the Labor and Delivery (L&D) unit. The purpose of the program is to synthesize a wide range of maternal patient data and inform clinicians of potential problems. Once alerted, the clinician is instructed to refer to the primary monitoring device or EMR before making a clinical decision. AlertWatch:OB should only be connected to EMR systems that have been validated for use with AlertWatch:OB. AlertWatch, LLC performs the validation for each installation site.
Here's an analysis of the provided text regarding the AlertWatch:OB device, addressing the requested information:
Key Takeaway: The provided document is a 510(k) summary for AlertWatch:OB, which is primarily demonstrating substantial equivalence to a predicate device (AlertWatch:OR) based on similar intended use and technological characteristics. As such, it does not contain typical acceptance criteria and a study proving the device meets those criteria in the way one might expect for a de novo device or an AI/ML product with novel performance claims. Instead, the focus is on verification and validation of the software and human factors.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy targets) and corresponding reported device performance values in the context of clinical accuracy for the AlertWatch:OB's core function of identifying patient issues. This is because its claim is for "secondary monitoring" and not for primary diagnostic capabilities or automated decision-making.
However, the document does describe performance activities related to usability, software functionality, and the establishment of "default limits and thresholds."
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Software Functionality (V&V) | "Verification of AlertWatch:OR was conducted to ensure that the product works as designed, and was tested with both constructed data and data from the EMR. Validation was conducted to check the design and performance of the product." (Implies successful completion against internal specifications). |
| Wireless Co-existence | "Wireless Co-existence testing was performed to establish that the wireless components work effectively in the hospital environment." (Implies effective operation in the intended environment). |
| Usability / Human Factors | Formative Study: Conducted to identify and fix usability problems. Summative Study (17 users): "The results of the study showed that users with minimal training were able to successfully perform critical tasks and use the device for its intended purpose – to clarify clinical information and support information access." (Implies successful usability). |
| Clinical Validity of Default Limits | Phase 1 (References): "AlertWatch sought out definitive published studies that highlighted appropriate limits for certain patient conditions."Phase 2 (Expert Committee): "Obstetricians and OB anesthesia physicians at the University of Michigan Health System...reviewed the limits, provided feedback, and reviewed the final results."Phase 3 (External Experts): "External group of four anesthesiology and OB anesthesia experts...All approved the clinical limits." (Implies clinical consensus and validity of the set limits). |
2. Sample Size Used for the Test Set and Data Provenance
- Software Verification & Validation: The document mentions "constructed data and data from the EMR" for testing but does not specify sample sizes for these test sets or their provenance (e.g., country of origin, retrospective/prospective).
- Human Factors Study: 17 users were recruited for the summative usability study. The provenance of these users (e.g., hospital staff, general population) is not specified, nor is the origin of the data used in the usability testing (likely simulated or de-identified data).
- Clinical Data (main performance study): "Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device." This implies no specific clinical test set was used to establish performance metrics like sensitivity/specificity for identifying patient issues for the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Given the "Not applicable" for clinical data, there isn't a "ground truth" establishment in the traditional sense of a diagnostic or predictive AI device for a clinical test set.
However, for the establishment of default limits and thresholds:
- An "Expert Committee" of Obstetricians and OB anesthesia physicians from the University of Michigan Health System were involved. Their specific number is not given, but their roles are.
- An "External Experts" group of four anesthesiology and OB anesthesia experts provided final review. Their specific qualifications (e.g., years of experience) are not detailed beyond their specialty.
4. Adjudication Method for the Test Set
Not applicable, as no formal clinical "test set" with a need for adjudicated ground truth (e.g., for disease presence/absence) was used for direct device performance evaluation in the supplied document. For the "default limits and thresholds," the process involved multiple stages of expert review and approval, implying a consensus-based approach rather than formal adjudication of individual cases.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was done or reported in the document. The device is not making claims about improving human reader performance but rather providing secondary monitoring and information synthesis.
6. Standalone Performance Study (Algorithm Only)
The document primarily focuses on the software's functionality and its role as a "secondary monitoring system" that synthesizes existing EMR data and alerts clinicians. While "Software Verification and Validation Testing" was conducted to ensure it "works as designed" and "performed a series of calculations," it does not present a standalone performance study in terms of quantifiable clinical metrics (e.g., sensitivity, specificity for detecting specific conditions) for the algorithm itself. The device is intended to be used by clinicians who then refer to primary sources.
7. Type of Ground Truth Used
- For Software Verification & Validation: Likely internal functional specifications and expected outputs based on "constructed data and data from the EMR."
- For Default Limits and Thresholds:
- Published medical literature/references.
- Expert consensus among obstetricians, OB anesthesia physicians (University of Michigan Health System), and an external group of four anesthesiology and OB anesthesia experts. This is the closest to a "ground truth" described, but it pertains to the establishment of the parameters the device uses, not the evaluation of the device's output against a diagnostic truth.
8. Sample Size for the Training Set
The document does not mention a "training set" in the context of an AI/ML algorithm that learns from data to make predictions or categorizations. The AlertWatch:OB appears to be a rule-based or threshold-based system that processes data according to predefined clinical limits.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set or machine learning algorithm that requires a "ground truth" for learning in this 510(k) summary. The "ground truth" for the device's operational parameters (default limits) was established via literature review and expert consensus.
Ask a specific question about this device
(117 days)
AlertWatch:OR is intended for use by clinicians for secondary monitoring of patients within operating rooms. AlertWatch:OR combines data from networked physiologic monitors, anesthesia information management systems and patient medical records and displays them in one place. AlertWatch:OR can only be used with both physiological monitors and AIMS versions that have been validated by AlertWatch. Once alerted, you must refer to the primary monitor or device before making a clinical decision.
AlertWatch:OR is also intended for use by supervising anesthesiologists outside of operating rooms. Once alerted, the supervising anesthesiologist must contact the clinician inside the operating room or must return to the operating room before making a clinical decision. Once either clinician is alerted, they must refer to the primary monitor or device before making a clinical decision.
AlertWatch:OR is a display and secondary alert system used by the anesthesiology staff residents. CRNA's, and attending anesthesiologists - to monitor patients in operating rooms. The purpose of the program is to synthesize a wide range of patient data and inform clinicians of potential problems that might lead to immediate or long-term complications. Once alerted, the clinician is instructed to refer to the primary monitoring device before making a clinical decision. AlertWatch:OR should only be connected to AIMS systems and physiologic monitors that have been validated for use with AlertWatch:OR. AlertWatch, LLC performs the validation for each installation site.
The purpose of this 510(k) is for marketing clearance of AlertWatch:OR 2.50 which includes minor modifications to some display views, user features, indicators and alerts as well as compatibility with the iPad and the iPhone.
The provided text is a 510(k) summary for the medical device AlertWatch:OR, focusing on its substantial equivalence to a previously cleared device. The "Performance Data" section discusses a human factors study and the process for establishing default limits and thresholds, but it does not describe an in-depth study with quantitative acceptance criteria for device performance in detecting or alerting.
Specifically, the document does not contain a table of acceptance criteria and reported device performance in terms of diagnostic metrics (e.g., sensitivity, specificity, accuracy). It focuses more on usability and the validation of alert thresholds as part of its performance data.
Therefore, many of the requested details about a study proving the device meets acceptance criteria are not present in this document.
Here's a breakdown of what can be extracted and what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
Not available in the provided text. The document refers to the validation of default limits and thresholds and a human factors study, but not quantitative performance criteria for the device's alerting function.
2. Sample Size for Test Set and Data Provenance
- Human Factors Study: The text mentions a "comprehensive human factors study" for the iPhone version. It does not specify the sample size of users or the provenance of the data used in this study (e.g., retrospective/prospective, country of origin).
- Default Limits and Thresholds: The validation of these limits involved "Review of References," an "Expert Committee" (anesthesia physicians at the University of Michigan Health System), and "External Experts" (four anesthesiology experts). This is not a "test set" in the traditional sense of evaluating device performance on patient data, but rather a process for establishing system parameters.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Human Factors Study: Not specified.
- Default Limits and Thresholds:
- Expert Committee: An unspecified number of "anesthesia physicians at the University of Michigan Health System." Their specific qualifications beyond being "anesthesia physicians" are not detailed (e.g., years of experience).
- External Experts: Four "anesthesiology experts." Specific qualifications are not detailed.
4. Adjudication Method for the Test Set
Not applicable as a traditional "test set" with adjudicated ground truth for diagnostic performance is not described. The expert involvement for default limits was for "opinion and confirmation" and "final review," not for adjudicating individual cases on a test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study is mentioned. The document primarily discusses changes and validation of an existing secondary monitoring system, not its comparative effectiveness against human readers.
6. Standalone Performance
The device is described as a "secondary monitoring" and "secondary alert system." The Indications for Use explicitly state: "Once alerted, you must refer to the primary monitor or device before making a clinical decision." This indicates it's designed to assist human clinicians, not to operate autonomously as a standalone diagnostic tool. Therefore, a standalone performance study in the sense of demonstrating diagnostic accuracy independent of a human is not directly applicable or discussed for decision-making. The "Performance Data" section addresses usability and the clinical validity of its alert thresholds rather than standalone diagnostic performance metrics.
7. Type of Ground Truth Used
- Human Factors Study: Ground truth would relate to user task completion and usability issues, not clinical diagnosis.
- Default Limits and Thresholds: Based on "Review of References" (published studies), "Expert Committee" opinion/confirmation, and "External Experts" review. This is expert consensus/opinion based on clinical knowledge and literature rather than pathology or outcomes data on a specific patient cohort for device performance evaluation.
8. Sample Size for the Training Set
No training set is mentioned in the context of machine learning. This device appears to be a rule-based or threshold-based system rather than one that employs machine learning requiring a training set.
9. How Ground Truth for the Training Set Was Established
Not applicable, as no training set for machine learning is described.
Ask a specific question about this device
(353 days)
AlertWatch:OR is intended for use by clinicians for secondary monitoring of patients within operating rooms. AlertWatch:OR combines data from networked physiologic monitors, anesthesia information management medical records and displays them in one place. AlertWatch:OR can only be used with both physiological monitors and AIMS versions that have been validated by AlertWatch. Once alerted, you must refer to the primary monitor or device before making a clinical decision.
AlertWatch:OR is a display and secondary alert system used by the anesthesiology staff - residents, CRNA's, and attending anesthesiologists - to monitor patients in operating rooms. The purpose of the program is to synthesize a wide range of patient data and inform clinicians of potential problems that might lead to immediate or long-term complications. Once alerted, the clinician is instructed to refer to the primary monitoring device before making a clinical decision. AlertWatch:OR should only be connected to AIMS systems and physiologic monitors that have been validated for use with Alert Watch:OR. Alert Watch, Inc. performs the validation for each installation site.
Here's a breakdown of the acceptance criteria and the study details for the AlertWatch:OR device, based on the provided document:
The document does not explicitly state formal acceptance criteria with specific performance metrics (e.g., sensitivity, specificity, accuracy thresholds). Instead, the performance testing section describes verification and validation activities designed to ensure the product works as designed, meets its stated requirements, and is clinically useful.
1. Table of Acceptance Criteria and Reported Device Performance
As specific numerical acceptance criteria (e.g., sensitivity > X%, specificity > Y%) are not provided, the table below reflects the described performance testing outcomes.
| Acceptance Criterion (Implicit from Study Design) | Reported Device Performance (from "Performance Testing" section) |
|---|---|
| Verification: Analysis Output Accuracy | Produced desired output for each rule/algorithm using constructed data. |
| Verification: Data Display Accuracy | Produced desired display for each test case using constructed data. |
| Verification: Data Collector Functionality | Live Collector and Data Collector returned correct data from the EMR. |
| Verification: Product Functionality with Historical Data | Product worked as designed using a set of cases from actual patients. |
| Validation: Design Review & Software Requirements Specification (SRS) Accuracy | Process and various inputs for creating the product design (SRS) were reviewed. SRS was reviewed for clinical accuracy. |
| Validation: Clinical Utility | Clinical utility of the product was validated by analyzing case outcomes. |
| Validation: Human Factors | Summative Human Factors study conducted to demonstrate the device meets user needs. |
| Overall Performance Claim | The results of the verification and validation activities demonstrate that the AlertWatch:OR complies with its stated requirements and meets user needs and intended uses. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Verification (Step 4: Historical Data): "a set of cases from actual patients" - The exact number of cases is not specified.
- Data Provenance: "data from the EMR" and "a set of cases from actual patients." The document does not specify the country of origin, nor explicitly whether it was retrospective or prospective, though "historical data" strongly implies retrospective data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not specify the number of experts used or their qualifications for establishing ground truth, and it does not explicitly describe a ground truth establishment process involving experts in the traditional sense (e.g., for diagnostic accuracy). The "clinical accuracy" review of the software requirements specification implies expert involvement, but details are missing.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method (such as 2+1, 3+1) for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or described. The study focused on the device's functionality and utility rather than a direct comparison of human readers with and without AI assistance to quantify improvement.
6. Standalone (Algorithm Only) Performance Study
The verification steps, particularly "Verify the analysis output" and "Verify the data display" using "constructed data," and "Verify the product with historical data," indicate that the algorithm's performance was evaluated in a standalone manner (without human-in-the-loop) to ensure it performs "as designed" and produces "desired output/display." However, these are functional verifications rather than a typical standalone diagnostic performance study with metrics like sensitivity, specificity, or PPV/NPV.
7. Type of Ground Truth Used
The ground truth for the performance testing appears to be established by:
- "Desired output" based on the "Software Requirements Specification" for constructed data tests (functional verification).
- "Works as designed" when tested with "a set of cases from actual patients" (implies comparison to expected system behavior based on its design, rather than a clinical outcome or expert diagnosis acting as a gold standard).
- "Clinical utility... by analyzing case outcomes" suggests that real-world patient outcomes were used to assess the value of the alerts generated. This hints at an outcome-based ground truth for the validation of clinical utility, but details are scarce.
8. Sample Size for the Training Set
The document does not specify the sample size for a training set. The descriptions of verification and validation do not refer to machine learning model training. The device seems to operate based on "rules/algorithms in the Software Requirements Specification" rather than a trained AI model.
9. How the Ground Truth for the Training Set Was Established
As there's no mention of a dedicated training set or a machine learning model requiring such, this information is not applicable based on the provided text. The device likely relies on predefined rules and algorithms.
Ask a specific question about this device
(44 days)
Tooth colored posterior restorative material
For usage in a similar situation that would be applicable to an amalgam restoration.
Not Found
This document is a 510(k) clearance letter from the FDA for a dental restorative material named "ALERT." As such, it does not contain the detailed information required to answer your specific questions about acceptance criteria and a study proving those criteria.
A 510(k) clearance primarily establishes substantial equivalence to a predicate device, rather than requiring extensive clinical trials with detailed performance metrics and statistical analyses as might be found in a Premarket Approval (PMA) application or a more in-depth clinical study report.
Therefore, I cannot provide the requested information from this document. The document primarily confirms that the device can be marketed because it is substantially equivalent to existing devices and provides information on regulatory compliance.
Ask a specific question about this device
Page 1 of 1